Isocapnic hyperpnea with a portable device in Cystic Fibrosis: an agreement study between two different set-up modalities

Journal of Clinical Monitoring and Computing Aims and scope Submit manuscript

Abstract

To evaluate the bias and precision of the respiratory muscle training device formulas to predict respiratory minute volume (RMV) and volume of the reservoir bag (BV) on a cohort of subjects with Cystic Fibrosis (CF). CF patients with available pulmonary function tests and maximal voluntary manoeuvres were included in the study. Vital capacity and maximal voluntary ventilation were extracted from subjects’ records and then inserted to the manufacturer’s formulas to obtain RMV and BV (measured setting). RMV and BV were compared according to standard and measured formulas in males and females. Sample was described and then processed using Bland–Altman analysis. Bland–Altman analysis for RMV revealed a bias and precision of 8.8 ± 29 L/min in males and 28.8 ± 16 L/min in females; 0.4 ± 0.5 L in males and 0.7 ± 0.4 L in females for BV. Concordance correlation coefficients for RMV were −0.03 in males and 0.02 in females; 0.22 in males and 0.03 in females for BV, reinforcing an unsatisfactory concordance between measured and manufacturer setting. This study shows considerable discrepancies between the two methods, making the degree of agreement not clinically acceptable. This might cause inappropriate setting and disservice to patients with CF.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Leddy JJ, Limprasertkul A, Patel S, Modlich F, Buyea C, Pendergast DR, et al. Isocapnic hyperpnea training improves performance in competitive male runners. Eur J Appl Physiol. 2007;99:665–76.

    Article  PubMed  Google Scholar 

  2. Sk Illy, Held U, Frank I, Spengler CM. Effect of respiratory muscle training on exercise performance in healthy individuals. Sports Med. 2012;42(8):707–24.

    Article  Google Scholar 

  3. Brusasco V, Crapo R, Viegi G. Coming together: the ATS/ERS consensus on clinical pulmonary function testing. Eur Respir J. 2005;26:1–2.

    Article  CAS  PubMed  Google Scholar 

  4. URL: http://spirotiger.it/spirotiger-medical/.

  5. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8:135–60.

    Article  CAS  PubMed  Google Scholar 

  6. Lin LI-K. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45:255–68.

    Article  CAS  PubMed  Google Scholar 

  7. Lin LI-K. A note on the concordance correlation coefficient. PDA J Pharm Sci Technol. 2000;54:23–6.

    Google Scholar 

  8. Keens TG, Krastins IRB, Wannameker EM, Levison H, Crozier DN, Bryan AC. Ventilatory muscle endurance training in normal subjects and patients with cystic fibrosis. Am Rev Respir Dis. 1977;116(5):853–60.

    CAS  PubMed  Google Scholar 

  9. Sartori R, Barbi E, Poli F, Ronfani L, Marchetti F, Amaddeo A, et al. Respiratory training with a specific device in cystic fibrosis: a prospective study. J Cyst Fibros. 2008;7:313–9.

    Article  CAS  PubMed  Google Scholar 

  10. Kamin W. Der SpiroTiger Medical-ein neues Traininggerät für die Atemmuskulatur von Mukoviszidose-Patienten. Eur Respir J. 2006;28(Suppl. 50):7169.

    Google Scholar 

Download references

Conflict of interest

The authors have no conflict of interest to declare. The authors have not received any financial support for the preparation of this manuscript. No commercial party having a direct financial interest in the results of the research supporting this article has conferred or will confer a benefit on the authors or on any organization with which the authors are associated.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Simone Gambazza.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gambazza, S., Ceruti, C., Brivio, A. et al. Isocapnic hyperpnea with a portable device in Cystic Fibrosis: an agreement study between two different set-up modalities. J Clin Monit Comput 29, 569–572 (2015). https://doi.org/10.1007/s10877-014-9633-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10877-014-9633-z

Keywords

Navigation