Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube

  • Alexander Dullenkopf
  • Andreas Gerber
  • Markus Weiss
Technical Note

Abstract

Objective

This study compared the recently introduced Microcuff endotracheal tube HVLP ICU featuring an ultrathin (7-µm) polyurethane cuff membrane with endotracheal tubes from different manufacturers regarding fluid leakage past the tube cuff.

Design

In vitro setup.

Measurements and results

The following endotracheal tubes (ID 7.5 mm) were compared: Mallinckrodt HiLo, Microcuff HVLP ICU, Portex Profile Soft Seal, Rüsch Super Safety Clear, and Sheridan CF. A vertical PVC trachea model (ID 20 mm) was intubated, and cuffs were inflated to 10, 15, 20, 25, 30, and 60 cmH2O. Colored water (5 ml) was added to the top of the cuff. The amount of leaked fluid past the tube cuff within 5, 10, and 60 min was recorded. Experiments were performed four times using two examples of each tube brand. Fluid leakage past tube cuffs occurred in all conventional endotracheal tubes at cuff pressures from 10 to 60 cmH2O. In the Microcuff tube cuff pressure fluid leakage was observed within 10 min only at 10 cmH2O. Results with the Microcuff tube were significantly better than all other tube brands at cuff pressures of 10–30 cmH2O.

Conclusions

Within the acceptable upper limit for tracheal cuff pressure (25–30 cmH2O) the Microcuff endotracheal tube was the only one of the tested tubes to prevent fluid leakage in our in vitro setup. In vivo studies are required to confirm these findings.

Keywords

Cuff Tube, tracheal Leakage Aspiration, fluid Sealing, high volume Low-pressure tracheal cuff 

References

  1. 1.
    Latto P (1997) The cuff. In: Latto IP, Vaughan RS (ed) Difficulties in tracheal intubation. Saunders, Philadelphia, pp 51–78Google Scholar
  2. 2.
    Seegobin RD, van Hasselt GL (1984) Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. BMJ 288:965–968PubMedGoogle Scholar
  3. 3.
    Asai T, Shingu K (2001) Leakage of fluid around high-volume, low-pressure cuffs. Anaesthesia 56:38–42CrossRefPubMedGoogle Scholar
  4. 4.
    Young PJ, Blunt MC (1999) Improving the shape and compliance characteristics of a high-volume, low-pressure cuff improves tracheal seal. Br J Anaesth 83:887–889PubMedGoogle Scholar
  5. 5.
    Young PJ, Basson C, Hamilton D, Ridley SA (1999) Prevention of tracheal aspiration using the pressure-limited tracheal tube cuff. Anaesthesia 54:559–563CrossRefPubMedGoogle Scholar
  6. 6.
    Young PJ, Ridley SA, Downward G (1998) Evaluation of a new design of tracheal tube cuff to prevent leakage of fluid to the lungs. Br J Anaesth 80:796–799PubMedGoogle Scholar
  7. 7.
    Young PJ, Rollinson M, Downward G, Henderson S (1997) Leakage of fluid past the tracheal tube cuff in a benchtop model. Br J Anaesth 78:557–562PubMedGoogle Scholar
  8. 8.
    Oikkonen M, Aromaa U (1997) Leakage of fluid around low-pressure tracheal tube cuffs. Anaesthesia 52:567–569PubMedGoogle Scholar
  9. 9.
    Reali-Forster C, Kolobow T, Giacomini M, Hayashi T, Horiba K, Ferrans V (1996) New ultrathin-walled endotracheal tube with a novel laryngeal seal design: long-term evaluation in sheep. Anesthesiology 84:162–172PubMedGoogle Scholar
  10. 10.
    Blunt MC, Young PJ, Patil A, Haddock A (2001) Gel lubrication of the tracheal tube cuff reduces pulmonary aspiration. Anesthesiology 95:377–381PubMedGoogle Scholar
  11. 11.
    Metha S, Myat HM (1984) The cross sectional shape and circumference of the human trachea. Ann R Coll Surg Engl 66:356–358PubMedGoogle Scholar
  12. 12.
    Inglis TJJ, Sproat LJ, Sherrat MJ, Hawkey PM, Gibson JS, Shah MV (1992) Gastroduodenal dysfunction as a cause of gastric bacterial overgrowth in patients undergoing mechanical ventilation of the lungs. Br J Anaesth 68:499–502PubMedGoogle Scholar
  13. 13.
    Ho AM, Aun CS, Karmakar MK (2002) The margin of safety associated with the use of cuffed paediatric tracheal tubes. Anaesthesia 57:173–175CrossRefPubMedGoogle Scholar
  14. 14.
    James I (2001) Cuffed tubes in children. Paediatr Anaesth 11:259–263CrossRefPubMedGoogle Scholar
  15. 15.
    Satishkumar S, Young P (2002) Tracheal cuff pressure–a survey of clinical practice. Br J Anaesth 88:456PubMedGoogle Scholar

Copyright information

© Springer-Verlag 2003

Authors and Affiliations

  • Alexander Dullenkopf
    • 1
  • Andreas Gerber
    • 1
  • Markus Weiss
    • 1
  1. 1.Department of AnesthesiaUniversity Children's HospitalZurichSwitzerland

Personalised recommendations