Advertisement

Surgical Endoscopy

, Volume 22, Issue 2, pp 430–435 | Cite as

Miniature bending manipulator for fetoscopic intrauterine laser therapy to treat twin-to-twin transfusion syndrome

  • H. YamashitaEmail author
  • K. Matsumiya
  • K. Masamune
  • H. Liao
  • T. Chiba
  • T. Dohi
Article

Abstract

Background

Recent typical therapy for twin-to-twin transfusion syndrome (TTTS) is selective laser photocoagulation of anastomotic communicating vessels on the placenta using the fetoscopic approach. The difficulty of a conventional laser device approach for this procedure depends significantly on the placental location, so a new robotized device is required to bend the direction of laser irradiation flexibly within the narrow uterus.

Methods

The authors designed a miniature bending mechanism impelled by a wire-guided linkage driving method that provides a stable procedure for bending laser irradiation from −90° to 90°. Using this bending mechanism, the authors developed a bending manipulator with a diameter of 3.5 mm and a hollow central channel with a diameter of 0.8 mm for passing a glass fiber for neodymium:yttrium–aluminum-garnet (Nd:YAG) laser photocoagulation. The bending mechanism is motorized by an electrical actuator and controlled by a grip-type interface with a small joystick. The robotized tip’s part and the actuator’s part are easily separable for cleaning and sterilization.

Results

In performance evaluations of the manipulator, the bending characteristics with a glass fiber were examined. The bending range was −52.6° to 80°, with a very small hysteresis error, and the bending repeatability error was 0.5° ± 0.2°, which corresponds with the high accuracy of 0.2 ± 0.1-mm positioning error at the glass fiber’s tip. In the evaluation of Nd:YAG laser photocoagulation, the study confirmed that the manipulator performed effective laser photocoagulation of the placental phantom surface (underwater chicken liver). The large bending range, reaching 80°, enabled a flexible approach from various directions with a high irradiation efficiency of no less than 96.6%.

Conclusions

The authors’ original miniature bending manipulator can change the laser irradiating direction with highly repeatable positioning accuracy for speedy, safe, and effective vessel occlusion in clinical practice.

Keywords

Miniature bending manipulator Selective laser photocoagulation Twin-to-twin transfusion syndrome 

Notes

Acknowledgments

A part of this work was supported by Health and Labour Sciences Research Grants in 2005 (research on medical devices for analyzing, supporting, and substituting the function of human body; H17-Physi-006), by FY2005 JSPS Grants-in-Aid for Scientific Research Development of Next Generation End Effecter and Navigation System for Computer Aided Surgery (17100008), and by Research Fellowships of the Japan Society for the Promotion of Science for Young Scientists (041400000227).

References

  1. 1.
    Abbou CC, Hoznek A, Salomon L, Olsson LE, Lobontiu A, Saint F, Cicco A, Antiphon P, Chopin D (2001) Laparoscopic radical prostatectomy with a remote controlled robot. Urology 165: 1964–1966CrossRefGoogle Scholar
  2. 2.
    Arata J, Mitsuishi M, Warisawa S, Tanaka K, Yoshizawa T, Hashizume M (2005) Development of a dexterous minimally-invasive surgical system with augmented force feedback capability. Proceedings of the. 2005 IEEE/RSJ International Conference on Intelligent Robots and Systems, pp 3207–3212Google Scholar
  3. 3.
    Dario P, Carrozza MC, Marcacci M, D’Attanasio S, Magnami B, Tonet O, Megali G (2000) A novel mechatronic tool for computer-assisted arthroscopy. IEEE Trans Inform Technol Biomed 4: 15–28CrossRefGoogle Scholar
  4. 4.
    Evans MI, Adzick NS, Holzgreve W, Harrison MR (2001) The unborn patient: the art and science of fetal therapy, 3rd ed. Saunders, PhiladelphiaGoogle Scholar
  5. 5.
    Feldstein VA, Machin GA, Albanese CT, Sandberg P, Farrell JA, Farmer DL, Harrison MR (2000) Twin-twin transfusion syndrome: the “select” procedure. Fetal Diagn Ther 15: 257–261CrossRefPubMedGoogle Scholar
  6. 6.
    Ikuta K, Sasaki K, Yamamoto K, Shimada T (2002) Remote microsurgery system for deep and narrow space: development of new surgical procedure and microrobotic tool. Proceedings of the 5th International Conference on Medical Image Computing and Computer-Assisted Intervention, pp 163–172Google Scholar
  7. 7.
    Peirs J, Reynaerts D, Van Brussel H (2001) A miniature manipulator for integration in a self-propelling endoscope. Sens Actuators A: 343–349Google Scholar
  8. 8.
    Quintero RA, Comas C, Bornick PW, Allen MH, Kruger M (2000) Selective versus nonselective laser photocoagulation of placental vessels in twin-to-twin transfusion syndrome. Ultrasound Obstet Gynecol 16: 230–236CrossRefPubMedGoogle Scholar
  9. 9.
    Senat MV, Deprest J, Boulvain M, Paupe A, Winer N, Ville Y (2004) Endoscopic laser surgery versus serial amnioreduction for severe twin-to-twin transfusion syndrome. N Engl J Med. 351: 136–144CrossRefPubMedGoogle Scholar
  10. 10.
    Yamashita H, Hata N, Hashizume M, Dohi T (2004) Handheld laparoscopic forceps manipulator using multislider linkage mechanisms. Lect Notes Computer Sci 3217: 121–128CrossRefGoogle Scholar
  11. 11.
    Yamashita H, Kim D, Hata N, Dohi T (2003) Multislider linkage mechanism for endoscopic forceps manipulator. Proceedings of the 2003 IEEE/RSJ International Conference on Intelligent Robots and Systems, pp 2577–2582Google Scholar

Copyright information

© Springer Science+Business Media, LLC 2007

Authors and Affiliations

  • H. Yamashita
    • 1
  • K. Matsumiya
    • 1
  • K. Masamune
    • 1
  • H. Liao
    • 2
  • T. Chiba
    • 3
  • T. Dohi
    • 1
  1. 1.Department of Mechano-Informatics, Graduate School of Information Science and TechnologyThe Univercity of TokyoTokyoJapan
  2. 2.Department of Precision Engineering, Graduate School of EngineeringThe University of TokyoTokyoJapan
  3. 3.Department of Strategic MedicineNational Center for Child Health and DevelopmentTokyoJapan

Personalised recommendations