Comparison between orogastric tube/bougie and a suction calibration system for effects on operative duration, staple-line corkscrewing, and esophageal perforation during laparoscopic sleeve gastrectomy
- 1.1k Downloads
- 1 Citations
Abstract
Introduction
Placement of a bougie for sleeve sizing during laparoscopic sleeve gastrectomy (LSG) is recommended. We compared this standard with a suction calibration system (SCS) that performs all functions with one insertion, and measured each step’s duration.
Methods
Primary LSG was performed using a bougie and SCS in alternating order. Number of tube movements to achieve optimal placement, durations of decompression, leak testing, and overall operative time, and remnant linear measurements were obtained.
Results
LSG was performed in 26 patients (15 women, 11 men; mean age 36.8 years; mean BMI 45.3 kg/m2). The mean number of tube movements was significantly greater for the bougie than for the SCS (8.13 vs. 3.58; p < 0.0001). Percent reductions achieved using the SCS were: time to full decompression of the stomach, 62 % (21 vs. 8 s; p < 0.138); tube placement, 51 % (101 vs. 49 s; p < 0.0001); leak testing, 78 % (119 vs. 26 s; p < 0.0003); and mean operative duration (from tube insertion to end of stapling), 21 % (875 vs. 697 s; p < 0.019). Variance of the staple-line distance, measured from the greater curvature to the staple line, was 1.64 and 0.92 for the bougie and SCS, respectively, indicating a reduction in corkscrewing, for a 43.9 % straighter sleeve.
Conclusion
SCS maintained the gastric wall in place, thereby preventing corkscrewing, and reducing total operating time. Reducing the number of tube insertions may prevent esophageal damage and accidental tube stapling.
Keywords
Sleeve gastrectomy Calibration system Bougie ViSiGi 3D™ Corkscrewing DelineationA critical step in the internationally recognized laparoscopic sleeve gastrectomy (LGS) is ensuring sleeve-size consistency. According to the International Sleeve Gastrectomy Expert Panel Consensus Statement [1], 100 % of the surgeons surveyed the sleeve size with a bougie. However, clinical disadvantages of using a bougie include accidental stapling across the tip, esophageal perforation, and staple-line corkscrewing.
With super-obese patients, visualization of a bougie or an intraluminal tube placement is often difficult (e.g., orogastric [OG] tube and temperature probe). Most surgeons rely on tactile feedback or ask the anesthesiologist to move the bougie to confirm placement. This difficulty in locating and securing intraluminal tubes inside bariatric patients often leads to accidental stapling across these tubes. A multicenter study reported the rate of this complication to be 1 in 132 [2]. Such accidents could result in conversion of LSG to gastric bypass, or even leakage. Also, the tungsten-filled and gravity-driven bougie has a tendency to spring away from the lesser curvature with the patient in reverse Trendelenburg position. Often the anesthesiologist has to hold the bougie in place throughout the stapling of the stomach.
LSG also requires the insertion of an OG tube, for stomach decompression before sizing and leak testing after sizing. Multiple tubes traversing the esophagus of a patient in reverse Trendelenburg position amplify the risk of esophageal perforation, which can result in treatment costs of over $100,000 [3].
Surgeons are instructed to keep LSG staple lines within the same frontal plane to avoid sleeve corkscrewing and uneven anterior and posterior traction on the staple line, both of which can result in a functional obstruction not easily remedied endoscopically [4]. Before each staple firing, surgeons often flip the stomach back and forth to check for excess tissue on the posterior side. Sliding tissues around the bougie can cause inconsistent tension on the anterior or posterior side of the stomach during and after stapling.
The patient safety profile can be improved by reducing the number of steps and clinical risks with LSG. The aim of this study was to compare aspects of the procedure, including visual confirmation of device position, intraoperative duration of all steps, staple line, and number of device movements to achieve optimal placement, between current practice and a new calibration system, the ViSiGi 3D™ (Boehringer Labs, LLC, Phoenixville, PA, USA), which employs a safe level of suction and performs all functions with one insertion.
