Breath analysis as promising indicator of hemodialysis efficiency

  • B. Grabowska-PolanowskaEmail author
  • P. Miarka
  • M. Skowron
  • G. Chmiel
  • A. Pietrzycka
  • I. Śliwka
Original article
  • 132 Downloads

Abstract

Background

The measurement of trimethylamine and isoprene in exhaled breath collected from dialysed patients indicates the changes in concentration of both compounds during dialysis. The aim of the presented study was to confirm diagnostic usefulness of TMA and isoprene detected in breath, as potential biomarkers of hemodialysis efficiency.

Methods

The samples of exhaled breath were collected from 22 dialyzed patients (9 women, 13 men) before and after hemodialysis (HD). All analyses were carried out using a gas chromatograph equipped with a mass spectrometer. Thermal desorption was used as breath sample enrichment method.

Results

Chromatographic analysis of breath samples indicated statistically significant differences in trimethylamine (TMA) and 2-methyl-1,3-butadiene (isoprene) concentrations in patients’ breath collected before and after HD. TMA concentrations measured in breath samples, before dialysis, ranged from 0.024 to 0.461 nmol/L. After dialysis, the values of detected TMA were lower versus output values and ranged from 0.008 to 0.050 nmol/L. Isoprene concentrations before dialysis were present in the range from 0.236 to 9.718 nmol/L, after dialysis in the range from 0.478 to 26.182 nmol/L. Additionally, the dependences of TMA and isoprene concentrations, detected in breath with renal efficiency parameters detected in blood, were studied. The relationships between TMA and urea (r = 0.67; p < 0.00001) and creatinine (r = 0.61; p = 0.00002) were checked. In case of isoprene considerably higher concentrations were observed after dialysis, but no statistically significant correlation of isoprene with blood parameters was noticed.

Conclusion

The observed decrease of TMA concentrations during dialysis could be useful as a measure of dialysis efficiency. The explanation of isoprene increase in breath during dialysis requires further investigation.

Keywords

Breath test Gas chromatography Trimethylamine Isoprene End stage renal disease (ESRD) Hemodialysis 

Notes

Acknowledgements

We are greatly indebted to Prof. W. Sułowicz, Department of Nephrology, University Hospital, Medical College of Jagiellonian University, Krakow, Poland, for patients’ recruitment and enabling of clinical studies.

Compliance with ethical standards

Conflict of interest

The authors have declared that no conflict of interest exists.

Human and animal rights (with IRB approval number)

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institution at which the studies were conducted. Ethical approval for this study were obtained from the Bioethics Committee IRB 00002280.

Informed consent

Informed consent of participation in described studies was obtained from all individual participants taking part in this study.

