Corrections for the Marinelli geometry

  • P. Dryák
  • P. Kovář
  • L. Plchová
  • J. Šurâň
Article

Abstract

The Marinelli beaker of the volume of 450 ml for the measurement of the radionuclide activity has been built in the Institute of Research, Production, and Application of Radioisotopes, according to the norm IEC 697/1981. For the determination of the two main correction factors, i.e. of the corrections for true summations and for the self-absorption, simple procedures have been worked out that make possible to find the correction factor for true summations with a precision better than 0.3% for the self-absorption with the precision better than 1%.

Keywords

Physical Chemistry Inorganic Chemistry Radionuclide Correction Factor Simple Procedure 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    P. Dryák, P. Kovář,Radioisotopy, 27 /1986/ 288.Google Scholar
  2. 2.
    F.J. Schima,Intern. J. Appl. Radiation Isotopes, 34 /1983/ 1109.CrossRefGoogle Scholar
  3. 3.
    D.S. Andreev,Izv. Akad. Nauk SSSR, ser. fiz., 37 /1973/ 1609.Google Scholar
  4. 4.
    V.P. Mashkovich, Zashchita ot ionisiruyushchikh izluchenii Energoatomizdat, Moscow, 1982.Google Scholar
  5. 5.
    J.H. Hubbell,Intern. J. Appl. Radiation Isotopes, 33 /1982/ 1269.CrossRefGoogle Scholar

Copyright information

© Akadémiai Kiadó 1989

Authors and Affiliations

  • P. Dryák
    • 1
  • P. Kovář
    • 1
  • L. Plchová
    • 1
  • J. Šurâň
    • 1
  1. 1.Institute of Research, Production, and Application of RadioisotopesPrague 10Czechoslovakia

Personalised recommendations