European Spine Journal

, Volume 16, Issue 7, pp 1055–1061

Metal levels in corrosion of spinal implants

Original Article

Abstract

Corrosion affects spinal instrumentations and may cause local and systemic complications. Diagnosis of corrosion is difficult, and nowadays it is performed almost exclusively by the examination of retrieved instrumentations. We conducted this study to determine whether it is possible to detect corrosion by measuring metal levels on patients with posterior instrumented spinal fusion. Eleven asymptomatic patients, with radiological signs of corrosion of their stainless steel spinal instrumentations, were studied by performing determinations of nickel and chromium in serum and urine. Those levels were compared with the levels of 22 patients with the same kind of instrumentation but without evidence of corrosion and to a control group of 22 volunteers without any metallic implants. Statistical analysis of our results revealed that the patients with spinal implants without radiological signs of corrosion have increased levels of chromium in serum and urine (P < 0.001) compared to volunteers without implants. Corrosion significantly raised metal levels, including nickel and chromium in serum and urine when compared to patients with no radiological signs of corrosion and to volunteers without metallic implants (P < 0.001). Metal levels measured in serum have high sensibility and specificity (area under the ROC curve of 0.981). By combining the levels of nickel and chromium in serum we were able to identify all the cases of corrosion in our series of patients. The results of our study confirm that metal levels in serum and urine are useful in the diagnosis of corrosion of spinal implants and may be helpful in defining the role of corrosion in recently described clinical entities such as late operative site pain or late infection of spinal implants.