Materials and methods
This was a single-center, single-surgeon prospective study. Informed consent was obtained from all individual participants included in the study. Primary LSG was performed using either a 40-Fr bougie or a 40-Fr suction calibration system (SCS), the ViSiGi 3D™, in alternating order, with the decision to use either device in the first case being determined at random. Sex, age, body mass index (BMI), and comorbidity information were recorded for each patient.
Time measurement criteria for each step involving an intraluminal tube
| Measurement | Start | End |
|---|---|---|
| Stomach decompression | When anesthesiologist was inserting the OG tube or SCS. In the case of anesthesiologist already placed OG tube before insufflation, the decompression time starts if the surgeon needs additional time/measure to ensure complete decompression | When the surgeon noted that the stomach was fully decompressed |
| Positioning time | When the anesthesiologist was instructed to advance the SCS, or to remove the OG tube and insert the bougie | When the surgeon was ready to fire the first staple load |
| Stapling time | When the stapler was inserted | When last staple load was fired |
| Leak testing | When the anesthesiologist was instructed to remove the bougie and insert the OG tube, or to inject methylene blue into the SCS | When the anesthesiologist was instructed to remove all intraluminal tubes |
To determine the staple-line straightness, three measurements were taken at three predetermined locations on the inflated, excised gastric specimen: circumference, and distances from the greater curvature to the staple line anteriorly and posteriorly (Figs. 2, 3). The three predetermined locations were the midpoint of the staple line and the points 5 cm above and below the midpoint. The variance of deviation from the midline at these locations was calculated for each specimen. To standardize the amount of inflation, each specimen was connected to the same gas source for 5 s through an insufflation needle.
The number of device movements was recorded in both groups. This included the number of intraluminal devices inserted, as well as the number of times the surgeon inquired about the status/location of the devices; e.g., asking the anesthesiologist “Is it in?” or “Is it out?” or to move or hold the device in place, or to insert or remove any intraluminal tube, and any repositioning of the device.
Survey questions. (A) ViSiGi case #2, (B) bougie case #1, (C) bougie case #23, (D) ViSiGi case #4, (E) bougie case #12, (F) ViSiGi case #9, (G) ViSiGi case #13, (H) bougie case #6, (I) ViSiGi case #19, (J) bougie case #14
Statistical analysis
Statistical analysis was performed with a two-sample t test assuming unequal variance. In addition, staple-line straightness analysis was performed with one-way analysis of variance (ANOVA) using StatPlus® software (AnalystSoft, Inc., Alexandria, VA, USA).
Diagram depicting measurement to determine straightness of staple line. Measurements of the circumference, anterior, and posterior distance were taken at three locations along the inflated excised gastric specimen: location 1: 5 cm above the midpoint of the staple line, location 2: midpoint of the staple line, location 3: 5 cm below the midpoint of the staple line
Measuring the straightness of the staple line
Staple lines. (A) Corkscrew and (B) straight staple lines
New technology
(A) One-step ViSiGi 3D™ versus multi-step orogastric tube/bougie system. (B) Circumferential fenestration pattern enabling flexibility of the ViSiGi 3D™. (C) Integral regulator safely decompresses the stomach at 125 mmHg regulated suction
Results
Patient demographics
| Female (N = 15) | Male (N = 11) | |
