The Nexfin device allows for non-invasive beat-to-beat blood pressure monitoring (BPNXF). Perioperative hypotension and hypertension have been shown to be associated with poor clinical outcomes. The goal of the present study was to assess the ability of this device to decrease the duration of significant intraoperative hypo- or hypertension compared to standard BP monitoring by cuff (BPCUFF). We studied25 patients (ASA I–III) undergoing either abdominal or orthopedic surgery. BPCUFF was monitored every 5 min from the introduction of anesthesia, while BPNXF was monitored continuously on the opposite arm. When systolic BPNXF (SBPNXF) decreased or increased more than 20% relative to baseline SBPNXF, a standard BPCUFF measurement was taken to compare values. In addition, the time interval between the 20% change in SBPNXF and the next scheduled standard SBPCUFF measurement was recorded for each event. The mean length of surgery was 3.0 ± 0.3 h. Patients presented with 11 ± 4 episodes of hypotension and 12 ± 4 episodes of hypertension during the surgery. If BPCUFF had been used, this would have resulted in 21 ± 7 min of hypotension and 20 ± 10 min of hypertension. If hemodynamic changes seen by SBPNXF were appropriately treated, an average of 7 ± 1 min/h of hypotension time, 7 ± 2 min/h of hypertension time and 14 ± 3 min per hour of hypo- or hypertension time may have been identified. The Nexfin BP has the potential to decrease the time of hypotension and hypertension compared to conventional intermittent BPCUFF monitoring. Therefore, this device has the potential to positively impact clinical outcomes.

Keywords

Cuff blood pressure Hypertension Hypotension Introperative hemodynamic management Nexfin 

