Canadian Anaesthetists’ Society Journal

, Volume 27, Issue 2, pp 106–110 | Cite as

An evaluation of gas density dependence of anaesthetic vaporizers

  • L. Prins
  • J. Strupat
  • J. Clement
  • R. L. Knill
Article

Abstract

Four commonly used vaporizers were studied for the effect of carrier gas density on vaporizer output. Vapour concentrations from a halothane Cyprane (Fluotec) Mark 2 increased in relation to the density of carrier gas, whereas the concentrations delivered by an enflurane Ohio vaporizer decreased. The halothane Cyprane (Fluotec) Mark 3 and enflurane Cyprane vaporizers were largely independent of density. Of clinical importance, nitrous oxide/oxygen (75/25), compared with oxygen alone, increased the vapour concentration outputs of the halothane Mark 2 up to 30 per cent and decreased the outputs of the enflurane Ohio unit up to 20 per cent.

Keywords

Nitrous Oxide Halothane Enflurane Vapour Concentration Anaesthetic Potency 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Résumé

Quatre vaporisateurs ďusage courant ont été étudiés en rapport avec les effets de la densité du gaz vecteur sur ľefficacité de la vaporisation. La concentration de vapeur obtenue ďun vaporisateur Cyprane (Fluotec) Mark 2 a augmenté proportionnellement à la densité du gaz vecteur alors que la concentration des vapeur fournie par le vaporisateur Ohio pour ľenflurane a diminué. Les vaporisateurs Cyprane (Fluotec) Mark 3 et Cyprane pour enflurane sont indépendants de la densité du gaz vecteur. Il est ďimportance clinique de noter que le mélange protoxyde ďazote-oxygène (75:25) comparé a ľoxygène seul a fait augmenter la concentration des vapeurs ďhalothane à la sortie du vaporisateur Mark 2 à des valeurs allant jusqu’à 30 pour cent et diminuer les concentrations ďenflurane à des valeurs allant jusqu’à 20 pour cent à la sortie du vaporisateur Ohio.

References

  1. 1.
    Stoelting, R.K. The effect of nitrous oxide on halothane output from fluotec Mark 2 vaporizers. Anesthesiology35: 215 (1971).PubMedCrossRefGoogle Scholar
  2. 2.
    Nawaf, K., &Stoelting, R.K. Nitrous oxide increases enflurane concentrations delivered by Ethrane vaporizers. Anesth. Analg.58: 30 (1979).PubMedCrossRefGoogle Scholar
  3. 3.
    Clark, J.B. Physical and Mathematical Tables. Edinburgh. Oliver and Boyd (1968).Google Scholar
  4. 4.
    Paterson, G.M., Hulands, G.H. &Nunn, J.F. Evaluation of a new halothane vaporizer: the Cyprane Fluotec Mark 3. British Journal of Anaesthesia41: 109 (1969).PubMedCrossRefGoogle Scholar
  5. 5.
    Hill, D.W. The design and calibration of vaporizers for volatile anaesthetic agents, pp. 386. In: C. Scurr & S. Feldman, Scientific Foundations of Anaesthesia, 1st Ed., London (1970).Google Scholar
  6. 6.
    Ohio Ethraneê Technical Manual.Google Scholar
  7. 7.
    McDowall, D.G. Anaesthesia in a pressure chamber. Anaesthesia19 (3): 321 (1964).PubMedCrossRefGoogle Scholar

Copyright information

© Canadian Anesthesiologists 1980

Authors and Affiliations

  • L. Prins
    • 1
  • J. Strupat
    • 1
  • J. Clement
    • 1
  • R. L. Knill
    • 1
  1. 1.Departments of Anaesthesia and Respiratory TechnologyUniversityHospital, University of Western OntarioLondon

Personalised recommendations