1 Introduction

Owing to sensitivity requirements, EPR measurements are almost always performed using a resonator, where the properties of the resonator are tuned for the particular type of EPR experiment. In continuous wave (CW) EPR measurements low-loss (high-Q) resonators are used, as they efficiently concentrate the available microwave magnetic field at the sample, enhancing sensitivity. As such, resonators used in classical CW EPR applications typically only have a small resonator bandwidth. In contrast, pulse EPR measurements require a sufficiently large resonator bandwidth. Many pulse EPR measurements drive off-resonance transitions, which can only be efficiency pumped if they fall within the resonator bandwidth, regardless of the bandwidth of the source or the duration of the microwave pulse(s). How bandwidth limitations affect pulse EPR experiments (ESEEM, ENDOR, DEER, EDNMR, etc.) is described in detail in the supplementary information (SI), see also [1, 2]. In addition, the resonator bandwidth places restrictions on the pulse lengths and timings. This includes the minimum pulse length that can be achieved without distortion, and in many circumstances, the dead time of the pulse experiment, with the dead time being the delay between the applied microwave pulse (excitation power in the range of Watts to kilowatts) and subsequent signal detection (signal power in the nanoWatt range). As such, resonators used for pulse EPR ideally need large bandwidths. Current EPR technology routinely allows minimum microwave pulse lengths of the order of 10 ns (excitation bandwidth corresponds to 120 MHz [3]). Consequently, the ideal resonator bandwidth needs to be at least ± 100 MHz, i.e., 200 MHz, regardless of the operating frequency.

The mismatch between the resonator requirements needed for CW and pulse EPR measurements makes the construction of a general purpose resonator challenging. This problem has been overcome at low frequencies (< 10 GHz) by developing mechanisms to vary the coupling of the EPR resonator to the microwave source via an intervening transmission line. Such coupling mechanisms allow for two modes of operation: (i) a high-Q regime for CW measurements in which the resonator is ‘critically coupled’ to the transmission line; and (ii) a broad bandwidth regime for pulse EPR in which the resonator is described as ‘overcoupled’; see supplementary information. There are many different types of coupling mechanisms (couplers). A key criterion of the coupler is the range of coupling strengths it allows. In CW EPR, the coupling range determines the maximum sample size of samples with high dielectric losses. In pulse EPR experiments, it limits the maximum resonator bandwidth. At low microwave frequencies such as X-band (9.5 GHz), dielectric or loop-gap resonators, which are typically coupled by an antenna coupling mechanism, perform well in both regimes.

At higher microwave frequencies, i.e., Q-band (34 GHz) and above, resonators are typically connected to the source via a hollow rectangular waveguide. Microwaves pass from the rectangular waveguide to the resonator through a thin metal diaphragm, which accommodates an aperture of fixed geometry, i.e., a slot or circular iris. The variability of the coupling is achieved using several approaches which regulate the transmittance of the iris and thus the cavity coupling (see SI). Cylindrical cavities, operating in the TE011 mode, typically use a compact sliding short in the waveguide to couple the transmission line and resonator. This design is suitable for CW and pulsed EPR experiments and is realized in several commercially available (Bruker Biospin) and laboratory-developed EPR probeheads [4,5,6]. One example is the general purpose Q-band probehead developed at the TU Berlin, which is now used in several EPR laboratories worldwide [5]. The original probehead allowed CW EPR and a large variety of pulsed EPR experiments including ENDOR, but had limited ESEEM, DEER and EDNMR performance because the maximum bandwidth that can be achived by the short coupler is limited to < 50 MHz [5].

In this work, we describe an improved short coupler design that overcomes this limitation. The new probehead design allows for precise, critical coupling of high-Q cavities and for very strong overcoupling, yielding a cavity bandwidth > 200 MHz, ideal for pulse EPR experiments. Strong overcoupling is achieved by modifying the iris self-resonance properties using a dielectric material. In addition, it is shown that the performance of the resonator-coupler-transmission line unit can be accurately predicted by electromagnetic calculations, allowing future development.

