Advertisement

Microfluidic device with a push–pull sequential solution-exchange function for affinity sensing

  • Shishir Kanti Pramanik
  • Hiroaki SuzukiEmail author
Research Paper
  • 31 Downloads

Abstract

A simple valveless microfluidic device consisting of a reaction chamber and eight flow channels was fabricated for the detection of proteins. The efficient exchange of solutions in the reaction chamber was achieved by sequentially introducing solutions into the reaction chamber and removing them from the reaction chamber through the same flow channels. The influence of the hydrophobicity of the reaction chamber and the viscosity of a blocking solution on solution exchange was investigated. The analytical performance of the device was examined by detecting human interleukin 2 (IL-2) using a sandwich fluorescence immunoassay, with the amount of immuno-complex formed in the reaction chamber monitored by fluorescence microscopy. The approach facilitated the rapid detection of IL-2, in which the total assay time was less than 30 min. A clear dependence of fluorescence intensity on IL-2 concentration was observed over the 125 pg/mL–2.0 ng/mL range, with a detection limit of 105 pg/mL. To promote automation, a simple mechanism that exchanges solutions by the push–pull movement of a PDMS diaphragm was also demonstrated.

Keywords

Valveless microfluidics Solution exchange Immunoassay Human Interleukin-2 

Notes

Acknowledgements

We would like to thank Professor Masatoshi Yokokawa and Dr. Gokul Chandra Biswas of University of Tsukuba for valuable discussions.

Compliance with ethical standards

Conflict of interest

There are no conflicts to declare.

Supplementary material

10404_2019_2188_MOESM1_ESM.docx (3 mb)
Supplementary material 1 (DOCX 3083 KB)

Supplementary material 2 (WMV 688 KB)

Supplementary material 3 (WMV 1354 KB)

