Ceramic femoral head fracture with massive metallosis and severe wear of the cone: a case report and review of the literature

  • A PanagopoulosEmail author
  • P Megas
  • M Tyllianakis
  • P Dimakopoulos
  • E Lambiris
Case Report

Abstract

A case of a fractured ceramic femoral head without subsequent trauma, 14 months after a primary total hip arthroplasty, in a 40-year-old woman with hip dysplasia is reported. Intraoperative findings were multifragment femoral head and a damaged polyethylene insert along with diffuse metallosis and excessive wear of the cone of the stem. Both the stem and the acetabular component were stable. After the removal of ceramic fragments, metallotic tissue excision and careful lavage of the joint, the inlay was replaced by a similar one and a cobalt–chromium femoral head was placed to the existing notched taper of the firmly incorporated stem. At 22-month follow-up examination, the patient had no pain, used no walking aids and had normal activity.

Keywords

Fractured ceramic head Metallosis Cone wear Hip arthroplasty 

Fracture de tête fémorale en céramique avec métallose massive

Résumé

Nous rapportons un cas de fracture de tête céramique survenue sans notion traumatique 14 mois après une arthroplastie totale de hanche de première intention pour dysplasie chez une jeune femme de 40 ans. L’exploration per-opératoire a mis en évidence une fracture multifragmentaire de la tête, un insert polyéthylène endommagé, une métallose diffuse et une usure du cône métallique. La tige et l’implant acétabulaire étaient stables. Après ablation des fragments de céramique, excision de la métallose, et un lavage soigneux de l’articulation, un insert du même type a été mis en place et une tête métallique placée sur le cône de la tige. Au recul de 22 mois le patient est asymptomatique, marche sans canne et mène une activité normale.

