European Radiology

, Volume 23, Issue 8, pp 2279–2287 | Cite as

Optimised low-dose multidetector CT protocol for children with cranial deformity

  • Jose Luis VazquezEmail author
  • Miguel Angel Pombar
  • Jose Manuel Pumar
  • Victor Miguel del Campo
Pediatric

Abstract

Objective

To present an optimised low-dose multidetector computed tomography (MDCT) protocol for the study of children with cranial deformity.

Methods

Ninety-one consecutive MDCT studies were performed in 80 children. Studies were performed with either our standard head CT protocol (group 1, n = 20) or a low-dose cranial deformity protocol (groups 2 and 3). Group 2 (n = 38), initial, and group 3 (n = 33), final and more optimised. All studies were performed in the same 64-MDCT equipment. Cranial deformity protocol was gradationally optimised decreasing kVp, limiting mA range, using automatic exposure control (AEC) and increasing the noise index (NI). Image quality was assessed. Dose indicators such us CT dose index volume (CTDIvol), dose-length product (DLP) and effective dose (E) were used.

Results

The optimised low-dose protocol reached the following values: 80 kVp, mA range: 50–150 and NI = 23. We achieved a maximum dose reduction of 10-22 times in the 1- to 12-month-old cranium in regard to the 2004 European guidelines for MDCT.

Conclusion

A low-dose MDCT protocol that may be used as the first diagnostic imaging option in clinically selected patients with skull abnormalities.

Key Points

MDCT is a very useful tool in the study of skull lesions

Low-dose MDCT minimises child exposure to ionising radiation while maintaining image quality

Low-dose MDCT should be considered as the first imaging option in selected patients

Keywords

Multidetector computed tomography Low dose Image quality Cranial deformity Child 

Abbreviations

AEC

Automatic exposure control

NI

Noise index

VR

Volume-rendered

CTDIvol

Computed tomography dose index volume

DLP

Dose-length product

E

Effective dose

AAP

American Academy of Paediatrics

Notes

Acknowledgements

This study was presented as Scientific Exhibit at ECR 2012.

