MR imaging of the human hand and wrist at 7 T

  • Björn BehrEmail author
  • Jörg Stadler
  • Henrik J. Michaely
  • Hans-Georg Damert
  • Wolfgang Schneider
Technical Report

Abstract

Objective

The purpose of this study was to evaluate the feasibility, quality, and possible future implications of magnetic resonance imaging (MRI) of the human hand and wrist at 7 T.

Materials and methods

Images of the left hand of a healthy volunteer were acquired with a 7- and a 1.5-T whole body system and comparatively analyzed. Axial and coronal two-dimensional gradient echo (GRE) images with inflow saturation, coronal 3D GRE images, and time-of-flight angiographies were obtained without averaging. Image details were related to the complex hand anatomy.

Results

With the 7-T protocols established in this study, high-quality and high-resolution images of the hand and wrist were obtained. In the 2D GRE images at 7 T, small anatomical structures of the hand were depicted in vivo with superior detail and resolution, compared to 1.5 T and published studies at lower field strength. Signal-to-noise ratios (SNRs) were approximately five times higher at 7 T compared to 1.5 T. Additionally, thin 3D GRE images with good quality of the whole hand were obtained in a short acquisition time. Moreover, time-of-flight angiographies of the small hand arteries have been acquired without the application of contrast agents.

Conclusion

Seven-tesla imaging of the hand can be used in vivo with ultra-high resolution and sufficient SNR. It allows for exact delineation of most anatomical structures including nerves, muscles, tendons, ligaments, cartilage, and blood vessels.

Keywords

MRI 7 Tesla Magnetic resonance angiography Hand Wrist Finger 

Notes

Acknowledgments

The authors thank Özlem Krischek for 3D processing of the ToF data and Oliver Speck for critically reading the manuscript.

References

  1. 1.
    Steinbach LS, Smith DK. MRI of the wrist. Clin Imaging. 2000; 24: 298–322.CrossRefGoogle Scholar
  2. 2.
    Garcia J, Bianchi S. Diagnostic imaging of tumors of the hand and wrist. Eur Radiol. 2001; 11: 1470–1482.CrossRefGoogle Scholar
  3. 3.
    Mosher TJ. Musculoskeletal imaging at 3T: current techniques and future applications. Magn Reson Imaging Clin N Am. 2006; 14: 63–76.CrossRefGoogle Scholar
  4. 4.
    Schmitt F, Grosu D, Mohr C, et al. 3 Tesla MRI: successful results with higher field strengths. Radiologe. 2004; 44: 31–47.CrossRefGoogle Scholar
  5. 5.
    Hoult DI, Phil D. Sensitivity and power deposition in a high-field imaging experiment. J Magn Reson Imaging. 2000; 12: 46–67.CrossRefGoogle Scholar
  6. 6.
    Abduljalil AM, Kangarlu A, Yu Y, Robitaille PM. Macroscopic susceptibility in ultra high field MRI. II. Acquisitions of spin echo images from the human head. J Comput Assist Tomogr. 1999; 23: 842–844.CrossRefGoogle Scholar
  7. 7.
    Saupe N, Prüssmann KP, Luechinger R, Bösiger P, Marincek B, Weishaupt D. MR imaging of the wrist: comparison between 1.5- and 3-T MR imaging—preliminary experience. Radiology. 2005; 234: 256–264.CrossRefGoogle Scholar
  8. 8.
    Lenk S, Ludescher B, Martirosan P, Schick F, Claussen CD, Schlemmer HP. 3.0T high-resolution MR imaging of carpal ligaments and TFCC. Rofo. 2004; 176: 664–667.CrossRefGoogle Scholar
  9. 9.
    Tanaka T, Yoshioka H, Ueno T, Shindo M, Ochiai N. Comparison between high-resolution MRI with a microscopy coil and arthroscopy in triangular fibrocartilage complex injury. J Hand Surg [Am]. 2006; 31: 1308–1314.CrossRefGoogle Scholar
  10. 10.
    Farooki S, Ashman CJ, Yu JS, Abduljalil A, Chakeres D. In vivo high-resolution MR imaging of the carpal tunnel at 8.0 tesla. Skelet Radiol. 2002; 31(8): 445–450.CrossRefGoogle Scholar
  11. 11.
    Ashman CJ, Farooki S, Abduljalil AM, Chakeres DW. In vivo high resolution coronal MRI of the wrist at 8.0 tesla. J Comput Assist Tomogr. 2002; 26: 387–391.CrossRefGoogle Scholar
  12. 12.
    De Maeseneer M, Van Roy P, Jacobson JA, Jamadar DA. Normal MR imaging findings of the midhand and fingers with anatomic correlation. Eur J Radiol. 2005; 56: 278–285.CrossRefGoogle Scholar
  13. 13.
    Ragheb D, Stanley A, Gentili A, Hughes T, Chung CB. MR imaging of the finger tendons: normal anatomy and commonly encountered pathology. Eur J Radiol. 2005; 56: 296–306.CrossRefGoogle Scholar
  14. 14.
    Clavero JA, Alomar X, Monill JM, et al. MR imaging of ligament and tendon injuries of the fingers. Radiographics. 2002; 22: 237–256.CrossRefGoogle Scholar
  15. 15.
    Hauger O, Chung CB, Lektrakul N, et al. Pulley system in the fingers: normal anatomy and simulated lesions in cadavers at MR imaging, CT, and US with and without contrast material distention of the tendon sheath. Radiology. 2000; 217: 201–212.CrossRefGoogle Scholar
  16. 16.
    Lee VS, Lee HM, Rofsky NM. Magnetic resonance angiography of the hand. A review. Invest Radiol. 1998; 33: 687–698.CrossRefGoogle Scholar
  17. 17.
    Rofsky NM. MR angiography of the hand and wrist. Magn Reson Imaging Clin N Am. 1995; 3: 345–359.PubMedGoogle Scholar

Copyright information

© ISS 2009

Authors and Affiliations

  • Björn Behr
    • 1
    Email author
  • Jörg Stadler
    • 2
  • Henrik J. Michaely
    • 3
  • Hans-Georg Damert
    • 1
  • Wolfgang Schneider
    • 1
  1. 1.Department of Plastic, Reconstructive and Hand SurgeryUniversity of MagdeburgLeipziger Str. 44MagdeburgGermany
  2. 2.Leibniz Institute for NeurobiologyUniversity of MagdeburgMagdeburgGermany
  3. 3.Institute of Clinical Radiology and Nuclear MedicineUniversity Medical Center MannheimMannheimGermany

Personalised recommendations