Imaging responses of on-site CsI and Gd2O2S flat-panel detectors: Dependence on the tube voltage

  • Published:
Journal of the Korean Physical Society Aims and scope Submit manuscript

A Correction to this article was published on 26 April 2020

This article has been updated

Abstract

One of the emerging issues in radiography is low-dose imaging to minimize patient’s exposure. The scintillating materials employed in most indirect flat-panel detectors show a drastic change of X-ray photon absorption efficiency around their K-edge energies that consequently affects image quality. Using various tube voltages, we investigated the imaging performance of most popular scintillators: cesium iodide (CsI) and gadolinium oxysulfide (Gd2O2S). The integrated detective quantum efficiencies (iDQE) of four detectors installed in the same hospital were evaluated according to the standardized procedure IEC 62220-1 at tube voltages of 40 - 120 kVp. The iDQE values of the Gd2O2S detectors were normalized by those of CsI detectors to exclude the effects of image postprocessing. The contrast-to-noise ratios (CNR) were also evaluated by using an anthropomorphic chest phantom. The iDQE of the CsI detector outperformed that of the Gd2O2S detector over all tube voltages. Moreover, we noted that the iDQE of the Gd2O2S detectors quickly rolled off with decreasing tube voltage under 70 kVp. The CNRs of the two scintillators were similar at 120 kVp. At 60 kVp, however, the CNR of Gd2O2S was about half that of CsI. Compared to the Gd2O2S detectors, variations in the DQE performance of the CsI detectors were relatively immune to variations in the applied tube voltages. Therefore, we claim that Gd2O2S detectors are inappropriate for use in low-tube-voltage imaging (e.g., extremities and pediatrics) with low patient exposure.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Change history

  • 26 April 2020

    The affiliation of Yongkan Ki should be corrected as follow:

    Yongkan Ki

    Department of Radiation Oncology, Pusan National University Hospital and Pusan National University School of Medicine, Busan 46241, Korea

References

  1. E. Samei and M. J. Flynn, Med. Phys. 30, 608 (2003).

    Article  Google Scholar 

  2. M. J. Yaffe and J. A. Rowlands, Phys. Med. Biol. 42, 1 (1997).

    Article  Google Scholar 

  3. H. K. Kim, I. A. Cunningham, Z. Yin and G. Cho, Int. J. Precis. Eng. Manuf. 9, 86 (2008).

    Google Scholar 

  4. W. Zhao, I. Blevis, S. Germann, J. Rowlands, D. Waechter and Z. Huang, Med. Phys. 24, 1834 (1997).

    Article  Google Scholar 

  5. J. Siewerdsen, L. Antonuk, Y. El-Mohri, J. Yorkston, W. Huang, J. Boudry and I. Cunningham, Med. Phys. 24, 71 (1997).

    Article  Google Scholar 

  6. P. G. Barten, Contrast sensitivity of the human eye and its effects on image quality (SPIE Press, WA, 1999).

    Book  Google Scholar 

  7. D. J. Brenner and E. J. Hall, New Engl. J. Med. 357, 2277 (2007).

    Article  MATH  Google Scholar 

  8. IEC62220-1, Medical electrical equipment-characteristics of digital X-ray imaging devices-Determination of the detective quantum efficiency (IEC, Geneva, 2002).

  9. G. Borasi, A. Nitrosi, P. Ferrari and D. Tassoni, Med. Phys. 30, 1719 (2003).

    Article  Google Scholar 

  10. IEC61267, Medical diagnostic X-ray equipment- Radiation conditions for use in the determination of characteristics (IEC, Geneva, 2005).

  11. H. K. Kim, C. H. Lim, J. Tanguay, S. Yun and I. A. Cunningham, Med. Phys. 39, 2478 (2012).

    Article  Google Scholar 

  12. I. M. Blevis, D. C. Hunt and J. A. Rowlands, Med. Phys. 25, 638 (1998).

    Article  Google Scholar 

  13. C. W. van Eijk, Phys. Med. Biol. 47, R85(2002).

    Article  Google Scholar 

  14. A. Sarapata, J. W. Stayman, M. Finkenthal, J. H. Siewerdsen, F. Pfeiffer and D. Stutman, Med. Phys. 41, 021904 (2014).

    Article  Google Scholar 

  15. R. E. Burgess, Discuss. Faraday Soc. 28, 151 (1959).

    Article  Google Scholar 

  16. M. C. Teich, B. E. A. Saleh and D. Stoler, Optics Comm. 46, 244 (1983).

    Article  ADS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Myung Jin Chung.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Jeon, H., Chung, M.J., Youn, S. et al. Imaging responses of on-site CsI and Gd2O2S flat-panel detectors: Dependence on the tube voltage. Journal of the Korean Physical Society 67, 264–269 (2015). https://doi.org/10.3938/jkps.67.264

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.3938/jkps.67.264

Keywords

Navigation