Evaluation of a minimally invasive surgical fixation technique for young children with the Concerto Pin cochlear implant system
- 196 Downloads
- 2 Citations
Abstract
In 2011, Med-El (Innsbruck, Austria) introduced a new cochlear implant system, designed to require a minimally invasive surgical technique and allow greater positional flexibility for its fixation on the skull. The Concerto Pin implant system is a good option for patients with thinner bone, such as children and elderly. The aim of this study was to investigate the implant’s stability in children using our minimally invasive surgical technique. This was a prospective, longitudinal study with a single-subject, repeated-examination design. Six children, implanted with a Concerto Pin using our minimally invasive surgical technique between October 2011 and September 2012, were assessed 1, 3 and 6 months after surgery. In each case, the implant remained in a stable position and no adverse events or problems with healing were observed at any time during the investigation. The minimally invasive technique and the method of implant fixation that bypass drilling a deep implant bed constitute a good option for patients with thinner bone, such as children. This clinical study shows the safety and stability of the Concerto Pin implant system using a minimally invasive surgical technique.
Keywords
Cochlear implant Minimally invasive surgery Concerto Pin ChildrenNotes
Conflict of interest
All authors declare that the manuscript has not been published previously or under review by another journal. The paper has not been presented to any professional society. Furthermore, all authors declare that no financial interest, commercial financial support or personal relationship with other people or organizations that could influence this work exist.
References
- 1.Alexander NS, Caron E, Woolley AL (2011) Fixation methods in pediatric cochlear implants: retrospective review of an evolution of 3 techniques. Otolaryngol Head Neck Surg 144:427–430PubMedCrossRefGoogle Scholar
- 2.Cuda D (2013) A simplified fixation of the new thin cochlear implant receiver-stimulators in children: long term results with the “back-pocket” technique. Int J Pediatr Otorhinolaryngol 77:1158–1161PubMedCrossRefGoogle Scholar
- 3.Adunka OF, Buchman CA (2007) Cochlear implant fixation in children using periosteal sutures. Otol Neurotol 28:768–770PubMedCrossRefGoogle Scholar
- 4.Cuda D (2009) Cochlear implantation with Pulsar Med El: a novel small incision technique. Acta Otorhinolaryngol Ital 29:76–78PubMedCentralPubMedGoogle Scholar
- 5.Dalchow CV, Werner JA (2005) A new instrument for minimal access surgery in cochlear implantation. Otol Neurotol 26:678–679PubMedCrossRefGoogle Scholar
- 6.Davis BM, Labadie RF, McMenomey SO, Haynes DS (2004) Cochlear implant fixation using polypropylene mesh and titanium screws. Laryngoscope 114:2116–2118PubMedCrossRefGoogle Scholar
- 7.Molony TB, Giles JE, Thompson TL, Motamedi KK (2010) Device fixation in cochlear implantation: is bone anchoring necessary? Laryngoscope 120:1837–1839PubMedCrossRefGoogle Scholar
- 8.Djalilian HR, King T, Faust RA, Smith S, Levine SC (2001) Securing cochlear implants to the skull: two alternate methods. Ear Nose Throat J 80:171–173PubMedGoogle Scholar
- 9.Guevara N, Sterkers O, Bebear JP, Meller R, Magnan J, Mosnier I, Amstutz I, Lerosey Y, Triglia JM, Roman S, Gahide I (2010) Multicenter evaluation of the digisonic SP cochlear implant fixation system with titanium screws in 156 patients. Ann Otol Rhinol Laryngol 119:501–505PubMedCrossRefGoogle Scholar
- 10.Gekeler J, Luers JC, Lang-Roth R, Beutner D (2013) Cochlear implant fixation: are sutures really necessary? Eur Arch Otorhinolaryngol 270:483–487PubMedCrossRefGoogle Scholar
- 11.Guldiken Y, Orhan KS, Yigit O, Basaran B, Polat B, Gunes S, Acoglu E, Deger K (2011) Subperiosteal temporal pocket versus standard technique in cochlear implantation: a comparative clinical study. Otol Neurotol 32:987–991PubMedCrossRefGoogle Scholar
- 12.Costain N, Cavanagh J, Bance M, Morris DP, Hong P (2013) A musculoperiosteal flap cochlear implant device fixation technique: our experience in one hundred and seventy nine patients. Clin Otolaryngol 38:86–89PubMedCrossRefGoogle Scholar
- 13.Balkany TJ, Whitley M, Shapira Y, Angeli SI, Brown K, Eter E, Van De Water T, Telischi FF, Eshraghi AA, Treaba C (2009) The temporalis pocket technique for cochlear implantation: an anatomic and clinical study. Otol Neurotol 30:903–907PubMedCrossRefGoogle Scholar
- 14.James AL, Papsin BC (2004) Device fixation and small incision access for pediatric cochlear implants. Int J Pediatr Otorhinolaryngol 68:1017–1022PubMedCrossRefGoogle Scholar
- 15.de Varebeke SP, Govaerts P, Cox T, Deben K, Ketelslagers K, Waelkens B (2012) Fixation of cochlear implants: an evidence-based review of literature. B-ENT 8:85–94PubMedGoogle Scholar
- 16.Schnabl J, Markl A, Hormann R, Wolf-Magele A, Schartinger V, Sprinzl G (2012) Concerto Pin: a novel concept of cochlear implant fixation. Otol Neurotol 33:1525–1529PubMedCrossRefGoogle Scholar
- 17.Loh C, Jiang D, Dezso A, Fitzgerald O’Connor A (2008) Non-sutured fixation of cochlear implants using a minimally-invasive approach. Clin Otolaryngol 33:255–264Google Scholar
- 18.Otto RA, Lane AP, Carrasco VN (1999) A new technique for securing cochlear implants. Otolaryngol Head Neck Surg 120:897–898PubMedCrossRefGoogle Scholar
- 19.O’Donoghue GM, Nikolopoulos TP (2002) Minimal access surgery for pediatric cochlear implantation. Otol Neurotol 23:891–894PubMedCrossRefGoogle Scholar
- 20.Campisi P, Hayward L, Papsin B (2005) Mitek QuickAnchor fixation of cochlear implants using a minimal access technique. Int J Pediatr Otorhinolaryngol 69:1655–1658PubMedCrossRefGoogle Scholar
- 21.Black B (2009) Keyhole cochlear implantation surgery. Cochlear Implants Int 10:150–159PubMedCrossRefGoogle Scholar
- 22.Jethanamest D, Channer GA, Moss WJ, Lustig LR, Telischi FF (2013) Cochlear implant fixation using a subperiosteal tight pocket without either suture or bone-recess technique. Laryngoscope. doi: 10.1002/lary.24426
- 23.Boscolo-Rizzo P, Muzzi E, Barillari MR, Trabalzini F (2011) Non-sutured fixation of the internal receiver-stimulator in cochlear implantation. Eur Arch Otorhinolaryngol 268:961–965PubMedCrossRefGoogle Scholar
- 24.Dirr F, Hempel JM, Krause E, Muller J, Berghaus A, Ertl-Wagner B, Braun T (2013) Value of routine plain X-ray position checks after cochlear implantation. Otol Neurotol 34:1666–1669PubMedCrossRefGoogle Scholar
- 25.Bremke M, Luers JC, Stenner M, Bovenschulte H, Dorn F, Unkel C, Hollering J, Beutner D (2013) Radiologic examinations in human temporal bone specimens using digital volume tomography and high-resolution computed tomography after implantation of middle ear prosthesis and cochlear implant electrode array. Otol Neurotol 34:1321–1328PubMedCrossRefGoogle Scholar