Non-invasive measurement of systolic blood pressure on the arm utilising photoplethysmography: development of the methodology

  • C. Laurent
  • B. Jönsson
  • M. Vegfors
  • L. -G. Lindberg
Article

Abstract

Photoplethysmography (PPG) can be used to measure systolic blood pressure at the brachial artery. With a specially designed probe, positioned in the most distal position beneath a pressure cuff on the upper arm, this is possible. The distance between the light source (880 nm) and the photodetector was 20 mm. A test was performed on neuro-intensive care patients by determining blood pressure from the PPG curves, and, when it was compared with systolic blood pressure obtained from inserted indwelling arterial catheters, a correlation factor of r=0.95 was achieved. The difference between blood pressure obtained using PPG and invasive blood pressure measurement was 3.9±9.1 mmHg (mean±SD), n=19. The depth to the brachial artery was 13.9±4.1 mm (mean±SD), n=18. A digital PPG system utilising pulsating light was also developed.

Keywords

Photoplethysmography Blood pressure Occlusion cuff Non-invasive Arm PPG 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Bukhari, A. R. S. (1993): ‘Skin blood flow measurements—a review of noninvasive methods’,J. Clin. Eng.,18, pp. 53–65Google Scholar
  2. Cederholm, I., Sørensen, J., andCarlsson, C. (1986): ‘Thrombosis following percutaneous radial artery cannulation’,Acta Anaesthesiol. Scand.,30, pp. 227–230Google Scholar
  3. Chawla, R., Kumarvel, V., Girdhar, K. K., Sethi, A. K., Indrayan, A., andBhattacharya, A. (1992): ‘Can pulse oximetry be used to measure systolic blood pressure?’,Anesth. Analg.,74, pp. 196–200Google Scholar
  4. Frezza, E. E., andMezghebe, H. (1998): ‘Indications and complications of arterial catheter use in surgical or medical intensive care units: Analysis of 4932 patients’,Am. Surgeon,64, pp. 127–131Google Scholar
  5. Fronek, A., Blazek, V., andCurran, B. (1994): ‘Toe pressure determination by audiophotoplethysmography’,J. Vasc. Surg.,20, pp. 267–270Google Scholar
  6. Guelen, I., Westerhof, B. E., Van Der Sar, G. L., Van Montfrans, G. A., Kiemeneij, F., Wesseling, K. H., andBos, W. J. (2003): ‘Finometer, finger pressure measurements with the possibility to reconstruct brachial pressure’,Blood Press. Monit.,8, pp. 27–30CrossRefGoogle Scholar
  7. Hertzman, A. B., andSpealman, C. R. (1937): ‘Observations on the finger volume pulse recorded photo-electrically’,Am. J. Physiol.,119, pp. 334–335Google Scholar
  8. Hunyor, S. N. (1991): ‘Blood pressure in clinical practice’, inO'Brien, E., andO'Malley, K. (Eds.) ‘Blood pressure measurement, vol. 14’, in ‘Handbook of hypertension’ (Elsevier, Amsterdam, New York, 1991), chap. 4, pp. 95–111Google Scholar
  9. Johansson, A., Öberg, P. Å., andSedin, G. (1999): ‘Monitoring of heart and respiratory rates in newborn infants using a new photoplethysmographic technique’,J. Clin. Monit. Comput.,15, pp. 461–467CrossRefGoogle Scholar
  10. Jönsson, B., Lindberg, L. G., Skau, T., andThulesius, O. (2001): ‘Is oscillometric ankle pressure reliable in leg vascular disease?’,Clin. Physiol.,21, pp. 155–163Google Scholar
  11. Kamal, A. A. R., Harness, J. B., Irving, G., andMearns, A. J. (1989): ‘Skin photoplethysmography—a review’,Comput. Methods Programs Biomed.,28, pp. 257–269CrossRefGoogle Scholar
  12. Langbaum, M., andEyall, F. G. (1994): ‘A practical and reliable method of measuring blood pressure in the neonate by pulse oximetry’,J. Pediatr. 125, pp. 591–595Google Scholar
  13. Lindberg, L.-G., Ugnell, H., andÖberg, P. Å. (1992): ‘Monitoring of respiratory and heart rates using a fibre-optic sensor’,Med. Biol. Eng. Comput.,30, pp. 533–537Google Scholar
  14. O'Brien, E., Waeber, B., Parati, G., Staessen, J., andMyers, M. G. (2001): ‘Blood pressure measuring devices: recommendations of the European Society of Hypertension (Clinical review)’,Br. Med. J.,322, pp. 531–536Google Scholar
  15. Sapinski, A. (1996): ‘Standard algorithm for blood pressure measurement by sphygmo-oscillographic method’,Med. Biol. Eng. Comput.,34, pp. 82–83Google Scholar
  16. Tremper, K. K., andBarker, S. J. (1989): ‘Pulse oximetry’,Anesthesiology,70, pp. 98–108Google Scholar
  17. Weinman, J. (1967): ‘Photoplethysmography’, inVenables, P., andMartin, I. (Eds.): ‘A manual of psychophysiological methods’ (Wiley, New York, 1967), pp. 185–217Google Scholar
  18. Yamakoshi, K.-I., andKamiya, A. (1987): ‘Noninvasive measurement of arterial blood pressure and elastic properties using photoelectric plethysmography technique’,Med. Progr. Technol.,12, pp. 123–143Google Scholar
  19. Yamakoshi, K.-I. (1991): ‘Non-invasive techniques for ambulatory blood pressure monitoring and simultaneous cardiovascular measurements’,J. Ambulat. Monit.,4, pp. 123–143Google Scholar
  20. Yao, S. T. (1993): ‘Pressure measurements in the extremity’, inBernstein, E. F. (Ed.): ‘Vascular diagnosis’ (Mosby, St. Louis, 1993), pp. 169–175Google Scholar
  21. Zhang, Q. (2002): ‘Non-invasive measurement of skeletal muscle blood using photoplethysmography—Experimental studies on the human leg’. Medical dissertation, Dept of Orthopaedics, Institute of Surgical Sciences, Göteborgs University, Sweden, p. 17Google Scholar
  22. Zourabian, A., Siegel, A., Chance, B., Ramanujan, N., Rode, M., andBoas, D. A. (2000): ‘Trans-abdominal monitoring of fetal arterial blood oxygenation using pulse oximetry’,J. Biomed. Opt.,5, pp. 391–405CrossRefGoogle Scholar

Copyright information

© FMBE 2005

Authors and Affiliations

  • C. Laurent
    • 1
  • B. Jönsson
    • 2
  • M. Vegfors
    • 3
  • L. -G. Lindberg
    • 1
  1. 1.Department of Biomedical EngineeringLinköpings universitetSweden
  2. 2.Department of Cardiovascular Surgery & AnaesthesiaLinköping University HospitalSweden
  3. 3.Department of AnaesthesiologyLinköping University HospitalSweden

Personalised recommendations