Compact devices for generation of reference trace VOC mixtures: a new concept in assuring quality at chemical and biochemical laboratories

  • Alessia Demichelis
  • Céline Pascale
  • Maricarmen Lecuna
  • Bernhard Niederhauser
  • Guido Sassi
  • Maria Paola Sassi
Research Paper

Abstract

Volatile organic compounds (VOCs) in gas mixtures at trace level (nmol/mol) are routinely measured by chemical and biochemical laboratories as climate indicators, indoor air quality pollutants from building materials emissions, contaminants in food and beverages, and biomarkers in body fluids (blood, urine, breath) of occupational exposure or human diseases. Current analytical instruments used for measurements are gas chromatographs equipped with various injector and detector configurations. The assurance of measurement quality is done by using a huge amount of certified liquid VOC standard solutions (or gaseous VOC standard cylinders) with multiple dilutions to reach the required trace level. This causes high standard uncertainty in instrument calibrations, high cost, and high consumption of analysis and laboratory personal time. In this paper, we present the implementation of portable generators producing VOC gas standards at trace level for automatic and direct calibration of VOC detectors employed in various contexts, removing the need for preparation of matrix calibration standards in cylinders. Two compact devices in-house developed by two national metrology institutes—the Istituto Nazionale di Ricerca Metrologica (INRIM) and the Federal Institute of Metrology (METAS)—are here used to dynamically generate reference gas mixtures in an SI traceable way. The two devices are based on different technologies: diffusion and permeation, for INRIM and METAS, respectively. A metrological characterization is given and the practical implementation at chemical and biochemical laboratories is discussed.

Graphical abstract

Onsite calibration with transportable generation system with similar performances to primary laboratory devices.

Keywords

Reference VOC gas mixtures Dynamic generation Metrological traceability GC calibration Chemical laboratory Biochemical laboratory 

Notes

Compliance with ethical standards

Conflict of interest

The authors declare that they have no conflict of interest.