References

  1. 1.
    Robinson S, Proctor M (2009) Diagnosis and management of deformational plagiocephaly. J Neurosurg Pediatr 3:284–295PubMedCrossRefGoogle Scholar
  2. 2.
    Esparza J, Hinojosa J, Munoz MJ, Romance A, Garcia-Recuero I, Munoz A (2007) Positional plagiocephaly. Diagnosis and treatment. Neurocirugia (Astur) 18:457–467Google Scholar
  3. 3.
    Kirmi O, Lo SJ, Johnson D, Anslow P (2009) Craniosynostosis: a radiological and surgical perspective. Semin Ultrasound CT MR 30:492–512PubMedCrossRefGoogle Scholar
  4. 4.
    Fernbach SK (1998) Craniosynostosis 1998: concepts and controversies. Pediatr Radiol 28:722–728PubMedCrossRefGoogle Scholar
  5. 5.
    Kirks DR, Griscom NT (1998) Practical pediatric imaging: diagnostic radiology of infants and children, 3rd edn. Lippincott-Raven, PhiladelphiaGoogle Scholar
  6. 6.
    Kotrikova B, Krempien R, Freier K, Muhling J (2007) Diagnostic imaging in the management of craniosynostoses. Eur Radiol 17:1968–1978PubMedCrossRefGoogle Scholar
  7. 7.
    Pattisapu JV, Gegg CA, Olavarria G, Johnson KK, Ruiz RL, Costello BJ (2010) Craniosynostosis: diagnosis and surgical management. Atlas Oral Maxillofac Surg Clin North Am 18:77–91PubMedCrossRefGoogle Scholar
  8. 8.
    Simanovsky N, Hiller N, Koplewitz B, Rozovsky K (2009) Effectiveness of ultrasonographic evaluation of the cranial sutures in children with suspected craniosynostosis. Eur Radiol 19:687–692PubMedCrossRefGoogle Scholar
  9. 9.
    Benson ML, Oliverio PJ, Yue NC, Zinreich SJ (1996) Primary craniosynostosis: imaging features. AJR Am J Roentgenol 166:697–703PubMedCrossRefGoogle Scholar
  10. 10.
    Medina LS, Richardson RR, Crone K (2002) Children with suspected craniosynostosis: a cost-effectiveness analysis of diagnostic strategies. AJR Am J Roentgenol 179:215–221PubMedCrossRefGoogle Scholar
  11. 11.
    Pilgram TK, Vannier MW, Hildebolt CF et al (1989) Craniosynostosis: image quality, confidence, and correctness in diagnosis. Radiology 173:675–679PubMedGoogle Scholar
  12. 12.
    Vannier MW, Hildebolt CF, Marsh JL et al (1989) Craniosynostosis: diagnostic value of three-dimensional CT reconstruction. Radiology 173:669–673PubMedGoogle Scholar
  13. 13.
    Álvarez Zallo N, Sanchez-Carpintero R (2010) 3D-CT in the diagnosis of craniosynostosis. Ann Pediatr (Barc) 72:355–356CrossRefGoogle Scholar
  14. 14.
    Bongartz G, Golding S, Jurik A et al (2004) European guidelines for multislice computed tomography. European Commission; Contract number FIGM-CT2000-20078-CT-TIPGoogle Scholar
  15. 15.
    Low E, O’Driscoll M, MacEneaney P, O’Mahony O (2008) Sedation with oral chloral hydrate in children undergoing MRI scanning. Ir Med J 101:80–82PubMedGoogle Scholar
  16. 16.
    Ziegler MA, Fricke BL, Donnelly LF (2003) Is administration of enteric contrast material safe before abdominal CT in children who require sedation? Experience with chloral hydrate and pentobarbital. AJR Am J Roentgenol 180:13–15PubMedCrossRefGoogle Scholar
  17. 17.
    Vade A, Sukhani R, Dolenga M, Habisohn-Schuck C (1995) Chloral hydrate sedation of children undergoing CT and MR imaging: safety as judged by American Academy of Pediatrics guidelines. AJR Am J Roentgenol 165:905–909PubMedCrossRefGoogle Scholar
  18. 18.
    Wong ET, Yu SK, Lai M, Wong YC, Lau PC (2001) MAPD—an objective way to select mAs for paediatric brain CT. Br J Radiol 74:932–937PubMedGoogle Scholar
  19. 19.
    ICRP (2007) Publication 102. Managing patient dose in multi-detector computed tomography. Chapter and appendix A. Ann ICRP 37:59–79CrossRefGoogle Scholar
  20. 20.
    Bracken J, Heaslip I, Ryan S (2012) Chloral hydrate sedation in radiology: retrospective audit of reduced dose. Pediatr Radiol 42:349–354PubMedCrossRefGoogle Scholar
  21. 21.
    Edwards AD, Arthurs OJ (2011) Paediatric MRI under sedation: is it necessary? What is the evidence for the alternatives? Pediatr Radiol 41:1353–1364PubMedCrossRefGoogle Scholar
  22. 22.
    Hanley M, Koonce JD, Bradshaw ML (2009) www.X-rayRisk.com: an online calculator for cancer risk. J Am Coll Radiol 6:475–476PubMedCrossRefGoogle Scholar
  23. 23.
    Alessio AM, Phillips GS (2010) A pediatric CT dose and risk estimator. Pediatr Radiol 40:1816–1821PubMedCrossRefGoogle Scholar
  24. 24.
    AAP (1992) American Academy of Pediatrics AAP task force on infant positioning and SIDS: positioning and SIDS. Pediatrics 89:1120–1126Google Scholar
  25. 25.
    AAP (2005) American Academy of Pediatrics AAP task force on sudden infant death syndrome: the changing concept of sudden infant death syndrome: diagnostic coding shifts, controversies regarding the sleeping environment, and new variables to consider in reducing risk. Pediatrics 116:1245–1255CrossRefGoogle Scholar
  26. 26.
    Sze RW, Parisi MT, Sidhu M et al (2003) Ultrasound screening of the lambdoid suture in the child with posterior plagiocephaly. Pediatr Radiol 33:630–636PubMedCrossRefGoogle Scholar
  27. 27.
    Brenner DJ, Hall EJ (2007) Computed tomography—an increasing source of radiation exposure. N Engl J Med 357:2277–2284PubMedCrossRefGoogle Scholar
  28. 28.
