Background

Frusemide or furosemide (FUR), chemically known as 4-chloro-2-[(furan-2-ylmethyl)-amino]-5-sulfamoyl-benzoic acid (Fig. 1), as a potent diuretic, is used to treat excessive fluid accumulation and swelling of the body caused by heart failure, cirrhosis, chronic kidney failure, and nephritic syndrome [1].

Fig. 1
figure 1

Structure of FUR

The drug FUR is monographed in United States Pharmacopeia [2] for assay. The procedure describes titration of FUR (about 600 mg) in dimethylformamide against standardised 0.1 M NaOH solution using bromothymol blue as indicator.

The literature survey revealed that several analytical techniques were employed to determine FUR in pharmaceuticals, and they include titrimetry [3, 4], ultraviolet (UV) spectrophotometry [5, 6], visible spectrophotometry [7,8,9,10,11,12,13,14,15,16,17,18,19,20], spectrofluorimetry [21,22,23], and high-pressure liquid chromatography (HPLC) technique with different detection systems like UV [24], fluorescence [25, 26], amperometric [27], diode array [28], and mass spectrometric (MS) [29]. Besides, gas chromatography–MS [30], micellar-LC [31], derivative spectroscopy with ratio-spectral analysis [32], and diffuse reflectance spectroscopic [33] techniques were also used for FUR.

A couple of reports were found in the literature on potentiometric sensors used to quantify FUR. Santini et al. [34] described the construction and analytical application of a potentiometric sensor using platinum, mercury, mercuric nitrate, and graphite to estimate FUR. The article devoted by Dias et al. [35] describes the construction and use of polymeric electrode by incorporating the FUR–tricaprylmethylammonium chloride (FUR-aliquat 336S) ion pair complex into a membrane for assaying FUR. The reported potentiometric sensors are not selective to determine FUR in pharmaceuticals because of their incomplete validation and scanty competing analytical results. Therefore, in this paper, the preparation, validation, and application of two new membrane sensors are presented for FUR determination in pharmaceuticals.

Methods

Apparatus

A PICO model digital potentiometer (Chennai-32, India) was used in all potential measurements. An Elico (Mumbai, India) pH metre was used in pH recordings. The electrochemical cell utilised Ag/AgCl as a reference electrode and a copper wire as a conducting device.

High-purity distilled water and analytical grade chemicals were used. The reagents were prepared using analytical grade substances. Jubilant Life Sciences Ltd, Nanjangud, Mysore, India gifted the pure FUR (99.8%), and the sample was used as received. Lasix tablets (40 mg FUR/tablet) were obtained from the local market. Bromophenol blue dye (BPB), bromocresol green (BCG), tetrahydrofuran (THF), chloroform (CHCl3), dibutyl phthalate (DBP), and polyvinyl chloride (PVC) were supplied from SD Fine Chem Ltd., Mumbai, India. Glacial acetic acid (HOAc) was purchased from Merck (Mumbai, India) and used to prepare a 2:3 (v/v) solution in water.

Solutions of 4 M sodium acetate (NaOAc) and 0.1% each of acetic acid (CH3COOH), sodium hydroxide (NaOH), sodium carbonate (Na2CO3), sodium bicarbonate (NaHCO3), ammonium acetate (CH3COONH4), potassium thiocyanate (KSCN), fructose, sucrose, maltose, glucose, lactose, starch, glycine, ammonium chloride (NH4Cl), cadmium chloride (CdCl2), calcium chloride (CaCl2), cobalt chloride (CoCl2), sodium fluoride (NaF), nickel chloride (NiCl2), potassium chloride (KCl), sodium formate, sodium benzoate, sodium salicylate, sodium citrate, sodium tartarate, sodium sulphate, sodium oxalate, and ammonium borate were prepared using pure compounds purchased from S.D. Fine Chem Pvt Ltd, in water.

A stock of 3 mmol/L FUR solution was prepared in 2:3 HOAc using a calibration flask.

