  • Aaron R. PurchaseEmail author
  • Tadeusz Pałasz
  • Hongwei Sun
  • Jonathan C. Sharp
  • Boguslaw Tomanek
Research Article
  • 170 Downloads

Abstract

Objective

A radiofrequency (RF) power amplifier is an essential component of any magnetic resonance imaging (MRI) system. Unfortunately, no commercial amplifier exists to fulfill the needs of the transmit array spatial encoding (TRASE) MRI technique, requiring high duty cycle, high RF output power and independently controlled multi-channel capability. Thus, an RF amplifier for TRASE MRI is needed.

Materials and methods

A dual-channel RF power amplifier dedicated for TRASE at 0.22 T (9.27 MHz) was designed and constructed using commercially available components. The amplifier was tested on the bench and used a 0.22 T MRI system with a twisted solenoid and saddle RF coil combination capable of a single-axis TRASE.

Results

The amplifier is capable of sequential, dual-channel operation up to 50% duty cycle, 1 kW peak output and highly stable 100 μs RF pulse trains. High spatial resolution one-dimensional TRASE was obtained with the power amplifier to demonstrate its capability.

Conclusion

The constructed amplifier is the first prototype that meets the requirements of TRASE rectifying limitations of duty cycle and timing presented by commercial RF amplifiers. The amplifier makes possible future high resolution in vivo TRASE MRI.

Keywords

MRI TRASE RF RF power amplifiers 

Notes

Acknowledgements

An abstract for this work was accepted for poster presentation at the International Society for Magnetic Resonance in Medicine (ISMRM) 27th Annual Meeting and Exhibition in Montreal, Canada (2019). We thank Curtis Osinchuk and Lance Spiridon, Cross Cancer Institute, Edmonton, Alberta, Canada for constructing enclosure for the TRASE amplifier and Vyacheslav Volotovskyy and the Technology Management Group (TMG), Cross Cancer Institute, Edmonton, Alberta, Canada for technical support.

Author contributions

Design: ARP, TP, JCS and BT. Construction and bench performance testing: ARP, TP and BT. MRI testing and TRASE acquisition: ARP, HS and JCS. Manuscript drafting and revision: ARP, TP, JCS and BT. Critical revisions: ARP, JCS and BT.

Funding

This study was funded by the Natural Sciences and Engineering Research Council of Canada (N-SERC), Grant/Award number: RGPIN-03992.

Compliance with ethical standards

Conflict of interest

Aaron R. Purchase declares that he has no conflict of interest. Tadeusz Pałasz declares that he has no conflict of interest. Hongwei Sun declares that he has no conflict of interest. Jonathan C. Sharp declares that he has no conflict of interest. Boguslaw Tomanek declares that he has no conflict of interest.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

