Smartphone versus knee ligament arthrometer when size does not matter

  • Ferretti Andrea
  • Valeo Luigi
  • Mazza DanieleEmail author
  • Muliere Luca
  • Iorio Paolo
  • Giovannetti Giovanni
  • Conteduca Fabio
  • Iorio Raffaele
Original Paper

Abstract

Purpose

The use of available mechanical methods to measure anterior tibial translation (ATT) in anterior cruciate ligament (ACL)-deficient knees are limited by size and costs. This study evaluated the performance of a portable device based on a downloadable electronic smartphone application to measure ATT in ACL-deficient knees.

Methods

A specific smartphone application (SmartJoint) was developed for this purpose. Two independent observers nonsequentially measured the amount of ATT during execution of a maximum manual Lachman test in 35 patients with an ACL-deficient knee using KT 1000 and SmartJoint on both involved and uninvolved knees. As each examiner performed the test three times on each knee, a total of 840 measurements were collected. Statistical analysis compared intertest, interobserver and intra-observer reliability using the interclass correlation coefficient (ICC). An ICC > 0.75 indicates excellent reproducibility among measurements.

Results

Mean amount of ATT on uninvolved knees was 6.1 mm [standard deviation (SD = 2)] with the KT 1000 and 6.4 mm (SD = 2) with SmartJoint. Mean side-to-side difference was 8.1 mm. (SD = 4) with KT 1000 and 8.3 mm (SD = 3) with SmartJoint. Intertest reliability between the two methods yielded an ICC 0.797 [95 % confidence interval (CI) 0.717–0.857] for the uninvolved knee and of 0.987 (CI 0.981–0.991) for the involved knee. Interobserver ICC for SmartJoint and KT 1000 was 0.957 (CI 0.927–0.976) for the uninvolved knee and 0.992 (CI 0.986–0.996) for the involved knee and 0.973 (CI 0.954–0.985) for the uninvolved knee and 0.989 (CI 0.981–0.994) for involved knee, respectively.

Conclusion

The performance of SmartJoint is comparable and highly correlated with measurements obtained from KT 1000. SmartJoint may provide a truly portable, noninvasive, accurate, reliable, inexpensive and widely accessible method to characterize ATT in ACL-deficient knee

Keywords

Smartphone Anterior cruciate ligament Anterior tibial translation KT 1000 

References

  1. 1.
    Torg JS, Conrad W, Kalen V (1976) Clinical diagnosis of anterior cruciate ligament instability in the athlete. Am J Sports Med 4:84–93PubMedCrossRefGoogle Scholar
  2. 2.
    His RS, Hotaling JM, Hartzier AL, Holt SK, Walsh TJ (2013) Validity and reliability of a smartphone application for the assessment of penile deformity in Peyronie’s disease. J Sex Med 10:1867–1873CrossRefGoogle Scholar
  3. 3.
    Jenny Jean-Yves (2013) Measurement of knee flexion angle with a smartphone application is precise and accurate. The Journal of Arthroplasty 28:784–787Google Scholar
  4. 4.
    Boyer P, Djian P, Christel P, Paoletti X (2004) Reliability of the KT 1000 arthrometer (Medmetric) for measuring anterior knee laxity: comparison with Telos in 147 knees. Rev Chir Orthop Reparatrice App Mot 90:757–764CrossRefGoogle Scholar
  5. 5.
    Monaco E, Labianca L, Maestri B, de Carli A, Conteduca F, Ferretti A (2009) Instrumented measurements of knee laxity: KT-1000 vs navigation. Knee Surg Sports Traum Arthrosc 17:617–621CrossRefGoogle Scholar

Copyright information

© SICOT aisbl 2014

Authors and Affiliations

  • Ferretti Andrea
    • 1
  • Valeo Luigi
    • 1
  • Mazza Daniele
    • 1
    Email author
  • Muliere Luca
    • 1
  • Iorio Paolo
    • 1
  • Giovannetti Giovanni
    • 1
  • Conteduca Fabio
    • 1
  • Iorio Raffaele
    • 1
  1. 1.Orthopaedic Unit and “Kirk Kilgour” Sports Injury Centre S. Andrea HospitalUniversity of RomeRome (RM)Italy

Personalised recommendations