Platelet measurements in rat, dog and mouse blood samples using the Sysmex XT-2000iV

  • R. A. Mathers
  • G. O. Evans
  • J. Bleby
Original Article

Abstract

In this study, blood platelet measurements were evaluated for three laboratory species—rat, dog and mouse—using the Sysmex XT-2000iV haematology analyser, which offers two alternative methods for counting platelets—electrical impedance and a combination of light scattering and fluorescence. The analytical performance was satisfactory in terms of imprecision, carry-over between samples and linearity. The rat platelet counts were stable for at least 24 h at 4 °C, but not at room temperature: There were marked reductions at 24 and 48 h for dogs at both 4 °C and room temperature. Mean platelet volumes increased with time for both rats and dogs. Using the XT-2000iV, results were obtained for platelet counts (×109/L) together with the mean platelet volume (femtolitres), platelet distribution width (femtolitres), platelet large cell ratio (percent) and plateletcrit (percent), and these were compared with results obtained with the Siemens Advia 120™ with its multispecies software. Differences were observed for platelet counts and other platelet measurements made with the two analysers and between counts made by light scattering vs. fluorimetry. Some of these differences are due to the technologies employed in gating the cell populations and expressions used to describe the platelet populations.

Keywords

Haematology analyser Sysmex XT-2000iV Advia 120 Rat blood Dog blood Mouse blood Animal platelets Platelet parameters 