|---|---|---|
| Min BMI | 39.0 | 35.0 |
| Max BMI | 61.0 | 60.6 |
| Mean BMI ± SD | 45.7 ± 6.7 | 44.7 ± 8.2 |
| Min age | 22 | 14 |
| Max age | 61 | 74 |
| Mean age | 36 ± 10.8 | 38 ± 17.0 |
Survey responses
| “In which of the two images below can you better identify the calibration system along the lesser curvature (bougie, etc.)?” | ||
|---|---|---|
| Left | Right | |
| Question 1 | 96.3 % (ViSiGi) | 3.7 % (Bougie) |
| Question 2 | 1.1 % (Bougie) | 98.9 % (ViSiGi) |
| Question 3 | 7.0 % (Bougie) | 93.0 % (ViSiGi) |
| Question 4 | 81.6 % (ViSiGi) | 18.4 % (Bougie) |
| Question 5 | 87.5 % (ViSiGi) | 12.5 % (Bougie) |
Mean time to complete each step of sleeve gastrectomy with an intraluminal tube
| Mean time (sec) | Bougie | SCSa | % Save | p value |
|---|---|---|---|---|
| Fully decompressed stomach | 21.0 | 8.0 | 61.9 | 0.138 |
| Tube placement | 101.0 | 48.6 | 51.9 | 0.000 |
| Each staple firing | 137.6 | 119.5 | 13.2 | 0.085 |
| Leak test | 119.4 | 26.5 | 77.8 | 0.000 |
| Total operative time (excluding complications, hernia, oversew, etc.) | 875.3 | 697.2 | 20.3 | 0.019 |
Summary of staple line straightness measurements
| Location 1—top | Location 2—middle | Location 3—bottom | Mean variance ± SD | ||||||
|---|---|---|---|---|---|---|---|---|---|
| D anterior | D posterior | V top | D anterior | D posterior | V middle | D anterior | D posterior | V tottom | |
| Bougie (cm) | |||||||||
| 12.0 | 10.5 | 1.5 | 8.5 | 9.5 | −1.0 | 6.0 | 4.0 | 2.0 | 2.85 ± 0.83 |
| 13.5 | 12.0 | 1.5 | 9.0 | 8.5 | 0.5 | 6.0 | 5.0 | 1.0 | 0.25 ± 0.5 |
| 8.0 | 9.0 | 0.1 | 11.0 | 10.0 | 1.0 | 13.0 | 10.5 | 2.5 | 3.08 ± 1.76 |
| 14.5 | 12.5 | 2.0 | 11.0 | 9.0 | 2.0 | 9.5 | 9.5 | 0.0 | 1.33 ± 1.15 |
| 12.0 | 12.0 | 0.0 | 8.5 | 10.5 | −2.0 | 7.0 | 8.0 | −1.0 | 1.00 ± 1.0 |
| 12.0 | 11.0 | 1.0 | 9.0 | 9.0 | 0.0 | 7.5 | 9.0 | −1.5 | 1.58 ± 1.26 |
| SCS (cm) | |||||||||
| 12.0 | 10.8 | 1.2 | 11.2 | 9.0 | 2.2 | 7.7 | 8.8 | −1.1 | 2.86 ± 1.69 |
| 10.0 | 12.5 | −2.5 | 5.0 | 10.0 | −2.0 | 5.0 | 6.5 | −1.5 | 0.25 ± 0.5 |
| 10.8 | 10.2 | 0.6 | 6.5 | 8.0 | 1.5 | 7.5 | 5.5 | 2.0 | 0.50 ± 0.71 |
| 13.0 | 11.5 | 1.5 | 10.0 | 8.0 | 2.0 | 6.0 | 4.0 | 2.0 | 0.08 ± 0.29 |
The number of device movements for both groups was recorded. The average number of tube movements was 8.1 (range 8–11) for the OG-bougie-OG system versus 3.6 (range 3–4) for the ViSiGi 3D™ (p < 0.0001). Typical bougie movements included introducing the OG tube for stomach decompression, removing the OG tube, inserting the bougie for sleeve sizing, surgeon asking the question: “Is there anything else inside the stomach?” holding the bougie in place during stapling, removing the bougie, inserting the OG tube for leak testing, and removing the OG tube.
Discussion
According to the most recent International Sleeve Gastrectomy Expert Panel Consensus statement [5], stricture formation should be avoided during sleeve gastrectomy. Additionally, among concepts found to be most important to experts: 75 % felt that “maintaining symmetrical lateral traction while stapling will reduce potential strictures”; 82 % felt the “use of a bougie when stapling the incisura angularis will result in decreased incidence of strictures,” and all experts surveyed agreed that “the incisura angularis is a potential stricture site.”
However, a bougie used for calibration has no internal stability. If gastric tissues are overextended during stapling, the elasticity of the tissues will lead to tissue retraction, and may result in stricture, especially near the incisura angularis. Uniform circumferential suction allows the anterior and posterior walls to envelope the SCS, which guides staple placement, decreasing the incidence of strictures.