    Sze RW, Hopper RA, Ghioni V et al (2005) MDCT diagnosis of the child with posterior plagiocephaly. AJR Am J Roentgenol 185:1342–1346PubMedCrossRefGoogle Scholar
  29. 29.
    Shah NB, Platt SL (2008) ALARA: is there a cause for alarm? Reducing radiation risks from computed tomography scanning in children. Curr Opin Pediatr 20:243–247PubMedCrossRefGoogle Scholar
  30. 30.
    McCollough CH, Primak AN, Braun N, Kofler J, Yu L, Christner J (2009) Strategies for reducing radiation dose in CT. Radiol Clin N Am 47:27–40PubMedCrossRefGoogle Scholar
  31. 31.
    Ledenius K, Gustavsson M, Johansson S, Stalhammar F, Wiklund LM, Thilander-Klang A (2009) Effect of tube current on diagnostic image quality in paediatric cerebral multidetector CT images. Br J Radiol 82:313–320PubMedCrossRefGoogle Scholar
  32. 32.
    Brix G, Nissen-Meyer S, Lechel U et al (2009) Radiation exposures of cancer patients from medical X-rays: how relevant are they for individual patients and population exposure? Eur J Radiol 72:342–347PubMedCrossRefGoogle Scholar
  33. 33.
    Casal MD, Jiménez M, Ureña A et al (2010) Estudio dosimétrico en equipos de TC monocorte en pacientes adultos y pediátricos, y su aplicación en niños con enfermedades neuronales crónicas. Radioprotección 17:19–24Google Scholar
  34. 34.
    España ML, Gómez G, García P, Bermúdez R, García E, Solís I (2010) Protección radiológica en radiodiagnóstico pediátrico. Radioprotección 17:13–18Google Scholar
  35. 35.
    Nievelstein RA, van Dam IM, van der Molen AJ (2010) Multidetector CT in children: current concepts and dose reduction strategies. Pediatr Radiol 40:1324–1344PubMedCrossRefGoogle Scholar
  36. 36.
    Fox SH, Toth T (2002) Dose reduction on GE CT scanners. Pediatr Radiol 32:718–723, discussion 751–714PubMedCrossRefGoogle Scholar
  37. 37.
    Alibek S, Brand M, Suess C, Wuest W, Uder M, Greess H (2011) Dose reduction in pediatric computed tomography with automated exposure control. Acad Radiol 18:690–693PubMedCrossRefGoogle Scholar
  38. 38.
    Goldman LW (2008) Principles of CT: multislice CT. J Nucl Med Technol 36:57–68, quiz 75–56PubMedCrossRefGoogle Scholar
  39. 39.
    Hopper KD, Neuman JD, King SH, Kunselman AR (2001) Radioprotection to the eye during CT scanning. AJNR Am J Neuroradiol 22:1194–1198PubMedGoogle Scholar
  40. 40.
    Heaney DE, Norvill CA (2006) A comparison of reduction in CT dose through the use of gantry angulations or bismuth shields. Australas Phys Eng Sci Med 29:172–178PubMedCrossRefGoogle Scholar
  41. 41.
    Keil B, Wulff J, Schmitt R et al (2008) Protection of eye lens in computed tomography–dose evaluation on an anthropomorphic phantom using thermo-luminescent dosimeters and Monte-Carlo simulations. Rofo 180:1047–1053PubMedCrossRefGoogle Scholar
  42. 42.
    Geleijns J, Salvado Artells M, Veldkamp WJ, Lopez Tortosa M, Calzado Cantera A (2006) Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality. Eur Radiol 16:2334–2340PubMedCrossRefGoogle Scholar
  43. 43.
    Chang KH, Lee W, Choo DM, Lee CS, Kim Y (2009) Dose reduction in CT using bismuth shielding: measurements and Monte Carlo simulations. Radiat Prot Dosim 138:382–388CrossRefGoogle Scholar
  44. 44.
    Mukundan S Jr, Wang PI, Frush DP et al (2007) MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children. AJR Am J Roentgenol 188:1648–1650PubMedCrossRefGoogle Scholar
  45. 45.
    Catuzzo P, Aimonetto S, Fanelli G et al (2009) Dose reduction in multislice CT by means of bismuth shields: results of in vivo measurements and computed evaluation. Radiol Med 115:152–169PubMedCrossRefGoogle Scholar
  46. 46.
    Raissaki M, Perisinakis K, Damilakis J, Gourtsoyiannis N (2010) Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction. Pediatr Radiol 40:1748–1754PubMedCrossRefGoogle Scholar
  47. 47.
    Birnbaum S (2010) Radiation protection in the era of helical CT: practical patient based programs for decreasing patient exposure. Semin Ultrasound CT MR 31:46–52PubMedCrossRefGoogle Scholar
  48. 48.
    Medina LS (2000) Three-dimensional CT maximum intensity projections of the calvaria: a new approach for diagnosis of craniosynostosis and fractures. AJNR Am J Neuroradiol 21:1951–1954PubMedGoogle Scholar
  49. 49.
    Mulliken JB, Vander Woude DL, Hansen M, LaBrie RA, Scott RM (1999) Analysis of posterior plagiocephaly: deformational versus synostotic. Plast Reconstr Surg 103:371–380PubMedCrossRefGoogle Scholar
  50. 50.
    Rao P, Bekhit E, Ramanauskas F, Kumbla S (2011) CT head in children. Eur J Radiol. doi: 10.1016/j.ejrad.2011.11.038

Copyright information

© European Society of Radiology 2013

Authors and Affiliations

  • Jose Luis Vazquez
    • 1
    Email author
  • Miguel Angel Pombar
    • 2
  • Jose Manuel Pumar
    • 3
  • Victor Miguel del Campo
    • 4
  1. 1.Department of RadiologyComplejo Hospitalario Universitario de VigoVigoSpain
  2. 2.Department of RadiophysicsComplejo Hospitalario Universitario de SantiagoSantiago de CompostelaSpain
  3. 3.Department of RadiologyComplejo Hospitalario Universitario de SantiagoSantiago de CompostelaSpain
  4. 4.Department of Public HealthComplejo Hospitalario Universitario de VigoVigoSpain

Personalised recommendations