References

  1. 1.
    Sułowicz W. Practical aspects of dialysis quality assessment. Forum Nefrologiczne. 2010;3(3):185–92.Google Scholar
  2. 2.
    Charra B, Calemard E, Ruffet M, Chazot C, Terrat J-C, Vanel T, et al. Survival as an index of adequacy of dialysis. Kidney Int. 1992;41:1286–91.CrossRefGoogle Scholar
  3. 3.
    Locatelli F. Dose of dialysis, convection and haemodialysis patients outcome—what the HEMO study doesn’t tell us: the European viewpoint. Nephrol Dial Transplant. 2003;18:1061–5.CrossRefGoogle Scholar
  4. 4.
    Eknoyan G, Beck GJ, Cheung AK, Daugirdas JT, Greene T, Kusek JW, et al. Effect of dialysis dose and membrane flux in maintenance hemodialysis. N Engl J Med. 2002;347(25):2010–9.CrossRefGoogle Scholar
  5. 5.
    Sułowicz W. Practical aspects of dialysis quality assessment. Forum Nefrologiczne. 2010;3:185–92.Google Scholar
  6. 6.
    Basile C, Casino F, Lopez T. Percent reduction in blood urea concentration duringdialysis estimates Kt/V in a simple and accurate way. Am J Kidney Dis. 1990;15(1):40–5.CrossRefGoogle Scholar
  7. 7.
    Kessler M, Canaud B, Pedrini LA. European best practice guidelines for hemodialysis (part 1). Section II. Hemodialysis adequacy. Nephrol Dial Transplant. 2002;17(Suppl 7):17–31.Google Scholar
  8. 8.
    Simenhoff ML, Burke JF, Suakkonen JJ. Biochemical profile or uremic breath. N Engl J Med. 1977;297:132–5.CrossRefGoogle Scholar
  9. 9.
    Bain MA, Faullb R, Fornasinic G, Milnea RW, Schumannd R, Evans AM. Quantifying trimethylamine and trimethylamine-N-oxide in human plasma: interference from endogenous quaternary ammonium compounds. Anal Biochem. 2004;334:403–5.CrossRefGoogle Scholar
  10. 10.
    Bain MA, Faull R, Fornasini G, Milne RW, Evans AM. Accumulation of trimethylamine and trimethylamine-N-oxide in end-stage renal disease patients haemodialysis. Nephrol Dial Transplant. 2006;21:1300–4.CrossRefGoogle Scholar
  11. 11.
    Davies SJ, Spanel P, Smith D. Breath analysis ammonia, volatile organic compounds and deuterated watervapour in chronic kidney disease and during dialysis. Bioanalysis. 2014;6(6):843–57.CrossRefGoogle Scholar
  12. 12.
    Weiner ID, Mitch WE, Sands JM. Urea and ammonia metabolism and the controlof renal nitrogen excretion. Clin J Am Soc Nephrol 2014:13.Google Scholar
  13. 13.
    Endre ZH, Pickering JW, Storer MK, Hu W-P, Moorhead KT, Allardyce R, et al. Breath ammonia and trimethylamine allow real-time monitoring of haemodialysis efficacy. Physiol Meas. 2011;32(1):115–30.CrossRefGoogle Scholar
  14. 14.
    Lirk P, Bodrogi F, Raifer H, Greiner K, Ulmer H, Rieder J. Elective haemodialysis increases exhaled isoprene. Nephrol Dial Transplant. 2003;18:937–41.CrossRefGoogle Scholar
  15. 15.
    Capodicasa E, Trovarelli G, Medio GED, Pelli MA, Lippi G, Verdura C, et al. Volatile alkanes and increased concentrations of isoprene in exhaled air during hemodialysis. Nephron. 1999;82:331–7.CrossRefGoogle Scholar
  16. 16.
    Davies S, Španel P, Smith D. A new “online” method to measure increased exhaled isoprene in end-stage renal failure. Nephrol Dial Transplant. 2001;16(4):836–9.CrossRefGoogle Scholar
  17. 17.
    Lowis S, Eastwood MA, Brydon WG. The influence of creatinine, lecithin and choline feeding on aliphatic amine production and excretion in the rat. Br J Nutr. 1985;54:43–51.CrossRefGoogle Scholar
  18. 18.
    Zeisel SH, daCosta KA. Increase in human exposure to methylamine precursors of N-nitrosamines after eating fish. Cancer Res. 1986;46(12 Pt 1):6136–8.Google Scholar
  19. 19.
    Zhang AQ, Mitchell S, Smith R. Fish odour syndrome: verification of carrier detection test. J Inherit Metab Dis. 1995;18:669–74.CrossRefGoogle Scholar
  20. 20.
    Ramezani A, Raj DS. The gut microbiome, kidney disease, and targeted interventions. J Am Soc Nephrol. 2014;25(4):657–70.CrossRefGoogle Scholar
  21. 21.
    Rydzewska-Rososłowska A, Kakareko K, Rosołowski M, Hryszko T, Brzósko S, Myśliwiec M, et al. Microbiom and the kidney. Post N Med. 2016;XXIX(9):659–63.Google Scholar
  22. 22.
    Manolis A. The diagnostic potential of breath analysis. Clin Chem. 1983;29(1):5–15.Google Scholar
  23. 23.
    Deneris ES, Stein RA, Mead JF. In vitro biosynthesis of isoprene from mevalonate utilizing a rat liver cytosolic fraction. Biochem Biophys Res Commun. 1984;123:691–6.CrossRefGoogle Scholar
