Chronic nonpulsatile blood flow is compatible with normal end-organ function: implications for LVAD development

  • Satoshi SaitoEmail author
  • Tomohiro Nishinaka
REVIEW

Abstract

Evolving blood pump technology has produced user-friendly continuous-flow left ventricular assist devices, but uncertainty exists about the safety of chronic nonpulsatile circulation. Recent experimental and clinical evidence suggest that pulse pressure is not required from a blood pump. End-organ function is well maintained with nonpulsatile systems, although pulse pressure may accelerate recovery from cardiogenic shock. Form follows function, so the effects of reduced pulse pressure on the arterial wall are not surprising. The ability to alter aortic wall morphology by reducing pulse pressure may have important implications for the future treatment of arterial pathology. Both centrifugal and axial-flow pumps can be miniaturized and are reliable and silent. Doubts about the feasibility of long-term circulation with reduced pulse pressure are disappearing.

Key words

Mechanical circulatory support Continuous flow LVAD Nonpulsatile circulation 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Rose, EA, Gelijns, AC, Moskowitz, AJ,  et al. 2001Long-term use of a left ventricular assist device for end stage heart failureN Engl J Med34514351443Google Scholar
  2. 2.
    Wesolowski, SA, Sauvage, LR, Pinc, RD 1955Extracorporeal circulation: the role of the pulse in maintenance of the systemic circulation during heart–lung bypassSurgery37663Google Scholar
  3. 3.
    Ead, HW, Green, JH, Nell, EA 1952Comparison of the effects of pulsatile and non-pulsatile blood flow through the carotid sinus on the reflexogenic activity of the sinus baroreceptor in the catJ Physiol118509519Google Scholar
  4. 4.
    Marry, M, Ciron, F, Birrwell, WC,  et al. 1969Effects of de-pulsation of renal blood flow upon renal function and renin secretionSurgery66242249Google Scholar
  5. 5.
    Buket, S, Alayunt, A, Ozbaran, M,  et al. 1994Effects of pulsatile flow during cardiopulmonary bypass on the thyroid hormone metabolismAnn Thorac Surg579396Google Scholar
  6. 6.
    Gaer, JAR, Shaw, ADS, Wild, R,  et al. 1994Effects of cardiopulmonary bypass on gastrointestinal perfusion and functionAnn Thorac Surg57371375Google Scholar
  7. 7.
    Minami, K, Komer, MM, Vyska, K,  et al. 1990Effects of pulsatile perfusion on plasma catecholamine levels and haemodynamics during and after cardiopulmonary bypassJ Thorac Cardiovasc Surg998291Google Scholar
  8. 8.
    Ciardullo, RC, Schaff, HV, Flaherty, JT,  et al. 1978Comparison of regional myocardial blood flow and metabolism distal to a critical coronary stenosis in a fibrillating heart during alternative periods of pulsatile and nonpulsatile perfusionJ Thorac Cardiovasc75371375Google Scholar
  9. 9.
    Minami, K, El-Banayosy, A, Posival, H,  et al. 1992Improvement of survival in patients with cardiogenic shock by using nonpulsatile and pulsatile ventricular assist devicesInt J Artif Organs14715721Google Scholar
  10. 10.
    Taylor, KM, Bain, WH, Davidson, KG,  et al. 1982Comparative clinical study of pulsatile and nonpulsatile perfusion in 350 consecutive patientsThorax37324330Google Scholar
  11. 11.
    Jett, GK 1996Abiomed BVS 5000: experience and potential advantagesAnn Thorac Surg61301304Google Scholar
  12. 12.
    Wikens, H, Regelson, W, Hoffmeister, FS 1962The physiologic importance of pulsatile blood flowN Eng J Med267443446Google Scholar
  13. 13.
    Golding, LR, Murakami, G, Harasaki, H,  et al. 1982Chronic nonpulsatile blood flowTrans ASAIO288185Google Scholar
  14. 14.
    Konishi, H, Yland, MI, Brown, M,  et al. 1996Effects of pulsatility and haemodynamic power on recovery of renal functionASAIO J42M720723Google Scholar
  15. 15.