References

  1. 1.
    Brown RW, Cheng YN, Haacke EM, Thompson MR,  Venkatesan R (2014) Multi-dimensional fourier imaging and slice selection, Chap 10. In: Magnetic resonance imaging: physical principles and sequence design, 2nd edn. Wiley Blackwell, Hoboken, pp 166–174Google Scholar
  2. 2.
    Sharp JC, King SB, Deng Q, Volotovskyy V, Tomanek B (2013) High-resolution MRI encoding using radiofrequency phase gradients. NMR Biomed 26:1602–1607CrossRefGoogle Scholar
  3. 3.
    Deng Q, King SB, Volotovskyy V, Tomanek B, Sharp JC (2013) B1 transmit phase gradient coil for single-axis TRASE RF encoding. Magn Reson Imaging 31:891–899CrossRefGoogle Scholar
  4. 4.
    Sharp JC, King SB (2010) MRI using radiofrequency magnetic field phase gradients. Magn Reson Med 63:151–161PubMedGoogle Scholar
  5. 5.
    Sun H, Yong S, Sharp JC (2018) The twisted solenoid RF phase gradient transmit coil for TRASE imaging. J Magn Reson 299:135–150.  https://doi.org/10.1016/j.jmr.2018.12.015 CrossRefPubMedGoogle Scholar
  6. 6.
    Stockmann JP, Cooley CZ, Guerin B, Rosen MS, Wald LL (2016) Transmit array spatial encoding (TRASE) using broadband WURST pulses for RF spatial encoding in inhomogenous B0 fields. J Magn Reson 268:36–48.  https://doi.org/10.1016/j.jmr.2016.04.005 CrossRefPubMedPubMedCentralGoogle Scholar
  7. 7.
    Myer DP (2007) Radiofrequency power amplifiers for NMR and MRI. eMagRes. Wiley, New York.  https://doi.org/10.1002/9780470034590.emrstm1138 CrossRefGoogle Scholar
  8. 8.
    Ivanov EN, Pogromsky AY, Van Den Brink JS, Rooda JE (2010) Optimization of duty cycles for MRI scanners. Concepts Magn Reson Part B (Magn Reson Eng) 37B(3):180–192CrossRefGoogle Scholar
  9. 9.
    Goldman M, Johannesson H, Axelsson O, Karlsson M (2006) Design and implementation of 13C hyper polarization from para-hydrogen, for new MRI contrast agents. C.R. Chimie 9:357–363CrossRefGoogle Scholar
  10. 10.
    Holland GN, Heysmond E (1979) A solid state high-power RF amplifier for pulsed NMR. J Phys E Sci Instrum 12:480CrossRefGoogle Scholar
  11. 11.
    Horowitz P, Hill W (2015) Foundations, Chap 1. In: The art of electronics, 3rd edn. Cambridge University Press, New York, pp 26–27Google Scholar
  12. 12.
    ZHL-1-2W+ Coaxial amplifier. 50 Ω High Power 2W 5 to 500 MHz, Mini-Circuits. https://ww2.minicircuits.com/pdfs/ZHL-1-2W.pdf. Accessed 1 Jan 2018
  13. 13.
    Klitzing J (2016) A 1 KW SSPA for 1.8–54 MHz. Amateur ratio station W6PQL. https://www.w6pql.com/. Accessed 1 Jan 2018
  14. 14.
    Ampleon (2015) BLF188XR; BLF188XRS. Power LDMOS transistor product data sheet, Rev. 6–1 September. https://www.ampleon.com/products/broadcast/0-500-mhz-rf-power-transistors/BLF188XR.html. Accessed 1 Jan 2018
  15. 15.
    Southam M (2013) TDK-Lambda Americas Inc., User Manual for GenesysTM 750W/1500W Programable DC Power Supplies. https://www.us.tdk-lambda.com/hp/pdfs/Product_manuals/83507013.pdf. Accessed 1 Jan 2018
  16. 16.
    Sharkov SB, Krasnov AA, Polikhov SA (2016) 100 kW very compact pulsed solid-state RF amplifier: development and tests. In: Proceedings of NAPA2016, Chicago, IL, USA. ISBN: 978-3-95450-180-9Google Scholar
  17. 17.
    Sharp JC, Yin D, Bernhardt RH, Deng Q, Procca AE, Tyson RL, Lo K, Tomanek B (2009) The integration of real and virtual magnetic resonance imaging experiments in a single instrument. Am Inst Phys Rev Sci Instrum 80:093709CrossRefGoogle Scholar
  18. 18.
    Walker J (2012) Amplifier measurements, Chap 12. In: Handbook of RF and microwave power amplifiers, 3rd edn. Cambridge University Press, New York, pp 636. ISBN: 978-0-521-76010-2 (sec. 12.7)Google Scholar
  19. 19.
