Advertisement

Design of internal MRI coils using ultimate intrinsic SNR

  • Yiğitcan Eryaman
  • Yusuf Öner
  • Ergin Atalar
Research Article

Abstract

Object

Internal MRI coils have important applications in diagnostic and interventional studies. Since they can be placed very close to the region of interest in the body, they are favored over external coils in applications where high-resolution images are required. In this paper it is demonstrated that ultimate intrinsic SNR (UISNR) and the optimum coil sensitivity solutions can be used to make new coil designs with higher intrinsic SNR.

Materials and methods

In this study, UISNR, which is the maximum attainable value of the intrinsic SNR, is used as a measure of performance and as a design criterion. As an example, a novel endorectal MRI coil is designed. The design is tested with phantom and patient studies.

Results

An endorectal coil is built to demonstrate the effectiveness of the design strategy. ISNR of the endorectal coil approximates the UISNR to 72%.

Conclusion

An internal coil design method that takes advantage of the UISNR and optimum coil sensitivity calculations was presented. This method can also be used to design better internal MRI coils for different applications.

Keywords

Internal MRI coils Intrinsic SNR Ultimate intrinsic SNR 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Hurst GC, Hua J, Duerk JL, Cohen AM (1992) Intravascular (catheter) NMR receiver probe Preliminary design analysis and application to canine iliofemoral imaging. Magn Reson Med 24(2): 343–357PubMedCrossRefGoogle Scholar
  2. 2.
    Ogan O, Atalar E (1997) Intravascular magnetic resonance imaging using a loopless catheter antenna. Magn Reson Med 37(1): 112–118CrossRefGoogle Scholar
  3. 3.
    Shunk KA, Lima JAC, Heldman AW, Atalar E (1999) Transesophageal magnetic resonance imaging. Magn Reson Med 41(4): 722–726PubMedCrossRefGoogle Scholar
  4. 4.
    Quick HH, Serfaty J-M, Pannu HK, Genadry R, Yeung CJ, Atalar E (2001) Endourethral MRI. Magn Reson Med 45(1): 138–146PubMedCrossRefGoogle Scholar
  5. 5.
    Schnall M, Lenkinski R, Pollack H, Imai Y, Kressel H (1989) Prostate: MR imaging with an endorectal surface coil. Radiology 172(2): 570–574PubMedGoogle Scholar
  6. 6.
    Krieger A, Susil RC, Menard C, Coleman JA, Fichtinger G, Atalar E, Whitcomb LL (2005) Design of a novel MRI compatible manipulator for image guided prostate interventions. IEEE Trans BME 52(2): 306–313CrossRefGoogle Scholar
  7. 7.
    Meyer CR, Bland PH, Pipe J (1995) Retrospective correction of intensity inhomogeneities in MRI. IEEE Trans Med Imaging 14(1): 36–41PubMedCrossRefGoogle Scholar
  8. 8.
    Sled JG, Zijdenbos AP, Evans AC (1998) A nonparametric method for automatic correction of intensity nonuniformity in MRI data. IEEE Trans Med Imaging 17(1): 87–97PubMedCrossRefGoogle Scholar
  9. 9.
    Edelstein WA, Glover GH, Hardy CJ, Redington RW (1986) The intrinsic signal-to-noise ratio in NMR imaging. Magn Reson Med 3(4): 604–618PubMedCrossRefGoogle Scholar
  10. 10.
    Ocali O, Atalar E (1998) Ultimate intrinsic signal-to-noise ratio in MRI. Magn Reson Med 39(3): 462–473PubMedCrossRefGoogle Scholar
  11. 11.
    Hoult DI, Richards RE (1976) The signal-to-noise ratio of the nuclear magnetic resonance experiment. J Magn Reson Imaging 24: 71–85Google Scholar
  12. 12.
    Celik H, Eryaman Y, Altintas A, Abdel-Hafez IA, Atalar E (2004) Evaluation of internal MRI coils using ultimate intrinsic SNR. Magn Reson Med 52(3): 640–649PubMedCrossRefGoogle Scholar
  13. 13.
    Yung AC, Öner AY, Serfaty J-M, Feneley M, Yang X, Atalar E (2003) Phased-array MRI of canine prostate using endorectal and endourethral coils. Magn Reson Med 49(4): 710–715PubMedCrossRefGoogle Scholar
  14. 14.
    Pozar DM (1998) Microwave engineering. Wiley, NY, pp 94Google Scholar

Copyright information

© ESMRMB 2009

Authors and Affiliations

  1. 1.Department of Electrical and Electronics EngineeringBilkent UniversityAnkaraTurkey
  2. 2.Department of RadiologyGazi UniversityAnkaraTurkey

Personalised recommendations