The risk of notching the anterior femoral cortex with the use of navigation systems in total knee arthroplasty

  • Yukihide MinodaEmail author
  • Akio Kobayashi
  • Hiroyoshi Iwaki
  • Ikebuchi Mitsuhiko
  • Yoshinori Kadoya
  • Hirotsugu Ohashi
  • Kunio Takaoka
  • Hiroaki Nakamura
Knee

Abstract

Use of navigation systems has recently been introduced in total knee arthroplasty (TKA) to achieve more reliable prosthetic alignment. In the sagittal plane, there are two important requirements for navigation systems: (1) perpendicular cut to the femoral mechanical axis and (2) prevention of notching of anterior femoral cortex. These two requirements, however, may conflict. The angles between the line of the anterior femoral cortex and four sagittal femoral mechanical axes for navigation systems using radiographs of the entire lower extremity, while standing were measured and compared. These four sagittal axes simulated on the radiographs in navigation systems were in extension relative to the line of the anterior femoral cortex in 40–85% of cases in male and 65–100% in elderly female. The present study showed that navigation systems have the potential risk for notching of anterior femoral cortex.

Keywords

Total knee arthroplasty Navigation system Notching of anterior femoral cortex 

References

  1. 1.
    Bathis H, Perlick L, Tingart M et al (2004) Alignment in total knee arthroplasty. A comparison of computer-assisted surgery with the conventional technique. J Bone Joint Surg Br 86:682–687CrossRefPubMedGoogle Scholar
  2. 2.
    Berger RA, Rubash HE, Seel MJ et al (1993) Determining the rotational alignment of the femoral component in total knee arthroplasty using the epicondylar axis. Clin Orthop Relat Res 286:40–47PubMedGoogle Scholar
  3. 3.
    Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMedGoogle Scholar
  4. 4.
    Bolognesi M, Hofmann A (2005) Computer navigation versus standard instrumentation for TKA: a single-surgeon experience. Clin Orthop Relat Res 440:162–169CrossRefPubMedGoogle Scholar
  5. 5.
    Bonutti PM, Dethmers DA, McGrath MS et al (2008) Navigation did not improve the precision of minimally invasive knee arthroplasty. Clin Orthop Relat Res 466:2730–2735CrossRefPubMedGoogle Scholar
  6. 6.
    Chauhan SK, Clark GW, Lloyd S et al (2004) Computer-assisted total knee replacement. A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT protocol). J Bone Joint Surg Br 86:818–823CrossRefPubMedGoogle Scholar
  7. 7.
    Clarke HD, Hentz JG (2008) Restoration of femoral anatomy in TKA with unisex and gender-specific components. Clin Orthop Relat Res 466:2711–2716CrossRefPubMedGoogle Scholar
  8. 8.
    Haaker RG, Stockheim M, Kamp M et al (2005) Computer-assisted navigation increases precision of component placement in total knee arthroplasty. Clin Orthop Relat Res 433:152–159CrossRefPubMedGoogle Scholar
  9. 9.
    Hart R, Janecek M, Cizmár I et al (2006) Minimally invasive and navigated implantation for total knee arthroplasty: X-ray analysis and early clinical results. Orthopade 35:552–557CrossRefPubMedGoogle Scholar
  10. 10.
    Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73:709–714PubMedGoogle Scholar
  11. 11.
    Jenny JY, Clemens U, Kohler S et al (2005) Consistency of implantation of a total knee arthroplasty with a non-image-based navigation system: a case–control study of 235 cases compared with 235 conventionally implanted prostheses. J Arthroplasty 20:832–839CrossRefPubMedGoogle Scholar
  12. 12.
    Lesh ML, Schneider DJ, Deol G et al (2000) The consequences of anterior femoral notching in total knee arthroplasty. A biomechanical study. J Bone Joint Surg Am 82:1096–1101PubMedGoogle Scholar
  13. 13.
    Lützner J, Krummenauer F, Wolf C et al (2008) Computer-assisted and conventional total knee replacement: a comparative, prospective, randomised study with radiological and CT evaluation. J Bone Joint Surg Br 90:1039–1044CrossRefPubMedGoogle Scholar
  14. 14.
    Middleton FR, Palmer SH (2007) How accurate is Whiteside’s line as a reference axis in total knee arthroplasty? Knee 14:204–207CrossRefPubMedGoogle Scholar
  15. 15.
    Minoda Y, Kobayashi A, Iwaki H et al (2008) Sagittal alignment of the lower extremity while standing in Japanese male. Arch Orthop Trauma Surg 128:435–442CrossRefPubMedGoogle Scholar
  16. 16.
    Minoda Y, Kobayashi A, Iwaki H et al (2009) TKA sagittal alignment with navigation systems and conventional techniques vary only a few degrees. Clin Orthop Relat Res 467:1000–1006CrossRefPubMedGoogle Scholar
  17. 17.
    Mizu-uchi H, Matsuda S, Miura H et al (2008) The evaluation of post-operative alignment in total knee replacement using a CT-based navigation system. J Bone Joint Surg Br 90:1025–1031CrossRefPubMedGoogle Scholar
  18. 18.
    Oswald MH, Jakob RP, Schneider E et al (1993) Radiological analysis of normal axial alignment of femur and tibia in view of total knee arthroplasty. J Arthroplasty 8:419–426CrossRefPubMedGoogle Scholar
  19. 19.
    Sparmann M, Wolke B, Czupalla H et al (2003) Positioning of total knee arthroplasty with and without navigation support. A prospective, randomised study. J Bone Joint Surg Br 85:830–835PubMedGoogle Scholar
  20. 20.
    Stulberg SD (2003) How accurate is current TKR instrumentation? Clin Orthop Relat Res 416:177–184CrossRefPubMedGoogle Scholar
  21. 21.
    Tew M, Waugh W (1985) Tibiofemoral alignment and the results of knee replacement. J Bone Joint Surg Br 67:551–556PubMedGoogle Scholar
  22. 22.
    Tria AJ Jr (2006) The evolving role of navigation in minimally invasive total knee arthroplasty. Am J Orthop 35(7 Suppl):18–22PubMedGoogle Scholar
  23. 23.
    Wasielewski RC, Galante JO, Leighty RM et al (1994) Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin Orthop Relat Res 299:31–43PubMedGoogle Scholar
  24. 24.
    Zalzal P, Backstein D, Gross AE et al (2006) Notching of the anterior femoral cortex during total knee arthroplasty characteristics that increase local stresses. J Arthroplasty 21:737–743CrossRefPubMedGoogle Scholar

Copyright information

© Springer-Verlag 2009

Authors and Affiliations

  • Yukihide Minoda
    • 1
    Email author
  • Akio Kobayashi
    • 1
    • 2
  • Hiroyoshi Iwaki
    • 1
  • Ikebuchi Mitsuhiko
    • 1
  • Yoshinori Kadoya
    • 3
  • Hirotsugu Ohashi
    • 4
  • Kunio Takaoka
    • 1
  • Hiroaki Nakamura
    • 1
  1. 1.Department of Orthopaedic SurgeryOsaka City University Graduate School of MedicineOsakaJapan
  2. 2.Department of Orthopaedic SurgeryOsaka General Medical CenterOsakaJapan
  3. 3.Department of Orthopaedic SurgeryOsaka Rosai HospitalSakai City, OsakaJapan
  4. 4.Department of Orthopaedic SurgerySaiseikai Nakatsu HospitalOsakaJapan

Personalised recommendations