Canadian Journal of Anesthesia

, 54:891

The GlideScope®-specific rigid stylet and standard malleable stylet are equally effective for GlideScope® use

  • Timothy P. Turkstra
  • Christopher C. Harle
  • Kevin P. Armstrong
  • Paidrig M. Armstrong
  • Richard A. Cherry
  • Jason Hoogstra
  • Philip M. Jones
Reports of Original Investigations

Abstract

Purpose

The GlideScope® videolaryngoscope usually provides excellent glottic visualization, but directing an endotracheal tube (ETT) through the vocal cords can be challenging. The goal of the study was to compare the dedicated GlideScope®-specific rigid stylet to the standard malleable stylet, assessed by time to intubation (TTI).

Methods

Eighty patients requiring orotracheal intubation for elective surgery were randomly allocated to either the GlideScope® rigid stylet (GRS) or a standard malleable stylet to facilitate intubation using the GlideScope®. Time to intubation was recorded by blinded assessors; operators were blinded until after laryngoscopy. The operator assessed the ease of intubation using a visual analogue scale (VAS). The number of intubation attempts, number of failures, glottic grades, and use of external laryngeal manipulation were documented.

Results

The median TTI was 42.7 sec (inter-quartile range (IQR) 38.9-56.7) for the GRS group compared to 39.9 sec (IQR 34.1-48.2) for the control group (P = 0.07). The median VAS score for ease of intubation was 20 (IQR 12.0-33.0) for the GRS group compared to 18 (IQR 9.5-29.5) for the control group (P = 0.21). There was no significant difference in TTI or VAS between stylets. The overall incidence of a Cormack-Lehane grade I or II glottic view was 98%.

Conclusions

In a group of experienced operators using the GlideScope®, the dedicated GRS and the standard malleable ETT stylet are equally effective in facilitating endotracheal intubation.

Le mandrin rigide spécifique au GlideScope® et le mandrin flexible standard sont tous deux aussi efficaces pour l’utilisation avec le GlideScope®

Résumé

Objectif

Le vidéolaryngoscope GlideScope® fournit en général une excellente visualisation glottique, mais diriger une sonde endotrachéale (SET) entre les cordes vocales peut être un défi. L’objectif de cette étude était de comparer le mandrin rigide spécifique au GlideScope® au mandrin flexible standard, évalués par le temps requis pour l’intubation (TTI).

Méthode

Quatre-vingts patients nécessitant une intubation orotrachéale pour une chirurgie élective ont été randomisés à être intubés à l’aide soit du mandrin rigide GlideScope® (GRS), soit du mandrin flexible standard pour faciliter l’intubation avec le GlideScope®. Le temps requis pour l’intubation a été mesuré par des évaluateurs en aveugle; les opérateurs étaient également en aveugle jusqu’à ce que la laryngoscopie soit terminée. L’opérateur a évalué la facilité d’intubation à l’aide d’une échelle visuelle analogique (EVA). Le nombre de tentatives d’intubation, le nombre d’échecs, le grade d’intubation et le recours à une manipulation laryngée externe ont été notés.

Résultats

Le TTI médian était de 42,7 sec (intervalle interquartile (IQR) 38,9-56,7) pour le groupe GRS comparé à 39,9 sec (IQR 34,1-48,2) pour le groupe témoin (P = 0,07). Le score EVA médian pour la facilité d’intubation était de 20 (IQR 12,0-33,0) pour le groupe GRS comparé à 18 pour le groupe témoin (P = 0,21). Il n’y a pas eu de différence significative dans le TTI ou l’EVA entre les mandrins. L’incidence globale de la visualisation glottique selon l’échelle de Cormack-Lehane de grade I ou II était de 98 %.

Conclusion

Dans un groupe d’opérateurs expérimentés se servant du GlideScope®, le mandrin GRS spécial et le mandrin flexible standard ont la même efficacité pour faciliter l’intubation endotrachéale.

