Improved signal-to-noise ratio in EPI sequences with highly asymmetric spin echo and highly asymmetric STEAM preparations (HASE-EPI and HASTEAM-EPI)

  • Manoj ShresthaEmail author
  • Ulrike Nöth
  • Ralf Deichmann
Research Article
  • 36 Downloads

Abstract

Objective

Alternative versions of echo-planar imaging (EPI) sequences with standard spin echo (SE) or stimulated echo acquisition mode (STEAM) preparations are proposed, allowing for improved signal-to-noise ratio (SNR), especially in high-resolution imaging. The suitability of the proposed sequences, dubbed “highly asymmetric SE-EPI” (HASE-EPI) and “highly asymmetric STEAM-EPI” (HASTEAM-EPI), is tested in vivo, for anatomical imaging with \(T_{2}\) weighting and diffusion-weighted imaging (DWI).

Materials and methods

In HASE-EPI and HASTEAM-EPI, echo formation occurs prior to the EPI readout, rather than at k-space centre as in standard SE-EPI and STEAM-EPI. This allows for a constant preparation period, independent of the spatial resolution. The proposed sequences are compared to their standard counterparts, via simulations and experimentally via in vivo anatomical imaging and DWI.

Results

HASE-EPI and HASTEAM-EPI yield SNR improvements for large matrix sizes and fully sampled EPI readout. Simulations and in vivo results show a signal gain in HASE-EPI versus SE-EPI or HASTEAM-EPI versus STEAM-EPI for white and gray matter, particularly for higher spatial resolution with full readout in anatomical imaging and DWI. However, simulations also show that in the case of very long EPI-readout trains, HASE-EPI and HASTEAM-EPI are more prone to pixel broadening due to relaxation effects.

Discussion

In contrast to commonly used SE-EPI or STEAM-EPI, the proposed sequences facilitate the acquisition of anatomical imaging and DWI data with improved SNR, especially for full EPI readouts. However, the applicability of HASE-EPI and HASTEAM-EPI should be carefully assessed: while signal gains are less pronounced when using parallel imaging and/or reduced matrix sizes, there may be pixel broadening for very long EPI-readout trains.

Keywords

Asymmetric echo Short echo time Signal-to-noise ratio Asymmetric imaging Diffusion-weighted imaging 

Notes

Authors’ contribution

Protocol/project development: MS, RD. Data collection and management: MS, UN. Data analysis: MS.

Compliance with ethical standards

Conflict of interest

The authors declare they have no conflict of interest.

Ethical approval

All procedures performed in the vivo study were approved by the local ethics committee of the University Hospital.

Informed consent

Written informed consent was obtained from all healthy participants before scanning.