References

  1. Acres RG, Ellis AV, Alvino J, Lenahan CE, Khodakov DA, Metha GF, Andersson GG (2012) Molecular structure of 3-aminopropyltriethoxysilane layers formed on silanol-terminated silicon surfaces. J Phys Chem C 116:6289–6297CrossRefGoogle Scholar
  2. Bange A, Halsall HB, Heineman WR (2005) Microfluidic immunosensor systems. Biosens Bioelectron 20:2488–2503CrossRefGoogle Scholar
  3. Barbosa AI, Reis NM (2017) A critical insight into the development pipeline of microfluidic immunoassay devices for the sensitive quantitation of protein biomarkers at the point of care. Analyst 142:858–882CrossRefGoogle Scholar
  4. Benard WL, Kahn H, Heuer AH, Huff MA (1998) Thin-film shape-memory alloy actuated micropumps. J Microelectromech Syst 7:245–251CrossRefGoogle Scholar
  5. Berktas M, Guducuoglu H, Bozkurt H, Onbasi K, Kurtoglu M, Andic S (2004) Change in serum concentrations of interleukin-2 and interferon-γ during treatment of tuberculosis. J Int Med Res 32:324–330CrossRefGoogle Scholar
  6. Bratten CDT, Cobbold PH, Cooper JM (1997) Micromachining sensors for electrochemical measurement in subnanoliter volumes. Anal Chem 69:253–258CrossRefGoogle Scholar
  7. Findlay JWA et al (2000) Validation of immunoassays for bioanalysis: a pharmaceutical industry perspective. J Pharm Biomed Anal 21:1249–1273CrossRefGoogle Scholar
  8. Giri B, Pandey B, Neupane B, Ligler FS (2016) Signal amplification strategies for microfluidic immunoassays. Trac Trends Anal Chem 79:326–334CrossRefGoogle Scholar
  9. Gubala V, Harris LF, Ricco AJ, Tan MX, Williams DE (2012) Point of care diagnostics: status and future. Anal Chem 84:487–515CrossRefGoogle Scholar
  10. Han SW, Koh W-G (2016) Hydrogel-framed nanofiber matrix integrated with a microfluidic device for fluorescence detection of matrix metalloproteinases-9. Anal Chem 88:6247–6253CrossRefGoogle Scholar
  11. Hua B et al (2014) An improved surface passivation method for single-molecule studies. Nat Methods 11:1233–1236CrossRefGoogle Scholar
  12. Huckle D (2008) Point-of-care diagnostics: an advancing sector with nontechnical issues. Expert Rev Mol Diagn 8:679–688CrossRefGoogle Scholar
  13. Iverson BD, Garimella SV (2008) Recent advances in microscale pumping technologies: a review and evaluation. Microfluid Nanofluidics 5:145–174CrossRefGoogle Scholar
  14. Juncker D et al (2002) Autonomous microfluidic capillary system. Anal Chem 74:6139–6144CrossRefGoogle Scholar
  15. Jung W, Han J, Choi J-W, Ahn CH (2015) Point-of-care testing (POCT) diagnostic systems using microfluidic lab-on-a-chip technologies. Microelectron Eng 132:46–57CrossRefGoogle Scholar
  16. Lai S, Wang S, Luo J, Lee LJ, Yang S-T, Madou MJ (2004) Design of a compact disk-like microfluidic platform for enzyme-linked immunosorbent assay. Anal Chem 76:1832–1837CrossRefGoogle Scholar
  17. Laxminarayan R et al (2006) Advancement of global health: key messages from the Disease Control Priorities Project. Lancet 367:1193–1208CrossRefGoogle Scholar
  18. Linder V, Sia SK, Whitesides GM (2005) Reagent-loaded cartridges for valveless and automated fluid delivery in microfluidic devices. Anal Chem 77:64–71CrossRefGoogle Scholar
  19. Liu Y, Wang H, Huang J, Yang J, Liu B, Yang P (2009) Microchip-based ELISA strategy for the detection of low-level disease biomarker in serum. Anal Chim Acta 650:77–82CrossRefGoogle Scholar
  20. Luppa PB, Müller C, Schlichtiger A, Schlebusch H (2011) Point-of-care testing (POCT): current techniques and future perspectives. Trends Anal Chem 30:887–898CrossRefGoogle Scholar
  21. Lynn NS, Henry CS, Dandy DS (2009) Evaporation from microreservoirs. Lab Chip 9:1780–1788.  https://doi.org/10.1039/b900556k CrossRefGoogle Scholar
  22. Moerman R, Knoll J, Apetrei C, van den Doel LR, van Dedem GWK (2005) Quantitative analysis in nanoliter wells by prefilling of wells using electrospray deposition followed by sample introduction with a coverslip method. Anal Chem 77:225–231CrossRefGoogle Scholar
  23. Nashida N, Satoh W, Fukuda J, Suzuki H (2007) Electrochemical immunoassay on a microfluidic device with sequential injection and flushing functions. Biosens Bioelectron 22:3167–3173.  https://doi.org/10.1016/j.bios.2007.02.010 CrossRefGoogle Scholar
  24. Ng AHC, Uddayasankar U, Wheeler AR (2010) Immunoassays in microfluidic systems. Anal Bioanal Chem 397:991–1007CrossRefGoogle Scholar
  25. Novo P, Volpetti F, Chu V, Conde JP (2013) Control of sequential fluid delivery in a fully autonomous capillary microfluidic device. Lab Chip 13:641–645CrossRefGoogle Scholar
  26. Obata H, Kuji T, Kojima K, Sassa F, Yokokawa M, Takekosh K, Suzuki H (2016) Electrochemical bubble-based bidirectional microfluidic transport. ACS Sens 1:190–196.  https://doi.org/10.1021/acssensors.5b00059 CrossRefGoogle Scholar
  27. Pai NP, Vadnais C, Denkinger C, Engel N, Pai M (2012) Point-of-care testing for infectious diseases: diversity, complexity, and barriers in low- and middle-income countries. PLoS Med 9:e1001306CrossRefGoogle Scholar
  28. Richter A, Klatt S, Paschew G, Klenke C (2009) Micropumps operated by swelling and shrinking of temperature-sensitive hydrogels. Lab Chip 9:613–618CrossRefGoogle Scholar
  29. Sassa F, Al-Zain Y, Ginoza T, Miyazaki S, Suzuki H (2012) Miniaturized shape memory alloy pumps for stepping microfluidic transport. Sens Actuators B 165:157–163CrossRefGoogle Scholar
  30. Shimizu Y, Takashima A, Satoh W, Sassa F, Fukuda J, Suzuki H (2009) Biochip with integrated pumps for plug-based sequential exchange of solutions. Sens Actuators B 140:649–655CrossRefGoogle Scholar
  31. Suzuki H, Yoneyama R (2003) Integrated microfluidic system with electrochemically actuated on-chip pumps and valves. Sens Actuators B 96:38–45.  https://doi.org/10.1016/S0925-4005(03)00482-9 CrossRefGoogle Scholar
  32. Suzuki H, Tokuda T, Kobayashi K (2002) A disposable “intelligent mosquito” with a reversible sampling mechanism using the volume-phase transition of a gel. Sens Actuators B 83:53–59CrossRefGoogle Scholar
  33. Usuba R et al (2016) Photonic Lab-on-a-Chip for rapid cytokine detection. ACS Sens 1:979–986CrossRefGoogle Scholar
  34. Vandenberg ET, Bertilsson L, Liedberg B, Uvdal K, Erlandsson R, Elwing H, Lundström I (1991) Structure of 3-aminopropyl triethoxy silane on silicon oxide. J Colloid Interface Sci 147:103–118CrossRefGoogle Scholar
  35. Vashist SK, Luppa PB, Yeo LY, Ozcan A, Luong JHT (2015) Emerging technologies for next-generation point-of-care testing. Trends Biotechnol 33:692–705CrossRefGoogle Scholar
  36. Yu L, Li CM, Zhou Q, Luong JHT (2007) Poly(vinyl alcohol) functionalized poly(dimethylsiloxane) solid surface for immunoassay. Bioconjugate Chem 18:281–284CrossRefGoogle Scholar
  37. Yu ZTF et al (2015) Rapid, automated, parallel quantitative immunoassays using highly integrated microfluidics and AlphaLISA. Sci Rep 5:11339CrossRefGoogle Scholar
  38. Zimmermann M, Delamarche E, Wolf M, Hunziker P (2005) Modeling and optimization of high-sensitivity low-volume microfluidic-based surface immunoassays. Biomed Microdevices 7:99–110CrossRefGoogle Scholar

Copyright information

© Springer-Verlag GmbH Germany, part of Springer Nature 2019

Authors and Affiliations

  1. 1.Graduate School of Pure and Applied SciencesUniversity of TsukubaTsukubaJapan

Personalised recommendations