In vivo molecular sensing in diabetes mellitus: an implantable glucose sensor with direct electron transfer

  • J. C. Pickup
  • G. W. Shaw
  • D. J. Claremont
Originals

Summary

Miniature, amperometric glucose sensors were constructed for implantation in the subcutaneous tissue of normal and insulin-dependent diabetic subjects. To minimise dependence on fluctuating tissue oxygen tension, we employed the technology of mediated electron transfer, with 1,1′-dimethylferrocene acting as the redox shuttle between immobilized glucose oxidase and a platinum base electrode. In 6 normal subjects, the subcutaneous sensor responses mirrored the simultaneously-measured changes in blood glucose concentration after a 75 g oral glucose load and after intravenous injection of 0.15 U/kg short-acting insulin, though increases and decreases in the sensor output were slower than the glycaemic changes. The mean peak delay in sensor response after the oral glucose was 40 min (range 0–45 min) and the delay in the hypoglycaemic nadir was 4 min (range 0–15 min). In 5 insulin-dependent diabetic subjects, spontaneous and induced hypoglycaemia was detectable by the implanted sensor. In addition, marked and frequent oscillations in the sensor current occurred in several normal and diabetic individuals as the blood glucose fell below about 1.9 mmol/l. These oscillations were present in a diabetic subject who had lost adrenergic warning symptoms to hypoglycaemia. Continuous metabolic monitoring in diabetes, particularly the detection of hypoglycaemia, may be possible with implanted sensors based on this technology.

Key words

Glucose sensor enzyme electrode biosensor glucose oxidase ferrocene diabetes mellitus 

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Pickup JC (1985) Biosensors: a clinical perspective. Lancet II: 817–820Google Scholar
  2. 2.
    Anon (1987) Awareness of hypoglycaemia in diabetes. Lancet II: 371–372Google Scholar
  3. 3.
    Claremont DJ, Sambrook IE, Penton C, Pickup JC (1986) Subcutaneous implantation of a ferrocene-mediated glucose sensor in pigs. Diabetologia 29: 817–821Google Scholar
  4. 4.
    Pickup JC, Shaw GW, Claremont DJ (1989) Potentially-implantable amperometric glucose sensors with mediated electron transfer: improving the operating stability. Biosensors in pressGoogle Scholar
  5. 5.
    Claremont DJ, Shaw GW, Pickup JC (1988) Improving the stability of potentially-implantable glucose sensors. Diabetic Med 5 [Suppl.]: 2AGoogle Scholar
  6. 6.
    Cass AEG, Davis G, Francis GD, Hill HAO, Aston WJ, Higgins IJ, Plotkin EV, Scott LDL, Turner APF (1984) Ferrocene-mediated enzyme electrode for amperometric determination of glucose. Anal Chem 56: 667–671Google Scholar
  7. 7.
    Matthews DR, Bown E, Beck TW, Plotkin E, Lock L, Gosden E, Wickham M (1988) An amperometric needle-type glucose sensor tested in rats and man. Diabetic Med 5: 248–252Google Scholar
  8. 8.
    Bessman SP, Thomas LJ, Kojiima H, Sayler DF, Layne EC (1981) The implantation of a closed-loop artifical beta cell in dogs. Trans Am Soc Art Int Org 27: 7–17Google Scholar
  9. 9.
    Abel P, Miller A, Fischer U (1984) Experience with an implantable glucose sensor as a prerequisite of an artificial beta cell. Biomed Biochim Acta 5: 577–584Google Scholar
  10. 10.
    Shichiri M, Kawamori R (1983) Feasibility of a needle-type glucose sensor and the wearable artifical endocrine pancreas system in human diabetic subjects. In: Irsigler K, Kritz, Lovett R (eds) Diabetes treatment with implantable insulin infusion systems, Urban & Schwarzenberg, Munich, pp 224–230Google Scholar
  11. 11.
    Fischer U, Ertle R, Abel P, Rebrin K, Brunstein E, Hahn von Dorsche H, Freyse EJ (1987) Assessment of subcutaneous glucose concentration: validation of the wick technique as a reference for implanted electrochemical sensors in normal and diabetic dogs. Diabetologia 30: 940–945Google Scholar
  12. 12.
    Velho G, Froguel P, Reach G, Thevenot D (1987) In vivo calibration of a needle-type glucose sensor to determine the kinetics of subcutaneous glucose concentration. Diabetologia 30: 592 AGoogle Scholar
  13. 13.
    Rebrin K, Fischer U, Woedtke TV, Brunstein E, Abel P (1988) Tissue glucose telemetry and control in normal and diabetic dogs. Diabetologia 31: 534 AGoogle Scholar
  14. 14.
    D'Costa EJ, Higgins IJ, Turner APF (1986) Quinoprotein glucose dehydrogenase and its application in an amperometric glucose sensor. Biosensors 2: 71–87Google Scholar
  15. 15.
    Kulys JJ (1986) Enzyme electrodes based on organic metals Biosensors 2: 3–13Google Scholar
  16. 16.
    Albery WJ, Bartlett PN, Craston DH (1985) Amperometric enzyme electrodes II. Conducting salts as electrode materials for the oxidation of glucose oxidase. J Electroanal Chem 194: 223–235Google Scholar

Copyright information

© Springer-Verlag 1989

Authors and Affiliations

  • J. C. Pickup
    • 1
  • G. W. Shaw
    • 1
  • D. J. Claremont
    • 1
  1. 1.Division of Chemical PathologyUnited Medical and Dental Schools of Guy's and St Thomas's Hospitals, Guy's HospitalLondonUK

Personalised recommendations