References

  1. 1.
    Hajjar I, Zhao P, Alsop D, Abduljalil A, Selim M, Novak P, Novak V. Association of blood pressure elevation and nocturnal dipping with brain atrophy, perfusion and functional measures in stroke and nonstroke individuals. Am J Hypertens. 2010;23:17–23.PubMedCrossRefGoogle Scholar
  2. 2.
    Molander L, Gustafson Y, Lovheim H. Low blood pressure is associated with cognitive impairment in very old people. Dement Geriatr Cogn Disord. 2010;29:335–41.PubMedCrossRefGoogle Scholar
  3. 3.
    Yocum GT, Gaudet JG, Teverbaugh LA, Quest DO, McCormick PC, Connolly ES Jr, Heyer EJ. Neurocognitive performance in hypertensive patients after spine surgery. Anesthesiology. 2009;110:254–61.PubMedGoogle Scholar
  4. 4.
    Saager L, Greenwald SD, Kelley SD, Schubert A, Sessler DI. Duration of a “triple low” of blood pressure, BIS, and anesthestic concentration predicts poor outcome. Anesthesiology. 2009;111:A880.Google Scholar
  5. 5.
    Monk TG, Saini V, Weldon BC, Sigl JC. Anesthetic management and one-year mortality after noncardiac surgery. Anesth Analg. 2005;100:4–10.PubMedCrossRefGoogle Scholar
  6. 6.
    Tassoudis V, Vretzakis G, Petsiti A, Stamatiou G, Bouzia K, Melekos M, Tzovaras G. Impact of intraoperative hypotension on hospital stay in major abdominal surgery. J Anesth. 2011;25:492–9.Google Scholar
  7. 7.
    Reich DL, Bennett-Guerrero E, Bodian CA, Hossain S, Winfree W, Krol M. Intraoperative tachycardia and hypertension are independently associated with adverse outcome in noncardiac surgery of long duration. Anesth Analg. 2002;95:273–7.Google Scholar
  8. 8.
    Aronson S, Dyke CM, Levy JH, Cheung AT, Lumb PD, Avery EG, Hu MY, Newman MF. Does perioperative systolic blood pressure variability predict mortality after cardiac surgery? An exploratory analysis of the ECLIPSE trials. Anesth Analg. 2011;113:19–30.Google Scholar
  9. 9.
    Aronson S, Stafford-Smith M, Phillips-Bute B, Shaw A, Gaca J, Newman M. Intraoperative systolic blood pressure variability predicts 30-day mortality in aortocoronary bypass surgery patients. Anesthesiology. 2010;113:305–12.PubMedCrossRefGoogle Scholar
  10. 10.
    American Society of Anesthesiologists. Standards of the American Society of Anesthesiologists: Standards for Basic Anesthetic Monitoring.Google Scholar
  11. 11.
    Imholz BP, Wieling W, van Montfrans GA, Wesseling KH. Fifteen years experience with finger arterial pressure monitoring: assessment of the technology. Cardiovasc Res. 1998;38:605–16.PubMedCrossRefGoogle Scholar
  12. 12.
    Penaz J. Photoelectric measurement of blood pressure, volume, and flow in the finger. In: Albert R, editor. Digest of the 10th international conference of medical and biological engineering. Dresden: International Federation for Medical/Biological Engineering; 1973. p. 104.Google Scholar
  13. 13.
    Wesseling KH. Finger arterial pressure measurement with Finapres. Z Kardiol. 1996;85(3):38–44.PubMedGoogle Scholar
  14. 14.
    Nowak RM, Sen A, Garcia AJ, Wilkie H, Yang JJ, Nowak MR, Moyer ML. Noninvasive continuous or intermittent blood pressure and heart rate patient monitoring in the ED. Am J Emerg Med. 2011;29:782–9.Google Scholar
  15. 15.
    Bogert LW, Wesseling KH, Schraa O, Van Lieshout EJ, de Mol BA, van Goudoever J, Westerhof BE, van Lieshout JJ. Pulse contour cardiac output derived from non-invasive arterial pressure in cardiovascular disease. Anaesthesia. 2010;65:1119–25.Google Scholar
  16. 16.
    Eeftinck Schattenkerk DW, van Lieshout JJ, van den Meiracker AH, Wesseling KR, Blanc S, Wieling W, van Montfrans GA, Settels JJ, Wesseling KH, Westerhof BE. Nexfin noninvasive continuous blood pressure validated against Riva-Rocci/Korotkoff. Am J Hypertens. 2009;22:378–83.Google Scholar
  17. 17.
    Akkermans J, Diepeveen M, Ganzevoort W, van Montfrans GA, Westerhof BE, Wolf H. Continuous non-invasive blood pressure monitoring, a validation study of Nexfin in a pregnant population. Hypertens Pregnancy. 2009;28:230–42.PubMedCrossRefGoogle Scholar
  18. 18.
    Martina JR, Westerhof BE, Van Goudoever J, De Jonge N, Van Lieshout JJ, Lahpor JR, De Mol BA. Noninvasive blood pressure measurement by the Nexfin monitor during reduced arterial pulsatility: a feasibility study. ASAIO J. 2010;56:221–7.PubMedCrossRefGoogle Scholar
  19. 19.
    Hofhuizen CM, Lemson J, Hemelaar AE, Settels JJ, Schraa O, Singh SK, van der Hoeven JG, Scheffer GJ. Continuous non-invasive finger arterial pressure monitoring reflects intra-arterial pressure changes in children undergoing cardiac surgery. Br J Anaesth. 2010;105:493–500.PubMedCrossRefGoogle Scholar
  20. 20.
    Maggi R, Viscardi V, Furukawa T, Brignole M. Non-invasive continuous blood pressure monitoring of tachycardic episodes during interventional electrophysiology. Europace. 2010;12:1616–22.PubMedCrossRefGoogle Scholar
  21. 21.
    Lemson J, Hofhuizen CM, Schraa O, Settels JJ, Scheffer GJ, van der Hoeven JG. The reliability of continuous noninvasive finger blood pressure measurement in critically ill children. Anesth Analg. 2009;108:814–21.PubMedCrossRefGoogle Scholar
  22. 22.
    Sipkens LM, Treskes K, Ariese-Beldman K, Veerman DP, Boer C. Application of Nexfin noninvasive beat-to-beat arterial blood pressure monitoring in autonomic function testing. Blood Press Monit. 2011;16:246–51.Google Scholar
  23. 23.
    Bartels SA, Stok WJ, Bezemer R, Boksem RJ, van Goudoever J, Cherpanath TG, van Lieshout JJ, Westerhof BE, Karemaker JM, Ince C. Noninvasive cardiac output monitoring during exercise testing: Nexfin pulse contour analysis compared to an inert gas rebreathing method and respired gas analysis. J Clin Monit Comput. 2011;25:315–21.Google Scholar
  24. 24.
    Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.PubMedCrossRefGoogle Scholar
  25. 25.