    Fukae, K, Tominaga, R, Tokunaga, S,  et al. 1996The effects of pulsatile and non-pulsatile systemic perfusion on renal sympathetic nerve activity in the anaesthetized dogJ Thorac Cardiovasc Surg111148184Google Scholar
  16. 16.
    Sadahiro, M, Haneda, K, Mohri, H,  et al. 1994Experimental study of cerebral autoregulation during cardiopulmonary bypass with or without pulsatile perfusionJ Thorac Cardiovasc Surg108446454Google Scholar
  17. 17.
    Saito, S, Westaby, S, Pigott, D,  et al. 2001Reliable long-term non-pulsatile circulatory support without anticoagulationEur J Cardiothorac Surg19678683Google Scholar
  18. 18.
    Valdes, F, Takatani, S, Jacobs, BG,  et al. 1980Comparison of haemodynamic changes in a chronic non-pulsatile biventricular bypass and total artificial heartTrans ASAIO26455Google Scholar
  19. 19.
    Tatsumi, E, Taenaka, Y, Sakaki, M,  et al. 1994Rapid conversion of systemic flow pattern from pulsatile to nonpulsatile in conscious goatsNiimi, Oda, MSawada, TXiu, RJ eds. Progress in microcirculation researchPergamonOxford463Google Scholar
  20. 20.
    Tatsumi, E, Toda, K, Taenaka, Y,  et al. 1995Acute phase responses of vasoactive hormones to nonpulsatile systemic circulationASAIO J41460465Google Scholar
  21. 21.
    Yozu, R, Golding, LAR, Jacobs, G,  et al. 1985Experimental results and future prospects for a nonpulsatile cardiac prosthesisWorld J Surg9116127Google Scholar
  22. 22.
    Golding, LR, Jacobs, G, Murakami, T,  et al. 1980Chronic nonpulsatile blood flow in an alive awake animal: 34 days survivalTrans ASAIO26251Google Scholar
  23. 23.
    Saito, S, Westaby, S, Pigott, D,  et al. 2002End-organ function during chronic non-pulsatile circulationAnn Thorac Surg7410801085Google Scholar
  24. 24.
    Toda, K, Tatsumi, E, Taenaka, Y,  et al. 1995Characteristics of sympathetic nerve activity during nonpulsatile circulationAkutsu, TKoyanagi, H eds. Artificial Heart 5Springer-VerlagTokyoGoogle Scholar
  25. 25.
    Nishinaka, T, Tatsumi, E, Nishimura, R,  et al. 1999Cardiac autonomic nervous function during nonpulsatile left heart bypassArtif Organs23500503Google Scholar
  26. 26.
    Nishinaka, T, Tatsumi, E, Taenaka, Y,  et al. 2000Changes of heart rate variability during left ventricular assist: Comparison between pulsatile device and centrifugal pumpASAIO J46170Google Scholar
  27. 27.
    Shoor, PM, Hammill, FS, Griffith, LD, Dilley, RB, Bernstein, EF 1980Haemodynamic response to exercise in the unanesthesized calf with pulseless arterial flowTrans ASAIO2617Google Scholar
  28. 28.
    Tominaga, R, Smith, W, Massiello, A, Harasaki, H, Golding, AR 1996Chronic nonpulsatile blood flow. II. Hemodynamic responses to progressive exercise in calves with chronic nonpulsatile biventricular bypassJ Thorac Cardiovasc Surg111857862Google Scholar
  29. 29.
    Bellotto, F, Johnson, RG, Watanabe, J,  et al. 1992Mechanical assistance of the left ventricle: acute effects on cardiac performance and coronary flow of different perfusion patternsJ Thorac Cardiovasc Surg104561Google Scholar
  30. 30.
    Bixer, TJ, Magee, PG, Flaherty, JT, Gardner, TJ, Gott, VL 1979Beneficial effects of pulsatile perfusion in the hypertrophied ventricle during ventricular fibrillationCirculation60141Google Scholar
  31. 31.
    Moores, WY, Gago, O, Morris, JD, Peck, CC 1977Serum and urinary amylase levels following pulsatile and continuous cardiopulmonary bypassJ Thorac Cardiovasc Surg747376Google Scholar
  32. 32.
    Chiang, BY, Ye, CX, Gau, DX,  et al. 1992Effects of pulsatile reperfusion on myocardial high-energy phosphate following global ischemiaASAIO Trans37485Google Scholar
  33. 33.
    Anstadt, MP, Tedder, M, Hegde, SS,  et al. 1993Pulsatile versus nonpulsatile reperfusion improves cerebral blood flow after cardiac arrestAnn Thorac Surg56453Google Scholar
  34. 34.