2 Materials and Methods

2.1 Microwave Resonator

The probehead model used for the electromagnetic calculations is depicted in Fig. 2a. The dimensions of the model correspond to those of the cavity used in the experiments (Table 1) and are also very similar to the dimensions of general purpose Q-band resonators described previously [5]. Dielectric plates were prepared from laminate Rogers 3010 materials. Pieces with thicknesses of 100, 210, 250 or 290 µm were prepared. The material was affixed to the resonator using ethyl 2-cyanoacrylate adhesive.

Table 1 Summary of final probehead model geometry and materials

2.2 Electromagnetic Simulations

These were performed using the CST microwave studio package. The reflection (S11) and transmission (S21) coefficients, and the electric and magnetic field amplitudes were calculated using a frequency domain solver with a tetrahedral mesh adjusted to yield reproducible results (typically about 106 tetrahedrons for the full cavity model). The transmission properties of the isolated iris were calculated using an aperture of different geometries (round, slot etc.) in a WR28 waveguide. Losses were neglected in the simulations of the isolated iris, i.e., a perfect conductor and lossless dielectric material were assumed.

2.3 Q-Band mw Bridge Extension

A home-built X-to-Q-band bridge extension compatible with a Bruker ELEXSYS-I X-band EPR spectrometer was used for all experiments. A simplified block diagram of the bridge extension is shown in Fig. 1. Further details are described in the supplementary information.

Fig. 1
figure 1

Basic scheme of the the pulse Q-band bridge extension: (1) DRO-based fixed frequency mw source; (2) voltage controlled attenuator; (3) up-converter mixer; (4) bandpass filter; (5) high-power mw amplifier; (6) rotary vane attenuator; (7) circulator; (8) fast PIN switch; (9) low-noise amplifier; (10) down-converter mixer. Full schematic details including power output measurements are presented in the SI

2.4 EPR Measurements

1,3-Bis(diphenylene)-2-phenylallyl (BDPA) radical (Sigma Aldrich) prepared in ortho-terphenyl (Sigma Aldrich) matrix (0.1% BDPA concentration) was used for all experiments.

The bandwidth of the critically coupled and overcoupled resonator was estimated by performing microwave nutation measurements across the full frequency range of the microwave bridge (33.5–34.5 GHz). The pulse sequence used for the nutation experiment was tpT– tπ/2 –\(\tau\) – 2tp π/2 – \(\tau\) – echo. The π/2 pulse length, tπ/2, was fixed to 40 ns, \(\tau\) was set to 1000 ns and the delay between the preparative pulse tp and echo detection was fixed at T = 10 μs. The length of the preparative pulse tp was incremented in 4 ns steps. The static magnetic field for different microwave frequencies was adjusted to maintain the resonance condition, i.e., achieve the maximum spin echo intensity at each frequency offset. At room temperature a shot repetition time of 5 ms was used. At low temperatures (20–50 K), the shot repetition time was extended to 400 ms. The resonator ringdown decay was measured using an external oscilloscope, by recording the response of the system following a low power pulse of 200 ns.

The procedure used to reproducibly overcouple the resonator was as follows: the lower and higher critically coupled short positions were identified, as visualized using the spectrometer’s in-built voltage controlled oscillator (VCO, i.e., tune mode). The short was then set at the position halfway between. At room temperature, the maximally overcoupled short position was further optimized using water as a microwave absorber, as described previously [7]. The sample tube was loaded with two thin-wall capillaries (0.8 mm OD, 0.6 mm ID) filled with water, offset from the bottom of the tube by a 5 mm length of quartz or cotton wool. The water sample was inserted into the resonator to the maximum length where it could still be critically coupled. The short position at which critical coupling for water is achieved corresponds to the maximally overcoupled position for the BDPA sample.