With a 1 % incidence, stricture is the second most prevalent early complication, and most symptomatic within 6 weeks postoperatively [6, 7, 8, 9, 10, 11, 12]. Treatment options vary, and some authors are proponents of a silicone-covered nitinol stent in an attempt to cause scarred permanent dilatation. One author has advocated several forms of gastroplasty to keep the sleeve intact, such as a stapled end-to-end gastrogastrostomy with transoral anvil passage, local excision with a vagus-preserving, hand-sewn stapled end-to-end gastrogastrostomy, or a transverse linear gastroplasty. However, Roux-en-Y gastric bypass remains the best surgical treatment for persistent strictures after endoscopic treatments have failed.
Corkscrewing of the staple line has been discussed at meetings and in books [4, 15]; but this form of functional kink or spiraling has not been well studied. Because of gastric tissue elasticity, it is possible to apply uneven tension on the anterior and posterior walls as the posterior stomach is examined for redundant tissue during stapling. Our results suggest that the ViSiGi 3D™ provides controlled and uniform suction, which may enable symmetrical lateral traction, thus preventing corkscrewing. This can potentially decrease the incidence of dysphagia and malnutrition, including water-soluble vitamin deficiencies. Prevention is preferable to excision of a corkscrewed sleeve and conversion to Roux-en-Y gastric bypass.
Perforation, which can occur if bougie or the tube used is too rigid, is a complication to both patients and hospitals [3]. The ViSiGi 3D™ incorporates a fenestration pattern that makes the distal tip more flexible, which may contribute to reducing the risk of perforation. Also, with the OG-bougie-OG system, the anesthesiologist has to insert and remove three intraluminal tubes during the surgery. Thus, consolidation of these three tasks into one using the ViSiGi 3D™ confers an additional threefold reduction in the risk of perforation.
Additionally, intraluminal tubes such as temperature probes, OG tubes, and bougies have been inadvertently transected during stapling and creation of the sleeve [2, 13, 14]. At any time during surgery, a bougie can slip out of the unfinished sleeve, after which it can be easily transected proximally. Accidental stapling of these tubes can cause significant morbidity and require prolonged operating time for repair. In certain cases, postoperative complications may include stricture or leak. A calibration system that is easily visualized and stays anchored along the lesser curvature provides a clinical advantage. In our survey of surgeons, 91.5 % found the ViSiGi 3D™ enabled clearer, more direct visualization of the device and made the stapling process easier and safer.
By integrating multiple steps into one, the ViSiGi 3D™ significantly reduced the operating time needed to complete each step of a sleeve gastrectomy and therefore the total operating time. Based on the clinical benefits found in the present study, we recommend incorporating the ViSiGi 3D™ into the LSG protocol.
Our study objectively examined and compared the degree of staple-line spiraling in LSG. Larger studies are needed to investigate the effect of suction on corkscrewing of the staple line. Despite our small sample size, the preliminary results of the present study appear encouraging. In addition, the limit of this study includes the inability to blind the surgeon to whether the patient was receiving a bougie or a SCS, which could potentially introduce bias. Due to the mechanism of SCS, which requires the surgeon to instruct the anesthesiologist turning on and off suction, the surgeon needs to know which device was using ahead of time. The blinded portion of the study was the survey conducted during IFSO. To ensure objectivity of the interpretation, the 273 surgeons were neither informed about the study nor aware of a new device. They were simply asked the question: “In which of the two images below can you better identify the calibration system along the lesser curvature (bougie, etc.)?”
Notes
Acknowledgments
This study was support by a research Grant from Boehringer Laboratories, LLC.
Compliance with Ethical Standards
Disclosures
Rose Huang is a clinical research and development engineer with Boehringer Laboratories, LLC. Dr. Gagner has received personal and other fees from Boehringer Laboratories, LLC, during the conduct of the study and personal fees from Ethicon, Covidien, Gore, Olympus, Transenterix, and MID outside the scope of the present study. Dr. Gagner has no equity or proprietary interest in the device.
