Objective

The aim of the study was to compare 4 cm with 16 cm Z-axis coverage in the assessment of brain CT perfusion (CTP) using. 320 slice multidetector CT

Methods

A retrospective non-randomised review of CTP performed on MD320 CT between September 2008 and January 2009 was undertaken. Two experienced readers reviewed the studies along with the 4 cm and 16 cm Z-axis CTP image data set. The outcome parameters assessed were the extent of the original finding, any additional findings and a change of diagnosis.

Results

14 out of 27 patients were found to have abnormal CTP (mean age 58.1 years, 9 male). The 16 cm Z-axis increased the accuracy of the infarct core in 78% and ischaemic penumbra quantification in 100% of the cases. It also diagnosed additional infarcts in the same vascular territory in 28% of cases and in a different vascular territory in 14%.

Conclusions

The increased field of view with MD320 better defines the true extent of the infarct core and ischaemic penumbra. It also identified other areas of infarction that were not identified on the 4 cm Z-axis.

Keywords

CT Brain Infarct Ischemia Perfusion 

References

  1. 1.
    World Health Organization. The top ten causes of death. Available via WHO www.who.int/mediacentre/factsheets/fs310/en/index.html. Accessed 6 Apr 2009
  2. 2.
    National Institute of Neurological Diseases and Stroke rt-PA Stroke Study Group (1995) Tissue plasminogen activator for acute ischemic stroke. N Engl J Med 333:1581–1587CrossRefGoogle Scholar
  3. 3.
    Wintermark M, Sesay M, Barbier E et al (2005) Comparative overview of brain perfusion imaging techniques. Stroke 36:e83–e99CrossRefPubMedGoogle Scholar
  4. 4.
    Latchaw RE, Yonas H, Hunter GJ et al (2003) Guidelines and recommendations for perfusion imaging in cerebral ischemia. Stroke 34:1084–1104CrossRefPubMedGoogle Scholar
  5. 5.
    Silvernnoinen HM, Hamberg LM, Lindberg PJ et al (2008) CT Perfusion identifies increased salvage of tissue in patients receiving intravenous recombinant tissue plasminogen activator within 3 hours of stroke onset. AJNR Am J Neuroradiol 29:1118–1123CrossRefGoogle Scholar
  6. 6.
    Hoeffner EG, Case I, Jain R et al (2004) Cerebral perfusion CT: technique and clinical applications. Radiology 231:632–644CrossRefPubMedGoogle Scholar
  7. 7.
    Konig K (2007) Diagnosis of acute stroke. In: Miles KA, Eastwood JD, Konig M (eds) Multidetector computed tomography in cerebrovascular disease, CT perfusion imaging. Informa Healthcare Ltd, Milton Park, pp 83–97Google Scholar
  8. 8.
    Miles KA, Menon DK (2007) Cerebrovascular physiology and pathophysiology. In: Miles KA, Eastwood JD, Konig M (eds) Multidetector computed tomography in cerebrovascular disease, CT perfusion imaging. Informa Healthcare, Milton Park, pp 71–82Google Scholar
  9. 9.
    Roberts HC, Roberts PL, Wade SS et al (2001) Multisection dynamic CT perfusion for acute cerebral ischemia: the “toggling-table” technique. AJNR Am J Neuroradiol 22:1077–1080PubMedGoogle Scholar
  10. 10.
    Wintermark M, Maeder P, Verdun FR et al (2000) Using 80 kVp versus 120 kVp in perfusion CT measurement of regional cerebral blood flow. AJNR Am J Neuroradiol 21:1881–1884PubMedGoogle Scholar
  11. 11.
    Konig M, Bultmann E, Bode-Schnurbus L et al (2007) Image quality in CT perfusion imaging of the brain, the role of iodine concentration. Eur Radiol 17:39–47CrossRefPubMedGoogle Scholar
  12. 12.
    Wiesmann M, Berg S, Bohner G et al (2008) Dose reduction in dynamic perfusion CT of the brain: effects of the scan frequency on measurements of cerebral blood flow. Cerebral blood volume and mean transit time. Eur Radiol 18:2967–2974CrossRefPubMedGoogle Scholar
  13. 13.
    Wintermark M, Smith WS, Nerissa U et al (2004) Dynamic perfusion CT: optimizing the temporal resolution and contrast volume calculation of perfusion CT parameters in stroke patients. AJNR Am J Neuroradiol 25:720–729PubMedGoogle Scholar
  14. 14.
    Soustiel JF, Mor N, Zaaroor M et al (2006) Cerebral perfusion computerized tomography: influence of reference vessels, regions of interest and interobserver variability. Neuroradiology 48:670–677CrossRefPubMedGoogle Scholar
  15. 15.
    Wintermark M, Fischbein NJ, Smith WS et al (2005) Accuracy of dynamic perfusion CT with deconvolution in detecting acute hemispheric stroke. AJNR Am J Neuroradiol 26:104–112PubMedGoogle Scholar
  16. 16.
    Eastwood JD, Lev MH, Wintermark M et al (2003) Correlation of early dynamic CT perfusion imaging with whole-brain MR diffusion and perfusion imaging in acute hemispheric stroke. AJNR Am J Neuroradiol 24:1869–1875PubMedGoogle Scholar
  17. 17.
    Ezzeddine MA, Lev MH, McDonald CT et al (2002) CT angiography with whole brain perfused blood volume imaging: added clinical value in the assessment of acute stroke. Stroke 33:959–966CrossRefPubMedGoogle Scholar
  18. 18.
    Klingebiel R, Siebert E, Diekmann S et al (2009) 4-D Imaging in cerebrovascular disorders by using 320-slice CT: feasibility and preliminary clinical experience. Acad Radiol 16:123–129CrossRefPubMedGoogle Scholar
  19. 19.
    Mori S, Obata T, Nakajima N et al (2005) Volumetric perfusion CT using prototype 256-detector row CT scanner: preliminary study with healthy porcine model. AJNR Am J Neuroradiol 26:2536–2541PubMedGoogle Scholar
  20. 20.
    Provenzale JM, Shah K, Patel U et al (2008) Systemic review of CT and MR perfusion imaging for assessment of acute cerebrovascular disease. AJNR Am J Neuroradiol 29:1476–1482CrossRefPubMedGoogle Scholar

Copyright information

© European Society of Radiology 2009

Authors and Affiliations

  • Mark Page
    • 1
  • Dee Nandurkar
    • 1
  • Marcus Peter Crossett
    • 1
  • Stephen L. Stuckey
    • 1
  • Kenneth P. Lau
    • 1
  • Nicholas Kenning
    • 1
  • John M. Troupis
    • 1
    Email author
  1. 1.Department of Diagnostic Imaging, Southern Health NetworkMonash Medical CentreClaytonAustralia

Personalised recommendations