3 Results

3.1 Electromagnetic Simulations

Figure 2 shows the electromagnetic calculation results for the cavity model with a circular iris of 2.6 mm diameter. The cavity height was adjusted such that the TE011 cavity mode appeared at 34 GHz. In the 28 GHz–40 GHz frequency range, only two additional modes can be identified: a TE211 mode at about 30.18 GHz, and a TE311 mode at 39.44 GHz (Fig. 2b). Additional eigenmode analysis reveals that all transverse magnetic modes (TM012, TM110, TM111) in this range are suppressed by the four ENDOR posts located symmetrically about the sample tube. Also note that the transverse electric TE112 mode and additional hybrid modes caused by the ENDOR posts and the sample tube are not excited by the sidewall iris. This is not the case for top/bottom wall irises [8] and represents a key advantage of the sidewall iris design.

Fig. 2
figure 2

a Model of the TE011 cavity used for electromagnetic simulations. The geometrical and material parameters of the model are summarized in Table 1. The coordinate system used in this work is annotated in the figure. b Reflection coefficient, S11, for a critically coupled cavity (short position L = − 9.7 mm). c Dependence of the coupling coefficient (resonator bandwidth) on the short position relative to the center of a circular iris of 2.6 mm diameter

Figure 2c shows the dependence of the coupling parameter (β) and cavity bandwidth (\(\Delta {\nu }_{1/2}\)) of the TE011 mode as a function of the short position, as evaluated from the reflection curves, S11, for different short positions. The short position regulates the waveguide-to-cavity transmission coefficient of the iris which is related to cavity coupling; see SI. Maximum transmission and coupling is achieved for short positions where the transverse electric (z-direction) and magnetic (x-direction) field amplitudes are maximal at the iris, i.e., short positioned at \(L(\beta =max.)=n\frac{{\lambda }_{g}}{2}+\frac{{\lambda }_{g}}{4}\)=(8.4 mm, 14.0 mm, …) with \({\lambda }_{g}\)=11.24 mm being the wavelength of the WR28 waveguide at 34 GHz and \(n=\) 0, 1, 2 …. Zero transmission (no coupling) takes place for the short positioned at \(L(\beta =0)=n\frac{{\lambda }_{g}}{2}\) with respect to the iris center, i.e., zero field amplitudes at the iris center. Here we opted for an iris diameter of 2.6 mm of the experimental probehead. While suitable for CW-EPR, this iris diameter yields only a maximum bandwidth of 18.7 MHz (\(\beta =1.7\), i.e., overcoupled), which is severely below the minimum bandwidth required for many pulse EPR experiments.

The diameter of the iris modulates the maximum waveguide-to-cavity transmittance of the iris, with a larger diameter resulting in a larger bandwidth; see Fig. S3 in the SI. The largest possible circular iris (3.55 mm) is constrained by the dimensions of the waveguide (WR28 waveguide dimensions of 7.116 × 3.556 mm2). Using a 3.55 mm iris diameter yields a maximum cavity bandwidth of 100 MHz, but also reduces the unloaded Q-value, and thus the maximum magnetic field amplitude for the critically coupled cavity, which is disadvantageous for CW EPR; see Fig S3 in the SI.

To obtain a higher cavity bandwidth while keeping the iris diameter small, the iris transmission coefficient to the cavity must be increased. This is best realized by adjusting the iris self-resonance frequency to the resonance frequency of the TE011 mode, which will result in both the highest wave transmission and the maximum cavity bandwidth. Positioning a dielectric plate with a high dielectric constant in front of the iris increases the electric and magnetic field amplitudes at the iris. This is equivalent to increasing of the iris inductance and capacitance, and leads to a downshift in the iris resonance frequency [9]. Calculations which included a dielectric plate (ε = 10.7) of 3 × 3 mm2 placed in front of the isolated iris of 2.6 mm diameter do indeed show a continuous downshift of the self-resonance frequency and a corresponding increase in the transmission coefficient at 34 GHz as a function of the dielectric plate thickness. A maximum transmission coefficient of T = 0.96 is predicted for a dielectric plate thickness of 0.32 mm; see Fig S5 in the SI.