General procedures

Designing of sensors

Accurately weighed 10.1 mg of FUR was dissolved in 25 mL 2:3 HOAc. To the same solution, 20.04 mg of BPB or 20.6 mg of BCG dye was added; the content was transferred into a separatory funnel and extracted thrice with 10 mL portions of chloroform. The organic layer was passed through anhydrous sodium sulphate and collected into a clean and dry beaker. After evaporating the solvent, the resulted residue in the form of ion pair was used. About 20.0 mg of FUR-BPB or FUR-BCG ion pair was dissolved in 10 mL THF. The solution took in a Petri dish (width 4 cm), 0.1 g each of DBP and PVC were added; it was dissolved, and the content was dried under laboratory temperature for 24 h. The obtained dry membrane was composed with 9% (w/w) of respective ion pair and 45.5% (w/w) each of DBP and PVC. The resultant sufficiently dried membrane was attached to one end of a plastic tube using THF and dried at room temperature for 24 h. A 3–5 mL internal solution of 0.005 mol/L FUR was filled, the lower terminal of copper wire (15-cm length and 2.0-mm diameter) coated with insulator (leaving 1.0 cm at the top end and 0.5 cm at the bottom not insulated) was immersed, and the upper terminal was taken into connection to a negative terminal of the potentiometer. At this moment, the resulted FUR-BPB (Sensor I) and FUR-BCG (Sensor II) sensors were soaked into FUR solution and were used for measuring potentials after 5 h.

Procedure for bulk drug

Construction of calibration curves

Using a microburet, different volumes (0.0, 0.08, 1.0, 2.0, 3.0, 4.0, 5.0, 6.0, 7.0, and 8.0 ml) of 3.0 mmol/L FUR solutions were availed in a series of 10-mL calibrated flasks. The volumes in each flask were raised to 8 mL with 2:3 HOAc, and the pH were adjusted to 3.71 with 4 M NaOAc solution. The volumes were brought to 10 mL with 2:3 HOAc and the potentials were measured with respective membrane sensor against Ag/AgCl reference electrode.

The calibration curves were constructed using the solutions’ potentials data versus log [FUR] were prepared. The FUR concentrations in unknowns were calculated using calibration curves or regression data derived using potential versus log [FUR].

Study of interference

A 2-ml each of 3 mmol/L FUR and 0.1% interferent solutions were taken in an array of 10-mL calibrated flasks. The solutions were made to the mark with 2:3 HOAc and the potentials of each measured using FUR-BPB (Sensor I) or FUR-BCG (Sensor II) indicative sensors in reference to Ag/AgCl standard electrode.

Analysis of tablets

A group of twenty pre-weighed tablets were ground into a fine powder. The homogeneous powder corresponding to 50 mg FUR was measured and taken in a 50-ml calibrated flask, added 20 ml HOAc and shaken for 20 min and the volume raised to calibration mark with water. The content after mixing was filtered with Whattman No. 41 paper. A suitably measured volume of the resultant solution was drawn out, and potential was measured as given under the calibration curve construction. The FUR concentration in the taken extract was computed.

Results

Optimisation of parameters

Effect of dye content

The quantity of BPB or BCG was varied in the preparation of ionic aggregate with FUR. Sensors prepared with differential amounts of dyes were employed to measure the potential, and consequently, slopes of the linear curves were found out. Calibration lines’ slopes were close to 59.16 mV/decade when the quantities of BPB and BCG added were 20.04 and 20.6 mg in Sensors I and II, respectively. At other quantities of dyes, varying slopes were observed. The resultant slopes from differential amounts of dyes are summarised in Table 1.

Table 1 Influence of the dye content in the sensor’s membrane on the electrode response

Response time and lifetime of the electrodes

The time required to achieve a steady-state potential (within ± 0.2 mV/min), for 2.4 × 10-5–2.4 × 10-3 mol/L FUR solutions at pH 3.71. The potentials were reached steady state in 8 s for FUR solutions at concentrations above 1.0 × 10-3 mol/L and about 20 s at concentrations near the limit of detection. The experimental results proved 60 days of lifetime with a total of 500 determinations using these FUR-BPB and FUR-BCG sensors. During this period, each sensor was in daily use and was stored in 0.010 mol/L FUR solution and pH adjusted at 3.71 when not in use. There were no significant changes in working concentration range, slope, and response time in this period.

