Discrepancies in volume calculations between different radiotherapy treatment planning systems

  • T. Ackerly
  • J. Andrews
  • D. Ball
  • M. Guerrieri
  • B. Healy
  • I. Williams
Technical Note

Abstract

It has been determined that, contrary to expectation, there is a clinically significant variation in the volume calculations of different RTPS (RadiotherapyTreatmentPlanningSystem) for identical contours. The situation was investigated prior to a multi-centre trial to determine whether tumour volume is an independent prognostic factor in NSCLC (nonsmall cell lung cancer) and included four of the commercially available RTPS. The four RTPS tested were, Theraplan Plus V3.0, Cadplan V6.2, Focus V2.6 and ADAC V3.0. Five randomly chosen clinical target related volumes (3 GTVs, one PTV and one CTV) from the trial database originally marked on the Cadplan system were transferred to the other four systems and the resulting volumes were calculated. It was found that Cadplan consistently underestimated the volume relative to the other three systems by 6–12%. This systematic underestimation was found to be caused by different assumptions made by the Cadplan system about the axial outer slice extension of the volume. Cadplan truncates the volume, while the other three systems extrapolate it by half the slice thickness at each end. A short program was written to apply the same method of volume extension to the Cadplan volume that is utilised by the other systems. This produced calculated volumes that were within ±1.0% of the average of the volumes calculated by the other three planning systems, and the maximum deviation from the average for any planning system was then reduced to 1.5%. This program was implemented at all participating trial centres utilising Cadplan, thus reducing the intersystem variability to a negligible factor in comparison to the estimates of inter-physician variation. This unexpected finding has significant implications for the validity of multi-centre trials using dose volume histograms, and indeed the adoption of any clinical protocol employing dose volume histogram constraints derived from experience at another centre employing a different RTPS.

Key words

radiotherapy treatment planning DVH clinical trials 

References

  1. 1.
    TROG 99.05 Tumour volume as an independent prognostic factor in patients with non-small cell lung cancer: A protocol for a prospective database. Chair,Ball, D.Google Scholar
  2. 2.
    Bowden P., Fisher R., Mac Manus M., et.al.,Measurement of Lung Tumour Volumes using thee-dimensional computer planning software, Int J. Rad. Onc. Biol. (53,3):566–574, 2002.CrossRefGoogle Scholar
  3. 3.
    Maguire, G.Q., and Noz, M.E.,Image formats: five years after the AAPM standard for digital image interchange, Med. Phys.,16: 818–23, 1989.CrossRefPubMedGoogle Scholar
  4. 4.
    Kernighan, B.W. and Ritchie, D. M.,The C Programming Language, 2nd ed., Prentice Hall, Englewood Cliffs New Jersey U.S.A., 1988Google Scholar
  5. 5.
    Breimen R. S., Beck J. W., Korobkin M., Gleny R., Akwari, O. E., Heaston D.K., Moore A. V., Ram P. C.,Volume Determinations Using Computed Tomography, Amer. J. Roent. 136: 329–333, 1982.Google Scholar
  6. 6.
    Stewart, J.,Calculus, 2nd ed., Wadsworth Inc., Belmont California U.S.A., p48, 1991Google Scholar
  7. 7.
    Cook, L.T., et. al.An algorithm for volume estimation based on polyhedral approximation IEEE Trans. Biomed. Eng. 27: 493, 1980.CrossRefPubMedGoogle Scholar
  8. 8.
    Focus Release 2.3.1 Calculation Physics and Clinical Applications Manual, Volume 2 Chapter 13 Page 7 Volume Computation Computerized Medical Systems, Inc. 1145 Corporate Lake Drive Suite 100 St. Louis Missouri USA.Google Scholar
  9. 9.
    ADAC Pinnacle version 6.0I notes. ADAC Laboratories 540 Alder Drive Milpitas CA USA.Google Scholar
  10. 10.
    Theratronics Theraplan Plus Users Manual Version 1.1/1.2 (9-55,56 External Planning:Setting the calculation matrix; 13–4 Dose Volume Histograms:To change the number of bins) MDS Nordion Therapy Systems 447 March Road Kanata, Ontaria Canada.Google Scholar

Copyright information

© Australasian College of Physical Scientists and Engineers in Medicine 2003

Authors and Affiliations

  • T. Ackerly
    • 1
  • J. Andrews
    • 1
  • D. Ball
    • 1
  • M. Guerrieri
    • 1
  • B. Healy
    • 2
  • I. Williams
    • 1
  1. 1.Physical Sciences DepartmentPeter MacCallum Cancer Institute Locked Bag 1Melbourne
  2. 2.Queensland Radium InstituteBrisbaneAustralia

Personalised recommendations