  24. 24.
    Thomas K, Prazeller P, Mayr D, Jordan A, Rieder J, Fall R, et al. Human breath isoprene and its relation to blood cholesterol levels: new measurements and modeling. J Appl Physiol. 2001;91:762–70.CrossRefGoogle Scholar
  25. 25.
    Stone BG, Thiel DHV, Besse TJ, Duane WC, Evans CD, DeMaster EG. Effect of regulating cholesterol biosynthesis on breath isoprene excretion in men. Lipids. 1993;28:705–8.CrossRefGoogle Scholar
  26. 26.
    Salermo-Kenedy R, Cashman KD. Potentil applications of breath isopene as a biomarker in modern medicine: a concose overview. Wien Klin Wochenschr. 2005;117(5–6):180–6.CrossRefGoogle Scholar
  27. 27.
    Turner C, Smith PSpanel. D. A longitudinal study of breath isoprene in healthy volunteers using selected ion flow tube mass sectrometry (SIFT-MS). Physiol Meas. 2006;27:13–22.CrossRefGoogle Scholar
  28. 28.
    Kushch I, Arendacka B, Stolc S, Mochalski P, Filipiak W, Schwarz K, et al. Breath isoprene—aspects of normal physiology related to age, gender and cholesterol profile as determined in a proton transfer reaction mass spectrometry study. Clin Chem Lab Med. 2008;46(7):1011–8.CrossRefGoogle Scholar
  29. 29.
    King JKH, Unterkofler K, Mochalski P, Kupferthaler A, Teschl G, Teschl S, Hinterhuber H, Amann A. Physiological modeling of isoprene dynamics in exhaled breath. J Theor Biol. 2010;267:626–37.CrossRefGoogle Scholar
  30. 30.
    King J, Koc H, Unterkofler K, Mochalski P, Kupferthaler A, Teschl G, et al. Physiological modeling of isoprene dynamics in exhaled breath. J Theor Biol. 2010;267:626–37.CrossRefGoogle Scholar
  31. 31.
    Wzorek B, Mochalski P, Śliwka I, Amann A. Application of GC-MS with a SPME and thermal desorption technique for determination of dimethylamine and trimethylamine in gaseous samples for medical diagnostic purposes. J Breath Res. 2010;4:1–6.CrossRefGoogle Scholar
  32. 32.
    Grabowska-Polanowska B, Faber J, Skowron M, Miarka P, Pietrzycka A, Śliwka I, et al. Detection of potential CKD marker in breath using GC/MS coupled with thermal desorption method. J Chromatogr A. 2013;1301:179–89.CrossRefGoogle Scholar
  33. 33.
    King J, Kupferthaler A, Unterkofler K, Koc H, Teschl S, Teschl G, Miekisch W, Schubert J, Hinterhuber H, Amann A. Isoprene and acetone concentration profiles during exercise on an ergometer. J Breath Res. 2009;3:027006.CrossRefGoogle Scholar
  34. 34.
    Holmar J, Fridolin I, Uhlin F, Fernström A, Lumana M. Proceedings of the Estonian Academy of Sciences, estimation of dialysis patients’ survival through combined approach of small molecule uremic markers. Biomed Eng. 2014;63(3):315–21.  https://doi.org/10.3176/proc.2014.3.0.Google Scholar
  35. 35.
    Grabowska-Polanowska B, Skowron M, Miarka P, Pietrzycka A, Śliwka I. The application of chromatographic breath analysis in the search of volatile biomarkers of chronic kidney disease and coexisting type 2 diabetes mellitus. J Chromatogr B. 2017;1060:103–10.CrossRefGoogle Scholar
  36. 36.
    Narasimhan LR, Goodman W, Kumar C, Patel N. Correlation of breath ammonia with blood urea nitrogen and creatinine during hemodialysis. PNAS. 2001;98:4617–21.CrossRefGoogle Scholar
  37. 37.
    Capodicasa E, Trovarelli G, Brunori F, Vecchi L, Carobi C, De M et al. Lack of isoprene overproduction during peritoneal dialysis. Perit Dial Int. 2002;22(1):48–52.Google Scholar
  38. 38.
    Takayama F, Tsutsui S, Horie M, Shimokata K, Niwa T. Glutathionyl hemoglobin in uremic patients undergoing hemodialysis and continuous ambulatory peritoneal dialysis. Kidney Int. 2001;59(Suppl.78):155–8.CrossRefGoogle Scholar
  39. 39.
    Scoppola A, de Paolis P, Menzinger G, Lala A, Glulio SD. Plasma mevalonate concentrations in uremic patients. Kidney Int. 1997;51:908–12.CrossRefGoogle Scholar

Copyright information

© Japanese Society of Nephrology 2018

Authors and Affiliations

  • B. Grabowska-Polanowska
    • 1
    Email author
  • P. Miarka
    • 2
  • M. Skowron
    • 1
  • G. Chmiel
    • 2
  • A. Pietrzycka
    • 3
  • I. Śliwka
    • 1
  1. 1.Institute of Nuclear Physics Polish Academy of SciencesKrakówPoland
  2. 2.Department of NephrologyThe Jagiellonian University Medical CollegeKrakówPoland
  3. 3.Faculty of PharmacyThe Jagiellonian University Medical CollegeKrakówPoland

Personalised recommendations