A non-invasive device to objectively measure tibial rotation: verification of the device
- 179 Downloads
- 28 Citations
Abstract
The purpose of this study was the correlation of the results of a new measurement device for tibial rotation (Rotameter) in comparison with the measurements of a knee navigation system as standard method. In a biomechanical laboratory study, all soft tissues were removed from 20 human cadaveric knees leaving only the intact capsule and the bone. Specific tracers were bicortically fixed in the bone in order to measure tibial rotation using a knee navigation system. The knees were fixed to a custom-made inside-boot to rule out undesirable rotation of the reconstruction inside the Rotameter measurement device. Internal and external rotation values were measured at an applied torque of 5, 10 and 15 Nm. The different methods to evaluate tibial rotation were compared using the Pearson correlation coefficient. The correlations were deemed to be reliable if a value of ≥0.80 was achieved. At 5 Nm of applied torque, high correlations for the internal rotation, external rotation and the entire rotational range were found in the Pearson correlation coefficient between the Rotameter testing device in comparison with the knee navigation system as invasive reference method. These results were also confirmed at an applied torque of 10 and 15 Nm. In conclusion, the Rotameter testing device showed high correlations compared with the knee navigation system as an invasive standard method. It might be used as a non-invasive and easy alternative to investigate tibial rotation.
Keywords
Knee rotation Measurement device Knee navigation Tibial rotation ACLNotes
Acknowledgments
The authors thank Stryker Medical Systems (Duisburg, Germany) for the provision of the knee navigation equipment and Dr. Thomas Georg for the statistical support of the study.
Conflict of interest statement
No potential conflicts of interests are declared.
References
- 1.Almquist PO, Arnbjornsson A, Zatterstrom R, Ryd L, Ekdahl C, Friden T (2002) Evaluation of an external device measuring knee joint rotation: an in vivo study with simultaneous Roentgen stereometric analysis. J Orthop Res 20:427–432PubMedCrossRefGoogle Scholar
- 2.Andriacchi TP, Alexander EJ (2000) Studies of human locomotion: past, present and future. J Biomech 33:1217–1224PubMedCrossRefGoogle Scholar
- 3.Colombet P, Robinson J, Christel P, Franceschi JP, Djian P (2007) Using navigation to measure rotation kinematics during ACL reconstruction. Clin Orthop Relat Res 454:59–65PubMedCrossRefGoogle Scholar
- 4.Czerniecki JM, Lippert F, Olerud JE (1988) A biomechanical evaluation of tibiofemoral rotation in anterior cruciate deficient knees during walking and running. Am J Sports Med 16:327–331PubMedCrossRefGoogle Scholar
- 5.Georgoulis AD, Ristanis S, Chouliaras V, Moraiti C, Stergiou N (2007) Tibial rotation is not restored after ACL reconstruction with a hamstring graft. Clin Orthop Relat Res 454:89–94PubMedCrossRefGoogle Scholar
- 6.Hsieh HH, Walker PS (1976) Stabilizing mechanisms of the loaded and unloaded knee joint. J Bone Joint Surg Am 58:87–93PubMedGoogle Scholar
- 7.Hsu WH, Fisk JA, Yamamoto Y, Debski RE, Woo SL (2006) Differences in torsional joint stiffness of the knee between genders: a human cadaveric study. Am J Sports Med 34:765–770PubMedCrossRefGoogle Scholar
- 8.Jarvela T (2007) Double-bundle versus single-bundle anterior cruciate ligament reconstruction: a prospective, randomize clinical study. Knee Surg Sports Traumatol Arthrosc 15:500–507PubMedCrossRefGoogle Scholar
- 9.Kubo S, Muratsu H, Yoshiya S, Mizuno K, Kurosaka M (2007) Reliability and usefulness of a new in vivo measurement system of the pivot shift. Clin Orthop Relat Res 454:54–58PubMedCrossRefGoogle Scholar