Effect of plasticizer

The sensing membranes were prepared separately by adding different amounts of dibutyl sebacate (DBS), dibutyl phthalate (DBP), dioctyl phthalate (DOP), and o-nitrophenyl octylether (NPOE) as plasticizers. The membranes were prepared using 0.1 g of DBP for both; the sensors were found to behave in Nernstian response. The sensors were found to perform satisfactorily with respect to stable potential readings, ease of conditioning, and less response time. Therefore, DBP was used as plasticizer in fabricating the FUR-BPB and FUR-BCG membrane sensors.

pH effect

The pH effect on electrode potential using electrodes over the range of 2.0–8.0 for 2.4 × 10-5–2.4 × 10-3 mol/L FUR solutions was investigated. The resulting solutions’ pHs were adjusted with either 2 M HNO3 or dilute NaOAc solutions. As it can be seen in Fig. 2, the potential response was found maximum and acceptable at pH 3.71 for both the sensors. However, at pHs which differ from 3.71, significant change in either the solution’s clarity or potentials was observed. Therefore, the pH of 3.71 was maintained as optimum in all potential measurements.

Fig. 2
figure 2

Effect variation of pH on potentials while using Sensors I and II at 25 ± 0.2 °C ([FUR] = 2.4 × 10-3 mol/L)

Evaluation of selectivity coefficients

The chemical interaction between the active sites of the membrane surface and the analyte is responsible for generation of membrane potential (Em). The poor selectivity of membranes towards a single species is found due to the factor of chemical process on the signal. Practically, Emem is in proportion to the concentration ions capable of interacting at the membrane’s active sites. The generalised Nernst equation to include the contribution of an interferent, I, can be written as follows [36].

$$ {E}_{\mathrm{cell}}=K+\frac{0.05916}{Z_A}\log \left(\left[A\right]+{K}_{A,I}{\left[I\right]}^{\frac{Z_A}{Z_I}}\right), $$
(1)

where K is a constant and includes reference electrode’s potential, liquid junction potential, and Em; ZA and ZI are the charges on the analyte and interferent, respectively, with concentrations [A] and [I]. The parameter KA,I is known as selectivity coefficient and is defined as

$$ {K}_{A,I}=\frac{{\left[A\right]}_E}{{{\left[I\right]}_E}^{\frac{Z_A}{Z_I}}} $$
(2)

where [A]E and [I]E are the analyte’s and interferent’s concentrations yielding identical cell potentials, i.e. E. When the KA,I is equal to unity, the membrane responds for both analyte and interferent. A selective membrane must always possess KA,I value of less than unity [36].

Thus, as an important feature, the value of KA,I can be evaluated using the data of potentials of solutions contained with known but constant amount of interferent, [I]add, and varying amounts of analyte. Two distinct linear regions in the plot of Ecell against logarithmic values of the analyte’s concentration are very informative in evaluating the value of KA,I. When the [A] is significantly larger than KA,I [I]add, Ecell is a linear function of log [A]. If KA,I [I]add is significantly larger than [A]; however, the Ecell remains constant. The analyte’s and interferent’s concentrations at the intersection of the two linear regions are used to calculate KA,I.

Thus, KA,I values were calculated in the presence of different interferents and are given below in Table 2. Both the sensors were used to measure the potentials of FUR along with various inorganic and organic compounds, anions, and cations. A 0.1% solution of each interferent was spiked into pre-analyzed FUR solution, and analysis was done [37, 38]. The summarised results indicated non-interference from the added species as showed by the calculated KA,I values in Table 2. Therefore, the developed sensors are the most suitable electrodes for assay of FUR in the presence of such studied interferents, too.

Table 2 The selectivity coefficients of sensors for different interferents

Validation of sensors

The developed FUR-BPB and FUR-BCG sensors were evaluated for linearity by following the IUPAC [37, 38] recommendations. The sensors were validated for accuracy, precision, robustness, and ruggedness by following the ICH Guidelines [39]. The validation results for individual parameters are presented in the following sections.

