Fiberoptic microindentation technique for early osteoarthritis diagnosis: an in vitro study on human cartilage

  • G. MarchiEmail author
  • P. Foehr
  • S. Consalvo
  • A. Javadzadeh-Kalarhodi
  • J. Lang
  • B. Hartmann
  • P. Alberton
  • A. Aszodi
  • R. Burgkart
  • J. Roths
Article

Abstract

In this study, the capability of a fiber optic microindenter sensor to discriminate between healthy and slightly degenerated human articular cartilage samples is demonstrated. The purely optical indenter sensor is characterized by extremely reduced dimensions (0.125 mm in diameter and 27 mm in length) in comparison to existing indenter probes offering advantages for endoscopic deployment. The indenter sensor is intended to assist the surgeon in the identification of damaged articular cartilage. From each of seven specimens of human tibia plateau three samples showing different Outerbridge grading were extracted. On each sample stress-relaxation measurements were performed with eight indentation steps, each step being 40 μm and the relaxation of the material was observed for 240 s after each step. A viscoelastic model was used to fit the relaxation and to extract the characteristic parameters according to the model. A highly significant difference in stiffness (p value <0.01) was observed between the native (grade 0) and early diseased (grade 1) human cartilage samples demonstrating the potential of the fiber optic indenter for the diagnosis of cartilage breakdown.

Keywords

Fiber Bragg grating Osteoarthritis Indentation Cartilage biomechanics 

Notes

Acknowledgements

The research was funded by the Deutsche Forschungsgemeinschaft (DFG) (grants RO 4145/4-1, CS 409/2-1 and AS 150/10.1). The authors thanks Rolf Kuttler for technical support, as well as Sabrina Wagner, Carmen Marthen and Jutta Tübel of the University Hospital Klinikum rechts der Isar for the Outerbridge grading of the human samples.

Compliance with ethical standards

Conflict of interests

The authors declare that no conflict of interest exists.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

