CardioVascular and Interventional Radiology

, Volume 32, Issue 4, pp 714–719 | Cite as

Telemetric Catheter-Based Pressure Sensor for Hemodynamic Monitoring: Experimental Experience

  • Andreas H. MahnkenEmail author
  • Ute Urban
  • Holger Fassbender
  • Uwe Schnakenberg
  • Felix Schoth
  • Thomas Schmitz-Rode
Laboratory Investigation

Abstract

The purpose of this study was to evaluate the technical and animal experimental feasibility of a percutaneously implantable pulmonary arterial implant for permanent hemodynamic monitoring. Two systems for measuring pulmonary artery pressure (PAP) as well as pulmonary artery occlusion pressure (PAOP) were developed by modifying a commercially available pulmonary artery catheter (PAC). First, a cable-bound catheter-based system was designed by implementation of a capacitive absolute-pressure sensor in the catheter tip. This system was developed further into a completely implantable telemetric system. The devices were tested in an acute setting in a total of 10 sheep. The implant was placed with its tip in the descending pulmonary artery via the right jugular approach. Results were compared with conventional PAC positioned in the contralateral pulmonary artery using Pearson’s correlation coefficients and Bland-Altman plots. Implantation of the monitoring systems was uneventful in 10 animals. Data from two fully functional cable-bound and telemetric pressure monitoring systems were available, with a total of 18,506 measurements. There was an excellent correlation between reference data and the data obtained with the implants (r = 0.9944). Bland-Altman plots indicated a very good agreement between the techniques. We report the development and successful initial test of an implantable catheter-based device for long-term measurement of PAP and PAOP. Both devices may be applicable for hemodynamic monitoring. Further long-term studies for assessing reliability and durability of the device are warranted.

