Abstract
Introduction
Lag screw cut-out is one of the major postoperative complications on femoral trochanteric fractures. However, precise analyses of excessive sliding and lag screw cut-out were limited. The purpose of this study was to investigate the factors that induce this unfavorable event.
Methods
From April 2010 to April 2013, 226 patients were operated in our institute using a short femoral nail. Among them, 177 patients (29 males and 148 females) with a mean age of 84 years (60–97 years), who were followed up >3 months, were included in this study. The postoperative sliding distance, fracture type (AO/OTA classification), tip-apex distance (TAD), reduction pattern in the postoperative X-ray (antero-posterior and lateral views), bone quality (canal flare and cortical indices), walking ability at the time of pre-injury and final follow-up, and complications were investigated retrospectively.
Results
The mean sliding distance was 3.7 mm, and one cut-out case (0.6 %) was observed. The sliding distance of the AO/OTA 31-A2 fractures was significantly longer than that of the A1 fractures (p < 0.0001). Regarding the reduction pattern, the sliding distance of the intramedullary type was longer than that of the extramedullary and anatomical types in the lateral view (p < 0.01, p < 0.001, respectively). Further, even in the medial and anatomical types, where the reduction patterns are recommended in the antero-posterior view, the sliding distance of the intramedullary type was significantly longer than that of the anatomical type in the lateral view. There was no correlation of bone quality with the sliding distance.
Conclusions
Because the sliding distance increased in the intramedullary type in the lateral view of unstable fractures, an accurate reduction in the lateral view at surgery is important, particularly in unstable fractures.
Similar content being viewed by others
References
Orimo H, Yaegashi Y, Onoda T, Fukushima Y, Hosoi T, Sakata K (2009) Hip fracture incidence in Japan: estimates of new patients in 2007 and 20-year trends. Arch Osteoporos 4:71–77
Japanese Orthopaedic Association (JOA) (2011) Clinical practice guideline on the diagnosis. Treatment of femoral neck and trochanteric fractures, 2nd edn. Nankodo, Tokyo, p 26 (In Japanese)
Williams WW, Parker BC (1992) Complications associated with the use of the gamma nail. Injury 23:91–92
Norris R, Bhattacharjee D, Parker MJ (2012) Occurrence of secondary fracture around intramedullary nails used for trochanteric hip fractures: a systematic review of 13,568 patients. Injury 43:706–711
Buciuto R, Hammer R (2001) RAB-plate versus sliding hip screw for unstable trochanteric hip fractures: stability of the fixation and modes of failure—radiographic analysis of 218 fractures. J Trauma 50:545–550
Parker MJ (1996) Trochanteric hip fractures. Fixation failure commoner with femoral medialization, a comparison of 101 cases. Acta Orthop Scand 67:329–332
Marsh JL, Slongo TF, Agel J, Broderick JS, Creevey W, DeCoster TA, Prokuski L, Sirkin MS, Ziran B, Henley B, Audigé L (2007) Fracture and dislocation classification compendium—2007: orthopaedic Trauma Association classification, database and outcomes committee. J Orthop Trauma 21:S1–133
Choo SK, Oh HK, Choi JY (2010) PFNA and PFN in intertrochanteric fractures—comparison study of sliding. J Korean Hip Soc 22:79–85
Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM (1995) The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am 77:1058–1064
Noble PC, Alexander JW, Lindahl LJ, Yew DT, Granberry WM, Tullos HS (1988) The anatomic basis of femoral component design. Clin Orthop Relat Res 235:148–165
Gruen T (1997) A simple assessment of bone quality prior to hip arthroplasty: cortical index revised. Acta Orthop Belg 63(Suppl 1):20–27
Salvati EA, Wilson PD Jr (1973) Long-term results of femoral-head replacement. J Bone Joint Surg Am 55:516–524
Davis TR, Sher JL, Horsman A, Simpson M, Porter BB, Checketts RG (1990) Intertrochanteric femoral fractures. Mechanical failure after internal fixation. J Bone Joint Surg Br 72:26–31
Larsson S, Friberg S, Hasson LI (1990) Trochanteric fractures. Influence of reduction and implant position on impaction and complications. Clin Orthop Relat Res 259:130–139
Ebbinghaus S, Mjöberg B (1991) Posterior angulation in trochanteric fractures detected with roentgen stereophotogrammetry. Ups J Med Sci 96:235–237
Tsukada S, Okumura G, Matsueda M (2012) Posterior stability on lateral radiographs in the surgical treatment of pertrochanteric hip fractures. Arch Orthop Trauma Surg 132:839–846
Carr JB (2007) The anterior and medial reduction of intertrochanteric fractures: a simple method to obtain a stable reduction. J Orthop Trauma 21:485–489
Conflict of interest
The authors declare that they have no conflicts of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Ito, J., Takakubo, Y., Sasaki, K. et al. Prevention of excessive postoperative sliding of the short femoral nail in femoral trochanteric fractures. Arch Orthop Trauma Surg 135, 651–657 (2015). https://doi.org/10.1007/s00402-015-2200-3
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-015-2200-3