References

  1. R.C. Appleyard, M.V. Swain, et al., The accuracy and reliability of a novel handheld dynamic indentation probe for analysing articular cartilage. Phys. Med. Biol. 46(2), 541 (2001)CrossRefGoogle Scholar
  2. W.C. Bae, M.M. Temple, et al., Indentation testing of human cartilage: Sensitivity to articular surface degeneration. Arthritis & Rheumatism 48(12), 3382–3394 (2003)CrossRefGoogle Scholar
  3. S. Beekmans, D. Iannuzzi, Characterizing tissue stiffness at the tip of a rigid needle using an opto-mechanical force sensor. Biomed. Microdevices 18(1), 15 (2016)CrossRefGoogle Scholar
  4. S. Chae, S.-W. Jung, et al., In vivo biomechanical measurement and haptic simulation of portal placement procedure in shoulder arthroscopic surgery. PLoS One 13(3), e0193736 (2018)CrossRefGoogle Scholar
  5. M. Charlebois, M.D. McKee, et al., Nonlinear tensile properties of bovine articular cartilage and their variation with age and depth. J. Biomech. Eng. 126(2), 129–137 (2004)CrossRefGoogle Scholar
  6. N.P. Cohen, R.J. Foster, et al., Composition and dynamics of articular cartilage: Structure, function, and maintaining healthy state. J. Orthop. Sports Phys. Ther. 28(4), 203–215 (1998)CrossRefGoogle Scholar
  7. C.C. Cychosz, J.N. Tofte, et al., Fundamentals of arthroscopic surgery training program improves knee arthroscopy simulator performance in arthroscopic trainees. Arthroscopy 34(5), 1543–1549 (2018)CrossRefGoogle Scholar
  8. T. Franz, E. Hasler, et al., In situ compressive stiffness, biochemical composition, and structural integrity of articular cartilage of the human knee joint. Osteoarthr. Cartil. 9(6), 582–592 (2001)CrossRefGoogle Scholar
  9. F. Guilak, D.L. Butler, et al., Functional tissue engineering: The role of biomechanics in articular cartilage repair. Clin. Orthop. Relat. Res. 391, S295–S305 (2001)CrossRefGoogle Scholar
  10. P. Hansma, H. Yu, et al., The tissue diagnostic instrument. Rev. Sci. Instrum. 80(5), 054303 (2009)CrossRefGoogle Scholar
  11. Jülich, F. (2009). Charakterisierung von Faser-Bragg-Gitter Sensorelementen. Master, University of Applied SciencesGoogle Scholar
  12. F. Jülich, L. Aulbach, et al., Gauge factors of fibre Bragg grating strain sensors in different types of optical fibres. Meas. Sci. Technol. 24(9), 094007 (2013)CrossRefGoogle Scholar
  13. P. Kiviranta, E. Lammentausta, et al., Indentation diagnostics of cartilage degeneration. Osteoarthr. Cartil. 16(7), 796–804 (2008)CrossRefGoogle Scholar
  14. R. Kleemann, D. Krocker, et al., Altered cartilage mechanics and histology in knee osteoarthritis: Relation to clinical assessment (ICRS grade). Osteoarthr. Cartil. 13(11), 958–963 (2005)CrossRefGoogle Scholar
  15. Litwic, A., M. H. Edwards, et al. (2013). Epidemiology and burden of osteoarthritis. British Medical Bulletin: lds038, 105, 185Google Scholar
  16. T. Lyyra, J. Jurvelin, et al., Indentation instrument for the measurement of cartilage stiffness under arthroscopic control. Med. Eng. Phys. 17(5), 395–399 (1995)CrossRefGoogle Scholar
  17. J. Machado, N. Viriato, et al., Experimental characterization of the mechanical properties of knee articular cartilages in compression: First approach with swine tissues. Rheumatology and Orthopedic Medicine 2(4), 1–4 (2017)Google Scholar
  18. J.T. Mäkelä, M.R. Huttu, et al., Structure–function relationships in osteoarthritic human hip joint articular cartilage. Osteoarthr. Cartil. 20(11), 1268–1277 (2012)CrossRefGoogle Scholar
  19. H.J. Mankin, D. H, L. Lippiello, A. Zarins, Biochemical and metabolic abnormalities in articular cartilagefrom osteoarthritic human hips. II. Correlation of morphology with biochemical and metabolic data. J. Bone Joint Surg 53, 523–537 (1971)CrossRefGoogle Scholar
  20. G. Marchi, V. Baier, et al., Microindentation sensor system based on an optical Fiber Bragg grating for the mechanical characterization of articular cartilage by stress-relaxation. Sensors Actuators B Chem. 252, 440–449 (2017)CrossRefGoogle Scholar
  21. G. Marchi, O. Canti, et al., Cartilage microindentation using cylindrical and spherical optical fiber indenters with integrated Bragg gratings as force sensors. SPIE photonics west. Optical Elastography and Tissue Biomechanics V 104960Z, 10496 (2018)Google Scholar
  22. M. Niederauer, A novel instrument for quantitatively measuring the stiffness of articular cartilage. Trans. Orthop. Res. Soc. 23, 905 (1998)Google Scholar
  23. Othonos, A. and K. Kalli (1999). Fiber Bragg Gratings: Fundamentals and Applications in Telecommunications and Sensing. Artech HouseGoogle Scholar
  24. R. Outerbridge, The etiology of chondromalacia patellae. The Journal of bone and joint surgery. British volume 43(4), 752–757 (1961)CrossRefGoogle Scholar
  25. K. Pritzker, S. Gay, et al., Osteoarthritis cartilage histopathology: Grading and staging. Osteoarthr. Cartil. 14(1), 13–29 (2006)CrossRefGoogle Scholar
  26. G. Smith, G. Knutsen, et al., A clinical review of cartilage repair techniques. Bone & Joint Journal 87(4), 445–449 (2005)Google Scholar
  27. G. Spahn, H.M. Klinger, et al., How valid is the arthroscopic diagnosis of cartilage lesions? Results of an opinion survey among highly experienced arthroscopic surgeons. Arch. Orthop. Trauma Surg. 129(8), 1117–1121 (2009)CrossRefGoogle Scholar
  28. M. Stolz, R. Gottardi, et al., Early detection of aging cartilage and osteoarthritis in mice and patient samples using atomic force microscopy. Nat. Nanotechnol. 4(3), 186 (2009)CrossRefGoogle Scholar
  29. A. Turkiewicz, I.F. Petersson, et al., Current and future impact of osteoarthritis on health care: A population-based study with projections to year 2032. Osteoarthr. Cartil. 22(11), 1826–1832 (2014)CrossRefGoogle Scholar
  30. R. J. Williams, L. Peterson, et al. (2007). Cartilage Repair Strategies. SpringerGoogle Scholar

Copyright information

© Springer Science+Business Media, LLC, part of Springer Nature 2019

Authors and Affiliations

  1. 1.Photonics Laboratory, Department of Applied Sciences and MechatronicsMunich University of Applied SciencesMunichGermany
  2. 2.Biomechanics Laboratory, Department of Orthopaedics and Sports Orthopaedics, Klinikum rechts der IsarTechnical University of MunichMunichGermany
  3. 3.Clinic and Policlinic for Orthopaedics and Sports Orthopaedics, Klinikum rechts der IsarTechnical University of MunichMunichGermany
  4. 4.Nanoanalitics and Biophysics Laboratory, Department of Applied Sciences and MechatronicsMunich University of Applied SciencesMunichGermany
  5. 5.Center for Applied Tissue Engineering and Regenerative Medicine-CANTERMunich University of Applied SciencesMunichGermany
  6. 6.Laboratory of Experimental Surgery and Regenerative Medicine, Clinic for General, Trauma and Reconstructive SurgeryLudwig Maximilian UniversityMunichGermany

Personalised recommendations