The impact of stent design on the structural mechanics of the crossing Y-stent: an in vitro study

  • Chang-Young Lee
  • Seong-Ho Park
  • Chang-Hyun Kim
  • Goetz Benndorf
Interventional Neuroradiology

Abstract

Introduction

Crossing Y-stent technique is a valid option for coiling wide-necked bifurcation aneurysms. Two main designs of stents, the closed- (CCS) and open-cell (OCS), are used in combination for a crossing Y-construct. This in vitro study was conducted to assess the mechanical characteristics of each stent-combination and to suggest an optimal combination for clinical practice.

Methods

The Enterprise and the Neuroform3 stents were used as closed-cell and open-cell stents, respectively. Four different Y-stent combinations; double CCSs (CCS-CCS; Enterprise-Enterprise), OCS followed by CCS (OCS-CCS; Neuroform-Enterprise), CCS followed by OCS (CCS-OCS; Enterprise-Neuroform), and double OCSs (OCS-OCS; Neuroform-Neuroform) were tested in a bifurcation aneurysm model. Images of the model were obtained by digital subtraction angiography (DSA), and morphological changes of Y-constructs caused by interaction between stents were compared.

Results

Double OCSs showed the best stent apposition to model branches and no collapse of the second stent at the intersection, whereas Y-stent combinations using a CCS as the second stent showed tubular collapse of the second stent at the crossing point. These combinations revealed unsatisfactory apposition to the model branch in which the second stent was deployed. Most narrowing of the second stent was noted in the double CCSs construct, which resulted in poorest stent apposition.

Conclusion

Based on a simple in vitro experiment, we suggest that double OCSs Y-construct is optimal for achieving best stent-wall apposition. Furthermore, our findings suggest that Y-stent combinations utilizing a CCS as the second stent may increase the risk of thromboembolic complications due to poor stent-wall apposition.