Figure 3a shows the reflection coefficient for a full cavity model calculated without a dielectric plate and the short position optimized for critical coupling. The frequency dependence of the magnetic field amplitude (x-component) at the cavity center is shown in Fig. 3c. Analysis of both plots yields a cavity bandwidth of \(\Delta {\nu }_{1/2}^{0}=\) 13.7 MHz, \({Q}_{U}=2\cdot\frac{{\nu }_{0}}{\Delta {\nu }_{1/2}^{0}}=\) 4960. Inspection of the magnetic field distribution at the resonance frequency shows the field distribution is very similar to that of the canonical TE011 mode, which is slightly distorted by the quartz tube and ENDOR posts (see inserts in Fig. 3c).

Fig. 3
figure 3

S11 parameters of the a critically coupled (DT = 0) and b max. overcoupled (DT = 0.25 mm, 3 × 3 mm2, fixed on iris) resonator, with 2.6 mm iris diameter. x-component of the magnetic field amplitude, \({H}_{x}(\mathrm{0,0},0)\), at the resonator center for c critically coupled and d max. overcoupled resonator. The calculated normalized distributions of the magnetic field x-amplitude, \({H}_{x}/{H}_{x}(\mathrm{0,0},0)\), in the (left) y,z-plane (x = 0) and (right) x,z-plane (y = 0) are shown in inserts on a logarithmic scale

The corresponding case for a dielectric plate of DT = 0.25 mm and a short position optimized for maximum overcoupling (waveguide-cavity transmittance) is shown in Fig. 3b, d. Both the S11 curve and magnetic field amplitude curves have asymmetrical profiles. The magnetic field distribution in the cavity is distorted as compared to that in the model without a dielectric plate (see inserts Fig. 3d). A cavity bandwidth of 206 MHz is evaluated from the frequency dependence of the magnetic field amplitude, Fig. 3d, which is already suitable for most pulse EPR applications. The maximum magnetic field amplitude is reduced by a factor of 2.9 compared to the critically coupled case without a dielectric plate (310 \(\frac{\text{A}}{{\text{m}}}\cdot\frac{1}{\sqrt{\text{W}}}\) vs. 909 \(\frac{\text{A}}{{\text{m}}}\cdot\frac{1}{\sqrt{\text{W}}}\)). This number is consistent with the theoretical value of 2.79, which assumes a “true” overcoupling regime where an increase of the cavity bandwidth (coupling coefficient) occurs without change of the unloaded Q-value, i.e., field distributions within the cavity remain constant (see SI and [10]):

$$\frac{{H}_{x}(\mathrm{0,0},0)}{{H}_{x}^{0}(\mathrm{0,0},0)}=\sqrt{2-\frac{\Delta {\nu }_{1/2}^{0}}{\Delta {\nu }_{1/2}}}\cdot \sqrt{\frac{\Delta {\nu }_{1/2}^{0}}{\Delta {\nu }_{1/2}}}.$$
(1)

However, if the short is adjusted to the critical coupling position, a value of \(\Delta {\nu }_{1/2}=\) 26.4 MHz and \({H}_{x}(\mathrm{0,0},0)\) = 640 \(\frac{\text{A}}{{\text{m}}}\cdot\frac{1}{\sqrt{\text{W}}}\) is obtained from both the S11 and \({H}_{x}\) frequency plots. This cavity bandwidth corresponds to \({Q}_{U}=2580\), which is about half the \({Q}_{U}\) without a dielectric plate, i.e., the power losses increase by a factor of 2. Additional losses in the system originate from changes in the magnetic field distribution, which cause an increase in current density at the cavity walls and iris, and dielectric losses in the dielectric plate. The additional losses from two sources are about equal, i.e., \({Q}_{cond}=10800\) and \({Q}_{diel}=\) 10,700 for materials used in the model. It is thus desirable to be able to introduce and remove the dielectric plate to optimize performance in both the critically coupled (CW-EPR) and overcoupled (pulse EPR) regime.

