A new procedure for simultaneous calibration of separation and axial dispersion in SEC
- 23 Downloads
- 45 Citations
Summary
A new computational procedure for the simultaneous calibration of separation and axial dispersion by using SEC coupled with a LALLS detector is presented. Its performance is tested by SEC/LALLS experiments with commercially available samples of broad and narrow polystyrene standards. The peakbroadening parameters were calculated by a novel method which utilizes the different effects of axial dispersion on polymers with narrow and broad molecular weight distribution. The dispersion function was assumed to be gaussian. The variation of the dispersion parameter with molecular weight is in good agreement with reports in the literature. Experimental chromatograms have been corrected for the axial dispersion effect.
Key Words
Size exclusion chromatography Lightscattering Calibration Axial dispersion Dispersion correctionPreview
Unable to display preview. Download preview PDF.
References
- [1]J. V. Dawkins, in ‘Steric Exclusion Liquid Chromatography of Polymers’, Chromatographic Science Series, Vol. 25, J. Janca, Ed., Dekker, New York, 1984, p. 53.Google Scholar
- [2]A. E. Hamielec, in ‘Steric Exclusion Liquid Chromatography of Polymers’, Chromatographic Science Series, Vol. 25, J. Janca, Ed., Dekker, New York, 1984, p. 117.Google Scholar
- [3]‘Modern Liquid Chromatography of Macromolecules’, Journal of Chromatography Library — vol. 25,B. G. Belenkii, L. Z. Vilenchik, Elsevier, New York, 1983, p. 117.Google Scholar
- [4]L. H. Tung, J. C. Moore, G. W. Knight, J. Appl. Polym. Sci.10, 1261 (1966).Google Scholar
- [5]D. Alba, G. R. Meira, J. Liq. Chromatog.,9 (6), 1141 (1986).Google Scholar
- [6]K. C. Berger, Makromol. Chem.176, 399 (1975).Google Scholar
- [7]Y. Chen, M. Ye, X. Li, L. Shi, J. Liq. Chromatog.,9 (6), 1163 (1986).Google Scholar
- [8]Z. D. He, X. Ch. Zhang, R. S. Cheng, J. Liq. Chromatogr.,5(7), 1209 (1982).Google Scholar
- [9]T. O. Nguyen, H.-H. Kausch, One Day Advanced Symposium on Size Exclusion Chromatography, Lausanne, 1987.Google Scholar
- [10]M. Netopilik, Polymer Bulletin7, 575 (1982).Google Scholar
- [11]K. Lederer, G. Imrich-Schwarz, M. Dunky, J. Appl. Polym. Sci.,32 4751 (1986).Google Scholar
- [12]G. Glöckner, J. Liq. Chromatog.7 (9), 1769 (1984).Google Scholar
- [13]L. H. Tung, J. Appl. Polym. Sci.13, 775 (1969).Google Scholar
- [14]K. C. Berger, Makromol. Chem.179, 719 (1978).Google Scholar
- [15]W. W. Yau, H. J. Stoklosa, D. D. Bly, J. Appl. Polym. Sci.21, 1911 (1977).Google Scholar
- [16]P. Fromment, A. Revillon, J. Liq. Chromatog.,10 (7), 1383 (1987).Google Scholar
- [17]R. Hooke, T. A. Jeeves, J. of the Assn. for Computing Machinery8, 212 (1961).Google Scholar
- [18]J. A. Nelder, R. Mead, The Computer Journal,7, 308 (1965).Google Scholar
- [19]T. Ishige, S.-I. Lee, A. E. Hamielec, J. Appl. Polym. Sci.,15, 1607 (1971).Google Scholar
- [20]G. Rois, Diplomarbeit Sept. 1987, Montanuniversität Leoben.Google Scholar
- [21]R. Brüssau, GIT Fachz. Lab. 5/87, 388.Google Scholar
- [22]R. Brüssau in ‘Liquid Chromatography of Polymers and Related Materials II’,J. Cazes andX. Delaware, Eds., Marcel Dekker, New York, 1980, p. 73.Google Scholar
- [23]R. Cheng, S. Bo in ‘Size Exclusion Chromatography, Methodology and Characterization of Polymers and Related Materials’, ACS Symposium Series 245, T. Provder, Ed., 1984, p. 125.Google Scholar
- [24]G. Eggink, J. A. P. P. van Dijk, J. A. M. Smit, Polymer Bulletin17, 531 (1987).Google Scholar
- [25]A. C. Ouano, J. Chromatogr.,118, 303–312 (1976).Google Scholar