Linearity and analytical and regression parameters

By following the IUPAC recommendations [37, 38], the linearity of the proposed analytical methods using FUR-BPB and FUR-BCG ISEs was evaluated. The results reflected that the ISEs respond rapidly and linearly for FUR concentration ranges presented in Table 3. The calibration graphs were linear (Fig. 3), and the regression equations were y = 58.73x + 322.8 and y = 57.65x + 336.89 for FUR-BPB and FUR-BCG ISEs, and the corresponding Nernstian slopes were 58.73 and 57.66 mV/decade. The regression coefficients, slopes, intercepts of the calibration lines and other relevant performance characteristics of both analytical methods using proposed sensors were calculated as described in the literature [37, 38, 40] and are mentioned in Table 3. The detection limits (LOD) were also calculated [37, 38] and reported in Table 3.

Table 3 Characteristics of performance of FUR sensors
Fig. 3
figure 3

Calibration curves

Intra- or same- and inter- or different-day precision and accuracy

Seven replicates each of three differently concentrated FUR solutions between the linear ranges were prepared; analyses were carried out on the same day and the %RSD values were calculated for intra- or same-day variations. However, inter- or between-day studies were done by analyzing five replicates of pure FUR solutions of three different concentrations during different days. The %RSD values were calculated for resultant FUR amounts. The accuracy was appraised by calculating the relative error (RE), the metric for accuracy, for each resultant concentration of FUR. The %RSD values of 1.13 to 3.11% declared the suitability of both sensors for precise determination of FUR. The accuracy was expressed using the metric of %RE and showed below in Table 4. The %RE values which ranged from 0.48 to 4.67 (Table 4) revealed the accurate performance of sensors. The sensors functioning without crossing the limits of accuracy range from 95.0 to 105.0% for pharmaceutical samples as recommended by ICH Guidelines [41] and USP [42] recommendations and made their applicability for use as quantification sensors for FUR.

Table 4 Results in FUR found with %RSD and %RE for showing intra- and inter-day accuracy and precisions

Robustness and ruggedness

The robust functioning of proposed ISEs was assessed by varying working pH and temperatures slightly. FUR solutions of 0.9, 1.5, and 2.1 mmol/L were used for analysis. The pH was varied by 0.02 [3.71 (± 0.02)] units for each sensor, and the calculated values of %RSD were from 1.78 to 3.33. Besides, the robustness was also been evaluated by varying the temperature by 1 °C during the measurement of potentials of FUR solutions of different concentrations. The temperatures of the FUR solutions were brought to 23 ± 1 and 37 ± 1 °C using both the sensors; potentials were measured; and %RSD values were reported. Range of %RSD from 2.68 to 3.52 indicated the robust functioning of proposed sensors.

Analyses were done using three different potentiometers for 5 days by the same and different analysts to assess the ruggedness. The variations expressed in %RSD values for inter-potentiometric and inter-analysts were < 5%, reflecting sensors’ rugged function. Summary of results of this study are presented in Table 5.

Table 5 Results showing robust and rugged functioning of sensors

Application of ISEs to tablets analysis

Five replicates of each of 0.9, 1.5 and 2.1 mmol/L in FUR from tablet extracts were analyzed with proposed Sensor I and Sensor II by using the described procedure for tablet analysis. The mean of the measured potentials obtained and the percent recoveries of FUR were evaluated. The results of the United States Pharmacopeia (USP) assay method [2] were compared with the obtained results using the proposed sensors. The procedure in reference method describes titration of FUR (about 600 mg) in dimethylformamide against standardised 0.1 M NaOH using bromothymol blue indicator. The Student’s t and and F values were intently obtained by calculation. The values of tcal and Fcal are tabulated in Table 6. The tcal and Fcal values are less than the values which are statistically tabulated. Therefore, the proposed sensors yielded accurate and precise results.