To understand the action of the dielectric plate in more detail, we calculated the dependence of the S11 coefficient on the dielectric plate thickness. Figure 4a shows the reflection coefficients in the frequency range of 27–42 GHz, represented as contour plots. For thin dielectric plates, well-isolated TE211, TE011 and TE311 modes are observed. The self-resonance frequency of the iris is shown using a dashed line. It initially appears off-scale, decreasing as the thickness of the dielectric plate increases, approaching the resonance frequency of the TE011 mode (34 GHz), at a thickness of 0.32 mm. Simultaneously, the TE011 mode begins to couple with both the TE211 mode and particularly the TE311 mode to produce mixed TE-modes. This is readily observed in the calculated magnetic field magnitude distributions, with the TE011 mode gaining TE311 character (see Fig. S6 in SI). An interesting situation is established as the iris self-resonance approaches the frequency of TE011 mode. Typical mode passage is observed with the closest approach of the two modes at DT = 0.34 mm, which is close to the 0.32 mm required for an isolated iris self-resonance frequency of 34 GHz.

Fig. 4
figure 4

a Dependence of the reflection coefficient, \(\left|\Gamma \right|\) on the dielectric plate thickness, shown as a contour plot, calculated for a circular iris diameter of 2.6 mm and short position of L = − 10.6 mm. The solid lines show the dependencies of cavity resonances below 42 GHz. The dashed curve shows the calculated resonance frequency of the circular iris in a WR28 waveguide. The horizontal dotted line indicates the TE011 mode resonance frequency of 34 GHz at DT = 0 mm. The vertical dotted line indicates the dielectric thickness DT = 0.34 mm, i.e., the point that most closely approaches the resonance modes indicated by red and yellow solid lines. b Reflection coefficient and c x-component of the magnetic field amplitude at the resonator center for a cavity with circular iris of 2.6 mm calculated for the critically coupled case without a dielectric plate, and for the maximally coupled case with dielectric plates of different thicknesses

From the perspective of optimizing EPR performance, the two significant parameters are the resonator bandwidth and the amplitude of the microwave magnetic field coincident with the sample, and thus only these parameters were analyzed. The reflection coefficient and x-component of the magnetic field amplitude were calculated at selected thicknesses of the dielectric plate between 0.20 and 0.34 mm and the results are shown in Fig. 4b, c and summarized in Table 2. The key observation to be made here is that the reduction in the maximum magnetic field amplitude at the sample position is lower than expected. The calculation for DT = 0.34 mm yields a reduction factor of 6.2, which is smaller compared to the factor calculated from bandwidths assuming a “true” overcoupling regime, i.e., 7.3; see Eq. (1) and SI. Thus, the mode mixing described above is advantageous, acting to concentrate the magnetic field in the region of space occupied by the sample.

Table 2 Parameters of cavity models for different thicknesses of dielectric plate

3.2 Experimental Results

The reliability of the electromagnetic simulations described above was tested experimentally. Cavity configurations, both without a dielectric plate and with plates of different thickness fixed at the iris, were investigated at room temperature and low temperature (20 K). A comparison of the cavity ringdown measurements and microwave nutation experiments with the electromagnetic simulations are shown in Fig. 5, and summarized in Table 3. The simulation parameters were adjusted to match the properties and dimensions of the real probehead, e.g., the iris diameter was reduced to 2.5 mm in accordance with the real cavity.

