Quantitative determination of particulate contamination in intravenous administration sets

  • E. R. Di Paolo
  • B. Hirschi
  • A. Pannatier
Articles

Abstract

Particle counts were performed on 1,000 intravenous administration sets from ten suppliers on the Swiss market using a HIAC/Royco electronic counter. The following main conclusions may be drawn from the results of this study:
  • the small particles were the most numerous, regardless of the type of set;

  • the differences in the counts obtained for the different suppliers' sets tended to level off for particles larger than 10μm;

  • the drip chamber and latex connector may be two important sources of particles;

  • the particulate contamination from the sets is relatively low compared with the amount of particles contained in the parenteral solutions.

Keywords

Equipment and supplies Infusions, parenteral Particle size Particulate matter 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Akers MJ. Parenteral quality control. New York: Marcel Dekker, 1985:143–97.Google Scholar
  2. 2.
    De Luca PP, Knapp JZ. Particulate matter. In: Avis KE, Lachman L, Lieberman HA, eds. Pharmaceutical dosage forms. Parenteral medication. New York: Marcel Dekker, 1984:217–307.Google Scholar
  3. 3.
    Anonymous. I.V. filters. Med Lett Drug Ther 1975;17:35–6.Google Scholar
  4. 4.
    Schubert GE, Reifferscheid P, Flach A. Mikroembolien von Fremdmaterial nach Angiographie und intravenösen Infusionen. Dtsch Med Wochenschr 1972;97:1745–8.Google Scholar
  5. 5.
    Dimmick JE, Bove KE, Mc Adams J, Benzing G. Fiber embolization: a hazard of cardiac surgery and catheterization. N Engl J Med 1975;292:685–7.Google Scholar
  6. 6.
    Klaus E. Materielle Verunreinigung in Infusionslösungen. Klin Anästh Intensivther 1977;14:54–69.Google Scholar
  7. 7.
    Ahnefeld FW, Klaus E. Quantitative Analyse über den Partikelgehalt von Infusionslösungen,-zubehör und Medikamenten. Anaesthesist 1977;26:476–84.Google Scholar
  8. 8.
    Gesler RM, Garvin PJ, Klamer B, et al. The biological effects of polystyrene latex particles administered intravenously to rats. Bull Parenter Drug Assoc 1973;27:101–13.Google Scholar
  9. 9.
    Schröder HG, Simmons GH, De Luca PP. Distribution of radiolabeled subvisible microspheres after intravenous administration to Beagle dogs. J Pharm Sci 1978;67:504–7.Google Scholar
  10. 10.
    Kanke M, Simmons GH, Weiss DL, Bivins BA, De Luca PP. Clearance of141Ce-labeled microspheres from blood and distribution in specific organs following intravenous and intraarterial administration in Beagle dog. J Pharm Sci 1980;69:755–62.Google Scholar
  11. 11.
    Ryan PB, Rapp RP, De Luca PP, Griffen WO, Clark JD, Cloys D. In-line final filtration. A method of minimizing contamination in intravenous therapy. Bull Parenter Drug Assoc 1973;27:1–14.Google Scholar
  12. 12.
    Bivins BA, Rapp RP, De Luca PP, Mc Kean H, Griffen WO. Final in-line fitraton: a mean of decreasing the incidence of infusion phlebitis. Surgery 1979;85:388–94.Google Scholar
  13. 13.
    Anonymous. Pharmacopoea Helvetica VII. Parenteralia. Bern: 1987.Google Scholar
  14. 14.
    Anonymous. Pharmacopée Européenne. Parenteralia. Ste-Ruffine: Maisonneuve SA, 1977:414.Google Scholar
  15. 15.
    Anonymous. United States Pharmacopeia XXI. Rockville: United States Pharmacopeial Convention, 1985.Google Scholar
  16. 16.
    Ho NFH, Church RL, Lee H. Particulate matter in parenteral solutions. II. Drug Intell 1967;1:356–61.Google Scholar
  17. 17.
    Williams A, Barnett MI. Particulate matter in intravenous fluids, administration sets and cannulae. Pharm J 1973;211:190–1.Google Scholar
  18. 18.
    Olson OT. Particulate generation from equipment used for the administration of infusion fluids. Acta Pharm Suec 1980;17:209–14.Google Scholar
  19. 19.
    Harrison MJ, Healy TEJ. Intravenous administration sets: the effect of flushing and filtration on particulate contamination. Br J Anaesth 1974;46:59–65.Google Scholar
  20. 20.
    Pavanetto F, Conti B, Modena T, Genta I, Ponci R. Particulate contamination in parenteral type medical device. Int J Pharm 1988;48:255–265.Google Scholar
  21. 21.
    Lantz RJ, Shami EG, Lachmann L. Application of the HIAC particle counter to monitoring particulate matter in parenteral solutions I. Bull Parenter Drug Assoc 1976;30:234–46.Google Scholar
  22. 22.
    Schröder HG, De Luca PP. Theoretical aspects of particulate matter monitoring by microscopic and instrumental methods. J Parenter Drug Assoc 1980;34:183–9.Google Scholar
  23. 23.
    Winer DJ. Statistical principles in experimental dessign. New York: McGraw Hill, 1971.Google Scholar
  24. 24.
    Bolton S. Pharmaceutical statistics. New York: Marcel Dekker, 1984.Google Scholar
  25. 25.
    Anonymous. Deutsches Institut für Normung (DIN). 58 362/1. Berlin: Beuth Verlag, 1988.Google Scholar

Copyright information

© Royal Dutch Association for Advancement of Pharmacy 1990

Authors and Affiliations

  • E. R. Di Paolo
    • 1
  • B. Hirschi
    • 1
  • A. Pannatier
    • 1
  1. 1.Service of PharmacyCentre Hospitalier Universitaire Vaudois (C.H.U.V.)LausanneSwitzerland

Personalised recommendations