Advertisement

Intensive Care Medicine

, Volume 29, Issue 3, pp 414–418 | Cite as

Temperature measurement in intensive care patients: comparison of urinary bladder, oesophageal, rectal, axillary, and inguinal methods versus pulmonary artery core method

  • J.-Y. LefrantEmail author
  • L. Muller
  • J. Emmanuel de La Coussaye
  • M. Benbabaali
  • C. Lebris
  • N. Zeitoun
  • C. Mari
  • G. Saïssi
  • J. Ripart
  • J.-J. Eledjam
Original

Abstract

Objective.

Comparisons of urinary bladder, oesophageal, rectal, axillary, and inguinal temperatures versus pulmonary artery temperature.

Design.

Prospective cohort study.

Setting.

Intensive Care Unit of a University-Hospital.

Patients.

Forty-two intensive care patients requiring a pulmonary artery catheter (PAC).

Intervention.

Patients requiring PAC and without oesophageal, urinary bladder, and/or rectal disease or recent surgery were included in the study. Temperature was simultaneously monitored with PAC, urinary, oesophageal, and rectal electronic thermometers and with axillary and inguinal gallium-in-glass thermometers. Comparisons used a Bland and Altman method.

Measurements and main results.

The pulmonary arterial temperature ranged from 33.7 °C to 40.2 °C. Urinary bladder temperature was assessed in the last 22 patients. A total of 529 temperature measurement comparisons were carried out (252 comparisons of esophageal, rectal, inguinal, axillary, and pulmonary artery temperature measurements in the first 20 patients, and 277 comparisons with overall methods in the last patients). Nine to 18 temperature measurement comparisons were carried out per patient (median = 13). The mean differences between pulmonary artery temperatures and those of the different methods studied were: oesophageal (0.11±0.30 °C), rectal (−0.07±0.40 °C), axillary (0.27±0.45 °C), inguinal (0.17±0.48 °C), urinary bladder (−0.21±0.20 °C).

Conclusion.

In critically ill patients, urinary bladder and oesophageal electronic thermometers are more reliable than the electronic rectal thermometer which is better than inguinal and axillary gallium-in-glass thermometers to measure core temperature.

Keywords

Core temperature Measurements Intensive Care 

Notes

Acknowledgments.

The authors very sincerely thank nurses and aid-nurses for their kind and skilled assistance in temperature measurements.

References

  1. 1.
    Cork RC, Vaughan RW, Humphrey LS (1983) Precision and accuracy of intraoperative temperature monitoring. Anesth Analg 62:211–214Google Scholar
  2. 2.
    Molnar GW, Read RC (1974) Studies during open-heart surgery on special characteristics of rectal temperature. J Appl Physiol 36:333–336PubMedGoogle Scholar
  3. 3.
    Robinson J, Charlton J, Seal R, Spady D, Joffres MR (1998) Esophageal, rectal, axillary, tympanic, and pulmonary artery temperatures during cardiac surgery. Can J Anaesth 45:317–323PubMedGoogle Scholar
  4. 4.
    Schmitz T, Bair N, Falk M, Levine C (1995) A comparison of five methods of temperature measurement in febrile intensive care unit patients. Am J Crit Care 4:186–192Google Scholar
  5. 5.
    Giuliano KK, Scott SS, Elliot S, Giuliano AJ (1999) Temperature measurement in critically ill orally intubated adults: a comparison of pulmonary artery core, tympanic, and oral methods. Crit Care Med 27:2188–2193PubMedGoogle Scholar
  6. 6.
    Lilly JK, Boland JP, Zekan S (1980) Urinary bladder temperature monitoring: a new index of core body temperature. Crit Care Med 8:742–744PubMedGoogle Scholar
  7. 7.
    Erickson KS, Kirklin SK (1993) Comparison of ear-based, bladder, oral, and axillary methods for core temperature measurement. Crit Care Med 21:1528–1534PubMedGoogle Scholar
  8. 8.
    Terndrup TE, Rajk J (1992) Impact of operator technique and device on infrared emission detection tympanic thermometry. J Emerg Med 10:683–687PubMedGoogle Scholar
  9. 9.
    Peterson MH, Hauge HN (1997) Can training improve the results with infrared tympanic thermometers? Acta Anaesthesiol Scand 41:1066–1070PubMedGoogle Scholar
  10. 10.
    Staven K, Saxholm H, Smith-Erichsen N (1997) Accuracy of infrared ear thermometry in adult patients. Intensive Care Med 23:100–105CrossRefPubMedGoogle Scholar
  11. 11.
    Eichna LW, Berger AR, Rader B, Buckaroo WH (1951) Comparison of intracardiac and intravascular temperatures with rectal temperatures in man. J Clin Invest 30:353–359Google Scholar
  12. 12.
    Knaus WA, Draper EA, Wagner DP, Zimmerman JE (1985) APACHE II: a severity of disease classification system. Crit Care Med 13:818–829PubMedGoogle Scholar
  13. 13.
    Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet i:307–310Google Scholar
  14. 14.
    Hasler ME, Cohen JA (1982) The effect of oxygen administration on oral temperature assessment. Nurs Res 31:265–268PubMedGoogle Scholar
  15. 15.
    Shinozaki T, Deane R, Perkins FM (1988) Infrared tympanic thermometer:,evaluation of a new clinical thermometer. Crit Care Med 16:148–150Google Scholar
  16. 16.
    Franchesi VT (1991) Accuracy and feasability of measuring oral temperature in critically ill adults. Focus Crit Care 18:221–228PubMedGoogle Scholar
  17. 17.
    Nierman DM (1991) Core temperature measurement in the intensive care unit. Crit Care Med 19:818–823PubMedGoogle Scholar
  18. 18.
    Erickson RS, Meyer LT (1994) Accuracy or infrared ear thermometry and other temperature methods in adults. Am J Crit Care 3:40–54Google Scholar
  19. 19.
    Romano MJ, Fortenberry JD, Autrey E, (1993) Infrared tympanic thermometry in the pediatric intensive care unit. Crit Care Med 21:1181–1185PubMedGoogle Scholar
  20. 20.
    Whitby JD, Dunkin LJ (1968) Temperature differences in the esophagus. Br J Anaesth 40:991–995PubMedGoogle Scholar

Copyright information

© Springer-Verlag 2003

Authors and Affiliations

  • J.-Y. Lefrant
    • 1
    Email author
  • L. Muller
    • 1
  • J. Emmanuel de La Coussaye
    • 1
  • M. Benbabaali
    • 1
  • C. Lebris
    • 1
  • N. Zeitoun
    • 1
  • C. Mari
    • 1
  • G. Saïssi
    • 1
  • J. Ripart
    • 1
  • J.-J. Eledjam
    • 1
  1. 1.Fédération Anesthésie-Douleur-Urgences-RéanimationCentre Hospitalier Universitaire de NîmesNîmesFrance