Fig. 5
figure 5

a Experimental mw power profile recorded for a cavity without a dielectric plate, applying a mw pulse of 200 ns at room temperature. The flat part during the pulse yields a return loss of 20 dB (|Γ|= 0.1). The linear approximation of the power decay after the mw pulse is switched off (dashed line) yields a power decay time of 11.5 ± 0.2 ns (voltage ringing time \({\tau }_{V}\) =23 ± 0.5 ns), which corresponds to a bandwidth of \(\Delta {\nu }_{1/2}=\frac{1}{\pi {\tau }_{V}}\) =13.8 ± 0.3 MHz. b Frequency dependence of the relative microwave nutation frequencies at room temperature for (i) a critically coupled cavity without a dielectric plate (blue dots) and (ii) for a maximally overcoupled cavity with a dielectric plate of DT = 0.25 mm (green dots) and 0.29 mm (black dots). The lines show the best fit result of the experimental data to the square root of a Lorentzian function. The dashed line indicates the level of 0.707. The absolute values of the π-pulse length obtained at the maximum evaluable mw power are summarized in Table 3

Table 3 Experimental and calculated cavity parameters

The parameters for a critically coupled cavity without a dielectric plate at room temperature are used as the reference parameters. A cavity bandwidth of 13.8 ± 0.6 MHz was evaluated from the ringdown trace and frequency dependence of the microwave nutation frequency (Fig. 4). This value is in a good agreement with the calculated value of 13.9 MHz, verifying the model parameters. The amplitude of the microwave magnetic field can be compared using the length of the π-pulse, \({t}_{p}(\pi)\), which is directly evaluated from the microwave nutation traces at cavity resonance. A value of \({t}_{p}(\pi)\)=15.6 ± 1 ns was easily achieved at the full microwave power (see SI for bridge outputs). The maximum mw magnetic field amplitude, \({H}_{x}\), is converted to the expected pulse length as \({t}_{p}\left(\pi \right)=\frac{2\pi }{\sqrt{W}{\mu }_{0}{\gamma }_{e}{H}_{x}}\), with \(W\), \({\mu }_{0}\) and \({\gamma }_{e}\) being the mw power at the cavity, permeability of a vacuum, and electron gyromagnetic ratio respectively. The theoretical value of \({t}_{p}\left(\pi \right)\) is 10 ns (\({H}_{x}(\mathrm{0,0},0)\)=890 \(\frac{\text{A}}{{\text{m}}}\cdot\frac{1}{\sqrt{\text{W}}}\) and \(W\)=10 W), which is about a factor of 1.5 lower than the experimental one. This discrepancy probably stems from the overestimation of mw power at the cavity which cannot be exactly determined experimentally. Therefore, for subsequent comparisons, we used the experimental and calculated pulse length normalized to the pulse length for a critically coupled cavity without a dielectric plate (Table 3).

Good agreement is seen between the calculated and experimental values in the overcoupling regime. Calculations predict slightly higher bandwidths at maximum overcoupling, which is most likely due to overestimation of the dielectric constant (Rogers 3010 material, \({\upvarepsilon}_{\text{r}}\)=10.7). If a slightly lower value (10.4) is used, which lies within the manufacturer’s specifications, a better match between calculation and experiment is obtained. The behavior of the magnetic field amplitude, \({t}_{p}\left(\pi \right),\) perfectly agrees with the behavior of the cavity bandwidth, see Table 3.

Agreement is less good between calculated and experimental bandwidths (and magnetic field amplitudes) in the critically coupled regime, which increases as a function of the dielectric thickness. This is likely due to the Rogers laminate material having a significantly higher dielectric loss factor than assumed. The calculation uses the loss tangent specified at 10 GHz (\(\mathrm{tan}\delta =2.3\cdot 1{0}^{-3}\)). There is no reliable data for \(\mathrm{tan}\delta\) at 34 GHz and recent studies have shown that Rogers laminate materials show significantly higher dielectric loss factors at high frequencies than those specified at lower frequencies (< 10 GHz) [11]. If the loss factor is increased to \(\mathrm{tan}\delta =1{0}^{-2}\), the calculated bandwidths for the critically coupled cavity are in good agreement with the experimental values.

