Fallout analysis of atmospheric water precipitations

Part I Determination of caesium-137 and strontium-89, 90 by ion-exchange
  • M. Senegačnik
  • Š. Paljk
Originalabhandlungen

Summary

A rapid ion-exchange method for the determination of 137Cs, 89Sr, and 90Sr in rain-, snow-, cistern-, and soft surface-waters is presented. The sorption of 2–30 l samples with a calcium content up to 300 mg is carried out on a specially shaped column filled with Dowex 50, X-8 or Dowex 50 W, X-8. Radiocaesium is washed from the column with 0,6 M ammonium glycolate/0.2 M NaCl, pH 5, radio-strontium with 1.0 M ammonium glycolate/0.3 M NaCl, pH 5. After their isolation from the respective eluates the radio-nuclides are counted on a low-background beta-counter: 137Cs as Cs-dipicrylaminate, 89Sr/90Sr as SrCO3, and 90Y as Y2O3 in cases of indirect 90Sr determination. The mean chemical recoveries of added carriers amount to 85% for caesium, 95% for strontium and 95% for yttrium. At levels of a few picocuries, in parallel determinations, the deviations of the results for 137Cs and 90Sr from respective mean values generally do not exceed ±5%. With a 30 l sample the limit of detection amounts to 0.006 μμCi/l for 137Cs and 0.005 μμCi/l for 90Sr.

Keywords

Strontium Yttrium Y2O3 Water Precipitation Caesium 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Zusammenfassung

Für die Bestimmung von 137Cs, 89Sr und 90Sr in Regen-, Schnee-, Zisternen- und weichem Oberflächenwasser wird eine schnelle Ionenaustausch-methode vorgeschlagen. Die Sorption von 2–30 l Wasserproben mit einem Calciumgehalt bis zu 300 mg wird mit einer besonders gestalteten, mit Dowex 50, X-8 oder Dowex 50 W, X-8, gefüllten Säule durchgeführt. Aus der Kolonne werden Radio-Caesium mit 0,6 M Ammoniumglykolat/ 0,2 M NaCl, pH 5 und Radio-Strontium mit 1,0 M Ammoniumglykolat/ 0,3 M NaCl, pH 5 eluiert. Nach ihrer Isolierung aus den entsprechenden Eluatsfraktionen werden die Radio-Nuklide in einem Beta-Antikoinzidenzzähler gemessen: 137Cs als Cs-Dipikrylaminat, 89Sr/90Sr als SrCO3, und 90Y als Y2O3 im Fall der indirekten 90Sr-Bestimmung. Die durchschnittliche Rückgewinnung der zugesetzten Träger beträgt 85% für Caesium, 95% für Strontium und 95% für Yttrium. In Parallelbestimmungen überschreiten bei Gehalten von einigen μμCi/l die Abweichungen der 137Cs- und 90Sr-Werte von den entsprechenden Mittelwerten im allgemein nicht ±5%. Bei Anwendung einer 30 l-Wasserprobe beträgt die untere Nachweisgrenze 0,006 μμCi/l für 137Cs und 0,005 μgmCi/l für 90Sr.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    For an extensive review of these procedures see Anal. Chem., Annual (Analyt.) Reviews, sections Ion-Exchange, Nucleonics 30, No. 4, Part II (1958); 32, No.5 (1960); 34, No. 5 (1962); 36, No. 5 (1964); 38, No. 5 (1966), and section Water 31, No. 4, Part II (1959); 33, No. 5 (1961); 35, No. 5 (1963); 37, No. 5 (1965).Google Scholar
  2. 2.
    Dowex-ion exchange. Midland, Michigan: The Dow Chemical Company 1964.Google Scholar
  3. 3.
    Dyrssen, D.: Acta Chem. Scand. 11, 1277 (1957); cf. Z. Anal. Chem. 163, 136 (1958).Google Scholar
  4. 4.
    —, and D. H. Liem: Acta Chem. Scand. 14, 1091 (1960).Google Scholar
  5. 5.
    Evans, E. A., and J. L. Huston: Anal. Chem. 24, 1482 (1952).Google Scholar
  6. 6.
    Friedlander, G., J. W. Kennedy, and J. M. Miller: Nuclear and radio-chemistry. New York: John Wiley & Sons 1964.Google Scholar
  7. 7.
    Herrmann, R., and C. T. J. Alkemade: Chemical analysis by flame photometry. New York: Interscience 1963.Google Scholar
  8. 8.
    Južnič, K., M. Senegačnik, and Z. Milavec: Z. Anal. Chem. 207, 193 (1965).Google Scholar
  9. 9.
    Kahn, B., D. K. Smith, and C. P. Straub: Anal. Chem. 29, 1210 (1957); cf. Z. Anal. Chem. 161, 381 (1958).Google Scholar
  10. 10.
    Knapstein, H.: Z. Anal. Chem. 175, 255 (1960).Google Scholar
  11. 11.
    Kunin, R.: Ion exchange resins, p. 322. New York: John Wiley & Sons, Inc. 1963.Google Scholar
  12. 12.
    McNaughton, G. S., and R. N. Woodward: New Zealand J. Sci. 4, 523 (1961).Google Scholar
  13. 13.
    Morrison, C. (ed.): Trace analysis (physical methods). New York: John Wiley & Sons 1965.Google Scholar
  14. 14.
    Osmond, R. G. D., T. W. Evett, J. W. Arden, M. B. Lovett, and B. Swenney: A.E.R.E.-AM 84, Harwell 1961.Google Scholar
  15. 15.
    Spedding, F. H., and A. H. Daane: The rare earths, p. 59 New York: John Wiley & Sons, Inc. 1961.Google Scholar
  16. 16.
    Welcher, F. J.: Organic analytical reagents IV, 8. New York: D. Van Nostrand 1948.Google Scholar

Copyright information

© Springer-Verlag 1967

Authors and Affiliations

  • M. Senegačnik
    • 1
    • 2
  • Š. Paljk
    • 1
    • 2
  1. 1.Nuclear Institute “J. Stefan”Ljubljana
  2. 2.Department of MetallurgyThe University of LjubljanaYugoslavia

Personalised recommendations