- 10.Kurosawa H, Yamakoshi K, Yasuda K, Sasaki T (1991) Simultaneous measurement of changes in length of the cruciate ligaments during knee motion. Clin Orthop Relat Res (265):233–240Google Scholar
- 11.Lorbach O, Wilmes P, Brockmeyer M, Maas S, Kohn D, Seil R (2008) Inter- and intra-observer reliability of a new measurement device for tibiofemoral rotation. In: Podium presentation, 13th ESSKA congress, Porto 16(Suppl 1), May 2008 (abstract)Google Scholar
- 12.Markolf KL, Graff-Radford A, Amstutz HC (1978) In vivo knee stability. A quantitative assessment using an instrumented clinical testing apparatus. J Bone Joint Surg Am 60:664–674PubMedGoogle Scholar
- 13.McPherson A, Karrholm J, Pinskerova V, Sosna A, Martelli S (2005) Imaging knee position using MRI, RSA/CT and 3D digitisation. J Biomech 38:263–268PubMedCrossRefGoogle Scholar
- 14.Muneta T, Koga H, Mochizuki T, Ju YJ, Hara K, Nimura A, Yagishita K, Sekiya I (2007) A prospective randomized study of 4-strand semitendinosus tendon anterior cruciate ligament reconstruction comparing single-bundle and double-bundle techniques. Arthroscopy 23:618–628PubMedCrossRefGoogle Scholar
- 15.Musahl V, Bell KM, Tsai AG, Costic RS, Allaire R, Zantop T, Irrgang JJ, Fu FH (2007) Development of a simple device for measurement of rotational knee laxity. Knee Surg Sports Traumatol Arthrosc 15:1009–1012PubMedCrossRefGoogle Scholar
- 16.Noyes FR, Grood ES, Cummings JF, Wroble RR (1991) An analysis of the pivot shift phenomenon. The knee motions and subluxations induced by different examiners. Am J Sports Med 19:148–155PubMedCrossRefGoogle Scholar
- 17.Pearle AD, Solomon DJ, Wanich T, Moreau-Gaudry A, Granchi CC, Wickiewicz TL, Warren RF (2007) Reliability of navigated knee stability examination: a cadaveric evaluation. Am J Sports Med 35:1315–1320PubMedCrossRefGoogle Scholar
- 18.Robinson J, Carrat L, Granchi C, Colombet P (2007) Influence of anterior cruciate ligament bundles on knee kinematics: clinical assessment using computer-assisted navigation. Am J Sports Med 35:2006–2013PubMedCrossRefGoogle Scholar
- 19.Shoemaker SC, Markolf KL (1982) In vivo rotatory knee stability. Ligamentous and muscular contributions. J Bone Joint Surg Am 64:208–216PubMedGoogle Scholar
- 20.Shultz SJ, Shimokochi Y, Nguyen AD, Schmitz RJ, Beynnon BD, Perrin DH (2007) Measurement of varus-valgus and internal-external rotational knee laxities in vivo–Part I: assessment of measurement reliability and bilateral asymmetry. J Orthop Res 25:981–988PubMedCrossRefGoogle Scholar
- 21.Steckel H, Murtha PE, Costic RS, Moody JE, Jaramaz B, Fu FH (2007) Computer evaluation of kinematics of anterior cruciate ligament reconstructions. Clin Orthop Relat Res 463:37–42PubMedGoogle Scholar
- 22.Wang CJ, Walker PS (1974) Rotatory laxity of the human knee joint. J Bone Joint Surg Am 56:161–170PubMedGoogle Scholar
- 23.Woo SL, Kanamori A, Zeminski J, Yagi M, Papageorgiou C, Fu FH (2002) The effectiveness of reconstruction of the anterior cruciate ligament with hamstrings and patellar tendon. A cadaveric study comparing anterior tibial and rotational loads. J Bone Joint Surg Am 84-A:907–914PubMedGoogle Scholar
- 24.Yagi M, Kuroda R, Nagamune K, Yoshiya S, Kurosaka M (2007) Double-bundle ACL reconstruction can improve rotational stability. Clin Orthop Relat Res 454:100–107PubMedCrossRefGoogle Scholar
- 25.Zarins B, Rowe CR, Harris BA, Watkins MP (1983) Rotational motion of the knee. Am J Sports Med 11:152–156PubMedCrossRefGoogle Scholar
- 26.Zelle BA, Vidal AF, Brucker PU, Fu FH (2007) Double-bundle reconstruction of the anterior cruciate ligament: anatomic and biomechanical rationale. J Am Acad Orthop Surg 15:87–96PubMedGoogle Scholar