3.3 Coupling Mechanism

The experimental and simulation results demonstrate that the cavity bandwidth can be increased without compromising the pulsed EPR performance; the resonance properties of the circular iris can be “tuned” using a dielectric material positioned at the front of the iris. While the dielectric plate coupling scheme allows for large, tunable overcoupling and therefore a large bandwidth, the use of a fixed plate has several disadvantages in the context of a general purpose cavity. Firstly, the resonance properties of the iris become fixed, which leads to the fixed, distorted-mode profile in the cylindrical cavity described above. The mixed-mode character results in higher losses in the cavity, reducing the amplitude of the microwave magnetic field at the sample position, which diminishes its performance in CW EPR experiments. Secondly, the very high iris transmittance requires precise adjustment of the short position to accurately set the critical coupling position. This leads to the cavity coupling being highly sensitive to mechanical vibrations and pressure variations in the cryostat, again compromising the resonator performance for CW EPR.

To overcome these limitations, we propose a coupling mechanism in which the dielectric plate is fixed onto the movable short, rather than onto the iris. The waveguide-cavity transmission has a periodic dependence on the distance of the short from the iris (Fig. 2c and Fig. S02 in the SI); hence, a short-fixed dielectric plate of sufficient length should allow for two interchangeable regimes of operation. These regimes are illustrated by the simulations in Fig. 6, calculated for a dielectric plate of 10 mm length (DT = 0.3 mm) attached to the movable short. Positioning the short between -14 mm to -11 mm from the iris center would allow the coupling of the waveguide–iris–cavity configuration to be varied without influence from the dielectric plate, i.e., without the dielectric plate changing the iris properties. This first regime does not allow for overcoupling, but is advantageous for critical coupling of a high-Q cavity. The dielectric plate starts to interact with the iris and increases the iris transmittance from short positions above -11 mm. This second regime is suitable for controlled overcoupling of the cavity, as required for pulse EPR experiments. This position can also be used for critically coupling a cavity that is loaded with a sample with substantial dielectric losses. In this case, the power losses from the sample will overcome the power losses caused by the dielectric plate, and mode distortions can be neglected.

Fig. 6
figure 6

a Dependence of the reflection coefficient, \(\left|\Gamma \right|\) on the position of the short carrying the dielectric plate (DT = 0.3 mm) of 10 mm length. The yellow dashed line shows the behavior of the cavity operation mode. b The cavity bandwidth evaluated from S11 curves. Two regimes of operation are indicated and visualized in (c)

Experimental results up to this point suggest that dielectric laminates are not an ideal material for the dielectric coupling plate. Their dielectric constant is strongly temperature-dependent and the laminate material displays poor mechanical properties. Ongoing work of our laboratories is the manufacture of a new short coupler made from sapphire. Sapphire has good mechanical properties and, importantly, low dielectric losses (\(\mathrm{tan}\delta\) ~10–5 at room temperature, ~ 10–7 at temperatures below 100 K) [12] such that they can be safely neglected. The choice of crystal cut allows us to choose between two temperature-stable dielectric constants (9.4 and 11.6 at room temperature) [12].