References

  1. 1.
    Cooper RM, Pacey JA, Bishop MJ, McCluskey SA. Early clinical experience with a new videolaryngoscope (GlideScope) in 728 patients. Can J Anesth 2005; 52:191–8.PubMedGoogle Scholar
  2. 2.
    Doyle DJ. The GlideScope video laryngoscope (Letter). Anaesthesia 2005; 60:414–5.PubMedCrossRefGoogle Scholar
  3. 3.
    Cuchillo JV, Rodriguez MA. Considerations aimed at facilitating the use of the new GlideScope videolaryngoscope (Letter). Can J Anesth 2005; 52:661–2.PubMedGoogle Scholar
  4. 4.
    Cooper RM. Complications associated with the use of the GlideScope videolaryngoscope. Can J Anesth 2007; 54:54–7.PubMedGoogle Scholar
  5. 5.
    Chin KJ, Arango MF, Paez AF, Turkstra TP. Palatal injury associated with the GlideScope®. Anaesth Intensive Care 2007; 35:449–50.PubMedGoogle Scholar
  6. 6.
    Hsu WT, Hsu SC, Lee YL, Huang JS, Chen CL. Penetrating injury of the soft palate during GlideScope intubation. Anesth Analg 2007; 104:1609–10.PubMedCrossRefGoogle Scholar
  7. 7.
    Malik AM, Frogel JK. Anterior tonsillar pillar perforation during Glidescope video laryngoscopy. Anesth Analg 2007; 104:1610–1.PubMedCrossRefGoogle Scholar
  8. 8.
    Bader SO, Heitz JW, Audu PB. Tracheal intubation with the GlidesScope videolaryngoscope, using a“J” shaped endotracheal tube (Letter). Can J Anesth 2006; 53:634–5.PubMedGoogle Scholar
  9. 9.
    Cooper RM. Considerations aimed at facilitating the use of the new GlideScope videolaryngoscope (Letter, reply). Can J Anesth 2005; 52:661–2.CrossRefGoogle Scholar
  10. 10.
    Jones PM, Turkstra TP, Armstrong KP, et al. Effect of stylet angulation and endotracheal tube camber on time to intubation with the GlideScope. Can J Anesth 2007; 54:21–7.PubMedGoogle Scholar
  11. 11.
    Mallampati SR, Gatt SP, Gugino LD, et al. A clinical sign to predict difficult tracheal intubation: a prospective study. Can Anaesth Soc J 1985; 32:429–34.PubMedCrossRefGoogle Scholar
  12. 12.
    Cormack RS, Lehane J. Difficult tracheal intubation in obstetrics. Anaesthesia 1984; 39:1105–11.PubMedCrossRefGoogle Scholar
  13. 13.
    Lim TJ, Lim Y, Liu EH. Evaluation of ease of intubation with the GlideScope or Macintosh laryngoscope by anaesthetists in simulated easy and difficult laryngoscopy. Anaesthesia 2005; 60:180–3.PubMedCrossRefGoogle Scholar
  14. 14.
    Turkstra TP, Craen RA, Pelz DM, Gelb AW. Cervical spine motion: a fluoroscopic comparison during intubation with lighted stylet, GlideScope, and Macintosh laryngoscope. Anesth Analg 2005; 101:910–5.PubMedCrossRefGoogle Scholar
  15. 15.
    Sun DA, Warriner CB, Parsons DG, Klein R, Umedaly HS, Moult M. The GlideScope Video Laryngoscope: randomized clinical trial in 200 patients. Br J Anaesth 2005; 94:381–4.PubMedCrossRefGoogle Scholar
  16. 16.
    Rai MR, Dering A, Verghese C. The Glidescope system: a clinical assessment of performance. Anaesthesia 2005; 60:60–4.PubMedCrossRefGoogle Scholar
  17. 17.
    Holden JD. Hawthorne effects and research into professional practice. J Eval Clin Pract 2001; 7:65–70.PubMedCrossRefGoogle Scholar

Copyright information

© Canadian Anesthesiologists 2007

Authors and Affiliations

  • Timothy P. Turkstra
    • 1
  • Christopher C. Harle
    • 1
  • Kevin P. Armstrong
    • 1
  • Paidrig M. Armstrong
    • 1
  • Richard A. Cherry
    • 1
  • Jason Hoogstra
    • 1
  • Philip M. Jones
    • 1
  1. 1.Department of Anesthesia & Perioperative MedicineLondon Health Sciences Centre - University HospitalLondonCanada