Rheologica Acta

, Volume 46, Issue 7, pp 913–919

A tool for rapid quenching of elongated polymer melts

Original Contribution

Abstract

In this article, we present a device for rapid quenching of elongated polymer melts. The tool is an accessory to the uniaxial elongational rheometer RME of Meissner and Hostettler. It is intended to be used for microscopic and other investigations of stretched polymers. The device allows us to solidify a polymer melt by pouring liquid nitrogen on it and to cut it at the nearly same instant of time. Then the sample can be easily removed from the stretching apparatus. Solving the heat diffusion equation for a polymer melt, which is cooled by liquid nitrogen, we theoretically estimate the quenching time of this method. To demonstrate that this quenching procedure indeed rapidly cools a polymer melt, the stress birefringence of elongated and subsequently quenched polystyrene melts is measured and the stress-optical coefficient C is determined. The experimental value of the stress-optical coefficient is |C|= 4.65×10−9 Pa−1, which agrees well with the data in literature. Using this tool for elongation experiments with the RME, polymer melts can be solidified in between approximately 0.2 and 2.0 s, depending on the thickness of the sample.

Keywords

Extensional rheology Rapid quenching Polystyrene Flow birefringence 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Bach A, Rasmussen HK, Hassager O (2003) Extensional viscosity for polymer melts measured in the filament stretching rheometer. J Rheol 47(2):429–441CrossRefGoogle Scholar
  2. Delaby I, Ernst B, Germain Y, Muller R (1994) Droplet deformation in polymer blends during uniaxial elongational flow: influence of viscosity ratio for large capillary numbers. J Rheol 38(6):1705–1720CrossRefGoogle Scholar
  3. Ferry JD (1980) Viscoelastic properties of polymers, 3rd edn. Wiley, New YorkGoogle Scholar
  4. Fornes TD, Baur JW, Sabba Y, Thomas EL (2006) Morphology and properties of melt-spun polycarbonate fibers containing single- and multi-wall carbon nanotubes. Polymer 47: 1704–1714CrossRefGoogle Scholar
  5. Freiburg Materials Research Center (2001) Service group scientific data processing, NLREG (non-linear-regularization). Software version Rheology 2.2, August 2001Google Scholar
  6. Gramespacher H, Meissner J (1997) Melt elongation and recovery of polymer blends, morphology, and influence of interfacial tension. J Rheol 41(1):27–44CrossRefGoogle Scholar
  7. Hachmann P, Meissner J (2003) Rheometer for equibiaxial and planar elongations of polymer melts. J Rheol 47(4): 989–1010CrossRefGoogle Scholar
  8. Handge UA (2005) Experimental investigation of equibiaxial extension and breakup of drops in a molten two phase polymer blend. Phys Rev E 72:011801CrossRefGoogle Scholar
  9. Handge UA, Pötschke P (2004,2005) Interplay of rheology and morphology in melt elongation and subsequent recovery of polystyrene/poly(methyl methacrylate) blends J Rheol 48:1103–1122; Erratum 49:1553CrossRefGoogle Scholar
  10. Handge UA, Pötschke P (2006) Deformation and orientation during shear and elongation of a molten polycar-bonate/carbon nanotubes composite. Rheol Acta (accepted)Google Scholar
  11. Heindl M, Sommer MK, Münstedt H (2004) Morphology development in polystyrene/polyethylene blends during uniaxial elongational flow. Rheol Acta 44:55–70CrossRefGoogle Scholar
  12. Kröger M (2004) Simple models for complex nonequilibrium fluids. Phys Rep 390(6):453–551CrossRefGoogle Scholar
  13. Lee WK, Kim HD, Kim EY (2006) Morphological reorientation by extensional flow deformation of a triblock copolymer styrene–isoprene–styrene. Current Applied Physics 6(4):718–722CrossRefGoogle Scholar
  14. Luap C, Müller C, Schweizer T, Venerus DC (2005) Simultaneous stress and birefringence measurements during uniaxial elongation of polystyrene melts with narrow molecular weight distribution. Rheol Acta 45:83–91CrossRefGoogle Scholar
  15. Maia JM, Covas JA, Nóbrega JM, Dias TF, Alves FE (1999) Measuring uniaxial extensional viscosity using a modified rotational rheometer. J Non-Newton Fluid Mech 80:183–197CrossRefGoogle Scholar
  16. Matsumoto T, Bogue DC (1977) Stress birefringence in amorphous polymers under nonisothermal conditions. J Polym Sci Polym Phys 15:1663–1674CrossRefGoogle Scholar
  17. McKinley GH, Sridhar T (2002) Filament-stretching rheometry of complex fluids. Annu Rev Fluid Mech 34:375–415CrossRefGoogle Scholar
  18. Meissner J, Hostettler J (1994) A new elongational rheometer for polymer melts and other highly viscoelastic liquids. Rheol Acta 33:1–21CrossRefGoogle Scholar
  19. Muller R, Froelich D (1985) New extensional rheometer for elongational viscosity and flow birefringence measurements: some results on polystyrene melts. Polymer 26:1477–1482CrossRefGoogle Scholar
  20. Münstedt H (1979) New universal extensional rheometer for polymer melts. Measurements on a polystyrene sample. J Rheol 23(4):421–436CrossRefGoogle Scholar
  21. Oosterlink F, Mours M, Laun HM, Moldenaers P (2005) Morphology development of a polystyrene/polymethylmethacrylate blend during start-up of uniaxial elongational flow. J Rheol 49(4):897–918CrossRefGoogle Scholar
  22. Schrader D (1999) Physical constants of poly(styrene). In: Brandrup J, Immergut EH, Grulke EA (eds) Polymer handbook, 4th edn. Wiley, New York, pp. V, 91Google Scholar
  23. Schweizer T (2000) The uniaxial elongational rheometer RME—six years of experience. Rheol Acta 39:428–443CrossRefGoogle Scholar
  24. Sentmanat ML (2004) Miniature universal testing platform: from extensional melt rheology to solid-state deformation behavior. Rheol Acta 43:657–669CrossRefGoogle Scholar
  25. Venerus DC, Zhu SH, Öttinger HC (1999) Stress and birefringence measurements during the uniaxial elongation of polystyrene melts. J Rheol 43(3):795–813CrossRefGoogle Scholar

Copyright information

© Springer-Verlag 2007

Authors and Affiliations

  1. 1.Department of MaterialsInstitute of PolymersZurichSwitzerland

Personalised recommendations