Advertisement

European Radiology

, Volume 19, Issue 3, pp 619–625 | Cite as

Visualisation of intracerebral haemorrhage with flat-detector CT compared to multislice CT: results in 44 cases

  • Tobias Struffert
  • Gregor Richter
  • Tobias Engelhorn
  • Marc Doelken
  • Philipp Goelitz
  • Willi A. Kalender
  • Oliver Ganslandt
  • Arnd Doerfler
Computer Tomography

Abstract

The aim of the study was to test the reliability of intracerebral haemorrhage (ICH) detection with C-arm-mounted flat-detector computed tomography (FD-CT) in the angio suite as compared to multislice CT (MSCT). In this study 44 patients with 45 ICH were included. All patients were investigated with MSCT and FD-CT during angiographic evaluation. As a control group we included 16 patients without ICH. In each haematoma we assessed volumetric data of the ICH and counted the numbers of ICH-positive slices. Using interobserver ratings, we additionally investigated the potential of FD-CT to serve as a diagnostic tool to detect ICH. In FD-CT three haematomas were not detected because of motion and beam-hardening artefacts in the region close to the skull base. The r value for the degree of interobserver agreement for the number of slices was 0.95 for MSCT and 0.94 for FD-CT. Measurements of the area and the calculated volume of the ICH showed high inter- and intraobserver agreement. Our results indicate that FD-CT is a helpful tool in the daily emergency management of ICH patients as detection of ICH was found to be nearly as reliable as in MSCT. Limitations of this technology are motion and beam-hardening artefacts that may mask small haematomas located in the posterior fossa or the skull base.

Keywords

Flat-detector computed tomography Multisclice computed tomography Intracerebral haemorrhage Computed tomography 

References

  1. 1.
    Sahni R, Weinberger J (2007) Management of intracerebral hemorrhage. Vasc Health Risk Manag 3(5):701–709PubMedGoogle Scholar
  2. 2.
    Hänggi D, Steiger HJ (2008) Spontaneous intracerebral haemorrhage in adults: a literature overview. Acta Neurochir 150:371–379CrossRefGoogle Scholar
  3. 3.
    Brisman JL, Niimi Y, Song JK, Berenstein A (2005) Aneurysmal rupture during coiling: low incidence and good outcomes at a single large volume centre. Neurosurgery 57:1103–1109CrossRefPubMedGoogle Scholar
  4. 4.
    Carvi y Nievas M, Haas E, Höllerhage HG (2007) Severe intracranial bleedings during endovascular procedures: outcome of surgically treated patients. Neurol Res 29(1):81–90CrossRefPubMedGoogle Scholar
  5. 5.
    Kalender WA, Kyriakou Y (2007) Flat-detector computed tomography (FD-CT). Eur Radiol 17:2767–2779CrossRefPubMedGoogle Scholar
  6. 6.
    Schumacher M, Kutluk K, Ott D (1989) Digital rotational radiography in neuroradiology. Am J Neuroradiol 10:644–649PubMedGoogle Scholar
  7. 7.
    Kalender WA (2003) [The use of flat-panel detectors for CT imaging]. Radiologe 43:379–387CrossRefPubMedGoogle Scholar
  8. 8.
    Loose R, Wucherer M, Brunner T, Adamus R, Simmler R (2005) Visualization of 3D low contrast objects by CT cone-beam reconstruction of a rotational angiography with a dynamic solid body detector. Roefo S1:PO 160Google Scholar
  9. 9.
    Heran NS, Song JK, Namba K, Smith W, Niimi Y, Berenstein A (2006) The utility of DynaCT in neuroendovascular procedures. Am J Neuroradiol 27:330–332PubMedGoogle Scholar
  10. 10.
    Arakawa H, Marks MP, Do HM, Bouley DM, Strobel N, Moore T, Fahrig R (2008) Experimental study of intracranial hematoma detection with flat panel detector C-arm CT. Am J Neuroradiol 29(4):766–772CrossRefPubMedGoogle Scholar
  11. 11.
    Kwak R, Kadoya S, Suzuke T (1983) Factors affecting the prognosis of thalamic hemorrhage. Stroke 14:493–500PubMedGoogle Scholar
  12. 12.
    Richter G, Engelhorn T, Struffert T, Doelken M, Ganslandt O, Hornegger J, Kalender WA, Doerfler A (2007) Flat panel detector angiographic CT for stent-assisted coil embolization of broad-based cerebral aneurysms. Am J Neuroradiol 28(19):1902–1908CrossRefPubMedGoogle Scholar
  13. 13.
    Engelhorn T, Rennert J, Richter G, Struffert T, Ganslandt O, Doerfler A (2007) Myelography using flat panel volumetric computed tomography: a comparative study in patients with lumbar spinal stenosis. Spine 32(18):E523–E527CrossRefPubMedGoogle Scholar
  14. 14.
    Doelken M, Struffert T, Richter G, Engelhorn T, Nimsky C, Ganslandt O, Hammen T, Doerfler A (2008) Flat-panel detector volumetric CT for visualization of subarachnoid hemorrhage and ventricles: preliminary results compared to conventional CT. Neuroradiology 50(6):517–523CrossRefPubMedGoogle Scholar

Copyright information

© European Society of Radiology 2008

Authors and Affiliations

  • Tobias Struffert
    • 1
  • Gregor Richter
    • 1
  • Tobias Engelhorn
    • 1
  • Marc Doelken
    • 1
  • Philipp Goelitz
    • 1
  • Willi A. Kalender
    • 2
  • Oliver Ganslandt
    • 3
  • Arnd Doerfler
    • 1
  1. 1.Department of NeuroradiologyUniversity of Erlangen-NuernbergErlangenGermany
  2. 2.Institute of Medical PhysicsUniversity of Erlangen-NuernbergErlangenGermany
  3. 3.Department of NeurosurgeryUniversity of Erlangen-NuernbergErlangenGermany

Personalised recommendations