    Bowick C, Byler J, Ajluni C (2008) RF circuit design. Elsevier Inc., Amsterdam, pp 174–176Google Scholar
  20. 20.
    Medical electrical equipment (2005) IEC 60601-1. International Standard, 3rd edn, 2005–12. Reference number: IEC 60601-1:2005(E)Google Scholar
  21. 21.
    Sarty G, Kontulainen S, Baxter-Jones A, Pierson R, Turek K, Obenaus A, Tomanek B, Sharp J, Scott A, Piche L (2012) Compact MRI for astronaut physiological research and medical diagnosis. In: Com Dev, AIAA SPACE conference & exposition, Pasadena, CaliforniaGoogle Scholar
  22. 22.
    Sarty GE, Obenaus A (2012) Magnetic resonance imaging of astronauts on the international space station and into the solar system. Can Aeronaut Space J 58:60–68CrossRefGoogle Scholar
  23. 23.
    Sarty GE, Vidarsson L (2018) Natural slice encoded magnetic resonance imaging. Magn Reson Imaging 46:47–55CrossRefGoogle Scholar
  24. 24.
    ANALOGIC: The World Resource for Health and Security Technology (2002) AN8110 RF Amplifier Installation Manual. Analogic Corporation Medical Imaging Division, Insert no. 16-00070 REV. 02., Binder no. 28-002011, Publication no. 10-62836-01Google Scholar
  25. 25.
    TOMCO Technologies (2002) 500 W Pulsed RF Amplifier Model BT00 BETA Operation Manual. Tomco Electronics Pty Ltd, Document No. T001591 Issue A, June 25Google Scholar
  26. 26.
    Communication Power Corporation (CPC) (2012) The world's leading supplier of broadband power amplifiers. Model 0.2T500M. https://www.cpcamps.com/sheets/0.2T500M.pdf. Accessed 1 Jan 2018
  27. 27.
    Harrington M, Heist W, Loriatti C, White J (N/A) Product Specification Document XX-S35-64. 18 kW, 1.5T Linear, Solid State RF Pulse. https://www.mksinst.com/f/63-87-mhz-mri-rf-amplifier. Accessed 1 Jan 2018
  28. 28.
    Tomco Technologies (2019) RF Amplifier Data Sheet BT01000-Delta. http://www.tomcorf.com/assets/standard_pulsed/DS006689B.pdf. Accessed 17 Jun 2019
  29. 29.
    Ultra Electronics Herley (N/A) RF Power Amplifier System Model 3900-1. 3061 Lancaster, PA, USA. https://www.ultraherley.com/uploads/herley/datasheets/lancaster/Medical/3900-1.pdf. Accessed 1 Jan 2018
  30. 30.
    ANALOGIC: The World Resource for Health and Security Technology (2016) 6 kW RF Power Amplifier for MRI Scanner Applications, Analogic Medical Imaging Group., ANALOGIC Corporation, Bulletin No. 16-00115 REV 1 3/04, Printed in USAGoogle Scholar
  31. 31.
    Qian D, El-Sharkawy AM, Bottomley PA, Edelstein WA (2013) An RF dosimeter for independent SAR measurement in MRI scanners. Med Phys 40(12):122303CrossRefGoogle Scholar
  32. 32.
    Zhen J, Dykstra R, Eccles C, Gouws G, Obruchkov S (2017) A compact Class D RF power amplifier for mobile nuclear magnetic resonance systems. Rev Sci Instrum 88:074704.  https://doi.org/10.1063/1.4994734 CrossRefPubMedGoogle Scholar
  33. 33.
    Zhen J, Gouws G, Dykstra R, Eccles C (2012) Class D RF amplifier for 20 MHz NMR MOUSE sensor. In: 2012 sixth international conference on sensing technology (ICST), Kolkata, pp 13–16. https://doi.org/10.1109/ICSensT.2012.6461657
  34. 34.
    Pozar DM (2012) Noise and nonlinear distortion, Chap 10. In: Microwave engineering, 2nd edn. Wiley, Hoboken, p 512 (sec. 10.3) Google Scholar

Copyright information

© European Society for Magnetic Resonance in Medicine and Biology (ESMRMB) 2019

Authors and Affiliations

  1. 1.Department of OncologyUniversity of AlbertaEdmontonCanada
  2. 2.Division of Medical PhysicsCross Cancer InstituteEdmontonCanada
  3. 3.Marian Smoluchowski Institute of Physics, Jagiellonian UniversityKrakówPoland
  4. 4.Institute of Nuclear PhysicsPolish Academy of SciencesKrakówPoland

Personalised recommendations