Accuracy of pleth variability index to predict fluid responsiveness in mechanically ventilated patients: a systematic review and meta-analysis

ReviewPaper

Abstract

To systemically evaluate the accuracy of pleth variability index to predict fluid responsiveness in mechanically ventilated patients. A literature search of PUBMED, OVID, CBM, CNKI and Wanfang Data for clinical studies in which the accuracy of pleth variability index to predict fluid responsiveness was performed (last update 5 April 2015). Related journals were also searched manually. Two reviewers independently assessed trial quality according to the modified QUADAS items. Heterogeneous studies and meta-analysis were conducted by Meta-Disc 1.4 software. A subgroup analysis in the operating room (OR) and in intensive care unit (ICU) was also performed. Differences between subgroups were analyzed using the interaction test. A total of 18 studies involving 665 subjects were included. The pooled area under the receiver operating characteristic curve (AUC) to predict fluid responsiveness in mechanically ventilated patients was 0.88 [95 % confidence interval (CI) 0.84–0.91]. The pooled sensitivity and specificity were 0.73 (95 % CI 0.68–0.78) and 0.82 (95 % CI 0.77–0.86), respectively. No heterogeneity was found within studies nor between studies. And there was no significant heterogeneity within each subgroup. No statistical differences were found between OR subgroup and ICU subgroup in the AUC [0.89 (95 % CI 0.85–0.92) versus 0.90 (95 % CI 0.82–0.94); P = 0.97], and in the specificity [0.84 (95 % CI 0.75–0.86) vs. 0.84 (95 % CI 0.75–0.91); P = 1.00]. Sensitivity was higher in the OR subgroup than the ICU subgroup [0.84 (95 % CI 0.78–0.88) vs. 0.56 (95 % CI 0.47–0.64); P = 0.00004]. The pleth variability index has a reasonable ability to predict fluid responsiveness.

Keywords

Pleth variability index Fluid responsiveness Meta-analysis Sensitivity Specificity AUC 

Notes

Compliance with ethical standards

Conflict of interest

The authors declare that they have no financial relationship with any organization and have no conflict of interest.