Keywords

Cerebral aneurysm Coiling Endovascular treatment Stent design Y-stent 

References

  1. 1.
    Chow MM, Woo HH, Masaryk TJ, Rasmussen PA (2004) A novel endovascular treatment of a wide-necked basilar apex aneurysm by using a Y-configuration, double-stent technique. AJNR Am J Neuroradiol 25:509–512PubMedGoogle Scholar
  2. 2.
    Rohde S, Bendszus M, Hartmann M, Hähnel S (2010) Treatment of a wide-necked aneurysm of the anterior cerebral artery using two Enterprise stents in “Y”-configuration stenting technique and coil embolization: a technical note. Neuroradiology 52:231–235PubMedCrossRefGoogle Scholar
  3. 3.
    Sani S, Lopes DK (2005) Treatment of a middle cerebral artery bifurcation aneurysm using a double neuroform stent “Y” configuration and coil embolization: technical case report. Neurosurgery 57(1 Suppl):E209PubMedCrossRefGoogle Scholar
  4. 4.
    Spiotta AM, Gupta R, Fiorella D et al (2011) Mid-term results of endovascular coiling of wide-necked aneurysms using double stents in a Y configuration. Neurosurgery 69:421–429PubMedCrossRefGoogle Scholar
  5. 5.
    Kanaan H, Jankowitz B, Aleu A et al (2010) In-stent thrombosis and stenosis after neck-remodeling device-assisted coil embolization of intracranial aneurysms. Neurosurgery 67:1523–1532PubMedCrossRefGoogle Scholar
  6. 6.
    Mocco J, Fargen KM, Albuquerque FC et al (2011) Delayed thrombosis or stenosis following enterprise-assisted stent-coiling: is it safe? Midterm results of the interstate collaboration of enterprise stent coiling. Neurosurgery 69:908–913PubMedCrossRefGoogle Scholar
  7. 7.
    Lee DH, Arat A, Morsi H, Shaltoni H, Harris JR, Mawad ME (2008) Dual antiplatelet therapy monitoring for neurointerventional procedures using a point-of-care platelet function test: a single-center experience. AJNR Am J Neuroradiol 29:1389–1394PubMedCrossRefGoogle Scholar
  8. 8.
    Yamada NK, Cross DT 3rd, Pilgram TK, Moran CJ, Derdeyn CP, Dacey RG Jr (2007) Effect of antiplatelet therapy on thromboembolic complications of elective coil embolization of cerebral aneurysms. AJNR Am J Neuroradiol 28:1778–1782PubMedCrossRefGoogle Scholar
  9. 9.
    Heller RS, Miele WR, Do-Dai DD, Malek AM (2011) Crescent sign on magnetic resonance angiography revealing incomplete stent apposition: correlation with diffusion-weighted changes in stent-mediated coil embolization of aneurysms. Neurosurgery 115:624–632CrossRefGoogle Scholar
  10. 10.
    Buchanan GL, Basavarajaiah S, Chieffo A (2012) Stent thrombosis: incidence, predictors and new technologies. Thrombosis 2012:956962PubMedCentralPubMedCrossRefGoogle Scholar
  11. 11.
    Capranzano P, Dangas G (2012) Late stent thrombosis: the last remaining obstacle in coronary interventional therapy. Curr Cardiol Rep 14:408–417PubMedCrossRefGoogle Scholar
  12. 12.
    Cook S, Wenaweser P, Togni M et al (2007) Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation. Circulation 115:2426–2434PubMedCrossRefGoogle Scholar
  13. 13.
    Tanaka N, Martin JB, Tokunaga K et al (2004) Conformity of carotid stents with vascular anatomy: evaluation in carotid models. AJNR Am J Neuroradiol 25:604–607PubMedGoogle Scholar
  14. 14.
    Cekirge HS, Yavuz K, Geyik S, Saatci I (2011) A novel “Y” stent flow diversion technique for the endovascular treatment of bifurcation aneurysms without endosaccular coiling. AJNR Am J Neuroradiol 32:1262–1268PubMedCrossRefGoogle Scholar
  15. 15.
    Lozen A, Manjila S, Rhiew R, Fessler R (2009) Y-stent-assisted coil embolization for the management of unruptured cerebral aneurysms: report of six cases. Acta Neurochir (Wien) 151:1663–1672CrossRefGoogle Scholar
  16. 16.
    Akgul E, Aksungur E, Balli T et al (2011) Y-stent-assisted coil embolization of wide-neck intracranial aneurysms. A single center experience. Interv Neuroradiol 17:36–48PubMedCentralPubMedGoogle Scholar
  17. 17.
    Kono K, Shintani A, Yoshimura R et al (2013) Triple antiplatelet therapy with addition of cilostazol to aspirin and clopidogrel for Y-stent-assisted coil embolization of cerebral aneurysms. Acta Neurochir (Wien) 155:1549–1557CrossRefGoogle Scholar
  18. 18.
    Thorell WE, Chow MM, Woo HH, Masaryk TJ, Rasmussen PA (2005) Y-configured dual intracranial stent-assisted coil embolization for the treatment of wide-necked basilar tip aneurysms. Neurosurgery 56:1035–1040PubMedGoogle Scholar
  19. 19.
    De Bock S, Iannaccone F, De Santis G et al (2012) Our capricious vessels: the influence of stent design and vessel geometry on the mechanics of intracranial aneurysm stent deployment. J Biomech 45:1353–1359PubMedCrossRefGoogle Scholar
  20. 20.
    Ebrahimi N, Claus B, Lee CY, Biondi A, Benndorf G (2007) Stent conformity in curved vascular models with simulated aneurysm necks using flat-panel CT: an in vitro study. AJNR Am J Neuroradiol 28:823–829PubMedGoogle Scholar
  21. 21.
    Sedat J, Chau Y, Mondot L, Vargas J, Szapiro J, Lonjon M (2009) Endovascular occlusion of intracranial wide-necked aneurysms with stenting (Neuroform) and coiling: mid-term and long-term results. Neuroradiology 51:401–409PubMedCrossRefGoogle Scholar
  22. 22.
    Peluso JP, van Rooij WJ, Sluzewski M, Beute GN (2008) A new self-expandable nitinol stent for the treatment of wide-neck aneurysms: initial clinical experience. AJNR Am J Neuroradiol 29:1405–1408PubMedCrossRefGoogle Scholar
  23. 23.
    Klisch J, Eger C, Sychra V, Strasilla C, Basche S, Weber J (2009) Stent-assisted coil embolization of posterior circulation aneurysms using solitaire ab: preliminary experience. Neurosurgery 65:258–266PubMedCrossRefGoogle Scholar
  24. 24.
    Valdivia y Alvarado M, Ebrahimi N, Benndorf G (2009) Study of conformability of the new leo plus stent to a curved vascular model using flat-panel detector computed tomography (DynaCT). Neurosurgery 64:130–134Google Scholar
  25. 25.
    Conrad MD, Brasiliense LB, Richie AN, Hanel RA (2013) Y stenting assisted coiling using a new low profile visible intraluminal support device for wide necked basilar tip aneurysms: a technical report. J Neurointerv Surg 0:1–5Google Scholar
  26. 26.
    Krischek O, Miloslavski E, Fischer S, Shrivastava S, Henkes H (2011) A comparison of functional and physical properties of self-expanding intracranial stents [Neuroform3, Wingspan, Solitaire, Leo+, Enterprise]. Minim Invasive Neurosurg 54:21–28PubMedCrossRefGoogle Scholar
  27. 27.
    Kadkhodayan Y, Rhodes N, Blackburn S, Derdeyn CP, Cross DT 3rd, Moran CJ (2013) Comparison of Enterprise with Neuroform stent-assisted coiling of intracranial aneurysms. AJR Am J Roentgenol 200:872–878PubMedCrossRefGoogle Scholar
  28. 28.
    Lijnen HR, Collen D (1997) Endothelium in hemostasis and thrombosis. Prog Cardiovasc Dis 39:343–350PubMedCrossRefGoogle Scholar
  29. 29.
    Fargen KM, Mocco J, Neal D et al (2013) A multicenter study of stent-assisted coiling of cerebral aneurysms with a Y configuration. Neurosurgery 73:466–472PubMedCrossRefGoogle Scholar

Copyright information

© Springer-Verlag Berlin Heidelberg 2014

Authors and Affiliations

  • Chang-Young Lee
    • 1
  • Seong-Ho Park
    • 1
  • Chang-Hyun Kim
    • 1
  • Goetz Benndorf
    • 2
  1. 1.Department of NeurosurgeryKeimyung University School of MedicineDaeguRepublic of Korea
  2. 2.Department of RadiologyBaylor College of MedicineHoustonUSA

Personalised recommendations