Keywords

Corrosion Spinal implants Metal levels 

References

  1. 1.
    Akazawa T, Minami S, Takahashi K, Kotani T, Hanawa T, Moriya H (2005) Corrosion of spinal implants retrieved from patients with scoliosis. J Orthop Sci 10:200–205PubMedCrossRefGoogle Scholar
  2. 2.
    Aulisa L, di Benedetto A, Vinciguerra A, Lorini G, Tranquilli-Leali P (1982) Corrosion of the Harrington’s instrumentation and biological behaviour of the rod-human spine system. Biomaterials 3:246–248PubMedCrossRefGoogle Scholar
  3. 3.
    Bago J, Ramirez M, Pellise F, Villanueva C (2003) Survivorship analysis of Cotrel–Dubousset instrumentation in idiopathic scoliosis. Eur Spine J 12:435–439PubMedCrossRefGoogle Scholar
  4. 4.
    Barceloux DG (1999) Chromium. J Toxicol Clin Toxicol 37:173–194PubMedCrossRefGoogle Scholar
  5. 5.
    Barceloux DG (1999) Nickel. J Toxicol Clin Toxicol 37:239–258PubMedCrossRefGoogle Scholar
  6. 6.
    Beguiristain J, del Rio J, Duart J, Barroso J, Silva A, Villas C (2006) Corrosion and late infection causing delayed paraparesis after spinal instrumentation. J Pediatr Orthop B 15:320–323PubMedGoogle Scholar
  7. 7.
    Clark CE, Shufflebarger HL (1999) Late-developing infection in instrumented idiopathic scoliosis. Spine 24:1909–1912PubMedCrossRefGoogle Scholar
  8. 8.
    Cook S, Asher M, Lai SM, Shobe J (2000) Reoperation after primary posterior instrumentation and fusion for idiopathic scoliosis. Toward defining late operative site pain of unknown cause. Spine 25:463–468PubMedCrossRefGoogle Scholar
  9. 9.
    Doorn PF, Mirra JM, Campbell PA, Amstutz HC (1996) Tissue reaction to metal on metal total hip prostheses. Clin Orthop Relat Res 329:S187–S205PubMedCrossRefGoogle Scholar
  10. 10.
    Dubousset J, Shufflebarger H, Wenger D (1994) Late “infection” with C-D instrumentation. Orthop Trans 18:121Google Scholar
  11. 11.
    Granchi D, Cenni E, Ciapetti G, Savarino L, Stea S, Gamberini S, Gori A, Pizzoferrato A (1998) Cell death induced by metal ions: necrosis or apoptosis? J Mater Sci Mater Med 9:31–37PubMedCrossRefGoogle Scholar
  12. 12.
    Hahn F, Zbinden R, Min K (2005) Late implant infections caused by Propionibacterium acnes in scoliosis surgery. Eur Spine J 14:783–788PubMedCrossRefGoogle Scholar
  13. 13.
    Hallab N, Merrit K, Jacobs JJ (2001) Metal sensitivity in patients with orthopaedic implants. J Bone Joint Surg Am 83-A:428–436PubMedGoogle Scholar
  14. 14.
    Harrington PR (1962) Treatment of scoliosis. Correction and internal fixation by spine instrumentation. J Bone Joint Surg Am 44:591–610PubMedGoogle Scholar
  15. 15.
    Jacobs JJ, Gilbert JL, Urban RM (1998) Corrosion of metal orthopaedic implants. J Bone Joint Surg Am 80:268–282PubMedGoogle Scholar
  16. 16.
    Jacobs JJ, Roebuck KA, Archibeck M, Hallab NJ, Glant TT (2001) Osteolysis: basic science. Clin Orthop Relat Res 393:71–77PubMedCrossRefGoogle Scholar
  17. 17.
    Jacobs JJ, Skipor AK, Campbell PA, Hallab NJ, Urban RM, Amstutz HC (2004) Can metal levels be used to monitor metal-on-metal hip arthroplasties? J Arthroplasty 19:S59–S65CrossRefGoogle Scholar
  18. 18.
    Jacobs JJ, Urban RM, Gilbert JL, Skipor AK, Black J, Jasty M, Galante JO (1995) Local and distant products from modularity. Clin Orthop Relat Res 319:94–105PubMedGoogle Scholar
  19. 19.
    Kim YJ, Kassab F, Berven SH, Zurakowski D, Hresko MT, Emans JB, Kasser JR (2005) Serum levels of nickel and chromium after instrumented posterior spinal arthrodesis. Spine 30:923–926PubMedCrossRefGoogle Scholar
  20. 20.
    Kraft CN, Burian B, Diedrich O, Wimmer MA (2001) Implications of orthopedic fretting corrosion particles on skeletal muscle microcirculation. J Mater Sci Mater Med 12:1057–1062PubMedCrossRefGoogle Scholar
  21. 21.
    MacDonald SJ, Brodner W, Jacobs JJ (2004) A consensus paper on metal ions in metal-on-metal hip arthroplasties. J Arthroplasty 19:S12–S16CrossRefGoogle Scholar
  22. 22.
    MacDonald SJ, McCalden RW, Chess DG, Bourne RB, Rorabeck CH, Cleland D, Leung F (2003) Metal-on-metal versus polyethylene in hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res 406:282–296PubMedCrossRefGoogle Scholar
  23. 23.
    Muschik M, Luck W, Schlenzka D (2004) Implant removal for late-developing infection after instrumented posterior spinal fusion for scoliosis: reinstrumentation reduces loss of correction. A retrospective analysis of 45 cases. Eur Spine J 13:645–651PubMedCrossRefGoogle Scholar
  24. 24.
    Richards BR, Emara KM (2001) Delayed infections after posterior TSRH spinal instrumentation for idiopathic scoliosis: revisited. Spine 26:1990–1996PubMedCrossRefGoogle Scholar
  25. 25.
    Takahashi S, Delecrin J, Passuti N (2001) Intraspinal metallosis causing delayed neurologic symptoms after spinal instrumentation surgery. Spine 26:1495–1498PubMedCrossRefGoogle Scholar
  26. 26.
    Tezer M, Kuzgun U, Hamzaoglu A, Ozturk C, Kabukcuoglu F, Sirvanci M (2005) Intraspinal metalloma resulting in late paraparesis. Arch Orthop Trauma Surg 125:417–421PubMedCrossRefGoogle Scholar
  27. 27.
    Vieweg U, van Roost D, Wolf HK, Schyma CA, Schramm J (1999) Corrosion on an internal spinal fixator system. Spine 24:946–951PubMedCrossRefGoogle Scholar
  28. 28.
    Waelchli B, Min K, Cathrein P, Boos N (2002) Vertebral body compression fracture after removal of pedicle screws: a report of two cases. Eur Spine J 11:504–506PubMedCrossRefGoogle Scholar
  29. 29.
    Wimmer C, Gluch H (1998) Aseptic loosening after CD instrumentation in the treatment of scoliosis: a report about eight cases. J Spinal Disord 11:440–443PubMedGoogle Scholar

Copyright information

© Springer-Verlag 2007

Authors and Affiliations

  • Javier del Rio
    • 1
  • Jose Beguiristain
    • 1
  • Julio Duart
    • 1
  1. 1.Department of Orthopedic Surgery and TraumatologyUniversity Clinic of NavarraPamplona, NavarraSpain

Personalised recommendations