Abstract

Background

Recurrent laryngeal nerve (RLN) palsy is a serious complication of thyroid surgery. During intraoperative neuromonitoring (IONM) of the RLN in thyroid surgery, repeated shifting between surgical instruments and the nerve stimulator is cumbersome and time-consuming. Therefore, we developed a simple detachable magnetic nerve stimulator that may be connected to all metallic surgical instruments. This study aimed to investigate the feasibility and efficacy of this detachable magnetic nerve stimulator for IONM in a porcine model and humans.

Methods

Eight RLNs in four pigs and thirteen in nine patients that underwent thyroidectomy were examined. We developed a detachable nerve stimulator that combined surgical instruments with the nerve-stimulating probe. We evaluated the electromyography (EMG) amplitudes of the RLNs in pigs and patients using conventional nerve probes and surgical instruments with the novel detachable magnetic nerve stimulator attached.

Results

The EMG amplitudes of the eight RLNs in pigs and thirteen in patients were analyzed. The detachable magnetic nerve stimulator was feasible and safe. There was no significant difference in the EMG amplitude between instruments (P = 0.423 in animals, P = 0.446 in humans).

Conclusions

The application of stimulating dissection using a detachable magnetic nerve stimulator during thyroidectomy with IONM is simple, convenient, and effective. It provides surgeons with real-time feedback of the EMG response during intermittent IONM. We propose that this novel device could be an essential guide for most surgeons, especially for less experienced head and neck surgeons.

Notes

Acknowledgements

The authors appreciate Prof. Jung- Hoon Ro at the Biomedical Engineering Department at Pusan National University Hospital for his help in developing the instrument.

Compliance with ethical standards

Conflict of interest

The authors have no conflicts of interest or financial ties to disclose.

References

  1. 1.
    Munch S, deKryger L (2001) A piece of my mind. Moral wounds: complicated complications. JAMA 285:1131–1132CrossRefPubMedGoogle Scholar
  2. 2.
    Chiang FY, Lu IC, Kuo WR et al (2008) The mechanism of recurrent laryngeal nerve injury during thyroid surgery—the application of intraoperative neuromonitoring. Surgery 143:743–749CrossRefPubMedGoogle Scholar
  3. 3.
    Dionigi G, Alesina PF, Barczynski M et al (2012) Recurrent laryngeal nerve injury in video-assisted thyroidectomy: lessons learned from neuromonitoring. Surg Endosc 26:2601–2608CrossRefPubMedGoogle Scholar
  4. 4.
    Randolph GW, Dralle H, Abdullah H et al (2011) Electrophysiologic recurrent laryngeal nerve monitoring during thyroid and parathyroid surgery: international standards guideline statement. Laryngoscope 121(Suppl 1):S1–16CrossRefPubMedGoogle Scholar
  5. 5.
    Chiang FY, Lu IC, Chen HC et al (2010) Intraoperative neuromonitoring for early localization and identification of recurrent laryngeal nerve during thyroid surgery. Kaohsiung J Med Sci 26:633–639CrossRefPubMedGoogle Scholar
  6. 6.
    Dionigi G, Wu CW, Kim HY et al (2016) Severity of recurrent laryngeal nerve injuries in thyroid surgery. World J Surg 40:1373–1381. doi: 10.1007/s00268-016-3415-3 CrossRefPubMedGoogle Scholar
  7. 7.
    Dionigi G, Donatini G, Boni L et al (2013) Continuous monitoring of the recurrent laryngeal nerve in thyroid surgery: a critical appraisal. Int J Surg 11(Suppl 1):S44–S46CrossRefPubMedGoogle Scholar
  8. 8.
    Wu CW, Dionigi G, Sun H et al (2014) Intraoperative neuromonitoring for the early detection and prevention of RLN traction injury in thyroid surgery: a porcine model. Surgery 155:329–339CrossRefPubMedGoogle Scholar
  9. 9.
    Almquist M, Thier M, Salem F (2016) Cardiac arrest with vagal stimulation during intraoperative nerve monitoring. Head Neck 38:E2419–E2420CrossRefPubMedGoogle Scholar
  10. 10.
    Chiang FY, Lu IC, Chang PY et al (2015) Stimulating dissecting instruments during neuromonitoring of RLN in thyroid surgery. Laryngoscope 125:2832–2837CrossRefPubMedGoogle Scholar
  11. 11.
    Yang S, Zhou L, Lu Z et al (2017) Systematic review with meta-analysis of intraoperative neuromonitoring during thyroidectomy. Int J Surg 39:104–113CrossRefPubMedGoogle Scholar
  12. 12.
    Potenza AS, Phelan EA, Cernea CR et al (2013) Normative intra-operative electrophysiologic waveform analysis of superior laryngeal nerve external branch and recurrent laryngeal nerve in patients undergoing thyroid surgery. World J Surg 37:2336–2342. doi: 10.1007/s00268-013-2148-9 CrossRefPubMedGoogle Scholar
  13. 13.
    Uludag M, Aygun N, Kartal K et al (2017) Is intraoperative neural monitoring necessary for exploration of the superior laryngeal nerve? Surgery 161:1129–1138CrossRefPubMedGoogle Scholar
  14. 14.
    Barczyński M, Konturek A, Pragacz K et al (2014) Intraoperative nerve monitoring can reduce prevalence of recurrent laryngeal nerve injury in thyroid reoperations: results of a retrospective cohort study. World J Surg 38:599–606. doi: 10.1007/s00268-013-2260-x CrossRefPubMedGoogle Scholar
  15. 15.
    Snyder SK, Lairmore TC, Hendricks JC et al (2008) Elucidating mechanisms of recurrent laryngeal nerve injury during thyroidectomy and parathyroidectomy. J Am Coll Surg 206:123–130CrossRefPubMedGoogle Scholar
  16. 16.
    Liu XL, Wu CW, Zhao YS et al (2016) Exclusive real-time monitoring during recurrent laryngeal nerve dissection in conventional monitored thyroidectomy. Kaohsiung J Med Sci 32:135–141CrossRefPubMedGoogle Scholar
  17. 17.
    Schneider R, Randolph GW, Sekulla C et al (2013) Continuous intraoperative vagus nerve stimulation for identification of imminent recurrent laryngeal nerve injury. Head Neck 35:1591–1598CrossRefPubMedGoogle Scholar
  18. 18.
    Phelan E, Schneider R, Lorenz K et al (2014) Continuous vagal IONM prevents recurrent laryngeal nerve paralysis by revealing initial EMG changes of impending neuropraxic injury: a prospective, multicenter study. Laryngoscope 124:1498–1505CrossRefPubMedGoogle Scholar
  19. 19.
    Lin YC, Dionigi G, Randolph GW et al (2015) Electrophysiologic monitoring correlates of recurrent laryngeal nerve heat thermal injury in a porcine model. Laryngoscope 125:E283–E290CrossRefPubMedGoogle Scholar

Copyright information

© Société Internationale de Chirurgie 2017

Authors and Affiliations

  • Eui-Suk Sung
    • 1
  • Jin-Choon Lee
    • 1
  • Sung-Chan Shin
    • 2
  • Sung-Won Choi
    • 2
  • Da-Woon Jung
    • 2
  • Byung-Joo Lee
    • 2
    Email author
  1. 1.Department of Otolaryngology - Head and Neck Surgery and Research Institute for Convergence of Biomedical Science and Technology, College of MedicinePusan National UniversityYangsanKorea
  2. 2.Department of Otolaryngology - Head and Neck Surgery, College of Medicine and Medical Research InstitutePusan National UniversityBusanKorea

Personalised recommendations