Analysis of the accuracy of continuous thermodilution cardiac output measurement

Comparison with intermittent thermodilution and Fick cardiac output measurement
  • L. Jacquet
  • G. Hanique
  • D. Glorieux
  • P. Matte
  • M. Goenen
Technique

Abstract

Objective

To evaluate the accuracy of cardiac output measurement obtained by a new continuous thermodilution cardiac output (CCO) pulmonary artery catheter compared to intermittent thermodilution (TCO) and the direct Fick method.

Design

Prospective open trial.

Setting

University hospital, intensive care unit.

Patients

23 patients (15 surgical, 8 non-surgical) were monitored with the Intellicath pulmonary catheter. Cardiac output was evaluated by the three methods every 4 to 6 h as long as the pulmonary artery catheter was necessary (8–96h).

Results

The correlation coefficient between CCO and TCO was 0.92, no systematic bias was observed, and the relative error increased from 13.9% for a cardiac output of 1 l/min to 23.7% for an output of 10 l/min. When comparing CCO and Fick, the correlation coefficient was 0.89, no bias was detected, and the relative error increased from 20.4% for outputs of 2 l/min to 27.2% for outputs of 10 l/min.

Conclusions

CCO provides clinically acceptable measurements. At high cardiac outputs, the difference with other methods increases and the results must be cautiously interpreted.

Key words

Continuous cardiac output Thermodilution Fick principle Pulmonary artery catheter 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Bernstein DP (1986) Continuous non-invasive real-time monitoring of stroke volume and cardiac output by the thoracic electrical bioimpedance. Crit Care Med 14:898–901Google Scholar
  2. 2.
    Perrino AC, Lippman A, Ariyan C, O'Connor TZ, Luther M (1994) Intraoperative cardiac output monitoring: comparison of impedance cardiography and thermodilution. J Cardiothorac Vasc Anesth 8:24–29Google Scholar
  3. 3.
    Davies G, Jebson P, Hess D (1986) Continuous Fick cardiac output compared to thermodilution. Crit Care Med 14:881–885Google Scholar
  4. 4.
    Keinanen O, Takala J, Kari A (1992) Continuous measurement of cardiac output by the Fick principle: clinical validation in intensive care. Crit Care Med 20:360–365Google Scholar
  5. 5.
    Jansen JR, Wesseling KH, Schreuder JJ (1990) Continuous cardiac output monitoring by pulse contour during cardiac surgery. Eur Heart J 11 (Suppl I):26–32Google Scholar
  6. 6.
    Abrams JH, Weber RE, Holmen KD (1989) Transtracheal Doppler: a new method of continuous cardiac output measurement. Anesthesiology 70:134–138Google Scholar
  7. 7.
    Miyasaka K, Takata M, Miyasaka K (1993) Flow velocity profile of the pulmonary artery measured by the continuous cardiac output monitoring catheter. Can J Anaesth 40:98–102Google Scholar
  8. 8.
    Yelderman M, Ramsay MA, Quinn MD, Paulsen AW, McKown RC, Gillman PH (1992) Continuous thermodilution cardiac output measurement in intensive care unit patients. J Cardiothorac Vasc Anesth 3:270–274Google Scholar
  9. 9.
    Boldt J, Menges T, Wollbrück M, Hammermann H, Hempelmann G (1994) Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient. Crit Care Med 22:1913–1918Google Scholar
  10. 10.
    Yelderman M (1990) Continuous measurement of cardiac output with the use of stochastic system identifications techniques. J Clin Monit 6:322–332Google Scholar
  11. 11.
    Mihaljevic T, von Segesser LK, Tonz M, Leskosek B, Jenni R, Turina M (1994) Continuous thermodilution measurement of cardiac output: in vitro and in vivo evaluation. J Thorac Cardiovasc Surg 42:32–35Google Scholar
  12. 12.
    Trush D, Downs JB, Smith RA (1995) Continuous thermodilution cardiac output: agreement with Fick and bolus thermodilution methods. J Cardiothorac Vasc Anesth 9:399–404Google Scholar
  13. 13.
    Jansen JRC, Versprille A (1986) Improvement of cardiac output estimation of the thermodilution method during mechanical ventilation. Intensive Care Med 12:71–79Google Scholar
  14. 14.
    Stevens JH, Raffin TA, Mihm FG, Rosenthal MH, Stetz CW (1985) Thermodilution cardiac output measurement. Effects of the respiratory cycle on its reproducbility. JAMA 253:2240–2242Google Scholar
  15. 15.
    Takala J, Keinanen O, Vaisanen P, Kari A (1989) Measurement of gas exchange in intensive care: laboratory and clinical validation of a new device. Crit Care Med 17:1041–1047Google Scholar
  16. 16.
    Makita K, Nunn JF, Royston B (1990) Evaluation of metabolic measuring instruments for use in critically ill patients. Crit Care Med 18:638–644Google Scholar
  17. 17.
    Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310Google Scholar
  18. 18.
    Yelderman M, Quinn MD, McKnown RC, Eberhart RC, Dollar ML (1992) Continuous thermodilution cardiac output measurement in sheep. J Thorac Cardiovasc Surg 104:315–320Google Scholar
  19. 19.
    Lichtenthal PR, Marchand B, Gordon D, Leissing N, Konno M (1992) Asafety comparison between a new continuous cardiac output monitoring system and a standard pulmonary artery catheter in sheep (abstract). Anesthesiology 77:A473Google Scholar

Copyright information

© Springer-Verlag 1996

Authors and Affiliations

  • L. Jacquet
    • 1
  • G. Hanique
    • 1
  • D. Glorieux
    • 1
  • P. Matte
    • 1
  • M. Goenen
    • 1
  1. 1.Cardio-thoracic Intensive Care UnitSaint-Luc University HospitalBrusselsBelgium

Personalised recommendations