Mots clés

fracture de tête céramique métallose usure du cône arthroplastie de hanche 

References

  1. 1.
    Allain J, Goutallier D, Voisin MC, Lemouel S (1998) Failure of a stainless steel femoral head of a revision total hip arthroplasty performed after a fracture of a ceramic femoral head. A case report. J Bone Joint Surg 80-A:1355–1360Google Scholar
  2. 2.
    Allain J, Roudot-Thoraval F, Delecrin J, Anract P, Migaud H, Goutalier D (2003) Revision total hip arthroplasty performed after fracture of a ceramic femoral head. J Bone Joint Surg 85-A:825–830PubMedGoogle Scholar
  3. 3.
    Arenas A, Tejero A, Garbayo A, Arias C, Barbadillo J, Lecumberri J (1999) Ceramic femoral head fractures in total hip replacement. Int Orthop 23:351–352CrossRefPubMedGoogle Scholar
  4. 4.
    Burckhardt A, Berberat C (1993) How safe are ceramic heads as hip endoprostheses? A series of three head fractures within 3 months. Arch Orthop Trauma Surg 112:215–219CrossRefPubMedGoogle Scholar
  5. 5.
    Callaway GH, Flynn W, Ranawat CS, Sculco TP (1995) Fracture of the femoral head after ceramic-on-polyethylene total hip arthroplasty. J Arthroplasty 10:855–859CrossRefPubMedGoogle Scholar
  6. 6.
    Connes H (2003) Des déclarations à la matériovigilance aux décisions de la CNM. Revue de Chirurgie Orthopédique 89:49Google Scholar
  7. 7.
    Dambreville A (2003) Conduite chirurgicale pratique proposée aux chirurgiens confrontés au problème. Revue de Chirurgie Orthopédique 89:51Google Scholar
  8. 8.
    Davidson JA (1993) Characteristics of metal and ceramic total hip bearing surfaces and their effect on long-term ultra high molecular weight polyethylene wear. Clin Orthop 294:361–378PubMedGoogle Scholar
  9. 9.
    Fritsch E, Gleitz M (1996) Ceramic femoral head fractures in total hip arthroplasty. Clin Orthop 328:129–136CrossRefPubMedGoogle Scholar
  10. 10.
    Hannouche D, Nich C, Bizot P, Meunier A, Nizard R, Sedel L (2003) Fractures of ceramic bearings. Clin Orthop 417:19–26PubMedGoogle Scholar
  11. 11.
    Higuchi F, Shiba N, Inoue A, Wakebe I (1995) Fracture of an alumina ceramic head in total hip arthroplasty. J Arthroplasty 10:851–854CrossRefPubMedGoogle Scholar
  12. 12.
    Holmer P, Nielsen PT (1993) Fracture of a ceramic femoral head in total hip arthroplasty. J Arthroplasty 8:567–571CrossRefPubMedGoogle Scholar
  13. 13.
    Kempf I, Semlitsch M (1990) Massive wear of a steel ball head by ceramic fragments in the polyethylene acetabular cup after revision of a total hip prosthesis with fractured ceramic ball. Arch Orthop Trauma Surg 109:284–287CrossRefPubMedGoogle Scholar
  14. 14.
    Krikler S, Schatzker J (1995) Ceramic head failure. J Arthroplasty 10:860–862CrossRefPubMedGoogle Scholar
  15. 15.
    Mangione P, Pascarel X, Vinciguerra B, Honton JL (1994) Fracture of the ceramic head in total hip prosthesis. Apropos of 2 cases. Int Orthop 18:359–362CrossRefPubMedGoogle Scholar
  16. 16.
    Matziolis G, Perka C, Dish A (2003) Massive metallosis after revision of a fractured ceramic head onto a metal head. Arch Orthop Trauma Surg 123:48–50PubMedGoogle Scholar
  17. 17.
    McLean CR, Dabis H, Mok D (2002) Delayed fracture of the ceramic femoral head after trauma. J Arthroplasty 17:503–504CrossRefPubMedGoogle Scholar
  18. 18.
    Michaud RJ, Rashad SY (1995) Spontaneous fracture of the ceramic ball in a ceramic–polyethylene total arthroplasty. J Arthroplasty 10:863–867CrossRefPubMedGoogle Scholar
  19. 19.
    Moreschini O, Braidotti P (1991) Traumatic breakage of the ceramic head of prosthesis in total hip replacement. Ital J Orthop Traumatol 17:339–343PubMedGoogle Scholar
  20. 20.
    Nizard RS, Sedel L, Christel P, Meunier A, Soudry M, Witvoet J (1992) Ten-year survivorship of cemented ceramic–ceramic total hip prosthesis. Clin Orthop 282:53–63PubMedGoogle Scholar
  21. 21.
    Otsuka NY, Schatzker J (1994) A case of fracture of a ceramic head in total hip arthroplasty. Arch Orthop Trauma Surg 113:81–82CrossRefPubMedGoogle Scholar
  22. 22.
    Peiro A, Pardo J, Navarrete R, Rodríguez-Alonso L, Martos F (1991) Fracture of the ceramic head in total hip arthroplasty. Report of two cases. J Arthroplasty 6:371–374PubMedCrossRefGoogle Scholar
  23. 23.
    Piconi C, Labanti M, Magnani G, Caporale M, Maccauro G, Magliocchetti G (1999) Analysis of a failed alumina THR ball head. Biomaterials 20:1637–1646CrossRefPubMedGoogle Scholar
  24. 24.
    Pulliam I, Trousdale R (1997) Fracture of a ceramic femoral head after a revision operation. J Bone Joint Surg 79-A:118–121PubMedGoogle Scholar
  25. 25.
    Wenda K, Ritter G, Rudig L (1993) Therapy of burst ceramic head of a hip prosthesis. Aktuelle Traumatol 23:294–296PubMedGoogle Scholar
  26. 26.
    Willmann G (2000) Ceramic femoral head retrieval data. Clin Orthop 379:22–28PubMedCrossRefGoogle Scholar

Copyright information

© Springer-Verlag 2005

Authors and Affiliations

  • A Panagopoulos
    • 1
    Email author
  • P Megas
    • 1
  • M Tyllianakis
    • 1
  • P Dimakopoulos
    • 1
  • E Lambiris
    • 1
  1. 1.Orthopaedic DepartmentPatras University HospitalAktaio-RioGreece

Personalised recommendations