Ethical Approval
All procedures in studies involving human participants were performed in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
References
- 1.Rosenthal RJ, International Sleeve Gastrectomy Expert Panel, Diaz AA, Arvidsson D, Baker RS, Basso N et al (2012) International Sleeve Gastrectomy Expert panel consensus statement: best practice guidelines based on experience of >12,000 cases. Surg Obes Relat Dis 8(1):8–19CrossRefPubMedGoogle Scholar
- 2.Abu-Gazala S, Donchin Y, Keidar A (2012) Nasogastric tube, temperature probe, and bougie stapling during bariatric surgery: a multicenter survey. Surg Obes Relat Dis 8(5):595–600CrossRefPubMedGoogle Scholar
- 3.Oyasiji T, Lujic M, Suarez L, Federico J, Katigbak M (2012) Cost of treatment due to perforation, esophageal perforation: etiology, outcome and cost analysis over a decade in a community teaching hospital. Surg Endosc 26(Suppl 1):249–430Google Scholar
- 4.Parachand V (2013) Chapter 28: laparoscopic duodenal switch and sleeve gastrectomy. In: Swanstrom L, Soper N (eds) Mastery of endoscopic and laparoscopic surgery. Lippincott Williams & Wilkins, PhiladelphiaGoogle Scholar
- 5.Gagner M, Deitel M, Erickson AL, Crosby RD (2013) Survey on laparoscopic sleeve gastrectomy (LSG) at the fourth international consensus summit on sleeve gastrectomy. Obes Surg 23(12):2013–2017CrossRefPubMedGoogle Scholar
- 6.Sarkhosh K, Birch DW, Sharma A, Karmali S (2013) Complications associated with laparoscopic sleeve gastrectomy for mordbid obesity: a Surgeon’s guide. Can J Surg 56(5):347–352CrossRefPubMedPubMedCentralGoogle Scholar
- 7.Vilallonga R, Himpens J, van de Vrande S (2013) Laparoscopic management of persistent strictures after laparoscopic Sleeve Gastrectomy. Obes Surg 23(10):1655–1661CrossRefPubMedGoogle Scholar
- 8.Burgos AM, Csendes A, Braghetto I (2013) Gastric stenosis after laparoscopic sleeve gastrectomy in morbidly obese patients. Obes Surg 23(9):1481–1486CrossRefPubMedGoogle Scholar
- 9.Parikh A, Alley JB, Peterson RM, Harnisch MC, Pfluke JM, Tapper DM et al (2012) Management options for symptomatic stenosis after laparoscopic vertical sleeve gastrectomy in the morbidly obese. Surg Endosc 26(3):738–746CrossRefPubMedGoogle Scholar
- 10.Scheffel O, Weiner RA (2011) Therapy of stenosis after sleeve gastrectomy: stent and surgery as alternatives—case reports. Obes Facts 4(Suppl 1):47–49CrossRefPubMedGoogle Scholar
- 11.Jones M, Healey AJ, Efthimiou E (2011) Early use of self-expanding metallic stents to relieve sleeve gastrectomy stenosis after intragastric balloon removal. Surg Obes Relat 7(5):e16–e17CrossRefGoogle Scholar
- 12.Zundel N, Hernandez JD, Galvao Neto M, Campos J (2010) Strictures after laparoscopic sleeve gastrectomy. Surg Laparosc Endosc Percutan Tech 20(3):154–158CrossRefPubMedGoogle Scholar
- 13.Péquignot A, Dhahria A, Mensah E, Verhaeghe P, Badaoui R, Sabbagh C et al (2011) Stapling the section of the nasogastric tube during sleeve gastrectomy: How to prevent and recover? Case Rep Gastroenterol 5(2):350–354CrossRefPubMedPubMedCentralGoogle Scholar
- 14.Meshikhes AW, Al-Saif OH (2014) Iatrogenic oesophageal transection during laparoscopic sleeve gastrectomy. BMJ Case Rep 3:2014Google Scholar
- 15.Szomstein S (2013) The Surgeon’s Lounge. General surgery news. (2013) Jan [cited 2015 Jan]; Vol 40. Available from: http://www.generalsurgerynews.com/ViewArticle.aspx?d=The+Surgeons+Lounge&d_id=72&i=January+2013&i_id=923&a_id=22381
Copyright information
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.