References

  1. 1.
    Jeong EK, Kim SE, Guo J, Kholmovski EG, Parker DL (2005) High-resolution DTI with 2D interleaved multislice reduced FOV single-shot diffusion-weighted EPI (2D ss-rFOV-DWEPI). Magn Reson Med 54(6):1575–1579CrossRefGoogle Scholar
  2. 2.
    Robson MD, Gatehouse PD, Bydder M, Bydder GM (2003) Magnetic resonance: an introduction to ultrashort TE (UTE) imaging. J Comput Asssist Tomogr 27(6):825–846CrossRefGoogle Scholar
  3. 3.
    Shrestha M, Mildner T, Schlumm T, Robertson SH, Möller H (2016) Three-dimensional echo-planar cine imaging of cerebral blood supply using arterial spin labeling. Magn Reson Mater Phy 29(6):799–810CrossRefGoogle Scholar
  4. 4.
    Stejskal EO, Tanner JE (1965) Spin diffusion measurements: spin echoes in the presence of a time-dependent field gradient. J Chem Phys 42(1):288–292CrossRefGoogle Scholar
  5. 5.
    Tanner JE (1970) Use of the stimulated echo in NMR diffusion studies. J Chem Phys 52(5):2523–2526CrossRefGoogle Scholar
  6. 6.
    Reese TG, Heid O, Weisskoff RM, Wedeen VJ (2003) Reduction of eddy-current-induced distortion in diffusion MRI using a twice-refocused spin echo. Magn Reson Med 49(1):177–182CrossRefGoogle Scholar
  7. 7.
    Steidle G, Schick F (2006) Echoplanar diffusion tensor imaging of the lower leg musculature using eddy current nulled stimulated echo preparation. Magn Reson Med 55(3):541–548CrossRefGoogle Scholar
  8. 8.
    Nöth U, Shrestha M, Schüre JR, Deichmann R (2017) Quantitative in vivo T2 mapping using fast spin echo techniques—a linear correction procedure. NeuroImage 157:476–485CrossRefGoogle Scholar
  9. 9.
    Wansapura JP et al (1999) NMR relaxation times in the human brain at 3.0 tesla. J Magn Reson Imaging 9(4):531–538CrossRefGoogle Scholar
  10. 10.
    Preibisch C, Deichmann R (2009) Influence of RF spoiling on the stability and accuracy of T1 mapping based on spoiled FLASH with varying flip angles. Magn Reson Med 61(1):125–135CrossRefGoogle Scholar
  11. 11.
    Jenkinson M, Beckmann CF, Behrens TEJ, Woolrich MW, Smith SM (2012) FSL. NeuroImage 62:782–790CrossRefGoogle Scholar
  12. 12.
    Smith SM, Jenkinson M, Woolrich MW, Beckmann CF, Behrens TE, Johansen-Berg H, Bannister PR, De Luca M, Drobnjak I, Flitney DE, Niazy RK, Saunders J, Vickers J, Zhang Y, De Stefano N, Brady JM, Matthews PM (2004) Advances in functional and structural MR image analysis and implementation as FSL. NeuroImage 23:S208–S219CrossRefGoogle Scholar
  13. 13.
    Andersson JLR, Skare S, Ashburner J (2003) How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging. NeuroImage 20:870–888CrossRefGoogle Scholar
  14. 14.
    Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, Kiefer B, Haase A (2002) Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 47(6):1202–1210CrossRefGoogle Scholar
  15. 15.
    Nöth U, Laufs H, Stoermer R, Deichmann R (2012) Simultaneous electroencephalography-functional MRI at 3 T: an analysis of safety risks imposed by performing anatomical reference scans with the EEG equipment in place. J Magn Reson Imag 35(3):561–571CrossRefGoogle Scholar
  16. 16.
    Shrestha M, Hok P, Nöth U, Lienerth B, Deichmann R (2018) Optimization of diffusion-weighted single-refocused spin-echo EPI by reducing eddy-current artifacts and shortening the echo time. Magn Reson Mater Phy 31(5):585–597CrossRefGoogle Scholar
  17. 17.
    Huber L, Guidi M, Goense J, Mildner T, Trampel R, Schulz J, Eichner C, Turner R, Möller HE (2015) The magnitude point spread function is an inadequate measure of T2*-blurring in EPI. In: Proceedings of the 23rd annual meeting of ISMRMGoogle Scholar
  18. 18.
    Renvall V, Witzel T, Wald LL, Polimeni JR (2016) Automatic cortical surface reconstruction of high-resolution T1 echo planar imaging data. NeuroImage 134:338–354CrossRefGoogle Scholar
  19. 19.
    Van der Zwaag W, Buur PF, Fracasso A, van Doesum T, Uludağ K, Versluis MJ, Marques JP (2018) Distortion-matched T1 maps and unbiased T1-weighted images as anatomical reference for high-resolution fMRI. NeuroImage 176:41–55CrossRefGoogle Scholar
  20. 20.
    Hetzer S, Mildner T, Möller HE (2011) A Modified EPI sequence for high-resolution imaging at ultra-short echo time. Magn Reson Med 65(1):165–175CrossRefGoogle Scholar
  21. 21.
    Huber L, Ivanov D, Krieger SN, Streicher MN, Mildner T, Poser BA, Möller HE, Turner R (2014) Slab-selective, BOLD-corrected VASO at 7 Tesla provides measures of cerebral blood volume reactivity with high signal-to-noise ratio. Magn Reson Med 72(1):137–148CrossRefGoogle Scholar

Copyright information

© European Society for Magnetic Resonance in Medicine and Biology (ESMRMB) 2019

Authors and Affiliations

  1. 1.Brain Imaging Center (BIC)Goethe University FrankfurtFrankfurt am MainGermany

Personalised recommendations