Inadvertent malposition of a transvenous pacing lead in the left ventricle

  • A. Engström
  • B. Holmberg
  • A. Månsson
  • J. Carlsson
CASE REPORT

Summary

We report the case of inadvertent malpositioning of a pacemaker lead in the left ventricle. The lead went through an open foramen ovale from the right to the left atrium and through the mitral valve into the left ventricle. After a review of the literature, we decided to anticoagulate the patient and leave the electrode in place. During a follow-up period of 16 months, there were no clinical complications.

Key words

Pacemaker therapy lead implantation complications left ventricular malposition 

Versehentliche Fehlpositionierung einer transvenösen Schrittmacher-Elektrode im linken Ventrikel

Zusammenfassung

Wir berichten über den Fall einer unbeabsichtigten Fehlpositionierung einer Schrittmacher-Elektrode im linken Ventrikel. Die Elektrode gelangte über ein offenes Foramen ovale aus dem rechten in das linke Atrium und durch die Mitralklappe in den linken Ventrikel. Nach Sichtung der Literatur zu dieser Komplikation der Schrittmacher-Implantation entschieden wir uns dazu, die Elektrode im linken Ventrikel zu belassen. Unter oraler Antikoagulation war der weitere Verlauf (insgesamt 16 Monate) komplikationsfrei.

Schlüsselwörter

Schrittmacher-Therapie Elektroden-Implantation Komplikationen Linksventrikuläre Malposition 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Nägele H, Azizi M, Hashagen S, Behrens S (2006) Long term follow-up of a malpositioned ventricular pacing lead via the aortic valve; case report. Clin Res Cardiol 95:488–491PubMedCrossRefGoogle Scholar
  2. 2.
    Chun Julian KR, Bode F, Wiegand UKH (2004) Left ventricular malposition of pacemaker lead in Chagas disease. PACE 27:1682–1685PubMedGoogle Scholar
  3. 3.
    Van Gelder BM, Bracke FA, Oto A, Yildirir A, Haas PC, Seger JJ, Stainback RF, Botman KJ, Meijer A (2000) Diagnosis and management of inadvertently placed pacing and ICD-leads in the left ventricle. PACE 23:877–883PubMedGoogle Scholar
  4. 4.
    Sharifi M, Sorkin R, Sharifi V, Lakier JB (1995) Inadvertent malposition of a transvenous inserted pacing lead in the left ventricular chamber. Am J Cardiol 76:92–95PubMedCrossRefGoogle Scholar
  5. 5.
    Ghani M, Thakur RK, Boughner D, Morillo CA, Yee R, Klein GJ (1993): Malposition of transvenous pacing lead in the left ventricle. PACE 16:1800–1807PubMedGoogle Scholar

Copyright information

© Steinkopff-Verlag 2006

Authors and Affiliations

  • A. Engström
    • 1
  • B. Holmberg
    • 1
  • A. Månsson
    • 1
  • J. Carlsson
    • 1
  1. 1.MedicinklinikenHjärtsektionen Kalmar LänssjukhusKalmarSweden

Personalised recommendations