Keywords

Pulmonary artery catheter Implant Hemodynamic monitoring 

References

  1. 1.
    Nieminen MS, Böhm M, Cowie MR et al (2005) Executive summary of the guidelines on the diagnosis and treatment of acute heart failure: the Task Force on Acute Heart Failure of the European Society of Cardiology. Eur Heart J 26:384–416PubMedCrossRefGoogle Scholar
  2. 2.
    Cecconi M, Reynolds TE, Al-Subaie N et al (2007) Haemodynamic monitoring in acute heart failure. Heart Fail Rev 12:105–111PubMedCrossRefGoogle Scholar
  3. 3.
    Connors AF Jr, Speroff T, Dawson NV et al (1996) The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 276:889–897PubMedCrossRefGoogle Scholar
  4. 4.
    Bernard GR, Sopko G, Cerra F et al (2000) Pulmonary artery catheterization and clinical outcomes: National Heart, Lung, and Blood Institute and Food and Drug Administration Workshop Report. Consensus Statement. JAMA 283:2568–2572PubMedCrossRefGoogle Scholar
  5. 5.
    American Society of Anesthesiologists Task Force on Pulmonary Artery Catheterization (2003) Practice guidelines for pulmonary artery catheterization: an updated report by the American Society of Anesthesiologists Task Force on Pulmonary Artery Catheterization. Anesthesiology 99:988–1014CrossRefGoogle Scholar
  6. 6.
    Hunt SA, Abraham WT, Chin MH et al (2005) ACC/AHA 2005 guideline update for the diagnosis and management of chronic heart failure in the adult: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Update the 2001 Guidelines for the Evaluation and Management of Heart Failure): developed in collaboration with the American College of Chest Physicians and the International Society for Heart and Lung Transplantation: endorsed by the Heart Rhythm Society. Circulation 112:e154–e235PubMedCrossRefGoogle Scholar
  7. 7.
    Yu CM, Wang L, Chau E et al (2005) Intrathoracic impedance monitoring in patients with heart failure. Correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation 112:841–848PubMedCrossRefGoogle Scholar
  8. 8.
    Androne AS, Hryniewwicz K, Hudaihed A et al (2004) Relation of unrecognized hypervolemia in chronic heart failure to clinical status, hemodynamics and patient outcomes. Am J Cardiol 93:1254–1259PubMedCrossRefGoogle Scholar
  9. 9.
    Verdejo HE, Castro PF, Concepción R et al (2007) Comparison of a radiofrequency-based wireless pressure sensor to Swan-Ganz catheter and echocardiography for ambulatory assessment of pulmonary artery pressure in heart failure. J Am Coll Cardiol 50:2375–2382PubMedCrossRefGoogle Scholar
  10. 10.
    Rozenman Y, Schwartz RS, Shah H et al (2007) Wireless acoustic communication with a miniature pressure sensor in the pulmonary artery for disease surveillance and therapy of patients with congestive heart failure. J Am Coll Cardiol 49:784–789PubMedCrossRefGoogle Scholar
  11. 11.
    Wagner JG, Leatherman JW (1998) Right ventricular end-diastolic volume as a predictor of the hemodynamic response to a fluid challenge. Chest 113:1048–1054PubMedCrossRefGoogle Scholar
  12. 12.
    Heresi GA, Arroliga AC, Wiedemann HP et al (2006) Pulmonary artery catheter and fluid management in acute lung injury and the acute respiratory distress syndrome. Clin Chest Med 27:627–635PubMedCrossRefGoogle Scholar
  13. 13.
    Dudaicevs H, Kandler M, Manoli Y et al (1994) Surface micromachined pressure sensors with integrated CMOS read-out electronics. Sensors Actuators A 43:157–163CrossRefGoogle Scholar
  14. 14.
    Schlierf R, Horst U, Ruhl M et al (2007) Fast telemetric pressure and temperature sensor system for medical applications. J Micromech Microeng 17:S98–S102CrossRefGoogle Scholar
  15. 15.
    Forrester JS, Diamond G, Chatterjee K et al (1976) Medical therapy of acute myocardial infarction by application of hemodynamic subsets. Part 1. N Engl J Med 295:1356–1362PubMedGoogle Scholar
  16. 16.
    Forrester JS, Diamond G, Chatterjee K et al (1976) Medical therapy of acute myocardial infarction by application of hemodynamic subsets. Part 2. N Engl J Med 295:1404–1413PubMedCrossRefGoogle Scholar
  17. 17.
    Valenzuela-García LF, Almendro-Delia M, González-Valdayo M et al (2007) Percutaneous retrieval of a pulmonary artery catheter knot in pacing electrodes. Cardiovasc Interv Radiol 30:1082–1084CrossRefGoogle Scholar
  18. 18.
    Binanay C, Califf RM, Hasselblad V et al (2005) Evaluation study of congestive heart failure and pulmonary artery catheterization effectiveness: the ESCAPE trial. JAMA 294:1625–1633PubMedCrossRefGoogle Scholar
  19. 19.
    Bowdle TA (2002) Complications of invasive monitoring. Anesthesiol Clin No Am 20:571–588CrossRefGoogle Scholar
  20. 20.
    Gottwalles Y, Wunschel-Joseph ME, Hanssen M (2000) Coil embolization treatment in pulmonary artery branch rupture during Swan-Ganz catheterization. Cardiovasc Interv Radiol 23:477–479CrossRefGoogle Scholar
  21. 21.
    Miller JA, Singireddy S, Maldjian P et al (1999) A reevaluation of the radiographically detectable complications of percutaneous venous access lines inserted by four subcutaneous approaches. Am Surg 65:125–130PubMedGoogle Scholar
  22. 22.
    Summerhill EM, Baram M (2005) Principles of pulmonary artery catheterization in the critically ill. Lung 183:209–219PubMedCrossRefGoogle Scholar
  23. 23.
    Adamson PB, Magalski A, Braunschweig F et al (2003) Ongoing right ventricular hemodynamics in heart failure: clinical value of measurements derived from an implantable monitoring system. J Am Coll Cardiol 41:565–571PubMedCrossRefGoogle Scholar
  24. 24.
    Tang WH, Girod JP, Lee MJ et al (2003) Plasma B-type natriuretic peptide levels in ambulatory patients with established chronic symptomatic systolic heart failure. Circulation 108:2964–2966PubMedCrossRefGoogle Scholar
  25. 25.
    Gackowski A, Isnard R, Golmard JL et al (2004) Comparison of echocardiography and plasma B-type natriuretic peptide for monitoring response to treatment in acute heart failure. Eur Heart J 25:1788–1796PubMedCrossRefGoogle Scholar
  26. 26.
    Bourge RC, Abraham WT, Adamson PB et al (2008) Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure: the COMPASS-HF study. J Am Coll Cardiol 51:1073–1079PubMedCrossRefGoogle Scholar
  27. 27.
    Teichgräber UK, Gebauer B, Benter T et al (2003) Central venous access catheters: radiological management of complications. Cardiovasc Interv Radiol 26:321–333CrossRefGoogle Scholar

Copyright information

© Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE) 2009

Authors and Affiliations

  • Andreas H. Mahnken
    • 1
    • 2
    • 4
    Email author
  • Ute Urban
    • 2
  • Holger Fassbender
    • 3
  • Uwe Schnakenberg
    • 3
  • Felix Schoth
    • 1
  • Thomas Schmitz-Rode
    • 2
  1. 1.Department of Diagnostic RadiologyAachen University HospitalAachenGermany
  2. 2.Applied Medical Engineering, Helmholtz InstituteRWTH Aachen UniversityAachenGermany
  3. 3.Institute of Materials in Electrical Engineering 1RWTH Aachen UniversityAachenGermany
  4. 4.University Hospital, RWTH Aachen UniversityAachenGermany

Personalised recommendations