Computed tomography phantom for radiochromic film dosimetry

  • R. Gotanda
  • T. Katsuda
  • T. Gotanda
  • M. Eguchi
  • S. Takewa
  • A. Tabuchi
  • H. Yatake
Scientific Papers

Abstract

To evaluate in detail the dose distribution during computed tomography (CT), a sheet roll CT dosimetry phantom (SRCT-P) with a radiochromic film (RF) was experimentally developed. The SRCT-P was made by rolling up a vinyl chloride sheet in a cylindrical shape to arbitrarily select the SRCT-P diameter, dose measurement position, and depth. The SRCT-P centre core consisted of a plastic hose in which a 10 mm acrylic bar with a RF was inserted. To determine the availability of the SRCT-P, the surface and centre doses (at a 5 mm radius) at each SRCT-P diameter (6–16 cm; every 2 cm) were measured. The ratios of the centre-to-surface doses (Dcentre/Dsurface) systematically increased, from 80 to 111%, for decreasing SRCT-P diameters, between 16 and 6 cm, respectively. The centre dose approached the surface dose as the SRCT-P diameter decreased. To use a RF for a CT dose measurement, further detailed research and analysis is necessary. However, this study has shown that a SRCT-P is useful and beneficial for the measurement of the dose distribution during a CT examination.

