Developing low-dose C-arm CT imaging for temporomandibular joint (TMJ) disorder in interventional radiology

  • Xiaowei Zhu
  • Marc Felice
  • Laura Johnson
  • Marily Sarmiento
  • Anne Marie Cahill
Original Article

Abstract

Background

Manufacturers have provided C-arm CT imaging technologies for applications in interventional radiology in recent years. However, clinical imaging protocols and radiation doses have not been well studied or reported.

Objective

The purpose of this study is to develop low-dose settings for clinically acceptable CT imaging of temporomandibular joint in interventional radiology suites, using a C-arm imaging angiography system.

Materials and methods

CT scans were performed with a flat-panel digital C-arm angiographic system on a 5-year-old anthropomorphic phantom. The CTDI was determined for various rotation times, dose settings and Cu filter selections. The CTDI values were compared with those of conventional low-dose CT for the same phantom. The effectiveness of using Cu filters to reduce dose was also investigated. Images were reviewed by a senior radiologist for clinical acceptance.

Results

The manufacturer’s default setting gave an equivalent CTDI of 4.8 mGy. Optimizing the dose settings and adding copper filtration reduced the radiation dose by 94%. This represents a 50% reduction from conventional CT.

Conclusion

Use of Cu filters and low-dose settings significantly reduced radiation dose from that of standard settings. This phantom study process successfully guided the clinical implementation of low-dose studies for all ages at our institution.

Keywords

C-arm CT Radiation dose Temporomandibular joint arthropathy 

References

  1. 1.
    Racadio JM, Drazenko B, Homann R et al (2007) Live 3D guidance in the interventional radiology suite. AJR 189:W357–W364PubMedCrossRefGoogle Scholar
  2. 2.
    Racadio JM, Bradley LF, Jones BV et al (2006) Three-dimensional rotational angiography of neurovascular lesions in pediatric patients. AJR 186:75–84PubMedCrossRefGoogle Scholar
  3. 3.
    Soulez G, Dubois J, Giroux MF et al (2006) Value of angiographic computed tomography (Dyna-CT) in the treatment of vascular anomalies. Education exhibit (poster). RSNA 92th Scientific Assembly and Annual Meeting, Chicago, IL, USAGoogle Scholar
  4. 4.
    Smyth JM, Sutton DG, Houston JG et al (2006) Evaluation of quality of CT-like images obtained using a commercial flat panel detector system. Biomed Imaging Interv J 2:e48CrossRefGoogle Scholar
  5. 5.
    Fahrig R, Dixon R, Payne T et al (2006) Dose and image quality for a cone-beam c-arm CT system. Med Phys 33:4541–4550PubMedCrossRefGoogle Scholar
  6. 6.
    Cahill AM, Baskin KM, Kaye RD (2007) CT-guided percutaneous steroid injection for management of inflammatory arthropathy of the temporomandibular joint in children. AJR 188:182–186PubMedCrossRefGoogle Scholar

Copyright information

© Springer-Verlag 2010

Authors and Affiliations

  • Xiaowei Zhu
    • 1
  • Marc Felice
    • 2
  • Laura Johnson
    • 3
  • Marily Sarmiento
    • 4
  • Anne Marie Cahill
    • 1
  1. 1.Department of RadiologyThe Children’s Hospital of PhiladelphiaPhiladelphiaUSA
  2. 2.Environmental Health and Radiation SafetyUniversity of PennsylvaniaPhiladelphiaUSA
  3. 3.Siemens Healthcare SectorComputed Tomography DivisionShanghaiChina
  4. 4.Angiography and X-ray DivisionSiemens Medical SolutionsHoffman EstatesUSA

Personalised recommendations