Efficacy of regional saturation of oxygen monitor using near-infrared spectroscopy for lower limb ischemia during minimally invasive cardiac surgery

  • Teppei Toya
  • Tomoyuki FujitaEmail author
  • Satuki Fukushima
  • Yusuke Shimahara
  • Yuta Kume
  • Kizuku Yamashita
  • Yorihiko Matsumoto
  • Naonori Kawamoto
  • Junjiro Kobayashi
Original Article Cardiopulmonary Bypass
  • 96 Downloads

Abstract

Lower limb ischemia with an occlusive cannula is a potential complication of minimally invasive cardiac surgery (MICS). We evaluated intraoperative local oxygen supply–demand balance by monitoring regional saturation of oxygen (rSO2) using near-infrared spectroscopy (NIRS), and analyzed the correlation between cannula size and fluctuation range of rSO2. Fifty-four patients undergoing MICS surgery using femoral artery cannulation from April 2015 to August 2016 were enrolled. The rSO2 of both the cannulated and uncannulated lower limbs were measured using NIRS. The association between the decline of rSO2 from baseline (delta-rSO2) and the ratio of the cannula diameter to the femoral artery diameter (Cd/FAd) was analyzed. Of the 54 patients, 16 (30%) (Group 1) showed values over 0.65 for Cd/FAd, and the remaining 38 (70%) (Group 2) showed values under 0.65. No patient developed postoperative lower limb ischemia. No patient was treated with an ipsilateral distal perfusion cannula. There were significant differences between Group 1 and Group 2 in the decrease of rSO2 at the point of cannulation on the cannulated limb. In the lower limb on the cannulated side, delta-rSO2 showed a significant decrease in Group 1 compared to Group 2 (Group 1 vs Group 2: 19.9 vs 11.0%; p < 0.001). Delta-rSO2 was significantly correlated with body surface aera (BSA), but not with gender or age. Decreasing rSO2 correlates with the Cd/FAd index. Low BSA, Cd/Fad > 0.65 is considered as the risk factor for decline of rSO2 in cannulated limb in MICS.

Keywords

Near-infrared spectroscopy Regional saturation of oxygen Minimally invasive cardiac surgery Lower limb ischemia 

Notes

Compliance with ethical standards

Conflict of interest

All authors declare that they have no conflict of interest.

Supplementary material

10047_2018_1057_MOESM1_ESM.doc (36 kb)
Supplementary material 1 (DOC 35 KB)