References

  1. 1.
    Barrie L, Braathen G. Highlights from the Global Atmosphere Watch Programme; 2017.Google Scholar
  2. 2.
    Penkett S, Barrie L. WMO/GAW Expert Workshop on Global Long-term Measurements of Volatile Organic Compounds, WMO TD No. 1373. WMO; 2007.Google Scholar
  3. 3.
    Wolkoff P. Trends in Europe to reduce the indoor air pollution of VOCs. Indoor Air. 2003;13(Suppl 6):5–11.CrossRefGoogle Scholar
  4. 4.
    European Metrology Research Programme. ENV56 KEY-VOCs research project 2014–2017. n.d. http://www.key-vocs.eu/.
  5. 5.
    Yu C, Crump D. A review of the emission of VOCs from polymeric materials used in buildings. Build Environ. 1998;33:357–74.  https://doi.org/10.1016/S0360-1323(97)00055-3.CrossRefGoogle Scholar
  6. 6.
    Ashley DL, Bonin MA, Cardinali FL, McCraw JM, Wooten JV. Blood concentrations of volatile organic compounds in a nonoccupationally exposed US population and in groups with suspected exposure. Clin Chem. 1994;40:1401–4.Google Scholar
  7. 7.
    Heinrich-Ramm R, Jakubowski M, Heinzow B, Christensen JM, Olsen E, Hertel O. Biological monitoring for exposure to volatile organic compounds (VOCs) (IUPAC recommendations 2000). Pure Appl Chem. 2000;72:385–436.  https://doi.org/10.1351/pac200072030385.CrossRefGoogle Scholar
  8. 8.
    Sklerov JH, Couper FJ. Calculation and verification of blood ethanol measurement uncertainty for headspace gas chromatography. J Anal Toxicol. 2011;35:402–10.CrossRefGoogle Scholar
  9. 9.
    Mathew T, Pownraj P, Abdulla S, Pullithadathil B. Technologies for clinical diagnosis using expired human breath analysis. Diagnostics. 2015;5:27–60.  https://doi.org/10.3390/diagnostics5010027.CrossRefGoogle Scholar
  10. 10.
    Schmidt K, Podmore I. Current challenges in volatile organic compounds analysis as potential biomarkers of cancer. J Biomarkers. 2015;2015:981458.  https://doi.org/10.1155/2015/981458.CrossRefGoogle Scholar
  11. 11.
    Pereira J, Porto-Figueira P, Cavaco C, Taunk K, Rapole S, Dhakne R, et al. Breath analysis as a potential and non-invasive frontier in disease diagnosis: an overview. Meta. 2015;5:3–55.  https://doi.org/10.3390/metabo5010003.Google Scholar
  12. 12.
    Taipale R, Ruuskanen TM, Rinne J, Kajos MK, Hakola H, Pohja T, et al. Technical note: quantitative long-term measurements of VOC concentrations by PTR-MS—measurement, calibration, and volume mixing ratio calculation methods. Atmos Chem Phys. 2008;8:6681–98.  https://doi.org/10.5194/acp-8-6681-2008.CrossRefGoogle Scholar
  13. 13.
    Höllbacher E, Ters T, Rieder-Gradinger C, Srebotnikb E. Emissions of indoor air pollutants from six user scenarios in a model room. Atmos Environ. 2017;150:389–94.  https://doi.org/10.1016/J.ATMOSENV.2016.11.033.CrossRefGoogle Scholar
  14. 14.
    Gras R, Luong J, Shellie RA. Gas chromatography and diode array detection for the direct measurement of carbon disulfide in challenging matrices. Anal Methods. 2017;9:3908–13.  https://doi.org/10.1039/C7AY00392G.CrossRefGoogle Scholar
  15. 15.
    Ross BM, Babgi R. Volatile compounds in blood headspace and nasal breath. J Breath Res. 2017;11:46001.  https://doi.org/10.1088/1752-7163/aa7d10.CrossRefGoogle Scholar
  16. 16.
    Spinelle L, Gerboles M, Kok G, Persijn S, Sauerwald T. Review of portable and low-cost sensors for the ambient air monitoring of benzene and other volatile organic compounds. Sensors. 2017;17:1520.  https://doi.org/10.3390/s17071520.CrossRefGoogle Scholar
  17. 17.
    Cristofanelli P, Brattich E, Decesari S, Landi TC, Maione M, Putero D, et al. Investigation of atmospheric reactive gases at Mt. Cimone. Springer: Cham; 2018. p. 45–73.  https://doi.org/10.1007/978-3-319-61127-3_3.Google Scholar
  18. 18.
    Mansour SA. Residual pesticides and heavy metals analysis in food. Analysis of Food Toxins and Toxicants. Chichester, UK: John Wiley & Sons, Ltd; 2017. p. 537–70.  https://doi.org/10.1002/9781118992685.ch18.CrossRefGoogle Scholar
  19. 19.
    Horn W, Richter M, Nohr M, Wilke O, Jann O. Application of a novel reference material in an international round robin test on material emissions testing. Indoor Air. 2017;  https://doi.org/10.1111/ina.12421.
  20. 20.
    Richter M, Mull B, Horn W, Brödner D, Mölders N, Renner M. Reproducibly emitting reference material on thermoplastic polyurethane basis for quality assurance/quality control of emission test chamber measurements. Build Environ. 2017;122:230–6.  https://doi.org/10.1016/J.BUILDENV.2017.06.005.CrossRefGoogle Scholar
  21. 21.
    Haider SI, Asif ES. Quality control training manual: comprehensive training guide for API, finished pharmaceutical and biotechnologies laboratories. Boca Raton: CRC Press; 2011.Google Scholar
  22. 22.
    Hwang R-J, Beltran J, Rogers C, Barlow J, Razatos G. Measurement of uncertainty for aqueous ethanol wet-bath simulator solutions used with evidential breath testing instruments. J Forensic Sci. 2016;61:1359–63.  https://doi.org/10.1111/1556-4029.13133.CrossRefGoogle Scholar