References

  1. Borkataky S, Jain R, Gupta R et al (2009) Role of platelet volume indices in the differential diagnosis of thrombocytopenia: a simple and inexpensive method. Hematology 14:182–186PubMedCrossRefGoogle Scholar
  2. Bowles KM, Cooke LJ, Richards EM et al (2005) Platelet size has diagnostic predictive value in patients with thrombocytopenia. Clin Lab Haematol 27:370–373PubMedCrossRefGoogle Scholar
  3. Briggs C, Kunka S, Machin SJ (2004) The most accurate platelet count on the Sysmex XE-2100. Optical or impedance? Clin Lab Haematol 26:157–158PubMedCrossRefGoogle Scholar
  4. Briggs C, Harrison P, Machin S (2007) Continuing developments with the automated platelet count. Int J Lab Haematol 29:77–91CrossRefGoogle Scholar
  5. Buttarello M, Plebani M (2008) Automated blood cell counts: state of the art. Am J Clin Pathol 130:104–116PubMedCrossRefGoogle Scholar
  6. Eason CT, Pattison A, Howells DD et al (1986) Platelet population profiles: significance of species variation and drug-induced changes. J Appl Toxicol 6:437–441PubMedCrossRefGoogle Scholar
  7. Endler G, Klimesch A, Sunder-Plassmann H et al (2002) Mean platelet volume is an independent risk factor for myocardial infarction but not for coronary artery disease. Br J Haematol 117:399–404PubMedCrossRefGoogle Scholar
  8. Evans GO, Smith DE (1991) Platelet measurements in healthy Beagles, Canis familiaris. Comp Haem Int 1:49–51CrossRefGoogle Scholar
  9. Farias MG, Schunck EG, Bo SD et al (2010) Definition of reference ranges for the platelet distribution width (PDW): a local need. Clin Chem Lab Med 48:255–257PubMedCrossRefGoogle Scholar
  10. Furlanello T, Tasca S, Caldin M et al (2006) Artifactual changes in canine blood following storage, detected using the Advia 120 hematology analyzer. Vet Clin Pathol 35:42–46PubMedCrossRefGoogle Scholar
  11. Giacomini A, Legovini P, Antico F et al (2001) Evaluation of platelet analysis on the Advia 120 hematology system. Lab Hematol 7:180–185Google Scholar
  12. Hayashi Y, Matsuzawa T, Unno T et al (1995) Effects of haematology parameters during cold storage and cold transport of rat and dog blood samples. Comp Haematol Int 5:251–255CrossRefGoogle Scholar
  13. Hekimsoy Z, Payzin B, Ornek T et al (2004) Mean platelet volume in type 2 diabetic patients. J Diabetes Complications 18:1688–1691CrossRefGoogle Scholar
  14. ICSH (International Council for the Standardization in Haematology) (1994) Guidelines for the evaluation of blood cell analysers including those used for differential leucocyte and reticulocyte counting and cell marker applications. Clin Lab Haematol 16:157–174Google Scholar
  15. International Council for Standardization in Haematology Expert panel on Cytometry and International Society of Laboratory Hematology task force on platelet counting (2001) Platelet counting by the RBC/platelet ratio method. Am J Clin Pathol 115:460–464CrossRefGoogle Scholar
  16. Italiano JE, Lecine P, Shivdasani RA et al (1999) Blood platelets are assembled principally at the ends of proplatelet processes produced by differentiated megakaryocytes. J Cell Biol 147:1299–1312PubMedCrossRefGoogle Scholar
  17. Jackson SR, Carter JM (1993) Platelet volume: laboratory measurement and clinical application. Blood Rev 7:104–113PubMedCrossRefGoogle Scholar
  18. Kaito K, Otsubo H, Usui N et al (2005) Platelet size deviation width, platelet large cell ratio and mean platelet volume have sufficient sensitivity and specificity in the diagnosis of immune thrombocytopenia. Br J Haem 128:698–702CrossRefGoogle Scholar
  19. Kling G and Zelmanovic D (2004) Advia multispecies PLT analysis. Presentation at Huddinge, SwedenGoogle Scholar
  20. Kim HK, Kim J-E, Ham CK et al (2008) Prognostic value of platelet indices as determined by Advia 120 in patients suspected of having disseminated intravascular coagulation. Int J Lab Hem 30:117–123CrossRefGoogle Scholar
  21. Mathers RA, Evans GO, Bleby J (2012) Reticulocyte measurements in rat, dog and mouse whole blood samples using the Sysmex XT-2000iV. Comp Clin Pathol. doi: 10.1007/s00580-010-1147-6
  22. Mathers RA, Evans GO, Bleby J et al (2008) Evaluation of the Sysmex XT-2000iV haematology analyser for rat, dog and mouse whole blood samples. Comp Clin Pathol 17:137–144CrossRefGoogle Scholar
  23. Ntaios G, Papadopoulos A, Chatzinikolaou A et al (2008) Increased values of mean platelet volume and platelet size deviation width may provide a safe positive diagnosis of Idiopathic Thrombocytopenic Purpura. Acta Haematol 119:173–177PubMedCrossRefGoogle Scholar
  24. O’Malley T, Langhorne P, Elton RA et al (1995) Platelet size in stroke patients. Am Heart Assoc 26:995–999Google Scholar
  25. Pankraz A, Bauer N, Moritz A (2009) Comparison of flow cytometry with the Sysmex XT2000iV automated analyzer for the detection of reticulated platelets in dogs. Vet Clin Pathol 38:30–38PubMedCrossRefGoogle Scholar
  26. Prins M, van Leeuwen MW, Teske E (2009) Stability and reproducibility of Advia-120 measured red blood cell and platelet parameters in dogs, cats and horses, and the use of reticulocyte haemoglobin (Chr) in the diagnosis of iron deficiency. Tijdschr Diergeneeskd 134:272–278PubMedGoogle Scholar
  27. Reardon DM, Hutchinson D, Preston FE et al (1985) The routine measurement of platelet volume: a comparison of aperture-impedance and flow cytometric systems. Clin Lab Haematol 7:251–257PubMedCrossRefGoogle Scholar
  28. Thompson CB, Diaz DD, Quinn PG et al (1983) The role of anticoagulant in the measurement of platelet volumes. Am J Clin Pathol 80:327–332PubMedGoogle Scholar
  29. Threatte GA (1993) Usefulness of the mean platelet volume. Clinics Lab Med 13:937–950Google Scholar
  30. Zandecki M, Genevieve F, Gerard J et al (2006) Spurious counts and spurious results on haematology analysers: a review. Part 1: platelets. Int J Lab Hem 29:4–20Google Scholar
  31. Zelmanovic D, Collela GM, Hetherington EJ et al (1998) Automated method and devices for identifying and quantifying platelets and determining platelet activation state using whole blood samples. US Patent Number 5,817,519 October 6,1998Google Scholar
  32. Zucker-Franklin D (1997) Platelet structure and function. In: Kuter D, Hunt P, Sheridan W, Zucker-Franklin D (eds) Thrombopoiesis and thrombopoietins. Humana, Totowa, pp 41–62CrossRefGoogle Scholar

Copyright information

© Springer-Verlag London Limited 2012

Authors and Affiliations

  1. 1.AstraZeneca R+D CharnwoodLoughboroughUK
  2. 2.Sysmex UK LtdMilton KeynesUK

Personalised recommendations