Key words

radiochromic film phantom computed tomography radiation dosimetry 

References

  1. 1.
    Miyazaki, O., Kitamura, M., Masaki, H., Nosaka, S., Miyasaka, M., Kashima, K., Okada, Y. and Tsutsumi, Y.,Current practice of pediatric MDCT in Japan: Survey results of demographics and age-based dose reduction, Nippon Igaku Hoshasen Gakkai Zasshi, 65:216–223, 2005. (In Japanese)PubMedGoogle Scholar
  2. 2.
    Nishizawa, K., Matsumoto, M., Iwai, K. and Maruyama, T.,Survey of CT practice in Japan and collective effective dose estimation, Nippon Igaku Hoshasen Gakkai Zasshi, 64:151–158, 2004. (In Japanese)PubMedGoogle Scholar
  3. 3.
    United Nations Scientific Committee on the Effects of Atomic Radiation. 2000 Report Volume 1. Sources, Annex D, Medical radiation exposure, Available at: www.unscear.org/docs/ zreports/annexd.pdf, (2007).Google Scholar
  4. 4.
    Cody, D.D.,AAPM/RSNA physics tutorial for residents: Topics in CT. Imaging Processing in CT, Radiographics, 22:1255–1268, 2002.PubMedGoogle Scholar
  5. 5.
    McNitt-Gray, M.F.,AAPM/RSNA physics tutorial for residents: Topics in CT. Radiation Dose in CT, Radiographics, 22:1541–1553, 2002.CrossRefPubMedGoogle Scholar
  6. 6.
    Nickoloff, E.,Current adult and pediatric CT doses, Pediatr Radiol, 32:250–260, 2002.CrossRefPubMedGoogle Scholar
  7. 7.
    Morgan, H.T.,Image quality improvement and dose reduction in CT pediatric imaging, Medicamundi, 46:16–21, 2002.Google Scholar
  8. 8.
    Nickoloff, E.L., Dutta, A.K. and Lu, Z.F.,Influence of phantom diameter, kVp and scan mode upon computed tomography dose index, Med Phys, 30:395–402, 2003.CrossRefPubMedGoogle Scholar
  9. 9.
    Brenner, D., Elliston, C., Hall, E. and Berdon, W.,Estimated risks of radiation-induced fatal cancer from pediatric CT, AJR Am J Roentgenol, 176:289–295, 2001.PubMedGoogle Scholar
  10. 10.
    Nishitani, H., Yasutomo, M., Tominaga, M., Fukui, H. and Yagi, H.,Radiation exposure in radiological clinics radiation expousure in CT, Nippon Igaku Hoshasen Gakkai Zasshi, 62:347–351, 2002. (In Japanese)PubMedGoogle Scholar
  11. 11.
    FDA Public Health Notification, Reducing Radiation Risk from Computed Tomography for Pediatric and Small Adult Patients, Available at: www.fda.gov/cdrh/safety/110201-ct.html, (2001).Google Scholar
  12. 12.
    Boone, J.M., Geraghty, E.M., Seibert, J.A. and Wootton-Gorges, S. L.,Dose reduction in pediatric CT: A rational approach. Radiology, 228:352–360, 2003.CrossRefPubMedGoogle Scholar
  13. 13.
    Kalra, M.K., Maher, M.M., Toth, T.L., Hamberg, L.M., Blake, M.A., Shepard, J.A. and Saini, S.,Strategies for CT radiation dose optimization, Radiology, 230:619–628, 2004.CrossRefPubMedGoogle Scholar
  14. 14.
    Siegel, M.J., Schmidt, B., Bradley, D., Suess, C. and Hildebolt, C.,Radiation dose and image quality in pediatric CT: Effect of technical factors and phantom size and shape, Radiology, 233:515–522, 2004.CrossRefPubMedGoogle Scholar
  15. 15.
    US FDA, Center for Devices and Radiological Health, Title 21. Food and Drugs Subchapter J. Radiological Health Part 1020; Performance Standards for Ionizing Radiation Emitting Products, Sec. 21CFR1020.33. Computed tomography [CT] equipment, Available at: www.accessdata.fda.gov/ scripts/cdrh/cfdocs/cfcfr/CFRSearch.cfm?fr=1020.33, (2006).Google Scholar
  16. 16.
    CIRS Web site, CT Dose Phantom. Available at: www.cirsinc.com/pdfs/007cp.pdf, (2007).Google Scholar
  17. 17.
    Fluke Biomedical Corporation web site, Nested CT Dose Phantom Kit for Pediatric/Adult Head and Body: Model 76–424–4156. Available at: global.flukebiomedical.com/busen/ products/76–419–4150.htm?catalog_name=FlukeUnitedStates &Category=CTPHANT(FlukeProducts), (2007).Google Scholar
  18. 18.
    Capintec, Inc. web site, CT Phantoms: CT Head and Body Dose Phantoms.. Available at: www.capintec.com/pdf/ ctheadandbodyphantom.pdf, (2007).Google Scholar
  19. 19.
    International Specialty Products web site. Available at: www.ispcorp.com/products/dosimetry/content/gafchromic/ind ex.html, (2007).Google Scholar
  20. 20.
    Gorny, K.R., Leitzen, S.L., Bruesewitz, M.R., Kofler, J.M., Hangiandreou, N.J. and McCollough, C.H.,The calibration of experimental self-developing Gafchromic HXR film for the measurement of radiation dose in computed tomography, Med Phys, 32:1010–10016, 2005.CrossRefPubMedGoogle Scholar
  21. 21.
    Butson, M.J., Cheung, T. and Yu, P.K.,Absorption spectra of irradiated XRCT radiochromic film, Phys Med Biol, 51:3099–3103, 2006.CrossRefPubMedGoogle Scholar
  22. 22.
    Thomas, G., Chu, R.Y. and Rabe, F.,A study of GafChromic XR Type R film response with reflective-type densitometers and economical flatbed scanners, J Appl Clin Med Phys, 4:307–14, 2003.CrossRefPubMedGoogle Scholar
  23. 23.
    Butson, M.J., Yu, P.K.N., Cheung, T. and Metcalfe, P.,Radiochromic film for medical radiation dosimetry, Materials Science and Engineering R, 41:61–120, 2003.CrossRefGoogle Scholar

Copyright information

© Australasian College of Physical Scientists and Engineers in Medicine 2007

Authors and Affiliations

  • R. Gotanda
    • 4
  • T. Katsuda
    • 1
  • T. Gotanda
    • 4
  • M. Eguchi
    • 4
  • S. Takewa
    • 4
  • A. Tabuchi
    • 4
    • 2
  • H. Yatake
    • 4
    • 3
  1. 1.Faculty of Health SciencesOkayama University Medical SchoolOkayamaJapan
  2. 2.Department of RadiologyKawasaki HospitalOkayamaJapan
  3. 3.Department of RadiologyKaizuka City HospitalOsakaJapan
  4. 4.Graduate School of Health SciencesOkayama-city, OkayamaOkayama UniversityJapan

Personalised recommendations