  23. 23.
    Sassi G, Demichelis A, Sassi MP. Uncertainty analysis of the diffusion rate in the dynamic generation of volatile organic compound mixtures. Meas Sci Technol. 2011;22:105104.  https://doi.org/10.1088/0957-0233/22/10/105104.CrossRefGoogle Scholar
  24. 24.
    Nelson GO. Gas mixtures: preparation and control. Lewis Publishers; 1992.Google Scholar
  25. 25.
    International Organization of Legal Metrology. Evidential breath analyzers: Ethylometers. OIML R 126 2012;126.Google Scholar
  26. 26.
    Demichelis A, Sassi G, Lecuna M, Sassi MP. Molar fraction stability in dynamic preparation of reference trace gas mixtures. IET Sci Meas Technol. 2016;10:414–9.  https://doi.org/10.1049/iet-smt.2015.0051.CrossRefGoogle Scholar
  27. 27.
    Sassi G, Demichelis A, Lecuna M, Sassi MP. Preparation of standard VOC mixtures for climate monitoring. Int J Environ Anal Chem. 2015;95:1195–207.  https://doi.org/10.1080/03067319.2015.1016015.CrossRefGoogle Scholar
  28. 28.
    Sassi G, Demichelis A, Sassi MP. Air flow rate thermal control system at low pressure drop. Flow Meas Instrum. 2014;35:44–7.  https://doi.org/10.1016/j.flowmeasinst.2013.12.001.CrossRefGoogle Scholar
  29. 29.
    Demichelis A, Sassi G, Sassi MP. A handy method for reproducible and stable measurements of VOC at trace level in air. Springer International Publishing; 2014, p. 109–13. doi: https://doi.org/10.1007/978-3-319-00684-0_21.
  30. 30.
    Demichelis A, Sassi G, Sassi MP. Metrological performances of mass flow controllers for dynamic gas dilution. Accred Qual Assur. 2013:181–6.  https://doi.org/10.1007/s00769-013-0974-y.
  31. 31.
    Gautrois M, Koppmann R. Diffusion technique for the production of gas standards for atmospheric measurements. J Chromatogr A. 1999;848:239–49.  https://doi.org/10.1016/S0021-9673(99)00424-0.CrossRefGoogle Scholar
  32. 32.
    Sassi G, Vernai AM, Ruggeri B. Quantitative estimation of uncertainty in human risk analysis. J Hazard Mater. 2007;145:296–304.  https://doi.org/10.1016/j.jhazmat.2006.11.020.CrossRefGoogle Scholar
  33. 33.
    European Metrology Research Programme, ENV01. MACPoll Metrology of Chemical Pollutants in Air. 2011–2014 n.d. http://www.macpoll.eu/.
  34. 34.
    Pascale C, Guillevic M, Ackermann A, Leuenberger D, Niederhauser BC. Two generators to produce SI-traceable reference gas mixtures for reactive compounds at atmospheric levels. Meas Sci Technol. 2017;  https://doi.org/10.1088/1361-6501/aa870c.
  35. 35.
    CMOS. Technology Working Principle and its Applications n.d. https://www.elprocus.com/cmos-working-principle-and-applications/. Accessed 30 Sept 2017.
  36. 36.
    Haerri H-P, Macé T, Waldén J, Pascale C, Niederhauser B, Wirtz K, et al. Dilution and permeation standards for the generation of NO, NO2 and SO2 calibration gas mixtures. Meas Sci Technol. 2017;28:35801.  https://doi.org/10.1088/1361-6501/aa543d.CrossRefGoogle Scholar
  37. 37.
    Zhang NF. Allan variance of time series models for measurement data. Metrologia. 2008;45:549–61.  https://doi.org/10.1088/0026-1394/45/5/009.CrossRefGoogle Scholar
  38. 38.
    Lucero DP. Performance characteristics of permeation tubes. Anal Chem. 1971;43:1744–9.  https://doi.org/10.1021/ac60307a005.CrossRefGoogle Scholar
  39. 39.
    International Organization for Standardization (ISO). ISO 6145-10:2002—Gas analysis—preparation of calibration gas mixtures using dynamic volumetric methods—part 10: permeation method. 2002.Google Scholar
  40. 40.
    International Organization for Standardization (ISO). ISO 6145-8:2005—Gas analysis—preparation of calibration gas mixtures using dynamic volumetric methods—part 8: diffusion method. 2005.Google Scholar
  41. 41.
    Stan H-J. Pesticide residue analysis in foodstuffs applying capillary gas chromatography with mass spectrometric detection: state-of-the-art use of modified DFG-multimethod S19 and automated data evaluation. J Chromatogr A. 2000;892:347–77.  https://doi.org/10.1016/S0021-9673(00)00308-3.CrossRefGoogle Scholar
  42. 42.
    Valverde A, Aguilera A, Valverde-Monterreal A. Practical and valid guidelines for realistic estimation of measurement uncertainty in multi-residue analysis of pesticides. Food Control. 2017;71:1–9.  https://doi.org/10.1016/J.FOODCONT.2016.06.017.CrossRefGoogle Scholar

Copyright information

© Springer-Verlag GmbH Germany, part of Springer Nature 2018

Authors and Affiliations

  • Alessia Demichelis
    • 1
  • Céline Pascale
    • 2
  • Maricarmen Lecuna
    • 3
  • Bernhard Niederhauser
    • 2
  • Guido Sassi
    • 1
    • 3
  • Maria Paola Sassi
    • 1
  1. 1.Istituto Nazionale di Ricerca Metrologica INRIMTorinoItaly
  2. 2.Federal Institute of Metrology METASBern-WabernSwitzerland
  3. 3.Dipartimento di Scienza Applicata e TecnologiaPolitecnico di TorinoTorinoItaly

Personalised recommendations