Indian Journal of Clinical Biochemistry

, Volume 18, Issue 1, pp 65–70 | Cite as

Investigations on protein adsorption onin-vivo sensor membrane surfaces

  • Namrata Rai
  • M. C. Chattopadhyaya
Article
  • 33 Downloads

Summary

In order to understand the mechanism of fauling of electrode surfaces ofin vivo electrochemical sensors used for monitoring whole blood, the investigations on protein adsorption on the membrane surfaces were undertaken. Using an oxygen sensor, the protein adsorption studies were carried out on two different polymeric surfaces-teflon (PTFE) and polyurethane (PU)-using Scanning Electron Microscopy and Fourier Transform Infrared Spectrophotometry. Investigations reveal that even a single protein like albumin or fibrinogen tend to adhere to the sensor membrane surface and influences the electrochemical characteristics of the sensor. The influence is more significant, in the case of PTFE membrane.

Keywords

Fibrinogen PTFE Protein Adsorption Clinical Biochemistry Electrochemical Sensor 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Jakobson, R. J. and Grendreau, R. M. Artif (1978) Organs, 2, 183.CrossRefGoogle Scholar
  2. 2.
    Ivarson, B. and Lundstrom, I. (1986) in “Critical Rev. in Biocompatibility”, (ed.) D.F. Williams, CRC Press, Florida.Google Scholar
  3. 3.
    Lyman, D. J., Knutson, K., McNeil, B. and Shibatane, K. (1975) Trans. Am. Soc. Artif. Intern. Organs 21, 49.PubMedGoogle Scholar
  4. 4.
    Roohk, H. V., Rick, J., Hill, R., Hung, E. and Bartlett, R. H. (1976) Trans. Am. Soc. Artif. Intern. Organs 22, 1.PubMedGoogle Scholar
  5. 5.
    Lee, R. G. and Kim, I. W. (1974) J. Biomed. Mater. Res. 8, 251.PubMedCrossRefGoogle Scholar
  6. 6.
    Grosse-Seistrup, G., Olsen, D. B., Lemm, W., Unger, D., Affeld, K., Krautzberger, W. and Buckerl, F. I. (1978) Trans. Soc. Biomater., 2, 102.Google Scholar
  7. 7.
    Lyman, D. J., Metcalf, L. E., Albo, D., Richards, K. F. and Lamb, J. (1974) Trans. Am. Soc. Artif. Inst. Organs. 20, 474–479.Google Scholar
  8. 8.
    Hartland, A. J., Neary, R. H. (1999) Clin. Chem. 45, 1091.PubMedGoogle Scholar
  9. 9.
    Ye, Q. S. Keresztes, Z., Hovvai, G. (1999) Electroanaly. 11, 729.CrossRefGoogle Scholar
  10. 10.
    Katsu, T., Nakashima, K. (1999) Analyst 124, 883 (1999).CrossRefGoogle Scholar
  11. 11.
    Olivina, C. H., Rebelo, R. R., Cameo, M. J. M. (1999) Electroanaly. 11, 825.CrossRefGoogle Scholar
  12. 12.
    Lutze, D., Mernva, R. K., Frielich, A., Meyerhoff, M. E. (1999) Freieniurs, J. Anal. Chem. 364, 41.CrossRefGoogle Scholar
  13. 13.
    Ramamurthy, N., Baliga, N., Wakofield, T. W., Andrews, P. C., Yang, V. C., Meyerhoff, H. E. (1999) Anal. Biochem. 266, 16.CrossRefGoogle Scholar
  14. 14.
    Hitchman, M. L. (1978) “Measurement of Dissolved Oxygen”, John Wiley & Sons, New York.Google Scholar

Copyright information

© Association of Clinical Biochemists of India 2003

Authors and Affiliations

  • Namrata Rai
    • 1
  • M. C. Chattopadhyaya
    • 1
  1. 1.Electrochemical Sensors Lab Department of ChemistryUniversity of AllahabadAllahabad