3.4 Alternative Iris Geometries

This work has so far only considered a circular iris. However, the original description of the TE011 cavity published in [5] used a slot-type iris. This raises the question: does the slot-type iris pose any advantages over a circular iris? The slot-type iris has the “native” geometry for the TE01 wave transfer in a rectangular waveguide. The self-resonance of the slot can be tuned to a particular frequency by choosing the slot geometry. The resonance frequency of the narrow rectangular slot is approximately equal to the cutoff frequency of a waveguide having the same cross-sectional shape as the aperture [13]. Thus, an empirical relationship for a resonant slot of height, \(h\), and width \(w\), centered in the transverse plane of a rectangular waveguide for small \(w/h\), reduces to \(h\approx {\lambda }_{0}/2\). For a narrow rounded slot, some correction has to be made, \(h\approx \frac{{\lambda }_{0}}{2}+0.273\cdot w\) [14]. This results in a rounded slot height of 4.55 mm for a self-resonance frequency at 34 GHz. Simulations predict an optimal height of h = 4.6 mm for \(w=\) 0.5 mm, which closely matches the above estimation; see Figs. S7 in the SI. The maximum iris transmission coefficient is very high, |\(T\)|= 1. However, the transmission curve of the isolated resonant slot is very broad, giving a bandwidth of 14.3 GHz, which efficiently couples at least all three fundamental modes, TE211, TE011 and TE311. This results in unwanted effects for the critically coupled cavity, i.e., the maximum magnetic field amplitude in mixed modes, and \({H}_{x}\), reduces by about a factor of 3 as compared to the small round aperture coupling. A non-resonant rounded slot of \(hw\)=3.0 × 0.5 mm2 on the other hand, has the same transmittance at 34 GHz as compared to the circular iris of 2.6 mm diameter (Fig. S9 in SI). The self-resonance frequency of this slot is about 50.8 GHz, but this frequency can also be successfully downshifted by means of a dielectric plate. There are, however, two crucial differences between the 3.0 × 0.5 mm2 slot iris and its 2.6 mm circular counterpart. Firstly, due to the higher electric field in the slot iris (higher capacitance), it requires significantly less dielectric (Fig. S8). For an isolated 3.0 × 0.5 mm2 slot iris, a dielectric plate of only 0.098 mm thickness will match the self-resonance frequency to 34 GHz. This problem, however, can be avoided by using dielectric materials with smaller dielectric constant. Secondly, the transmission bandwidth for the slot iris is 4.6 GHz, almost two times smaller than the 8.7 GHz transmission bandwidth of the circular iris. Additionally, the transmission properties of the slot iris were found to be more critically dependent on system parameters such as geometry of the slot, iris wall thickness, geometry of the dielectric plate and its dielectric constant. In practice, this means that a real slot iris coupling system would be less predictable than the circular iris. A significant advantage of the circular iris is also the simplicity and achievable precision of its fabrication, especially when extending the applicability of the coupling mechanism to higher microwave frequencies. At 95 GHz (W-band) or even 263 GHz, the circular iris of 0.9 mm diameter or 0.33 mm, respectively can be easily machined mechanically. The precise manufacturing of the slots (1.07 × 0.18 mm2 for 95 GHz, 0.39 × 0.06 mm2 for 263 GHz, which are scaled from the Q-band slot dimensions) becomes challenging for mechanical machining and would require more advanced fabrication techniques.

4 Conclusions

We have demonstrated that the resonance properties of the iris can be effectively manipulated using dielectric plates. The modifications to the general purpose Q-band probehead described in this work allow the precise adjustment of the iris’ transmission toward the TE011 cavity, and enable the construction of new types of coupling mechanisms. The new mechanism is based on the combined action of a short-type coupler and dielectric plate coupler, and operates in two regimes which allow both critical coupling of a high-Q cavity without diminishing CW EPR performance, as well as large cavity overcoupling to > 1 GHz bandwidths. This greatly expands the versatility of the original probehead, now fulfilling the bandwidth requirements for a wide range of different EPR experiments including CW EPR, pulse EPR, pulse ENDOR and DEER.

The possibility to manipulate the resonance properties of different iris types opens further directions for the development of new coupling mechanisms for different types of cavities. We have already developed such a mechanisms for end waveguide to cavity configurations, e.g., Fabry–Perot Q-band, cylindrical TE011 W-band and rectangular TE102 X-band cavities. At high microwave frequencies in particular, e.g., W-band and above, such coupling mechanisms allow for more flexible cavity applications compared to other types of iris-based couplings such as those described in [15,16,17,18]. We plan to describe these findings in forthcoming publications.