Table 6 Results of Lasix tablet (40 mg FUR/tablet) analysis with proposed sensors and statistical tests of obtained results of proposed and reference analytical methods

Recovery study

The accurate functioning of proposed ISEs was further assessed by recovery studies following the standard addition method. The aliquots of pre-analyzed tablet extracts were spiked with pure FUR solution in three different concentration levels. To a fixed amount of five replicates of FUR tablet extract, pure FUR in amounts corresponding to 50, 100, and 150% to that of the amount from tablets were spiked; pH was manipulated to optimum value, and content was diluted to 10 mL and potentials measured. Then, FUR concentrations were calculated using the regression data. The %FUR recovered was calculated for each level. The mean %FUR recovered was 98.33 with standard deviation of less than 3% revealed that good and acceptable recovery is possible with proposed sensors. The relevant results of recovery experiments are summarised below (Table 7).

Table 7 Results of recovery test for Lasix tablets

Discussion

The development of ISEs using the ion pair complexes of therapeutic compounds and sulphonthalein dyes through their application to quantify FUR is described for the first time. The action of two sulphonthalein dyes viz. BPB and BCG is en-cashed to construct the membrane sensors or ISEs for FUR. Chemically, the nitrogenous group of FUR will possibly be utilised to react with anionic dye, as chromogenic reagent. In CHCl3, the resulting yellow product of FUR-BPB or FUR-BCG is due to the transformation of the lactoid ring structure to quinoid form [43] in the anionic dyes, BPB and BCG. The two tautomers (lactoid and quinoid) are supposed to be present in equilibrium. However, due to the acidic property of the sulphonic acid group, the latter (quinoid) form is assumed predominant. Finally, protonated FUR (FURH+) forms ion-pair with the anionic dye (BPB- or BCG-). These possible reaction mechanisms are presented in Scheme 1. Anionic dye such as BPB or BCG forms an ionic aggregate or ion pair complex with a positively charged drug, and the product formed between the drug and dye is an electrically neutral ion pair complex molecule. This formed the basis to develop novel membranes to fabricate potentiometric sensors of FUR with BPB and BCG in THF and PVC.

Scheme 1
scheme 1

Probable reaction pathways showing protonation of FUR, ring opening of BPB or BCG and formation of FUR-BPB or FUR-BCG ion pair complexes

The acidic solution of FUR reacted with either BPB or BCG and formed a stable aqueous insoluble product of ion pair of yellow-coloured appropriate grain size. The probable structures of ion pair complexes are given below (Scheme 1). The aqueous insoluble ion pair complex of FUR-BPB or FUR-BCG is useful as a recombinant material to fabricate the membrane sensor. The membrane is formed effectively with uniform thickness all over its area when PVC is used as matrix and DBP is used as plasticizer. The membranes constructed here should therefore selectively respond to FUR, as the artificial ion-selective membrane sensors. The potentiometry will enable their use to confirm the selective functioning by generating the potential difference due to the varied concentrations of FUR solutions at opposite sides of the membrane [44].

The obtained product was used successfully to construct two sensors, namely FUR-BPB (Sensor I) and FUR-BCG (Sensor II) using THF as solvent and PVC. These acid–base indicator-based Sensor I and Sensor II were utilised for the first time to quantify FUR in pure form and in tablets.

The membranes developed were used to fabricate Sensors I and II and to measure the potentials of the drug solutions. The systematic representation of the electrochemical cell is represented as

Conclusion

For the first time, the two membrane sensors for potentiometric determination of frusemide using two acid dyes, namely bromophenol blue and bromocresol green are presented. The sensors are applicable to quantify frusemide over a wider linear range of concentration with the Nernestian response. The sensors have been successfully used to reliably determine frusemide in pure state and tablets. Good agreement evolved out of the results while using these sensors to determine frusemide accurately and precisely. Potentiometry with these sensors is the most selective technique to determine frusemide in pharmaceutical laboratories. Since, potentiometry is neither highly sophisticated nor have stringent steps to be followed; these could be used as a conventional technique in place of other high-cost techniques such as HPLC, LC-MS, GC-MS, etc.