References

  1. 1.
    Marik PE, Cavallazzi R, Vasu T, Hirani A. Dynamic changes in arterial waveform derived variables and fluid responsiveness in mechanically ventilated patients: a systematic review of the literature. Crit Care Med. 2009;37:2642–7.CrossRefPubMedGoogle Scholar
  2. 2.
    Michard F, Boussat S, Chemla D, Anguel N, Mercat A, Lecarpentier Y, Richard C, Pinsky MR, Teboul JL. Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med. 2000;162(1):134–8.CrossRefPubMedGoogle Scholar
  3. 3.
    Tavernier B, Makhotine O, Lebuffe G, Dupont J, Scherpereel P. Systolic pressure variation as a guide to fluid therapy in patients with sepsis-induced hypotension. Anesthesiology. 1998;89(6):1313–21.CrossRefPubMedGoogle Scholar
  4. 4.
    Marik PE, Baram M, Vahid B. Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares. Chest. 2008;134:172–8.CrossRefPubMedGoogle Scholar
  5. 5.
    Cannesson M, Musard H, Desebbe O, Boucau C, Simon R, Hénaine R, Lehot JJ. The ability of stroke volume variations obtained with Vigileo/FloTrac system to monitor fluid responsiveness in mechanically ventilated patients. Anesth Analg. 2009;108:513–7.CrossRefPubMedGoogle Scholar
  6. 6.
    Cannesson M, Delannoy B, Morand A, Rosamel P, Attof Y, Bastien O, Lehot JJ. Does the pleth variability index indicate the respiratory-induced variation in the plethysmogram and arterial pressure waveforms? Anesth Analg. 2008;106:1189–94.CrossRefPubMedGoogle Scholar
  7. 7.
    Sandroni C, Cavallaro F, Marano C, Falcone C, De Santis P, Antonelli M. Accuracy of plethysmographic indices as predictors of fluid responsiveness in mechanically ventilated adults: a systematic review and meta-analysis. Intensive Care Med. 2012;38(9):1429–37.CrossRefPubMedGoogle Scholar
  8. 8.
    Takeyama M, Matsunaga A, Kakihana Y, Masuda M, Kuniyoshi T, Kanmura Y. Impact of skin incision on the pleth variability index. J Clin Monit Comput. 2011;25:215–21.CrossRefPubMedGoogle Scholar
  9. 9.
    Fischer MO, Pellissier A, Saplacan V, Gérard JL, Hanouz JL, Fellahi JL. Cephalic versus digital plethysmographic variability index measurement: a comparative pilot study in cardiac surgery patients. J Cardiothorac Vasc Anesth. 2014;28(6):1510–5.CrossRefPubMedGoogle Scholar
  10. 10.
    Siswojo AS, Wong DM, Phan TD, Kluger R. Pleth variability index predicts fluid responsiveness in mechanically ventilated adults during general anesthesia for noncardiac surgery. J Cardiothorac Vasc Anesth. 2014;28(6):1505–9.CrossRefPubMedGoogle Scholar
  11. 11.
    Xu Z, Zhang J, Xia Y, Deng X. Accuracy of pulse oximeter perfusion index in thoracic epidural anesthesia under basal general anesthesia. Int J Clin Exp Med. 2014;7(7):1728–34.PubMedPubMedCentralGoogle Scholar
  12. 12.
    Høiseth LØ, Hoff IE, Hagen OA, Landsverk SA, Kirkebøen KA. Dynamic variables and fluid responsiveness in patients for aortic stenosis surgery. Acta Anaesthesiol Scand. 2014;58(7):826–34.CrossRefPubMedGoogle Scholar
  13. 13.
    Lu W, Dong J, Xu Z, Shen H, Zheng J. The pleth variability index as an indicator of the central extracellular fluid volume in mechanically ventilated patients after anesthesia induction: comparison with initial distribution volume of glucose. Med Sci Monit. 2014;20:386–92.CrossRefPubMedPubMedCentralGoogle Scholar
  14. 14.
    Vos JJ, Kalmar AF, Struys MM, Wietasch JK, Hendriks HG, Scheeren TW. Comparison of arterial pressure and plethysmographic waveform-based dynamic preload variables in assessing fluid responsiveness and dynamic arterial tone in patients undergoing major hepatic resection. Br J Anaesth. 2013;110(6):940–6.CrossRefPubMedGoogle Scholar
  15. 15.
    Fischer MO, Pelissier A, Bohadana D, Gérard JL, Hanouz JL, Fellahi JL. Prediction of responsiveness to an intravenous fluid challenge in patients after cardiac surgery with cardiopulmonary bypass: a comparison between arterial pulse pressure variation and digital plethysmographic variability index. J Cardiothorac Vasc Anesth. 2013;27(6):1087–93.CrossRefPubMedGoogle Scholar
  16. 16.
    Julien F, Hilly J, Sallah TB, Skhiri A, Michelet D, Brasher C, Varin L, Nivoche Y, Dahmani S. Plethysmographic variability index (PVI) accuracy in predicting fluid responsiveness in anesthetized children. Paediatr Anaesth. 2013;23(6):536–46.CrossRefPubMedGoogle Scholar
  17. 17.
    Monnet X, Guérin L, Jozwiak M, Bataille A, Julien F, Richard C, Teboul JL. Pleth variability index is a weak predictor of fluid responsiveness in patients receiving norepinephrine. Br J Anaesth. 2013;110(2):207–13.CrossRefPubMedGoogle Scholar
  18. 18.
    Haas S, Trepte C, Hinteregger M, Fahje R, Sill B, Herich L, Reuter DA. Prediction of volume responsiveness using pleth variability index in patients undergoing cardiac surgery after cardiopulmonary bypass. J Anesth. 2012;26(5):696–701.CrossRefPubMedGoogle Scholar
  19. 19.
    Fu Q, Mi WD, Zhang H. Stroke volume variation and pleth variability index to predict fluid responsiveness during resection of primary retroperitoneal tumors in Hans Chinese. Biosci Trends. 2012;6(1):38–43.PubMedGoogle Scholar
  20. 20.
    Hood JA, Wilson RJ. Pleth variability index to predict fluid responsiveness in colorectal surgery. Anesth Analg. 2011;113:1058–63.CrossRefPubMedGoogle Scholar
  21. 21.
    Desgranges FP, Desebbe O, Ghazouani A, Gilbert K, Keller G, Chiari P, Robin J, Bastien O, Lehot JJ, Cannesson M. Influence of the site of measurement on the ability of plethysmographic variability index to predict fluid responsiveness. Br J Anaesth. 2011;107:329–35.CrossRefPubMedGoogle Scholar