    Critchley LA, Lee A, Ho AM. Review article: a critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111:1180–92.PubMedCrossRefGoogle Scholar
  26. 26.
    Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25:536–46.Google Scholar
  27. 27.
    Varon J, Marik PE. Perioperative hypertension management. Vasc Health Risk Manag. 2008;4:615–27.PubMedGoogle Scholar
  28. 28.
    Aronson S, Varon J. Hemodynamic control and clinical outcomes in the perioperative setting. J Cardiothorac Vasc Anesth. 2011;25:509–25.Google Scholar
  29. 29.
    Gan TJ, Soppitt A, Maroof M, el-Moalem H, Robertson KM, Moretti E, Dwane P, Glass PS. Goal-directed intraoperative fluid administration reduces length of hospital stay after major surgery. Anesthesiology. 2002;97:820–6.PubMedCrossRefGoogle Scholar
  30. 30.
    Cannesson M. Arterial pressure variation and goal-directed fluid therapy. J Cardiothorac Vasc Anesth. 2010;24:487–97.PubMedCrossRefGoogle Scholar
  31. 31.
    Venn R, Steele A, Richardson P, Poloniecki J, Grounds M, Newman P. Randomized controlled trial to investigate influence of the fluid challenge on duration of hospital stay and perioperative morbidity in patients with hip fractures. Br J Anaesth. 2002;88:65–71.PubMedCrossRefGoogle Scholar
  32. 32.
    Hamilton MA, Cecconi M, Rhodes A. A systematic review and meta-analysis on the use of preemptive hemodynamic intervention to improve postoperative outcomes in moderate and high-risk surgical patients. Anesth Analg. 2011;112:1392–402.Google Scholar
  33. 33.
    Brzezinski M, Luisetti T, London MJ. Radial artery cannulation: a comprehensive review of recent anatomic and physiologic investigations. Anesth Analg. 2009;109:1763–81.PubMedCrossRefGoogle Scholar
  34. 34.
    Bogert LW, Harms MP, Pott F, Secher NH, Wesseling KH, van Lieshout JJ. Reconstruction of brachial pressure from finger arterial pressure during orthostasis. J Hypertens. 2004;22:1873–80.PubMedCrossRefGoogle Scholar
  35. 35.
    Stover JF, Stocker R, Lenherr R, Neff TA, Cottini SR, Zoller B, Bechir M. Noninvasive cardiac output and blood pressure monitoring cannot replace an invasive monitoring system in critically ill patients. BMC Anesthesiol. 2009;9:6.PubMedCrossRefGoogle Scholar
  36. 36.
    Luginbuhl M, Bieniok C, Leibundgut D, Wymann R, Gentilini A, Schnider TW. Closed-loop control of mean arterial blood pressure during surgery with alfentanil: clinical evaluation of a novel model-based predictive controller. Anesthesiology. 2006;105:462–70.PubMedCrossRefGoogle Scholar
  37. 37.
    Ngan Kee WD, Tam YH, Khaw KS, Ng FF, Critchley LA, Karmakar MK. Closed-loop feedback computer-controlled infusion of phenylephrine for maintaining blood pressure during spinal anaesthesia for caesarean section: a preliminary descriptive study. Anaesthesia. 2007;62:1251–6.Google Scholar
  38. 38.
    Fields AM, Fields KM, Cannon JW. Closed-loop systems for drug delivery. Curr Opin Anaesthesiol. 2008;21:446–51.PubMedCrossRefGoogle Scholar
  39. 39.
    Bijker JB, van Klei WA, Kappen TH, van Wolfswinkel L, Moons KG, Kalkman CJ. Incidence of intraoperative hypotension as a function of the chosen definition: literature definitions applied to a retrospective cohort using automated data collection. Anesthesiology. 2007;107:213–20.PubMedCrossRefGoogle Scholar
  40. 40.
    Goldberg ME, Larijani GE. Perioperative hypertension. Pharmacotherapy. 1998;18:911–4.PubMedGoogle Scholar
  41. 41.
    Cheung AT. Exploring an optimum intra/postoperative management strategy for acute hypertension in the cardiac surgery patient. J Card Surg. 2006;21(1):S8–14.PubMedCrossRefGoogle Scholar
  42. 42.
    Bos WJ, van den Meiracker AH, Wesseling KH, Schalekamp MA. Effect of regional and systemic changes in vasomotor tone on finger pressure amplification. Hypertension. 1995;26:315–20.PubMedGoogle Scholar
  43. 43.
    Natarajan P, Shennan AH, Penny J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia. Am J Obstet Gynecol. 1999;181:1203–10.PubMedCrossRefGoogle Scholar
  44. 44.
    de Greeff A, Shennan A. Blood pressure measuring devices: ubiquitous, essential but imprecise. Expert Rev Med Devices. 2008;5:573–9.PubMedCrossRefGoogle Scholar
  45. 45.
    Araghi A, Bander JJ, Guzman JA. Arterial blood pressure monitoring in overweight critically ill patients: invasive or noninvasive? Crit Care. 2006;10:R64.PubMedCrossRefGoogle Scholar
  46. 46.
    Bur A, Hirschl MM, Herkner H, Oschatz E, Kofler J, Woisetschlager C, Laggner AN. Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients. Crit Care Med. 2000;28:371–6.PubMedCrossRefGoogle Scholar
  47. 47.
    Park MK, Menard SW, Yuan C. Comparison of auscultatory and oscillometric blood pressures. Arch Pediatr Adolesc Med. 2001;155:50–3.PubMedGoogle Scholar
  48. 48.
    Pomini F, Scavo M, Ferrazzani S, De Carolis S, Caruso A, Mancuso S. There is poor agreement between manual auscultatory and automated oscillometric methods for the measurement of blood pressure in normotensive pregnant women. J Matern Fetal Med. 2001;10:398–403.PubMedCrossRefGoogle Scholar
  49. 49.
    Hasan MA, Thomas TA, Prys-Roberts C. Comparison of automatic oscillometric arterial pressure measurement with conventional auscultatory measurement in the labour ward. Br J Anaesth. 1993;70:141–4.PubMedCrossRefGoogle Scholar

Copyright information

© Springer Science+Business Media, LLC 2012

Authors and Affiliations

  • Guo Chen
    • 1
    • 2
  • Elena Chung
    • 2
  • Lingzhong Meng
    • 2
  • Brenton Alexander
    • 2
  • Trung Vu
    • 2
  • Joseph Rinehart
    • 2
  • Maxime Cannesson
    • 2
  1. 1.Department of AnesthesiologyWest China Hospital, Sichuan UniversityChengduPeople’s Republic of China
  2. 2.Department of Anesthesiology and Perioperative CareUniversity of California, IrvineOrangeUSA

Personalised recommendations