Oxygen delivery to the blood stream by the dialysis fluid during continuous renal replacement therapy ex vivo

  • Yoshihiro TangeEmail author
  • Shigenori Yoshitake
Original Article Artificial Lung / ECMO
  • 27 Downloads

Abstract

Continuous renal replacement therapy (CRRT) maintains a balance in body water and electrolytes. CRRT supplies a higher quantity of fluid than intravenous fluid therapy along with simultaneous fluid withdrawal. We hypothesized that use of a high-oxygen-containing solution for high-volume fluid exchange would improve oxygenation in the blood during CRRT. To start with, we prepared a solution containing high oxygen. The objective of this study was to determine if this solution would increase the partial pressure of oxygen (pO2) in the blood more than that using a conventional solution during CRRT. We compared the gas profile of the experimental fluid ex vivo in a simulated CRRT for 24 h, using 2-L batches of bovine blood. A significant increase in the pO2, pH, and total oxygen delivery, and a significant decrease in the partial pressure of carbon dioxide (pCO2) were estimated in the bovine blood using the experimental solution during the simulated CRRT. This method is simpler to apply for oxygenation than the conventional method, and will be beneficial to hypoxic patients in terms of improving their blood oxygenation during CRRT.

Keywords

Oxygen Carbon dioxide Hypoxia Dialysis fluid Renal replacement therapy 

Notes

Acknowledgements

The authors would like to thank all the colleagues who helped with our experiments.

Funding

This research has not received any specific grant from any funding agencies in public, commercial, or not-for-profit sectors.

Compliance with ethical standards

Conflict of interest

The authors report no proprietary or commercial interest for any product mentioned or any concept discussed in this article.

References

  1. 1.
    Henig NR, Pierson DJ. Mechanisms of hypoxemia. Respir Care Clin N Am. 2000;6:501–21.CrossRefGoogle Scholar
  2. 2.
    Prasad B, Urbanski M, Ferguson TW, Karreman E, Tangri N. Early mortality on continuous renal replacement therapy (CRRT): the Prairie CRRT study. Can J Kidney Health Dis. 2016;3:36.CrossRefGoogle Scholar
  3. 3.
    Han F, Sun R, Ni Y, et al. Early initiation of continuous renal replacement therapy improves clinical outcomes in patients with acute respiratory distress syndrome. Am J Med Sci. 2015;349:199–205.CrossRefGoogle Scholar
  4. 4.
    Yang W, Hong J, Zeng Q, et al. Improvement of oxygenation in severe acute respiratory distress syndrome with high-volume continuous veno-venous hemofiltration. Glob Pediatr Health. 2016.  https://doi.org/10.1177/2333794X16645699.CrossRefPubMedPubMedCentralGoogle Scholar
  5. 5.
    Zhao P, Zheng R, Xue L, Zhang M, Wu X. Early fluid resuscitation and high volume hemofiltration decrease septic shock progression in swine. Biomed Res Int. 2015;2015:181845.PubMedPubMedCentralGoogle Scholar
  6. 6.
    Sombolos KI, Bamichas GI, Christidou FN, et al. pO2 and pCO2 increment in post-dialyzer blood: the role of dialysate. Artif Organs. 2005;29:892–8.CrossRefGoogle Scholar
  7. 7.
    Tange Y, Yoshitake S, Takesawa S. Simple method to make a supersaturated oxygen solution. J Artif Organs. 2018;21:392–5.CrossRefGoogle Scholar
  8. 8.
    Tange Y, Yoshitake S. Data on a simple method for producing a solution that contains a high partial pressure of oxygen and a low partial pressure of carbon dioxide. Data Brief. 2018;18:176–9.CrossRefGoogle Scholar
  9. 9.
    Gehlbach JA, Rehder KJ, Gentile MA, et al. Intravenous oxygen: a novel method of oxygen delivery in hypoxemic respiratory failure. Expert Rev Respir Med. 2017;11:73–80.CrossRefGoogle Scholar
  10. 10.
    Matsuki N, Ishikawa T, Ichiba S, et al. Oxygen supersaturated fluid using fine micro/nanobubbles. Int J Nanomed. 2014;9:4495–505.CrossRefGoogle Scholar
  11. 11.
    Tange Y, Migita H, Yoshitake S, et al. Dialysate with High Partial Pressure of oxygen enhances oxygenation in blood during simulated circulation of hemodialysis. Adv Biomed Eng. 2012;1:43–6.CrossRefGoogle Scholar
  12. 12.
    Xu H, Li Y, Gao W, Zhang H, Xu LX. Hyperoxygenated solutions in clinical practice: Preventing or reducing hypoxic injury. J Int Med Res. 2011;39:1589–606.CrossRefGoogle Scholar
  13. 13.
    Lieberman JA, Weiskopf RB, Kelly SD, et al. Critical oxygen delivery in conscious humans is less than 7.3 ml O2 x kg(-1) x min(-1). Anesthesiology. 2000;92: 407–13.CrossRefGoogle Scholar
  14. 14.
    Grady DJ, Gentile MA, Riggs JH, Cheifetz IM. Improved arterial blood oxygenation following intravenous infusion of cold supersaturated dissolved oxygen solution. Clin Med Insights Circ Respir Pulm Med. 2014;8:11–6.CrossRefGoogle Scholar
  15. 15.
    Takahashi N, Nakada T, Oda S. Efficient CO2 removal using extracorporeal lung and renal assist device. J Artif Organs. 2018;21:427–34.CrossRefGoogle Scholar

Copyright information

© The Japanese Society for Artificial Organs 2019

Authors and Affiliations

  1. 1.Department of Medical EngineeringKyushu University of Health and WelfareNobeokaJapan

Personalised recommendations