  22. 22.
    Renner J, Broch O, Gruenewald M, Scheewe J, Francksen H, Jung O, Steinfath M, Bein B. Non-invasive prediction of fluid responsiveness in infants using pleth variability index. Anaesthesia. 2011;66:582–9.CrossRefPubMedGoogle Scholar
  23. 23.
    Loupec T, Nanadoumgar H, Frasca D, Petitpas F, Laksiri L, Baudouin D, Debaene B, Dahyot-Fizelier C, Mimoz O. Pleth variability index predicts fluid responsiveness in critically ill patients. Crit Care Med. 2011;39:294–9.CrossRefPubMedGoogle Scholar
  24. 24.
    Cai QF, Mi WD, Yuan WX. The ability of pleth variability index to predict fluid responsiveness in mechanically ventilated patients under general anaesthesia. Zhonghua Wai Ke Za Zhi. 2010;48:1628–32.PubMedGoogle Scholar
  25. 25.
    Zimmermann M, Feibicke T, Keyl C, Prasser C, Moritz S, Graf BM, Wiesenack C. Accuracy of stroke volume variation compared with pleth variability index to predict fluid responsiveness in mechanically ventilated patients undergoing major surgery. Eur J Anaesthesiol. 2010;27:555–61.PubMedGoogle Scholar
  26. 26.
    Cannesson M, Desebbe O, Rosamel P, Delannoy B, Robin J, Bastien O, Lehot JJ. Pleth variability index to monitor the respiratory variations in the pulse oximeter plethysmographic waveform amplitude and predict fluid responsiveness in the operating theatre. Br J Anaesth. 2008;101:200–6.CrossRefPubMedGoogle Scholar
  27. 27.
    Leeflang MM, Deeks JJ, Gatsonis C, Bossuyt PM. Cochrane diagnostic test accuracy working group. Systematic reviews of diagnostic test accuracy. Ann Intern Med. 2008;149:889–97.CrossRefPubMedPubMedCentralGoogle Scholar
  28. 28.
    Higgins JP1, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med. 2002; 21(11):1539-58.Google Scholar
  29. 29.
    Altman DG, Bland JM. Interaction revisited: the difference between two estimates. Br Med J. 2003;326:219.CrossRefGoogle Scholar
  30. 30.
    Cannesson M, de Backer D, Hofer CK. Using arterial pressure waveform analysis for the assessment of fluid responsiveness. Expert Rev Med Devices. 2011;8:635–46.CrossRefPubMedGoogle Scholar
  31. 31.
    Keller G, Cassar E, Desebbe O, Lehot JJ, Cannesson M. Ability of pleth variability index to detect hemodynamic changes induced by passive leg raising in spontaneously breathing volunteers. Crit Care. 2008;12(2):R37.CrossRefPubMedPubMedCentralGoogle Scholar
  32. 32.
    Shelley KH, Jablonka DH, Awad AA, Stout RG, Rezkanna H, Silverman DG. What is the best site for measuring the effect of ventilation on the pulse oximeter waveform? Anesth Analg. 2006;103:372–7.CrossRefPubMedGoogle Scholar
  33. 33.
    Broch O, Bein B, Gruenewald M, Höcker J, Schöttler J, Meybohm P, Steinfath M, Renner J. Accuracy of the pleth variability index to predict fluid responsiveness depends on the perfusion index. Acta Anaesthesiol Scand. 2011;55:686–93.CrossRefPubMedGoogle Scholar
  34. 34.
    Awad AA, Ghobashy MA, Stout RG, Silverman DG, Shelley KH. How does the plethysmogram derived from the pulse oximeter relate to arterial blood pressure in coronary artery bypass graft patients? Anesth Analg. 2001;93:1466–71.CrossRefPubMedGoogle Scholar
  35. 35.
    Golparvar M, Naddafnia H, Saghaei M. Evaluating the relationship between arterial blood pressure changes and indices of pulse oximetric plethysmography. Anesth Analg. 2002;95:1686–90.CrossRefPubMedGoogle Scholar
  36. 36.
    Lima AP, Beelen P, Bakker J. Use of a peripheral perfusion index derived from the pulse oximetry signal as a non-invasive indicator of perfusion. Crit Care Med. 2002;30:1210–3.CrossRefPubMedGoogle Scholar
  37. 37.
    Haluska BA, Jeffriess L, Fathi RB, Mottram PM, Carlier SG, Marwick TH. Pulse pressure vs. total arterial compliance as a marker of arterial health. Eur J Clin Invest. 2005;35:438–43.CrossRefPubMedGoogle Scholar
  38. 38.
    Yamaura K, Irita K, Kandabashi T, Tohyama K, Takahashi S. Evaluation of finger and forehead pulse oximeters during mild hypothermic cardiopulmonary bypass. J Clin Monit Comput. 2007;21:249–52.CrossRefPubMedGoogle Scholar
  39. 39.
    Biais M, Cottenceau V, Petit L, Masson F, Cochard JF, Sztark F. Impact of noradrenaline on the relationship between pleth variability index and pulse pressure variations in ICU adult patients. Crit Care. 2011;15(4):R168.CrossRefPubMedPubMedCentralGoogle Scholar
  40. 40.
    Vos JJ, Kalmar AF, Struys MM, Porte RJ, Wietasch JK, Scheeren TW, Hendriks HG. Accuracy of non-invasive measurement of haemoglobin concentration by pulse co-oximetry during steady state and dynamic conditions in patients undergoing liver surgery. Br J Anaesth. 2012;109:522–8.CrossRefPubMedGoogle Scholar
  41. 41.
    Bergek C, Zdolsek JH, Hahn RG. Accuracy of noninvasive haemoglobin measurement by pulse oximetry depends on the type of infusion fluid. Eur J Anaesth. 2012;29:586–92.Google Scholar
  42. 42.
    Landsverk SA, Hoiseth LO, Kvandal P, Hisdal J, Skare O, Kirkeboen KA. Poor agreement between respiratory variations in pulse oximetry photoplethysmographic waveform amplitude and pulse pressure in intensive care unit patients. Anesthesiology. 2008;109:849–55.CrossRefPubMedGoogle Scholar
  43. 43.
    Lima A, Bakker J. Noninvasive monitoring of peripheral perfusion. Intensive Care Med. 2005;31:1316–26.CrossRefPubMedGoogle Scholar

Copyright information

© Springer Science+Business Media New York 2015

Authors and Affiliations

  1. 1.The First Hospital of China Medical UniversityShenyangChina

Personalised recommendations