    Henze, T, Stephan, H, Sonntag, H 1990Cerebral dysfunction following extracorporeal circulation for aortocoronary bypass surgery: no difference in neuropsychological outcome after pulsatile versus nonpulsatile flowThorac Cardiovasc Surg3865Google Scholar
  35. 35.
    Tominaga, R, Smith, WA, Massiello, A, Harasaki, H, Golding, LA 1994Chronic non-pulsatile flow. Cerebral auto-regulation in chronic nonpulsatile biventricular bypass and fibrillating heart: carotid blood flow responses to hypercapniaJ Thorac Cardiovasc Surg108907912Google Scholar
  36. 36.
    Nishinaka, T, Tatsumi, E, Taenaka, Y, Takano, H, Koyanagi, H 2000Effect of reduced pulse pressure to the cerebral metabolism during prolonged nonpulsatile left heart bypassArtificial Organs24676679Google Scholar
  37. 37.
    Nojiri, C, Kijima, T, Maekawa, J,  et al. 2000Terumo implantable left ventricular assist system; results of long-term animal studyASAIO J46117122Google Scholar
  38. 38.
    Ohnishi, H, Nishinaka, T, Tatsumi, E,  et al. 2002Proliferative changes in the media of afferent arterioles under prolonged continuous-flow left heart bypassJ Artif Organs5204207Google Scholar
  39. 39.
    Kihara, S, Litwak, KN, Nichols, L,  et al. 2003Smooth muscle cell hypertrophy of renal cortex arteries with chronic continuous-flow left ventricular assistAnn Thorac Surg75178183Google Scholar
  40. 40.
    Stacy, DL, Prewitt, RL 1989Effects of chronic hypertension and its reversal on arteries and arteriolesCirc Res65869879Google Scholar
  41. 41.
    Wilson, E, Mai, Q, Sudhir, K, Weiss, RH, Ives, HE 1993Mechanical strain induces growth of vascular smooth muscle cells via autocrine of PDGFJ Cell Biol123741749Google Scholar
  42. 42.
    Nishimura, T, Tatsumi, E, Takaichi, S,  et al. 1998Prolonged nonpulsatile left heart bypass with reduced systemic pulse pressure caused morphological changes in the aortic wallArtif Organs22405410Google Scholar
  43. 43.
    Nishimura, T, Tatsumi, E, Takaichi, S,  et al. 1997Morphologic changes of the aortic wall due to reduced systemic pulse pressure in prolonged nonpulsatile left heart bypassASAIO J43691695Google Scholar
  44. 44.
    Nishimura, T, Tatsumi, E, Taenaka, Y,  et al. 1999Effects of long-term nonpulsatile left heart bypass on mechanical properties of the aortic wallASAIO J45455459Google Scholar
  45. 45.
    Nishimura, T, Tatsumi, E, Nishinaka, T 1999Prolonged nonpulsatile left heart bypass diminishes vascular contractilityInt J Artif Organs22492498Google Scholar
  46. 46.
    Nishinaka, T, Tatsumi, E, Nishimura, T, Shioya, K, Ohnishi, H, Taenaka, Y, Takano, H 2001Change in vasoconstrictive function during prolonged nonpulsatile left heart bypass: Letter to the Editor and replyArtificial Organs2510221024Google Scholar
  47. 47.
    Nishinaka, T, Tatsumi, E, Taenaka, Y, Takano, H, Koyanagi, H 2000The effect of prolonged nonpulsatile left heart bypass on vascular control statusJ Artif Organs33438Google Scholar
  48. 48.
    Frazier, OH, Myers, TJ, Gregoric, ID,  et al. 2002Initial clinical experience with the Jarvik 2000 implantable axial-flow ventricular assist systemCirculation10528552860Google Scholar
  49. 49.
    Westaby, S, Banning, AP, Saito, S,  et al. 2002Circulatory support for long-term treatment of heart failure: Experience with an intraventricular continuous-flow pumpCirculation10525882591Google Scholar
  50. 50.
    DeBakey, ME 1999A miniature implantable axial flow ventricular assist deviceAnn Thorac Surg68637640Google Scholar
  51. 51.
    Wieselthaler, GM, Schima, H, Hiesmayr, M,  et al. 2000First clinical experience with the DeBakey VAD continuous axial flow pump for bridge to transplantationCirculation101356Google Scholar

Copyright information

© The Japanese Society for Artificial Organs 2005

Authors and Affiliations

  1. 1.Department of Cardiovascular SurgeryTokyo Women's Medical UniversityTokyoJapan
  2. 2.Oxford Heart CentreThe John Radcliffe HospitalOxfordUK

Personalised recommendations