References

  1. 1.
    Glower DD, Landolfo KP, Clements F, et al. Mitral valve operation via port-access versus median sternotomy. Eur J Cardiothorac Surg. 1998;14:143–7.CrossRefGoogle Scholar
  2. 2.
    Grossi EA, Loulmet DF, Schwartz CF, et al. Evolution of operative techniques and perfusion strategies for minimally invasive mitral valve repair. J Thorac Cardiovasc Surg. 2012;143:68–70.CrossRefGoogle Scholar
  3. 3.
    Gander JW, Fisher JC, Reichstein AR, et al. Limb ischemia after common femoral artery cannulation for venoarterial extracorporeal membrane oxygenation: an unresolved problem. J Pediatr Surg. 2010;45:2136–40.CrossRefGoogle Scholar
  4. 4.
    Bisdas T, Beutel G, Wamecke G, et al. Vascular complications in patients undergoing femoral cannulation for extracorporeal membrane oxygenation support. Ann Thorac Surg. 2011;92:626–31.CrossRefGoogle Scholar
  5. 5.
    Gates JD, Bichell DP, Rizzo RJ, et al. Thigh ischemia complicating femoral vessel cannulation for cardiopulmonary bypass. Ann Thorac Surg. 1996;61:730–3.CrossRefGoogle Scholar
  6. 6.
    Madershahian N, Nagib R, Wippermann J, et al. A simple technique of distal limb perfusion during prolonged femorofemoral cannulation. J Card Surg. 2006;21:168–9.CrossRefGoogle Scholar
  7. 7.
    Ceviz M, Ozyazicioglu A, Kocak H. The prevention of leg ischemia during femoral cannulation. Ann Thorac Surg. 1999;67:1217.CrossRefGoogle Scholar
  8. 8.
    Riley JB, Hardin SB, Winn BA, et al. In vitro comparison of cavoatrial (dual stage) cannulae for use during cardiopulmonary bypass. Perfusion. 1986;1:197–204.CrossRefGoogle Scholar
  9. 9.
    Delius RF, Montoya JP, Merz SJ, et al. A new method for describing the performance of cardiac surgery cannulas. Ann Thorac Surg. 1992;33:278–81.CrossRefGoogle Scholar
  10. 10.
    Kasirajan V, Simmons I, King J, et al. Technique to prevent limb ischemia during peripheral cannulation for extracorporeal membrane oxygenation. Perfusion. 2002;17:427–8.CrossRefGoogle Scholar
  11. 11.
    Formica F, Avalli L, Colagrande L, et al. Extracorporeal membrane oxygenation to support adult patients with cardiac failure: predictive factors of 30-day mortality. Interact Cardiovasc Thorac Surg. 2010;10:721–6.CrossRefGoogle Scholar
  12. 12.
    Greenberg S, Murphy G, Shear T, et al. Extracranial contamination in the INVOS 5100C versus the FORE-SIGHT ELITE cerebral oximeter: a prospective observational crossover study in volunteers. Can J Anesth/J Can Anesth. 2016;63:512.CrossRefGoogle Scholar
  13. 13.
    Faber DJ, Mik EG, et al. Toward assessment of blood oxygen saturation by spectroscopic optical coherence tomography. Opt Lett. 2005;30:1015–7.CrossRefGoogle Scholar
  14. 14.
    Scholkmann F, Kleiser S, Metz AJ, et al. A review on continuous wave functional near-infrared spectroscopy and imaging instrumentation and methodology. Neuroimage. 2014;15:6–27.CrossRefGoogle Scholar
  15. 15.
    Fantini S, Hueber D, Franceschini MA, et al. Non-invasive optical monitoring of the newborn piglet brain using continuous-wave and frequency-domain spectroscopy. Phys Med Biol. 1999;44:1543–63.CrossRefGoogle Scholar
  16. 16.
    Delpy DT, Cope M. Quantification in tissue near-infrared spectroscopy. Philos Trans R Soc Lond B Biol Sci. 1997;352:649–59.CrossRefGoogle Scholar
  17. 17.
    Scholkmann F, Wolf M. General equation for the differential pathlength factor of the frontal human head depending on wavelength and age. J Biomed Opt. 2013;18:105004.CrossRefGoogle Scholar
  18. 18.
    Gao L, Elwell CE, Kohl-Bareis M, et al. Effects of assuming constant optical scattering on haemoglobin concentration measurements using NIRS during a Valsalva Manoeuvre. Adv Exp Med Biol. 2011;701:15–20.CrossRefGoogle Scholar
  19. 19.
    Yoshitani K, Kawaguchi M, Okuno T, et al. Measurements of optical pathlength using phase-resolved spectroscopy in patients undergoing cardiopulmonary bypass. Anesth Analg. 2007;104:341–6.CrossRefGoogle Scholar
  20. 20.
    Yoshitani K, Kuwajima K, Irie T, et al. Clinical validity of cerebral oxygen saturation measured by time-resolved spectroscopy during carotid endarterectomy. J Neurosurg Anesthesiol. 2013;25:248–53.CrossRefGoogle Scholar

Copyright information

© The Japanese Society for Artificial Organs 2018

Authors and Affiliations

  • Teppei Toya
    • 1
  • Tomoyuki Fujita
    • 1
    Email author
  • Satuki Fukushima
    • 1
  • Yusuke Shimahara
    • 1
  • Yuta Kume
    • 1
  • Kizuku Yamashita
    • 1
  • Yorihiko Matsumoto
    • 1
  • Naonori Kawamoto
    • 1
  • Junjiro Kobayashi
    • 1
  1. 1.Department of Cardiac SurgeryNational Cerebral and Cardiovascular CenterOsakaJapan

Personalised recommendations